The ReTrace Ureteral Access Sheath is used to establish a continuous conduit during urological endoscopic procedures facilitating the in and out passage of endoscopes and other instruments into the urinary tract.
Device Story
ReTrace Ureteral Access Sheath; reinforced polymeric tube/sheath with introducer/dilator; used to establish continuous conduit in urinary tract during urological endoscopic procedures. Facilitates in/out passage of endoscopes and instruments. Features hydrophilic coating on exterior/distal tip, lubricious inner surface, kink-resistant reinforcement, and radiopaque distal ring. Includes Luer connector, guidewire entry/exit eye, and fluid delivery holes. Used in clinical settings by physicians. Device provides kink-resistant access, potentially reducing trauma during instrument exchange. Single-use, sterile, disposable.
Clinical Evidence
No clinical data or animal studies provided. Substantial equivalence supported by bench testing, including visual inspection, dimensional verification, guidewire pullout force, device integrity/functionality, simulated use, viscous fluid testing, kink resistance, and shelf life validation.
Technological Characteristics
Reinforced polymeric sheath (PEBA, PTFE, PVC, polycarbonate, Nitinol, stainless steel). Hydrophilic coating. Ethylene oxide sterilization (SAL 10^-6). Dimensions: 10/12 Fr and 12/14 Fr diameters; 28 cm and 55 cm lengths. Single-use. Mechanical sensing/actuation via guidewire-assisted insertion.
Indications for Use
Indicated for patients undergoing urological endoscopic procedures requiring a conduit for instrument passage into the urinary tract.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
K123675 — RE-TRACE URETERAL ACCESS SHEATH, 10/12 FRENCH, LENGTH 35 AND 45 CM, URETERAL ACCESS SHEATH, 12/14 CH-FR, LENGTH 35CM, UR · Coloplast A/S · Mar 1, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 7, 2018
Coloplast A/S Cori L. Ragan Regulatory Affairs Manager 1601 West River Road North Minneapolis, MN 55411
Re: K181811
Trade/Device Name: ReTrace Ureteral Access Sheath Regulation Number: 21 CFR& 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FED Dated: August 7. 2018 Received: August 8, 2018
Dear Cori L. Ragan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K181811
Device Name ReTrace Ureteral Access Sheath
Indications for Use (Describe)
The ReTrace Ureteral Access Sheath is used to establish a continuous conduit during urological endoscopic procedures facilitating the in and out passage of endoscopes and other instruments into the urinary tract.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Confidential
{3}------------------------------------------------
# 7. SPECIAL 510(k) SUMMARY
### SUBMITTER I.
| 510(K) Owner's Name: | Coloplast A/S |
|-----------------------------|------------------------------------------------------------------------------------------------|
| Legal Manufacturer Address: | Holtedam 1<br>3050 Humlebaek, Denmark |
| Phone/Fax/Email: | Phone: (612) 597-5106<br>Email: usclr@coloplast.com |
| Name of Contact Person: | Cori L. Ragan<br>Regulatory Affairs Manager |
| Address/Contact: | 1601 West River Road<br>Minneapolis, MN 55411 |
| Date Prepared: | June 28, 2018 |
| II. DEVICE | |
| Trade or Proprietary Name: | ReTrace® Ureteral Access Sheath |
| Common or Usual Name: | Endoscopic Access Overtube, Gastroenterology-Urology |
| Classification Name: | Endoscope and Accessories<br>(21 CFR section 876.1500)<br>Product Code: FED<br>Device Class: 2 |
| Classification Panel: | Gastroenterology-Urology |
### PREDICATE DEVICE III.
The proposed additional lengths of the ReTrace Ureteral Access Sheath are substantially equivalent in performance, indication, design, and materials to the Coloplast's own predicate device, ReTrace Ureteral Access Sheaths cleared under 510(k) K140523 and K123675.
These predicates have not been subject to a design-related recall. No reference devices were used in this submission.
