GASTROSCOPE AND COLONOSCOPE NOZZLE REPLACEMENT KITS

K970372 · Precision Endoscopy of America, Inc. · FDF · Mar 11, 1997 · Gastroenterology, Urology

Device Facts

Record IDK970372
Device NameGASTROSCOPE AND COLONOSCOPE NOZZLE REPLACEMENT KITS
ApplicantPrecision Endoscopy of America, Inc.
Product CodeFDF · Gastroenterology, Urology
Decision DateMar 11, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

These endoscope nozzle replacement kits are intended for use in replacing the nozzle on 43 models of Olympus America, Inc. gastroscopes and colonoscopes.

Device Story

Endoscope nozzle replacement kit; includes 6 nozzles, 3 set screws, sealant, and tools (screwdriver, tweezers, magnifying glass). Used by service technicians to replace worn or damaged nozzles on Olympus gastroscopes and colonoscopes. Components (nozzle, screw, sealant) contact patient; tools are general-purpose accessories. Replacement process ensures proper nozzle alignment for continued endoscope function.

Clinical Evidence

No clinical data; bench testing/servicing experience cited.

Technological Characteristics

Kit components: nozzles, set screws, sealant. Tools: #1 screwdriver, tweezers, magnifying glass. No electronic, software, or energy-based components.

Indications for Use

Indicated for use in replacing nozzles on 43 specific models of Olympus America, Inc. gastroscopes and colonoscopes.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Submission Summary (Full Text)

{0} K 970372 Attachment E MAR 11 1997 # 510(k) Summary for Endoscope Nozzle Repair Kits Applicant: P. E. A. Products, Inc. 10969 McCormick Road Hunt Valley, Maryland 21031 Contact: Mr. Daniel M. Honeywell Phone: (800) 285-5959 ## Predicate Device: Olympus America, Inc. endoscope nozzles and set screws ## Device Description: These endoscope nozzle replacement kits are intended for use in replacing the nozzle on 43 models of Olympus America, Inc. gastroscopes and colonoscopes. Each kit includes 6 nozzles, three set screws, sealant, and the necessary tools (#1 screw driver, tweezers, and magnifying glass) to replace the nozzle. Each kit will include only the nozzle size(s) and set screw size(s) appropriate for use with the Olympus of America, Inc. endoscope models currently being used by the purchaser. ## Substantial Equivalence: The nozzles, set screws, and sealant are the only kit components that are incorporated into the endoscope and come in contact with the patient. The premarket notification contains information and documentation that these components are substantially equivalent to those used by the original manufacturer in that they are for the same intended use and have the same technological characteristics. The remaining kit components (i.e., tools) are general purpose articles not regulated by FDA as medical devices. They are provided both as a convenience and as an assurance that the proper tools are used to replace the nozzle. The applicant has extensive experience in the servicing and repair of endoscopes. This experience has demonstrated that nozzle replacement is easily performed and that proper nozzle alignment can readily be confirmed without the need for additional equipment.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...