EVIS EXERA COLONOVIDEOSCOPE MODEL XCF-Q160W1L/1

K043487 · Aizu Olympus Co., Ltd. · FDF · Jan 28, 2005 · Gastroenterology, Urology

Device Facts

Record IDK043487
Device NameEVIS EXERA COLONOVIDEOSCOPE MODEL XCF-Q160W1L/1
ApplicantAizu Olympus Co., Ltd.
Product CodeFDF · Gastroenterology, Urology
Decision DateJan 28, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic, 3rd-Party Reviewed

Indications for Use

This instrument has been designed to be used with the Olympus Video System Center, Light Source, Documentation Equipment, Video Monitor, Electrosurgical Unit, Endo-therapy accessories such as a Biopsy Forceps and other ancillary equipment for endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum, and sigmoid colon, colon, and ileocecal valve).

Device Story

EVIS EXERA Colonovideoscope XCF-Q160W1L/I is a flexible endoscope for lower GI tract visualization and surgery. It captures optical images of the colon and rectum, transmitting them to an external Olympus video system center, light source, and monitor for physician viewing. The device features a 170° field of view, wider than the predicate, to improve lesion detection and reduce procedure time. It includes an instrument channel for endo-therapy accessories like biopsy forceps and electrosurgical units. Used in clinical settings by trained physicians, the device allows for real-time visual inspection of the lower digestive tract, facilitating diagnosis and therapeutic intervention. The output is a live video feed, enabling the clinician to navigate the colon, identify abnormalities, and perform biopsies or surgeries. Benefits include enhanced visualization of the colonic mucosa and potential for shorter procedure durations.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on bench testing and design comparison. Biocompatibility testing was conducted per Japan's Ministry of Health and Welfare guidelines.

Technological Characteristics

Flexible colonovideoscope; 170° field of view; 13.2 mm distal end diameter; 3.7 mm inner channel diameter. Connects to Olympus video system center, light source, and monitor. Patient-contact materials evaluated for biocompatibility per YAKKI No. 99. Non-powered, fiber-optic/electronic imaging device.