{4}------------------------------------------------
### IV. DEVICE DESCRIPTION
The ReTrace Ureteral Access Sheath is comprised of the following components:
- Reinforced tube/sheath ●
- Introducer/dilator
- Connector ●
- Clip
The reinforced polymeric introducer sheath includes a hydrophilic coating on the exterior and a lubricious inner surface. The sheath is reinforced to provide kink resistance. The distal tip of the sheath is fitted with a radiopaque ring. A white connector and orange clip is fitted on the proximal end of the sheath.
The introducer is fitted with a Luer connector on the proximal end and has a hydrophilic coating on the distal end. A guidewire entry and exit eye and exit holes for fluid delivery are located at the distal end of the introducer.
This submission expands the product line to include 28 cm long and 55 cm long sheaths with inner diameters of 10 and 12 Fr.
#### V. INDICATIONS FOR USE
The modified ReTrace Ureteral Access Sheaths have the same intended use as the previously cleared models which is as follows:
The ReTrace Ureteral Access Sheath is used to establish a continuous conduit during urological endoscopic procedures facilitating the in and out passage of endoscopes and other instruments into the urinary tract.
{5}------------------------------------------------
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE
The modified ReTrace Ureteral Access Sheath is substantially equivalent in performance, indication, design, and materials to the previously cleared, ReTrace Ureteral Access Sheath (K140523 and K123675).
| Technological<br>Characteristic | ReTrace Ureteral Access<br>Sheath<br>(Subject Device) | Retrace Ureteral Access<br>Sheath - K140523<br>(Predicate Device) | Retrace Ureteral Access<br>Sheath - K123675<br>(Predicate Device) |
|------------------------------------|-------------------------------------------------------------------------------------|-------------------------------------------------------------------|-------------------------------------------------------------------|
| Design | Access Sheath with<br>introducer | Same | Same |
| Sizes | 10 FR Inner Diameter (ID) /<br>12 FR Outer Diameter (OD)<br>and 12 FR ID / 14 FR OD | 12 FR ID / 14 FR OD | 10 FR ID / 12 FR OD and<br>12 FR ID / 14 FR OD |
| Length | 28cm and 55cm | 35 and 45cm | 35 and 45cm |
| Length of dilator out<br>of sheath | 5 cm | 5 cm | 5 cm<br>2 cm |
| Insertion Technique | Insertion over a guidewire | Same | Same |
| Device Materials | PEBA, PTFE, stainless steel,<br>PVC, polycarbonate, Nitinol | Same | Same |
| Coating | Hydrophilic coating | Same | Same |
| Single use | Yes | Same | Same |
| Sterility | Ethylene Oxide Sterilized | Same | Same |
| Shelf Life | 5 years | Same | 2 years |
### PERFORMANCE DATA VII.
The following testing data was provided in support of the substantial equivalence determination and changes to the product line.
## Biocompatibility Testing
No additional biocompatibility testing was provided.
## Mechanical Testing
Mechanical testing was completed in support of the substantial equivalence determination and changes to the product line.
- Visual ●
- Dimensional verification ●
- Guidewire pullout force
- Device integrity and functionality ●
- . Simulated use
- Viscous fluid
- Kink resistance
- Shelf life testing to support 5 years ●
{6}------------------------------------------------
## Sterilization
The ReTrace Ureteral Sheaths are sterilized using ethylene oxide in a validated cycle demonstrating a microbial assurance level of 10-6.
## Packaging and Distribution
No additional packaging or distribution testing was provided.
No animal studies or clinical testing were provided to support substantial equivalence between the subject and predicate devices.
## VIII. CONCLUSIONS
The additional lengths of the ReTrace Ureteral Access Sheath have been demonstrated to be substantially equivalent to the predicate, ReTrace Ureteral Access Sheath, based on the nonclinical data provided. The test results demonstrate that the additional models do not raise new questions of safety or effectiveness and are substantially similar to the ureteral access sheaths cleared as ReTrace Ureteral Access Sheaths in K140523 and K123675.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches