# Pure-Vu System (K173392)

_Motus GI Medical Technologies , Ltd. · FDF · Dec 12, 2017 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDF/K173392

## Device Facts

- **Applicant:** Motus GI Medical Technologies , Ltd.
- **Product Code:** [FDF](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDF.md)
- **Decision Date:** Dec 12, 2017
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Intended Use

The Pure Vu System is intended to connect to standard colonoscopes to help facilitate intra-procedural cleaning of a poorly prepared colon by irrigating or cleaning the colon and evacuating the irrigation fluid (water), feces and other bodily fluids and matter, e.g. blood. It is for use only by trained medical personnel located in hospitals, clinics and doctor offices.

## Device Story

Pure-Vu System is an add-on device for standard colonoscopes; facilitates intra-procedural colon cleansing. System comprises an Oversleeve mounted on the colonoscope, a Workstation (WS), a WS Connector, and a Loading Fixture. Workstation performs cyclic irrigation (water/air mixture) and evacuation of fecal matter/fluids via pumps and regulators. Physician operates the cleansing process using external foot pedals. Device enables irrigation and suction at any time during the procedure without removing tools. Used in hospitals, clinics, and doctor offices by trained medical personnel. Benefits include improved visualization of the colon mucosa by removing debris, potentially enhancing diagnostic or therapeutic outcomes during colonoscopy.

## Clinical Evidence

No clinical data provided. Substantial equivalence is supported by bench testing, including biocompatibility, electrical safety, EMC, software validation, bond strength, pressure testing, system testing, and steering tests. All tests demonstrated the device functioned as intended.

## Technological Characteristics

System includes an Oversleeve (mounted on colonoscope), Workstation (WS) with Monitoring & Control Unit, Inlet Module (pumps/regulators), and Outlet Module (evacuation pumps). Connectivity via WS Connector. Materials re-tested per ISO 10993. Electrical input 100V-240V, 50/60 Hz. Safety standards: IEC 60601-1, IEC 60601-1-2. Irrigation/suction via cyclic air/water mixture. Disposable components; non-sterile.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Pure-Vu System ([K160015](/device/K160015.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

December 12, 2017

Motus GI Medical Technologies Ltd. % Hagit Ephrath VP of Health Economics, Regulatory and Clinical Affairs Hogan Lovells US LLP 555 Thirteenth Street, NW Washington, DC 20004

Re: K173392 Trade/Device Name: Pure Vu System Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FDF Dated: December 6, 2017 Received: December 8, 2017

Dear Hagit Ephrath:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/6 description: The image shows the name "Benjamin R. Fisher -S" in a large, bold font. The text is black and appears to be centered on a white background. The letters are clear and easy to read.

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K173392

Device Name Pure Vu System

Indications for Use (Describe)

The Pure Vu System is intended to connect to standard colonoscopes to help facilitate intra-procedural cleaning of a poorly prepared colon by irrigating or cleaning the colon and evacuating the irrigation fluid (water), feces and other bodily fluids and matter, e.g. blood.

It is for use only by trained medical personnel located in hospitals, clinics and doctor offices.

| Type of Use (Select one or both, as applicable)                                   |                                                                                  |
|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| <div> <span>     ☑ Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span>     ☐ Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for PURE-VU by MOTUS GI. The logo features a circular design on the left, composed of multiple blue dots arranged in a circular pattern. To the right of the circular design, the text "PURE-VU" is displayed in a stylized blue font. Below the text, the words "by MOTUS GI" are written in a smaller, gray font.

## 510(k) Summary

This special 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR 807.92.