Indications for Use

Indicated for use in endoscopy and endoscopic surgery within the lower digestive tract, including the anus, rectum, sigmoid colon, colon, and ileocecal valve, in patients requiring diagnostic or therapeutic lower GI procedures.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K043487 # SMDA 510(k) SUMMARY EVIS EXERA Colonovidoscope XCF Q160W1L/I This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21CFR, Part 807, Subpart E, Section 807.92. ## 1. GENERAL INFORMATION | Applicant | Aizu Olympus Co., Ltd<br>500 Aza Muranishi Ooaza<br>Niidera, Monden-machi<br>Aizuwakamatsu-shi, Fukushima, JAPAN 965-8520<br>Establishment Registration No.: 9610595 | |--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submission Correspondent | Akiko Fukagawa<br>Olympus Medical Systems Corp.<br>(Former name: Olympus Corporation)<br>2951 Ishikawa-cho,<br>Hachioji-shi, Tokyo, 192-8507, Japan<br>Phone: +81-426-2891<br>Fax: +81-426-3174<br>E-mail: akiko_fukagawa@ot.olympus.co.jp<br>Establishment Registration No.: 8010047 | | Official Correspondent | Laura Storms-Tyler<br>Director, Regulatory Affairs and Quality Assurance<br>Olympus America Inc.<br>Two Corporate Center Drive, Melville,<br>NY 11747-9058<br>Phone: 631-844-5688<br>Fax: 631-844-5554<br>Establishment Registration No.: 2429304 | #### 2. Device Identification | Trade Name: | EVIS EXERA Colonovideoscope XCF-Q160W1L/I | |--------------------|---------------------------------------------| | Common Name: | Gastrointestinalvideoscope/Colonovideoscope | | Regulation Number: | 21CFR 876.1500 | | Regulation Name: | Endoscope and accessories | | Class: | II | | Product Code: | 78 FDF | {1}------------------------------------------------ #### 3. Predicate Device Predicate Device Name: EVIS EXERA Colonovideoscope CF-Q160AL/I Manufacturer: Olympus Corporation 510(k) Number: K001241 #### 4. Device Description The subject devices, EVIS EXERA Colonovideoscope XCF-Q160W1L/I are identical to the predicate device, CF-Q160AL/I, in intended use. These instrument have been designed to be used with an Olympus video system center, light source, documentation equipment, video monitor, endo-therapy accessories such as biopsy forceps and ancillary equipment for endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum, sigmoid colon, colon and ileocecal valve). As for the device specifications, they are basically identical to the predicate devices, the CF-Q160AL/I, with the exception that the subject devices have wider field of views compared to the CF-Q160AL/I/. With this field of view, the XCF-Q160W1L/I may improve the detection of lesions in the colon and has been shown to reduce overall procedure time. #### 5. Intended Use of the device This instrument has been designed to be used with the Olympus Video System Center, Light Source, Documentation Equipment, Video Monitor, Electrosurgical Unit, Endo-therapy accessories such as a Biopsy Forceps and other ancillary equipment for endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum, and sigmoid colon, colon, and ileocecal valve). #### 6. Comparison of Technological Characteristics | Specifications | Subject Device<br>XCF-Q160W1L/I | Predicate Device<br>CF-Q160AL/I | |----------------------------------|---------------------------------|---------------------------------| | Field of View | 170° | 140° | | Distal end Outer<br>Diameter | $\phi$ 13.2 mm | $\phi$ 12.8 mm | | Insertion Tube Outer<br>Diameter | $\phi$ 12.8 mm | $\phi$ 12.8 mm | | Inner Channel Diameter | $\phi$ 3.7 mm | $\phi$ 3.7 mm | Below is the comparison table between the subject devices and predicate device. {2}------------------------------------------------ ### 7. Materials Some of the patient contact materials used in the subject devices are identical to those used in the devices cleared in the past 510(k) submissions. As the others are not identical to the predicate devices, biocompatibility testing was performed in accordance with Japan's Ministry of Health and Welfare notification "GUIDELINES FOR BASIC BIOLOGICAL EVALUATION OF MEDICAL DEVICES" (issued on June 27, 1995), YAKKI No.99. #### 8. Conclusion When compared to the predicate device, XCF-Q160W1L/I do not incorporate any significant when compared to the progration, material, or design that could affect the safety and effectiveness. Therefore, clinical data is not necessary for its evaluation of safety and efficacy. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features an abstract design resembling an eagle or bird in flight, composed of three curved lines. Encircling the design is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The seal is presented in black and white. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JAN 2 8 2005 OLYMPUS AMERICA INC. c/o Mr. Ned E. Devine, Jr. Responsible Third Party Official Entela, Inc. 3033 Madison Avenue, SE GRAND RAPIDS MI 49548 Re: K043487 K043487 Trade/Device Name: EVIS EXERA Colonovideoscope XCF-Q160WIL/I Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: 78 FDF Dated: January 12, 2005 Received: January 13, 2005 Dear Mr. Devine: We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your occaon of the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for the stated in the enclosine) to regally manctment date of the Medical Device American of Read Drug commerce prior to May 20, 1776, are enalerations of the Federal Food, Drug, devices mat have occh reclassino in accessfire approval of a premarket approval application (PMA). and Cosmetic Act (Act) that do not require approval of a premarket approval You may, therefore, market the device, subject to the general controls provisions of the Act. The You may, therefore, market the devices, basjon to the ments for annual registration, listing of general controls provisions of the frectiners and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (IPMA), If your device is classified (sec above) into existing major regulations affecting your device can be it may be suoject to additional controls. Entraling migres 800 to 898. In addition, FDA may loudia in the Code of I odelar is concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issuation of our device complies with other requirements of the Act that FDA has made a decermination an administered by other Federal agencies. You must of ally rederal statues and roganients including, but not limited to: registration and listing Comply with an the Act 3 requirements, as a manufacturing practice requirements as set (21 CFR Part 807), laboring (21 CFR Part 820); and 1f applicable, the electronic form in the quality systems (QD) roganitimes 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) The station of the support of a final four of substantial equivalence of your de This letter will allow you to begin marketing your acvice of your device of your device to a legally premarket notification. The FDA inding of substantial equivalence of you premarket notification. The FDA finding of substantial equivatelies of your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please r many of the filler of the fillering pumbers, hased on the regulation nu If you desire specific advice for your device on our labeling regaration (e virus). contact the Office of Compliance at one of the following numbers, based on the regulation the top of this letter: | 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |-----------------|----------------------------------|--------------| | 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 892.xxxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Also, please note the regulation entitled, "Misoration on your responsibilities under the Act from the 807.97). You may obtain other general Infornation on your resposistences at its toll-free number (800) Division of Small Manufacturers, International (1) var the sounder (80 Division of Small Manufacturers, International and Consultor Institution of the comments. 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/edrh/dsmarl Sincerely yours, Nancy C. Brogdon Nancy C. È rogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use 510(k) Number(if known): K043487 Device Name: EVIS EXERA Colonovideoscope XCF-Q160W1L/l Indications for Use: Indications for Osc. This instrument has been designed to be used with the Olympus Video System Center, This Instrument nation Equipment, Video Monitor, Electrosurgical Unit, Endo-Light Ooulos, Documents a Biopsy Forceps and other ancillary equipment for therapy 1 loosconos vithin the lower digestive tract (including the anus, rectum, and sigmoid colon, and ileocecal valve). Prescription Use***_****_* (21 CFR 801 Subpart D) OR Over-The-Counter Use_ (21 CFR 801 Subpart C) (Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of GDRH, Office of Device Evaluation (ODE) David h. Severson (Division Sign-Off) Page 1 of __ 1 (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number K043487
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...