ED-3410, VIDEO DUODENOSCOPE

K963056 · Pentax Precision Instrument Corp. · FDT · Oct 10, 1996 · Gastroenterology, Urology

Device Facts

Record IDK963056
Device NameED-3410, VIDEO DUODENOSCOPE
ApplicantPentax Precision Instrument Corp.
Product CodeFDT · Gastroenterology, Urology
Decision DateOct 10, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic, Pediatric

Intended Use

The ED-3410, Video Duodenoscope, is intended to provide optical visualization (via a video monitor) of, and therapeutic access to, the Upper Gastrointestinal Tract and Biliary Tract. The Upper Gastrointestinal Tract and Biliary Tract includes, but is not limited to, the organs; tissues; and subsystems: Esophagus, Stomach, Duodenum, Common Bile Duct, Hepatic Duct, and Cystic Duct. The instrument is introduced per orally when indications consistent with the requirement for the procedure are observed in Adult and Pediatric patient populations.

Device Story

Flexible video duodenoscope; used for visualization and therapeutic access to upper GI and biliary tracts. Input: light reflected from body cavity captured by CCD sensor. Operation: requires connection to external video processor; processor uses 300W xenon lamp with RGB color wheel; CCD collects sequential color strobes; processor compiles full-color frames for monitor display. Features: angulation controls, air/water delivery, suction, elevator control, accessory inlet for biopsy forceps. Used in clinical settings by physicians. Output: real-time video feed on monitor; assists physician in diagnostic and therapeutic interventions. Benefits: enables minimally invasive visualization and access to GI/biliary structures.

Clinical Evidence

No clinical data; submission based on bench testing and technical specifications comparison.

Technological Characteristics

Flexible insertion tube with CCD imaging sensor and fiber optic light bundles. Requires external video processor with 300W xenon lamp and RGB color wheel. Includes manual controls for angulation, suction, and elevator mechanism. Connectivity: proprietary umbilicus connection to video processor.

Indications for Use

Indicated for optical visualization and therapeutic access to the Upper Gastrointestinal Tract (esophagus, stomach, duodenum) and Biliary Tract (common bile, hepatic, and cystic ducts) in adult and pediatric patients requiring such procedures.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} DOCUMENT NUMBER: KS041.B7 DATE PREPARED: 07-23-96 PAGE 1 OF 1 PREPARED BY: PAUL SILVA 510(K) SUMMARY DEVICE: ED-3410, VIDEO DUODENOSCOPE K963056 OCT 10 1996 1. SUBMITTER INFORMATION: Pentax Precision Instrument Corporation 30 Ramland Road Orangeburg, NY, 10962 TEL: (914)-365-0700 2. NAME OF DEVICE: Trade Name: ED-3410, Video Duodenoscope Classification Name: Duodenoscope (Esophago Gastro) (78FDT), 876.1500 3. PREDICATED DEVICE(S) INFORMATION: 1. FD-34A, Fiber Duodenoscope Pentax K832006 2. VB-1530/ 1830, Video Bronchoscope Pentax K934920 3. EPM-3300, Video Processor Pentax K934918 4. DEVICE DESCRIPTION: The ED-3410, Video Duodenoscope, must be used with a Video Processor (a software controlled device). The endoscope has a flexible insertion tube, a control body, and umbilicus. The umbilicus provides connection to the video processor. The control body includes controls for up/ down/ left/ right angulation, air/water delivery, suction, elevator control, and an accessory inlet port. The device contains light carrying bundles, to illuminate the body cavity, and a charge couple device (CCD) to collect image data. The instrument contains a working channel through which biopsy devices, or other devices, may be introduced (the instrument is supplied with two biopsy forceps). The Video Processor contains a 300 watt xenon lamp which provides white light that is filtered, via a Red, Green, Blue color wheel, and is focused at the connected video endoscope lightguide prong. The endoscope light carrying bundles present the color strobes to the body cavity and the CCD collects image data for each strobe of color. The video processor stores the CCD information until all three color strobes are completed and a full color frame is compiled. Image data and other screen display information are formatted and presented to the video outputs of the video processor for display. 5. INTENDED USE: The ED-3410, Video Duodenoscope, is intended to provide optical visualization (via a video monitor) of, and therapeutic access to, the Upper Gastrointestinal Tract and Biliary Tract. The Upper Gastrointestinal Tract and Biliary Tract includes, but is not limited to, the organs; tissues; and subsystems: Esophagus, Stomach, Duodenum, Common Bile Duct, Hepatic Duct, and Cystic Duct. The instrument is introduced per orally when indications consistent with the requirement for the procedure are observed in Adult and Pediatric patient populations. 6. COMPARISON TO PREDICATED DEVICE(S): The submission for substantial equivalence included ED-3410 literature including specifications, the identification of standard set components, and identification of optional accessories, comparison tables were provided to illustrate the comparisons to the predicated devices. The submission for substantial equivalence was not based on an assessment of clinical performance data.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...