| Applicant Information: | Motus GI Medical Technologies Ltd.22 Keren<br>Ha'yesod Str.<br>Tirat Carmel, 3902638Israel<br>Tel.: +972-4-6214446<br>Fax: +972-4-6214442                                                                                                                                                                                                                                                                                                        |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:        | Hagit Ephrath                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Phone Number:          | +972-4-6214447                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Fax Number:            | +972-4-6214442                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Establishment Regist.: | 3011816755                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Date Prepared:         | October 26th, 2017                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Trade Name(s):         | Pure-Vu System                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| CommonName:            | Pure-Vu System                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Classification Name:   | Endoscope and accessories                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Classification:        | Regulation No: 876.1500Class: II<br>Panel: Gastroenterology and Urology                                                                                                                                                                                                                                                                                                                                                                          |
| Predicate Device(s):   | Pure-Vu System (K160015)                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Intended Use:          | The Pure-Vu System is intended to connect to standard colonoscopes to help<br>facilitate intra-procedural cleansing by irrigating or cleaning the colon and<br>evacuating the irrigation fluid (water), feces, and other bodily fluids and matter,<br>e.g. blood.<br>It is for use only by trained medical personnel located in hospitals, clinics, and<br>doctors' offices.                                                                     |
| Reason for Submission: | The purpose of this special 510(k) is to modify the Pure Vu System by:<br>1) Simplifying the process of unloading the colonoscope by using disposable<br>materials rather than an Unloading Fixture<br>2) Minor material modifications to the Oversleeve and Workstation Connector,<br>3) Minor modifications to Workstation, and Loading Fixture components<br>4) Minor software modifications<br>5) Modify slight the air/water specifications |

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for "PURE-VU by MOTUS GI". The logo features a circular design on the left, composed of many small blue dots arranged in a spiral pattern. To the right of the circular design is the text "PURE-VU" in a sans-serif font, with the letters in varying shades of blue. Below the "PURE-VU" text is the text "by MOTUS GI" in a smaller font size.

| Technological<br>Characteristics:         | The Pure-Vu system comprises the following components:                                                                                                                                                                                                                                                                                                                                            |  |  |  |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|
|                                           | <b>Oversleeve</b> - The Oversleeve is mounted on standard commercially<br>available colonoscopes allowing thephysician to clean the colon.                                                                                                                                                                                                                                                        |  |  |  |
|                                           | <b>Workstation (WS)</b> - The Workstation operates by usingcyclic irrigation and<br>evacuation of colon content. Irrigationis based on mixture of liquid and air. The<br>workstation includes:                                                                                                                                                                                                    |  |  |  |
|                                           | Monitoring & Control Unit that continuously monitors and controls<br>irrigation and evacuation. Inlet Module that includes pumps and regulators enabling water &<br>airflow into the cleansing device. Outlet Module that includes pumps to evacuate fecal matter and<br>fluids from the colon. External on\off foot pedals that operate the cleansing process to be<br>used by the physician.    |  |  |  |
|                                           | <b>WS Connector (WSC)</b> connects the Oversleeve to the WS and saline or water<br>bag.                                                                                                                                                                                                                                                                                                           |  |  |  |
|                                           | <b>Loading fixture</b> to aid the nurse in mounting the Oversleeve on to a<br>colonoscope.                                                                                                                                                                                                                                                                                                        |  |  |  |
| Performance Data:                         | Performance tests were conducted for all modifications to the Pure-Vu System.<br>Specifically, the company performed the following performance tests:                                                                                                                                                                                                                                             |  |  |  |
|                                           | Biocompatibility Electrical Safety and EMC Software validation Bench tests  Bond strength Pressure test System test Steering test                                                                                                                                                                                                                                                                 |  |  |  |
|                                           | In all instances, the Pure-Vu System functioned as intended.                                                                                                                                                                                                                                                                                                                                      |  |  |  |
| Substantial<br>Equivalence<br>Discussion: | The Pure-Vu System has the same indications and similar<br>technological characteristics and principles of operation as its predicate<br>device. The minor technological differences between the Pure-Vu<br>System and its predicate devices do not raise different issues of safety<br>or effectiveness.<br>Performance data demonstrate that the Pure-Vu System is<br>substantially equivalent. |  |  |  |

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image is a logo for "PURE-VU by MOTUS GI". The logo features a circular design on the left, composed of multiple blue dots arranged in a ring-like pattern. To the right of the circular design, the text "PURE-VU" is displayed in a stylized blue font. Below the "PURE-VU" text, the words "by MOTUS GI" are written in a smaller, gray font.

