SHAPE-LOCKABLE ENDOSCOPIC OVERTUBE

K023902 · Usgi Medical, Inc. · FED · Aug 20, 2003 · Gastroenterology, Urology

Device Facts

Record IDK023902
Device NameSHAPE-LOCKABLE ENDOSCOPIC OVERTUBE
ApplicantUsgi Medical, Inc.
Product CodeFED · Gastroenterology, Urology
Decision DateAug 20, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The USGI Endoscopic Overtube is intended to be used with an endoscope to facilitate intubation, e.g., colonoscopy, change of endoscopes, removal of multiple polyps and/or foreign bodies.

Device Story

USGI Endoscopic Overtube is an endoscopic accessory providing external support to an endoscope. Device consists of multiple nested, bowl-shaped links held by four tension cables connected to an atraumatic tip. Operation: practitioner inserts endoscope through overtube; once desired configuration is reached, tension is applied to cables via handle, increasing friction between links to 'shape lock' the tube. Locked configuration allows repeated advancement or withdrawal of the endoscope without losing the established shape. Used in clinical settings (e.g., endoscopy suites) by physicians to facilitate procedures like colonoscopy, endoscope changes, and polyp/foreign body removal. Benefits include improved scope stability and maneuverability during complex endoscopic interventions.

Clinical Evidence

Bench testing only. Mechanical and functional testing demonstrated that the device performs as intended. No clinical data provided.

Technological Characteristics

Shape-locking tube with nested links, four tension cables, atraumatic tip, slip-coated inner liner, and handle. Mechanical friction-based locking mechanism. Patient-contacting materials have established history of biocompatibility. Manual operation.

Indications for Use

Indicated for use with an endoscope to facilitate intubation, including colonoscopy, endoscope exchange, and removal of polyps or foreign bodies in patients requiring endoscopic procedures.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle emblem. The logo is presented in black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 JUL 2 7 2015 USGI Medical, Inc. % Judy F. Gordon, D.V.M. Regulatory Consultant ClingReg Consulting Services, Inc. 2 Delphinus IRVINE. CA 92612 Re: K023902 Trade/Device Name: USGI Shape Locking Endoscopic Oveertube Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FED, FDF Dated (Date on orig SE ltr): May 19, 2003 . Received (Date on orig SE Itr): May 22, 2003 Dear Dr. Gordon, This letter corrects our substantially equivalent letter of August 20, 2003. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be {1}------------------------------------------------ Page 2 - found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Benjamin R. Fisher -S Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Page _1_ of _1_ 510(k) Number (if known): Device Name: USG! Endoscopic Overtube Indications for Use: . 4. The USGI Endoscopic Overtube is intended to be used with an endoscope to facilitate intubation, e.g., colonoscopy, change of endoscopes, removal of multiple polyps and/or foreign bodies. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use | <div> <span style="text-decoration: overline;"></span> </div> | |----------------------|---------------------------------------------------------------| | (Per 21 CFR 801.109) | | | | OR | Over-The-Counter Use | <div> <span style="text-decoration: overline;"></span> </div> | |--|----|--------------------------|---------------------------------------------------------------| | | | (Optional Format 1-2-96) | | | (Division Sign-Off) | <div> <span style="text-decoration: overline;">Nancy C. Brogdon</span> </div> | |---------------------|-------------------------------------------------------------------------------| |---------------------|-------------------------------------------------------------------------------| Division of Reproductive, Abdominal, and Radiological Devices | 510(k) Number | K0 23902 | |---------------|----------| |---------------|----------| {3}------------------------------------------------ K023902 ## SUMMARY OF SAFETY & EFFECTIVENESS This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. | APPLICANT | USGI Medical, Inc.<br>940 Calle Amanecer, Suite A<br>San Clemente, CA 92674<br>Tel: (949) 369-3890<br>Fax: (949) 369-3891 | |---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | OFFICIAL<br>CORRESPONDENT | Judy F. Gordon, D.V.M.<br>ClinReg Consulting Services. Inc.<br>2 Delphinus<br>Irvine, CA 92612<br>Tel: (949) 854-6314<br>Fax : (949) 854-9652<br>e-mail: judygordon@earthlink.net | | TRADE NAME: | USGI Endoscopic Overtube | | COMMON NAME: | Overtube | | CLASSIFICATION<br>NAME: | Endoscope and accessories | | DEVICE<br>CLASSIFICATION: | Class 2 per 21 CFR §876.1500 | | PRODUCT CODE | 78 (KOG) | | PREDICATE DEVICE: | Olympus ST-C3 Overtube (K903842), the Bard Endoscopic<br>Overtube (K973500), and the NDO Surgical, Inc. Overtube<br>(K002018) | Description of the Device Subject to Premarket Notification: The USGI Endoscopic Overtube is an endoscopic accessory designed to provide The OOO! Endoscope without hindering its flexibility and maneuverability. It is an overtube guide that can conform to any bend configuration of the scope and then le an overlabor in that form. In the "shape locked" configuration, the scope can be further advanced or withdrawn repeatedly. ### INDICATION FOR USE: The USGI Endoscopic Overtube is an endoscopic accessory designed to provide external support to an endoscope without hindering its flexibility or maneuverability. It is intended to be used with an endoscope to facilitate intubation, e.g., colonoscopy, change of endoscopes, removal of multiple polyps and/or foreign bodies. {4}------------------------------------------------ Kaaao-v P-g a #### TECHNICAL CHARACTERISTICS: The USGI Endoscopic Overtube is a shape locking tube with an atraumatic tip, a slip coated inner liner, a smooth outer skin, and a handle. The overtube is constructed from multiple nested links that are held together by four cables. Each link is bowl shaped, but without a bottom and has four equally spaced annular holes running longitudinally through the wall thickness. The purpose of the holes is to accommodate the tension wires which pass through each link. The wires are connected to the tip but slide freely through all other links. Applying tension to the wires squeezes the links together. The friction between links maintains their relative positions, allowing the practitioner to "lock in" a shape. #### PERFORMANCE DATA: All components that come in direct contact with the patient have a long history of use in medical devices and are biocompatibile This 510(k) notice includes mechanical and functional bench testing that demonstrate that the USGI Endoscopic Overtube performs as intended. #### BASIS FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE: The USGI Endoscopic Overtube is substantially equivalent to the Olympus ST-C3 Overtube (cleared for market under K903842), the Bard Endoscopic Overtube (cleared for market under K973500), and the NDO Surgical, Inc. Overtube (cleared for market under K002018). All four devices have the same indication for use and are made of the same materials (patient-contacting materials). Bench testing demonstrates that the devices are functionally equivalent.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...