← Product Code [FDT](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDT) · K143153

# EVIS EXERA II DUODENOVIDEOSCOPE OLYMPUS TJF TYPE Q180V (K143153)

_Olympus Medical Systems Corp. · FDT · Jan 15, 2016 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDT/K143153

## Device Facts

- **Applicant:** Olympus Medical Systems Corp.
- **Product Code:** [FDT](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDT.md)
- **Decision Date:** Jan 15, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Intended Use

This instrument has been designed to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery within the duodenum.

## Device Story

Flexible video endoscope for endoscopic diagnosis and treatment within the duodenum. Device features sealed elevator wire channel and dual guidewire locking mechanism. Used in clinical settings by trained physicians; requires connection to Olympus video system center, light source, monitor, and documentation equipment. Physician operates endoscope controls to navigate duodenum; utilizes EndoTherapy accessories (e.g., biopsy forceps) for procedures. Output consists of real-time video feed displayed on monitor for clinical visualization and decision-making. Sealed channel and locking mechanism enhance procedural control and device integrity. Benefits include improved diagnostic and therapeutic access to the duodenum.

## Clinical Evidence

Bench testing only. Includes performance testing for sealed elevator wire channel and dual guidewire locking mechanism, reprocessing validation per FDA guidance, biocompatibility per ISO 10993, software validation, electromagnetic compatibility, electrical safety, thermal safety, and risk analysis per ISO 14971.

## Technological Characteristics

Flexible video endoscope. Features: sealed elevator wire channel, dual guidewire locking mechanism. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, ISO 11135-1, ISO 10993-1/5/7/10, ISO 14971. Sterilization: Ethylene oxide (implied by ISO 11135-1). Connectivity: Proprietary interface to Olympus video system center.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- XTJF Q160VF1 DUODENOVIDEOSCOPE ([K080403](/device/K080403.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of three human profiles facing right, layered on top of each other to create a sense of depth. The profiles are black and are positioned above the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the graphic.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 15, 2016

Olympus Medical Systems Corp. % Sheri Musgnung Manager, Regulatory Affairs Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610

Re: K143153

> Trade/Device Name: EVIS EXERA II Duodenovideoscope Olympus TJF Type Q180V Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: FDT. NWB Dated: December 8, 2015 Received: December 11, 2015

Dear Sheri Musgnung.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

K143153

Device Name

EVIS EXERA II DUODENOVIDEOSCOPE OLYMPUS TJF TYPE Q180V

### Indications for Use (Describe)

This instrument has been designed to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery within the duodenum.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

## FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Olympus. The word "OLYMPUS" is written in large, bold, blue letters. Underneath the word is a yellow line, and below that is the text "Your Vision, Our Future" in a smaller, blue font.

## 510(k) SUMMARY DUODENOVIDEOSCOPE EVIS MPUS TJF TYPE Q180

# January 6, 2016

#### General Information I.

| Applicant: | OLYMPUS MEDICAL SYSTEMS CORP.<br>2951 Ishikawa-cho, Hachioji-shi, Tokyo, Japan<br>192-8507<br>Establishment Registration No: 8010047 |
|------------|--------------------------------------------------------------------------------------------------------------------------------------|
|------------|--------------------------------------------------------------------------------------------------------------------------------------|

- Sheri L. Musgnung Official Correspondent: . Manager, Regulatory Affairs Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610, USA Phone: 484-896-3147 FAX: 484-896-7128 Email: sheri.musgnung@olympus.com
- AIZU OLYMPUS CO., LTD. . Manufacturer: 500 Aza Muranishi Ooaza-Iidera Monden-cho Aizuwakamatsu-Shi Fukushima, JAPAN 965-8520 Registration Number: 9610595

#### Device Identification II.

| ■ | Device Trade Name: | EVIS EXERA II DUODENOVIDEOSCOPE<br>OLYMPUS TJF TYPE Q180V |
|---|--------------------|-----------------------------------------------------------|
| ■ | Common Name:       | Duodenoscope                                              |
| ■ | Regulation Number: | 876.1500                                                  |
| ■ | Regulation Name:   | Endoscope and Accessories                                 |
| ■ | Regulatory Class:  | II                                                        |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image displays the logo for Olympus, a company specializing in optics and imaging. The word "OLYMPUS" is written in large, bold, blue letters. A thin, horizontal, yellow line is underneath the company name. Below the line, the company's slogan, "Your Vision, Our Future," is written in a smaller font.

- Classification Panel: Gastroenterology and urology . FDT; Duodenoscope, accessories, flexible/rigid ■ Product Code: NWB; Endoscope, accessories, narrow band spectrum
#### Predicate Device Information III.

- Device Trade Name: XTJF Q160VF1 DUODENOVIDEOSCOPE
- . 510(k) Number K080403
- . Decision Date 05/20/2008
- . Manufacturer AIZU OLYMPUS CO., LTD

#### Device Description IV.

The EVIS EXERA II DUODENOVIDEOSCOPE OLYMPUS TJF TYPE Q180V is a flexible video endoscope used for endoscopic diagnosis and treatment within the duodenum. The device has a sealed elevator wire channel and a dual guidewire locking mechanism.

#### V. Indications for Use

This instrument has been designed to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery within the duodenum.

## VI. Comparison of Technological Characteristics

The difference between the subject device and the predicate device is as follows:

- Consists of a sealed elevator wire channel ●
- Consists of a dual guidewire locking mechanism
- . Modifications to the device's labeled reprocessing procedure
- . Addition of recommended combination of equipment and accessories
- . Physical dimensions of components at distal end tip

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Olympus. The word "OLYMPUS" is written in large, bold, blue letters. Below the word is a thin, yellow line. Under the line, the words "Your Vision, Our Future" are written in a smaller, blue font.

# VII. Summary of non-clinical testing

The differences of technological characteristics between the predicate device and the subject device are confirmed that they are substantially equivalent through the following tests and standards.

- Performance testing and design analysis was carried out to demonstrate the safety and effectiveness of the sealed elevator wire channel and the dual guidewire locking mechanism.
- The validation test on the reprocessing is performed in accordance with the FDA guidance "Labeling Reusable Medical Devices for Reprocessing in Health Care Facilities: FDA Reviewer Guidance - April 1996". The FDA guidance "Draft Guidance for Industry and FDA Staff - Processing/Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling -May 2011" is also taken into consideration.
- Biocompatibility testing is performed in accordance with the FDA Guidance, Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" dated on May 1 1995.
- The software validation activities were performed in accordance with the FDA Guidance, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
- Electromagnetic compatibility, electric safety, and thermal safety had been confirmed.
- Risk analysis was carried out in accordance with established in-house acceptance . criteria based on ISO 14971:2007. The design verification tests and their acceptance criteria were identified and performed as a result of this risk analysis assessment.
- The following standards have been applied to the TJF-Q180V.
	- IEC 60601-1 -
	- IEC 60601-1-2 -
	- IEC 60601-2-18 -
	- ISO 11135-1 -
	- -ISO 10993-7
	- ISO 10993-1 -
	- ISO 10993-5 -
	- ISO 10993-10 -
	- ISO 14971 -

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the Olympus logo. The word "OLYMPUS" is written in large, bold, blue letters. Below the word is a yellow line. Underneath the line is the text "Your Vision, Our Future".

# VIII. Conclusion

Based on the 510(k) submission data, Olympus believes that the subject device and the predicate device selected are substantially equivalent and do not change the fundamental scientific technology and indications for use.

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDT/K143153](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDT/K143153)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
