The PeerScope System is intended for diagnostic visualization of the digestive tract. The system also provides access for therapeutic interventions using standard endoscopy tools. The PeerScope System is indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon and ileocecal valve) for adult patients. The PeerScope System consists of PeerMedical camera heads, endoscopes, video system, light source and other ancillary equipment.
Device Story
PeerScope System is a flexible colonoscopy system for diagnostic visualization and therapeutic intervention in the lower digestive tract. Input consists of optical images captured by the endoscope; system includes camera heads, light source, and video processing unit. Device provides 160° standard front field of view and 300° wide field of view. Used in healthcare facilities/hospitals by physicians. Output is a video feed displayed for the clinician to guide navigation and therapeutic procedures. System facilitates clinical decision-making by providing enhanced visualization of the colon. Benefits include improved diagnostic capability via wide-angle imaging.
Clinical Evidence
Clinical validation performed at Elisha Medical Center, Israel, on 50 adult patients of diverse gender, ethnicity, and age. Procedures performed by US, European, and Israeli physicians. No adverse events observed. Bench validation conducted at three US medical centers under simulated clinical conditions to evaluate 300° field of view. All testing passed criteria.
Technological Characteristics
Flexible endoscope system; 160° standard/300° wide field of view. Components: camera heads, MCU, light source. Standards: IEC 60601-1, IEC 60601-2-18, IEC 60601-1-2, ISO 10993 (biocompatibility), ISO 8600 (endoscope specs), ASTM E 1837. Factory-adjusted brightness control. Video output: Composite and Y/C channels. Reprocessing validated per May 2011 FDA draft guidance.
Indications for Use
Indicated for diagnostic visualization and therapeutic intervention of the lower digestive tract (anus, rectum, sigmoid colon, colon, ileocecal valve) in adult patients.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
SEP
2 8 2012
## 510(k) Summary of Safety and Effectiveness (in accordance to 21 CFR 807.92)
#### Submitter Details
PeerMedical Ltd. 2 Hatochen St., PO 3161 Business and Industrial Park (North), Caesarea, 38900, Israel Tel: +972 4 6327731, Fax: +972 4 6327734 Date summary prepared: June 2012 Submission Contact: Dan Laor, 6 Sireni St., Haifa, 32972, Israel.
#### Device Details
Proprietary Name: Common Name: Classification Name Product Code: Subsequent Product Code: Committee/Panel: Device Class:
PeerScope System Colonoscope and accessories, flexible / rigid Endoscope accessories, 21 CFR 876.1500 FDF KOG Gastroenterology/Urology II
Reason for 510(k) Submission: New Device
## Identification of Legally Marketed Predicate Devices
ProtectiScope CS -K081004 - Manufactured by Stryker GI
#### Device Description
The PeerScope system consists of endoscopic Main Control Unit (MCU), and of the PeerScope CS colonoscope. The MCU controls the endoscope. As other endoscopic legally marketed systems, it includes video system, light source, and interface to other ancillary equipment. The device labeled for use in healthcare facility/hospital for endoscopy and endoscopic treatment within the lower digestive tract. The operation principles of the PeerScope System are similar to those of other legally marketed standard colonoscopy systems. The PeerScope system model provides 160° standard front field of view and a 300° wide field of view.
#### Intended use and indications for Use
The PeerScope System is intended for diagnostic visualization of the digestive tract. The system also provides access for therapeutic interventions using standard endoscopy tools. The PeerScope System is indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon and ileocecal valve) for adult patients. The PeerScope System consists of PeerMedical camera heads, endoscopes, video system, light source and other ancillary equipment.
{1}------------------------------------------------
## Technological characteristics
The technological characteristics of PeerScope system were compared to the characteristics of the predicate device and were found very similar, in terms of: endoscope geometry, illumination type, energy source, lens irrigation, air flow and air pressure. Several differences were observed and are listed below:
K1202
| Category | ProtectiScope<br>CS K081004 | PeerScope System | Impact of the differences on<br>device performance |
|------------------------------------------------------|----------------------------------------------------|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Brightness Control | Automatic or<br>Manual | Factory Adjusted | Factory Adjusted method provides<br>reliable design without<br>compromising performance. |
| Output Video<br>signals | VBS 2 channels<br>Y/C 2 channels<br>DVI 2 channels | Composite 3<br>channels<br>Y/C 3 channels | Both devices utilized industry<br>accepted standard Output Video<br>signals. |
| Foot Switch | Included | Not Included | The foot switch is used to operate<br>the ProtectiScope CS unique<br>feature which is not included in the<br>PeerScope System specifications. |
| Water Flow rate for<br>Auxiliary Jet water<br>supply | 20 cc/sec | 4.0 cc/sec | The design of the reduced jet water<br>flow has been verified & validated.<br>The results demonstrated that the jet<br>irrigation is efficient. |
| Disposable Sleeve | Included | Not Included | The disposable sleeve is a unique<br>feature which is not included in<br>the PeerScope System<br>specifications. |
Based on the results of verification, validation and performance testing, the impact of the above differences is insignificant in terms of the device safety and effectiveness for its intended use.
