← Product Code [FDF](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDF) · K043487

# EVIS EXERA COLONOVIDEOSCOPE MODEL XCF-Q160W1L/1 (K043487)

_Aizu Olympus Co., Ltd. · FDF · Jan 28, 2005 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDF/K043487

## Device Facts

- **Applicant:** Aizu Olympus Co., Ltd.
- **Product Code:** [FDF](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDF.md)
- **Decision Date:** Jan 28, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, 3rd-Party Reviewed

## Intended Use

This instrument has been designed to be used with the Olympus Video System Center, Light Source, Documentation Equipment, Video Monitor, Electrosurgical Unit, Endo-therapy accessories such as a Biopsy Forceps and other ancillary equipment for endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum, and sigmoid colon, colon, and ileocecal valve).

## Device Story

EVIS EXERA Colonovideoscope XCF-Q160W1L/I is a flexible endoscope for lower GI tract visualization and surgery. It captures optical images of the colon and rectum, transmitting them to an external Olympus video system center, light source, and monitor for physician viewing. The device features a 170° field of view, wider than the predicate, to improve lesion detection and reduce procedure time. It includes an instrument channel for endo-therapy accessories like biopsy forceps and electrosurgical units. Used in clinical settings by trained physicians, the device allows for real-time visual inspection of the lower digestive tract, facilitating diagnosis and therapeutic intervention. The output is a live video feed, enabling the clinician to navigate the colon, identify abnormalities, and perform biopsies or surgeries. Benefits include enhanced visualization of the colonic mucosa and potential for shorter procedure durations.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on bench testing and design comparison. Biocompatibility testing was conducted per Japan's Ministry of Health and Welfare guidelines.

## Technological Characteristics

Flexible colonovideoscope; 170° field of view; 13.2 mm distal end diameter; 3.7 mm inner channel diameter. Connects to Olympus video system center, light source, and monitor. Patient-contact materials evaluated for biocompatibility per YAKKI No. 99. Non-powered, fiber-optic/electronic imaging device.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- EVIS EXERA Colonovideoscope CF-Q160AL/I ([K001241](/device/K001241.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K043487

# SMDA 510(k) SUMMARY EVIS EXERA Colonovidoscope XCF Q160W1L/I

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21CFR, Part 807, Subpart E, Section 807.92.

## 1. GENERAL INFORMATION

| Applicant                | Aizu Olympus Co., Ltd<br>500 Aza Muranishi Ooaza<br>Niidera, Monden-machi<br>Aizuwakamatsu-shi, Fukushima, JAPAN 965-8520<br>Establishment Registration No.: 9610595                                                                                                                  |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submission Correspondent | Akiko Fukagawa<br>Olympus Medical Systems Corp.<br>(Former name: Olympus Corporation)<br>2951 Ishikawa-cho,<br>Hachioji-shi, Tokyo, 192-8507, Japan<br>Phone: +81-426-2891<br>Fax: +81-426-3174<br>E-mail: akiko_fukagawa@ot.olympus.co.jp<br>Establishment Registration No.: 8010047 |
| Official Correspondent   | Laura Storms-Tyler<br>Director, Regulatory Affairs and Quality Assurance<br>Olympus America Inc.<br>Two Corporate Center Drive, Melville,<br>NY 11747-9058<br>Phone: 631-844-5688<br>Fax: 631-844-5554<br>Establishment Registration No.: 2429304                                     |

#### 2. Device Identification

| Trade Name:        | EVIS EXERA Colonovideoscope XCF-Q160W1L/I   |
|--------------------|---------------------------------------------|
| Common Name:       | Gastrointestinalvideoscope/Colonovideoscope |
| Regulation Number: | 21CFR 876.1500                              |
| Regulation Name:   | Endoscope and accessories                   |
| Class:             | II                                          |
| Product Code:      | 78 FDF                                      |

{1}------------------------------------------------

#### 3. Predicate Device

Predicate Device Name: EVIS EXERA Colonovideoscope CF-Q160AL/I Manufacturer: Olympus Corporation 510(k) Number: K001241

#### 4. Device Description

The subject devices, EVIS EXERA Colonovideoscope XCF-Q160W1L/I are identical to the predicate device, CF-Q160AL/I, in intended use. These instrument have been designed to be used with an Olympus video system center, light source, documentation equipment, video monitor, endo-therapy accessories such as biopsy forceps and ancillary equipment for endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum, sigmoid colon, colon and ileocecal valve). As for the device specifications, they are basically identical to the predicate devices, the CF-Q160AL/I, with the exception that the subject devices have wider field of views compared to the CF-Q160AL/I/. With this field of view, the XCF-Q160W1L/I may improve the detection of lesions in the colon and has been shown to reduce overall procedure time.

#### 5. Intended Use of the device

This instrument has been designed to be used with the Olympus Video System Center, Light Source, Documentation Equipment, Video Monitor, Electrosurgical Unit, Endo-therapy accessories such as a Biopsy Forceps and other ancillary equipment for endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum, and sigmoid colon, colon, and ileocecal valve).