| Characteristics Comparison:                    |                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                               |
|------------------------------------------------|-------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                | Modified Device                                                                           | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                              |
| Manufacturer                                   | Motus GI Medical Technologies Ltd.                                                        |                                                                                                                                                                                                                                                                                                                                                                                                               |
| Description                                    | Pure-Vu System                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                               |
| Intended Use                                   | Same                                                                                      | The Pure-Vu System is intended to connect<br>to standard colonoscopes to help facilitate<br>intra-procedural cleaning of a poorly prepared<br>colon by irrigating or cleaning the colon and<br>evacuating the irrigation fluid (water), feces<br>and other bodily fluids and matter, e.g. blood.<br>It is for use only by trained medical personnel<br>located in hospitals, clinics and doctors'<br>offices. |
| Environment of Use                             | Same                                                                                      | Hospitals, clinics and doctors' offices                                                                                                                                                                                                                                                                                                                                                                       |
| Prescriptive                                   | Same                                                                                      | Yes, only trained medical personnel                                                                                                                                                                                                                                                                                                                                                                           |
| Disposable                                     | Same                                                                                      | Single patient, single use                                                                                                                                                                                                                                                                                                                                                                                    |
| Distal tip design                              | Same                                                                                      | Multi irrigation holeTwo distal suction holes                                                                                                                                                                                                                                                                                                                                                                 |
| Principle of operation                         | Same                                                                                      | Distal attachment to an endoscope, sleeve<br>ensuring attachment along entire length,<br>suction and irrigation tubes running along the<br>endoscope, suction and irrigation head at the<br>distal tip. Enables irrigation and suction at any<br>time during the procedure without removing<br>any tools, which may be inserted in the                                                                        |
| Operational Procedures                         | Same;<br>Simplified procedure to<br>unload the colonoscope                                | Attachment to a standard colonoscopeIntra-procedure coloncleansing during<br>standard colonoscopyEvacuation of water and feces                                                                                                                                                                                                                                                                                |
| System Components                              | "Add-on" name change<br>to "Oversleeve""WS Cartridge" name<br>change to "WS<br>Connector" | Pure-Vu Add-onPure-Vu WS Cartridge                                                                                                                                                                                                                                                                                                                                                                            |
| Oversleeve outer<br>diameter                   | Same                                                                                      | 21 mm                                                                                                                                                                                                                                                                                                                                                                                                         |
| -                                              | Modified Device                                                                           | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                              |
| Irrigation & suction<br>system                 | Same                                                                                      | Irrigation: 4 nozzle x 0.7 mm<br>Suction: 2 nozzles x 12.5 $mm^2$                                                                                                                                                                                                                                                                                                                                             |
| Disposable length                              | Same                                                                                      | 167 cm attached to colonoscope                                                                                                                                                                                                                                                                                                                                                                                |
| Air / Water pressure<br>specification<br>(bar) | Up to 23 psi                                                                              | Up to 26 psi<br>Suction specifications: 0.5 Bar                                                                                                                                                                                                                                                                                                                                                               |
| Flow rate<br>(cc / min)                        | Water - Up to 645  cc/min<br>Air - Same                                                   | Water - Up to 630 cc/min<br>Air - up to 1350 cc/min                                                                                                                                                                                                                                                                                                                                                           |
| Dimensions Workstation                         | Same                                                                                      | W460 \ D 520 \ H340 mm                                                                                                                                                                                                                                                                                                                                                                                        |
| Weight Workstation                             | Same                                                                                      | 25Kg                                                                                                                                                                                                                                                                                                                                                                                                          |
| Electrical Input                               | Same                                                                                      | 100V-240V<br>50/60 Hz                                                                                                                                                                                                                                                                                                                                                                                         |
| Material                                       | Minor material modifications<br>Re-tested to ISO 10993                                    | Complies with ISO 10993                                                                                                                                                                                                                                                                                                                                                                                       |
| Sterilization                                  | Same                                                                                      | Clean, Non-sterile                                                                                                                                                                                                                                                                                                                                                                                            |
| System Components                              | Motus GI Workstation<br>Motus GI Oversleeve<br>Loading Fixture                            | Motus GI Workstation<br>Motus GI Add-on<br>Loading Apparatus<br>Unloading Apparatus                                                                                                                                                                                                                                                                                                                           |
| Safety Standards                               | Minor component<br>modifications<br>Re-tested to:<br>IEC 60601-1;<br>IEC 60601-1-2        | IEC 60601-1<br>IEC 60601-1-2                                                                                                                                                                                                                                                                                                                                                                                  |

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for "PURE-VU by MOTUS GI". The logo features a circular design made of blue dots on the left side. The text "PURE-VU" is in a stylized blue font, with the "by MOTUS GI" text in a smaller font below it.

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDF/K173392](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDF/K173392)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).