### Performance data
#### Bench data:
The device safety and performance were verified by tests by PeerMedical and accredited third party laboratories. List of standards was used / relied upon for testing:
IEC 60601-1: 1988 + Am. #1 (1991), #2 (1995) IEC 60601-2-18:1996 /EN 60601-2-18:1996 2nd Ed. +Am.1 (2000) IEC 60601-1-2:2007 ISO 10993:2009 Part #1 ISO 10993:2009 Part #5 ISO 10993:2010 Part #10 ISO 10993:2007 Part #12 ISO 8600-4 First edition 1997-07-01 ISO 8600-1 Second edition 2005-05-01 ISO 8600-3 First edition 1997-07-01 Am 1 ISO 8600-6 First edition 2005-03-15 ASTM E 1837- 96 (reapproved 2007)
{2}------------------------------------------------
K120289 page 3 of 3
Reprocessing validation was carried out in accordance with " Processing/Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling Draft Guidance / May 2011.
The device performance was validated by testing of a final production unit in simulated clinical conditions focusing on evaluation of the device wide 300° field of view. The procedure was performed by physicians at three US medical centers. Methods and procedures were identical to those used in clinical conditions. The results of the bench validation passed the experiment criteria. The device met its intended use and specifications. Hazardous conditions were not observed.
#### Clinical Data:
The device performance was validated by testing of a final production unit in clinical conditions. The procedure was performed at Elisha Medical Center, Haifa, Israel on a diverse adult population in terms of gender, ethnicity and age. The procedure was performed by US, European and Israeli physicians. Methods and procedures were identical to those used in USbased studies. Fifty (50) patients were enrolled in conformance with the device labeling. The results of the clinical validation passed the experiment criteria. No adverse events were observed.
The conclusions drawn from the bench and clinical tests demonstrate that the device meets its specifications, intended use and indication for use.
#### Substantial Equivalence
The above presented data demonstrate that:
- The predicate device ProtectiScope CS -K081004 is legally marketed. a.
- The PeerScope System and the predicate device have the same intended use. b.
- The PeerScope System does not raise different questions of safety and C. effectiveness.
- The data provided demonstrate equivalence and support the indications. d.
Conclusion: It is the opinion of PeerMedical that the PeerScope system is substantially equivalent to the predicate device, in terms of safety and effectiveness.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a circular seal or logo. The seal contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the circumference of the circle. In the center of the seal is a stylized image of an eagle or other bird with its wings spread, with the body and wings formed by thick, curved lines. The image is black and white.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609. Silver Spring, MD 20993-0002
2 8 2012
PeerMedical Ltd.
% Mr. Dan Laor
Quality & Regulatory Advisory
6 Sireni
HAIFA 32972
ISRAEL
Re: K120289
Trade/Device Name: PeerScope System Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FDF Dated: August 14, 2012 Received: August 15, 2012
Dear Mr. Laor:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your bootion 310(t) presidentially equivalent (for the indications ferenced above and nave colorizes ally marketed predicate devices marketed in interstate for the stated in the encrosule) to regars mannent date of the Medical Device American Cruze commerce prior to May 20, 1976, the excordance with the provisions of the Federal Food, Drug, devices that have been rochasinou in accessariou of a premarket approval application (PMA). alla Cosmetic Hot (110) market the device, subject to the general controls provisions of the Act. The I ou may, merciolo, market the device, books of the requirements for annual registration, listing of general controls provisions of the fisc labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract lability adulteration. Trease note. CDICT does inst bevice labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), If your device is classified (500 above) and in the major regulations affecting your device can be It may be subject to additional consols. Entrally - Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that I DA 3 issuation of a brander complies with other requirements of the Act
that FDA has made a determination that your device complies . You must that FDA has made a decornmance and regulations administered by other Federal agencies. You must or ally Federal statutes and regulations , including, but not limited to: registration and listing comply with an the Act 3 requirements, moradaily, strenging (reporting (reporting of medical
{4}------------------------------------------------
Page 2-
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set device-related adverse evones) (21 OFF 003), good many and 820); and if applicable, the electronic
forth in the quality systems (QS) regulation (21 CFR Part 820); and 1000 10 forth in the quality by brounds (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you desire specific actives for your actives/CDRH/CDRHOffices/ucm115809.htm for go to mp.nww.ida.gov/rtbound2123.cal Health's (CDRH's) Office of Compliance. Also, please the comer for DOY.cos and "Misbranding by reference to premarket notification" (21CFR Part note the regulations regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the I ou may obtain other general incomment one Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
K120289 510(k) Number (if known):
Device Name: PeerScope System
Indications For Use:
The PeerScope System is intended for diagnostic visualization of the digestive tract. The system also provides access for therapeutic interventions using standard endoscopy tools. The PeerScope System is indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon and ileocecal valve) for adult patients.
The PeerScope System consists of PeerMedical camera heads, endoscopes, video system, light source and other ancillary equipment.
× Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1
Page 1 of
n Sian-Off
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.