#### 6. Comparison of Technological Characteristics

| Specifications                   | Subject Device<br>XCF-Q160W1L/I | Predicate Device<br>CF-Q160AL/I |
|----------------------------------|---------------------------------|---------------------------------|
| Field of View                    | 170°                            | 140°                            |
| Distal end Outer<br>Diameter     | $\phi$ 13.2 mm                  | $\phi$ 12.8 mm                  |
| Insertion Tube Outer<br>Diameter | $\phi$ 12.8 mm                  | $\phi$ 12.8 mm                  |
| Inner Channel Diameter           | $\phi$ 3.7 mm                   | $\phi$ 3.7 mm                   |

Below is the comparison table between the subject devices and predicate device.

{2}------------------------------------------------

### 7. Materials

Some of the patient contact materials used in the subject devices are identical to those used in the devices cleared in the past 510(k) submissions. As the others are not identical to the predicate devices, biocompatibility testing was performed in accordance with Japan's Ministry of Health and Welfare notification "GUIDELINES FOR BASIC BIOLOGICAL EVALUATION OF MEDICAL DEVICES" (issued on June 27, 1995), YAKKI No.99.

#### 8. Conclusion

When compared to the predicate device, XCF-Q160W1L/I do not incorporate any significant when compared to the progration, material, or design that could affect the safety and effectiveness. Therefore, clinical data is not necessary for its evaluation of safety and efficacy.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features an abstract design resembling an eagle or bird in flight, composed of three curved lines. Encircling the design is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The seal is presented in black and white.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 2 8 2005

OLYMPUS AMERICA INC. c/o Mr. Ned E. Devine, Jr. Responsible Third Party Official Entela, Inc. 3033 Madison Avenue, SE GRAND RAPIDS MI 49548

Re: K043487

K043487
Trade/Device Name: EVIS EXERA Colonovideoscope XCF-Q160WIL/I Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: 78 FDF Dated: January 12, 2005 Received: January 13, 2005

Dear Mr. Devine:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your occaon of the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for the stated in the enclosine) to regally manctment date of the Medical Device American of Read Drug commerce prior to May 20, 1776, are enalerations of the Federal Food, Drug, devices mat have occh reclassino in accessfire approval of a premarket approval application (PMA).
and Cosmetic Act (Act) that do not require approval of a premarket approval You may, therefore, market the device, subject to the general controls provisions of the Act. The You may, therefore, market the devices, basjon to the ments for annual registration, listing of general controls provisions of the frectiners and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (IPMA), If your device is classified (sec above) into existing major regulations affecting your device can be it may be suoject to additional controls. Entraling migres 800 to 898. In addition, FDA may loudia in the Code of I odelar is concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issuation of our device complies with other requirements of the Act that FDA has made a decermination an administered by other Federal agencies. You must of ally rederal statues and roganients including, but not limited to: registration and listing Comply with an the Act 3 requirements, as a manufacturing practice requirements as set (21 CFR Part 807), laboring (21 CFR Part 820); and 1f applicable, the electronic form in the quality systems (QD) roganitimes 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k)
The station of the support of a final four of substantial equivalence of your de This letter will allow you to begin marketing your acvice of your device of your device to a legally
premarket notification. The FDA inding of substantial equivalence of you premarket notification. The FDA finding of substantial equivatelies of your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
r many of the filler of the fillering pumbers, hased on the regulation nu If you desire specific advice for your device on our labeling regaration (e virus).
contact the Office of Compliance at one of the following numbers, based on the regulation the top of this letter:

| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology)                      | 240-276-0120 |
| Other           |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Also, please note the regulation entitled, "Misoration on your responsibilities under the Act from the 807.97). You may obtain other general Infornation on your resposistences at its toll-free number (800)
Division of Small Manufacturers, International (1) var the sounder (80 Division of Small Manufacturers, International and Consultor Institution of the comments.
638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/edrh/dsmarl

Sincerely yours,

Nancy C. Brogdon

Nancy C. È rogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

510(k) Number(if known):

K043487

Device Name: EVIS EXERA Colonovideoscope XCF-Q160W1L/l

Indications for Use:

Indications for Osc.
This instrument has been designed to be used with the Olympus Video System Center, This Instrument nation Equipment, Video Monitor, Electrosurgical Unit, Endo-Light Ooulos, Documents a Biopsy Forceps and other ancillary equipment for therapy 1 loosconos vithin the lower digestive tract (including the anus, rectum, and sigmoid colon, and ileocecal valve).

Prescription Use***_****_*
(21 CFR 801 Subpart D)

OR

Over-The-Counter Use_ (21 CFR 801 Subpart C)

(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of GDRH, Office of Device Evaluation (ODE)

David h. Severson
(Division Sign-Off)

Page 1 of __ 1

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K043487

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDF/K043487](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDF/K043487)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
