← Product Code [FDS](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDS) · K971660

# OLYMPUS GF-UM130 ULTRASOUND GASTROVIDEOSCOPE, ITS ASSOCIATED ACCESSORIES AND ANCILLARY EQUIPMENT (K971660)

_Olympus America, Inc. · FDS · Jul 25, 1997 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDS/K971660

## Device Facts

- **Applicant:** Olympus America, Inc.
- **Product Code:** [FDS](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDS.md)
- **Decision Date:** Jul 25, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

Olympus GF-UM130 Ultrasound Gastroscope has been designed to be used with an Olympus Endoscopic Ultrasound Center, EVIS Video System Center, light source, documentation equipment, video monitor, Endo-Therapy accessories and other ancillary equipment for endoscopic ultrasound imaging of the gastrointestinal wall, biliary and pancreatic duct, and surrounding organs.

## Device Story

Olympus GF-UM130 is an ultrasound gastrovideoscope used for endoscopic ultrasound (EUS) imaging. Device integrates optical video endoscopy with ultrasonic pulsed echo imaging. Operates by inserting scope into gastrointestinal tract; ultrasound transducer captures internal images of GI wall, biliary/pancreatic ducts, and surrounding organs. Used in clinical settings by trained physicians. System requires connection to Olympus Endoscopic Ultrasound Center, EVIS Video System Center, light source, and video monitor. Output displayed on monitor for real-time visualization; assists physicians in diagnostic assessment of internal structures. Benefits include minimally invasive visualization of deep tissue and organs not accessible by standard endoscopy alone.

## Clinical Evidence

Bench testing only. No clinical data provided.

## Technological Characteristics

Combination ultrasound gastrovideoscope. Features ultrasonic pulsed echo imaging and optical video endoscopy. Connects to external ultrasound center, video system center, and light source. Form factor is a flexible endoscope. Sterilization via standard endoscopic reprocessing protocols.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Olympus GF-UM20 Ultrasound Gastroscope ([K926514](/device/K926514.md))
- Olympus GIF Type 140 Series Gastrovideoscopes ([K954451](/device/K954451.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image is a black and white logo for the Department of Health & Human Services USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a bird in flight, but upon closer inspection, it is composed of three human profiles facing to the right.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# JUL 2 5 1997

Re: K971660 .. . . . . . . . . . . .

- Olympus GF-UM130 Ultrasound Gastrovideoscope, its associated accessories and ancillary equipment Dated: May 2, 1997 Received: May 6, 1997 Regulatory Class: II 21 CFR 876.1500/Procode: 78 FDS 21 CFR 892.1560/Procode: 90 IYO
Subhash R. Patel Regulatory Affairs Associate Endoscope Division OLYMPUS AMERICA, INC. Two Corporate Center Drive Melville, New York 11747-3157

Dear Mr. Patel:

We have reviewed your section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug and Cosmetic Act. You may, therefore, market the device, subject to the general controls provisions Act (Act). The general controls provisions of the Act include requirements for registration, listing of devices, good manufacturing practices, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice requirement , as set forth in the Ouality System Regulation (OS) for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that. through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification does not affect any obligation you may have under sections 531 and 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This determination of substantial equivalence is granicd on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information. including acoustic output measurements based on production line devices, requested in Appendix G. (enclosed) of the Center's February 17, 1993 "Revised 510(k) Diagnostic Uitrasound Guidance for 1993." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.

{1}------------------------------------------------

Page 2 - Mr. Subhash Patel

The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:

> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard ... _ ... ... ... ... ... ... .... ... Rockville, Maryland 20850

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".

If you have any questions regarding the content of this letter, please contact Mary Joan Cornelius at (301) 594-2194.

Sincerely yours,

Robert R. Satting/

s

Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure(s)

{2}------------------------------------------------

K971660

### 510 (k) SUMMARY OLYMPUS GF-UM 130 ULTRASOUND GASTROVIDEOSCOPE

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR, Section 807.92.

| Device Name:                                 | Olympus GF-UM130 Ultrasound Gastrovideoscope                                                                                                     |
|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| Common/Usual Name:                           | Ultrasound Gastroscope                                                                                                                           |
| Classification Name:                         | Class II, 21 CFR 876.1500<br>Endoscope and Accessories.                                                                                          |
|                                              | Class II, 21 CFR 892-1560<br>Ultrasonic Pulsed Echo Imaging System                                                                               |
| Predicate Devices:                           | Olympus GF-UM20 Ultrasound Gastroscope (K926514)<br>Olympus GIF Type 140 Series Gastrovideoscopes (K954451)                                      |
| Prepared & Submitted By:<br>(Contact Person) | Mr. Subhash Patel<br>Olympus America Inc.<br>Endoscope Division<br>Two Corporate Center Drive<br>Melville, New York 11747-3157<br>(516) 844-5481 |
| Summary Preparation Date:                    | 05/02/97                                                                                                                                         |

#### Statement of Intended Use:

Olympus GF-UM130 Ultrasound Gastroscope has been designed to be used with an Olympus Endoscopic Ultrasound Center, EVIS Video System Center, light source, documentation equipment, video monitor, Endo-Therapy accessories and other ancillary equipment for endoscopic ultrasound imaging of the gastrointestinal wall, biliary and pancreatic duct, and surrounding organs.

{3}------------------------------------------------

K 971660

## Indications for Use Statement

510(k) Number (if known):

Device Name:.

Not assigned yet

Olympus GF-UM130 Ultrasound Gastrovideoscope, its . . . associated accessories and ancillary equipment.

#### Indications for Use:

Olympus GF-UM130 Ultrasound Gastroscope has been designed to be used with an Olympus Endoscopic Ultrasound Center, EVIS Video System Center, light source, documentation equipment, video monitor, Endo-Therapy accessories and other ancillary equipment for endoscopic ultrasound imaging of the gastrointestinal wall, biliary and pancreatic duct, and surrounding organs.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use (per 21CFR 801.109) | <div></div> |
|--------------------------------------|-------------|
|--------------------------------------|-------------|

OR

| Over-the Counter Use (Optional Format 1-2-96) | <div></div> |
|-----------------------------------------------|-------------|
|-----------------------------------------------|-------------|

| (Division Sign-Off)                                                | Duler R Sathy/ |
|--------------------------------------------------------------------|----------------|
| Division of Reproductive, Abdominal, ENT, and Radiological Devices | <div></div>    |
| 510(k) Number                                                      | K971660        |

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDS/K971660](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDS/K971660)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
