Ancora-SB

K231323 · Aspero Medical, Inc. · FDA · Aug 31, 2023 · Gastroenterology, Urology

Device Facts

Record IDK231323
Device NameAncora-SB
ApplicantAspero Medical, Inc.
Product CodeFDA · Gastroenterology, Urology
Decision DateAug 31, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The Ancora-SB is an accessory to an endoscope. The Ancora-SB is intended for use with any standard endoscope that has an outer diameter of 9.0 - 9.4 mm and a working length of 1680 mm or greater. The Ancora-SB is intended for use with any standard flexible endoscopy balloon inflation unit with a set pressure of 5.4 kPa (+2.6 kPa, -1.8 kPa). The device is indicated to ensure complete positioning of an endoscope in the small intestine by either oral or anal insertion, and assist with optical visualization, diagnosis, and endoscopic treatment.

Device Story

Ancora-SB is a single-use, close-fitting sleeve accessory for standard endoscopes (9.0–9.4 mm diameter). Used in clinical settings by physicians during small bowel enteroscopy. Device features a proprietary textured silicone balloon at the distal end; balloon inflates via standard flexible endoscopy balloon inflation units to 5.4 kPa. Textured surface improves stability and reduces mucosal slippage compared to smooth-balloon predicates, facilitating endoscope advancement. Device includes air supply tube assembly and Luer adaptor. Physician operates endoscope and inflation unit; output is improved endoscopic access and visualization of the small intestine. Benefits include increased procedure time efficiency and enhanced diagnostic/therapeutic capability.

Clinical Evidence

Bench testing included biocompatibility (ISO 10993-1, -5, -10, -23), distribution (ASTM D4169-22), shelf-life (ASTM F1980-21), and performance testing (dimensions, balloon reliability, friction force, anchor force, joint strength). A GLP animal study was also conducted. No human clinical trial data reported.

Technological Characteristics

Single-use hydrophilic-coated silicone sleeve with distal radiopaque material. Features a textured balloon (38 mm OD) for mucosal anchoring. Compatible with standard endoscopes (9.0-9.4 mm OD) and standard balloon inflation units (5.4 kPa). Non-sterile. Mechanical principle of operation via balloon inflation for endoscope stabilization.

Indications for Use

Indicated for patients undergoing small intestine endoscopy via oral or anal insertion to assist with positioning, optical visualization, diagnosis, and endoscopic treatment.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. August 31, 2023 Aspero Medical, Inc. % Pierre Bounaud Principal Consultant Ram+ 2251 San Diego Ave, Suite B-257 San Diego, California 92110 Re: K231323 Trade/Device Name: Ancora-SB Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: FDA, FED Dated: July 31, 2023 Received: July 31, 2023 Dear Pierre Bounaud: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {2}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Shanil P. Haugen -S Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. Submission Number (if known) K231323 Device Name Ancora-SB Indications for Use (Describe) The Ancora-SB is an accessory to an endoscope. The Ancora-SB is intended for use with any standard endoscope that has an outer diameter of 9.0 - 9.4 mm and a working length of 1680 mm or greater. The Ancora-SB is intended for use with any standard flexible endoscopy balloon inflation unit with a set pressure of 5.4 kPa (+2.6 kPa, -1.8 kPa). The device is indicated to ensure complete positioning of an endoscope in the small intestine by either oral or anal insertion, and assist with optical visualization, diagnosis, and endoscopic treatment. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Aspero Medical. The logo includes a graphic of three vertical bars of varying heights, followed by the word "aspero" in a bold, sans-serif font. Below the word "aspero" is the word "MEDICAL" in a smaller, blue, sans-serif font. The text "Page 1 of 5" and "K231323" are in the upper right corner of the image. # 510(k) Summary #### DATE PREPARED July 31, 2023 #### MANUFACTURER AND 510(k) OWNER Aspero Medical, Inc. 320 E. Vine Drive, Suite 101, Fort Collins, CO 80524, USA Telephone: Official Contact: Dr. Mark Rentschler, CEO #### REPRESENTATIVE/CONSULTANT Pierre Bounaud, Ph.D. Allison C. Komiyama, Ph.D., R.A.C. RQM+ Telephone: +1 (412) 816-8139 Email: pbounaud@rqmplus.com akomiyama@rqmplus.com #### DEVICE INFORMATION | Proprietary Name/Trade Name: | Ancora-SB | |------------------------------|------------------------------------------------------------------------------| | Regulation Name: | Endoscope and accessories | | Regulation Number: | 21 CFR 876.1500 | | Class: | II | | Product Codes: | FDA, FED | | Premarket Review: | OPEQ/OHT3/Renal, Gastrointestinal, Obesity and<br>Transplant Devices (DHT3A) | | Review Panel: | Gastroenterology/Urology | #### PREDICATE DEVICE IDENTIFICATION The Ancora-SB is substantially equivalent to the following predicates: | 510(k) Number | Predicate Device Name / Manufacturer | Primary<br>Predicate | Reference<br>Device | |---------------|------------------------------------------------------------------------------------------------------------|----------------------|---------------------| | K071254 | Splinting Tube ST-SB1 (accessory to Small Intestinal<br>Videoscope System) / Olympus Medical Systems Corp. | ✓ | | | K221452 | DiLumen C1, EZ1 and Tool Mount / Lumendi, LLC | | ✓ | The predicate devices have not been subject to a design related recall. #### DEVICE DESCRIPTION The Ancora-SB is a single-use, close-fitting sleeve that slides freely over a currently marketed standard endoscope having an outer diameter of 9.0 – 9.4 mm and a working length of 1680 {5}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Aspero Medical. The logo has a series of vertical bars of increasing height to the left of the word "aspero" in bold black letters. Below the word "aspero" is the word "MEDICAL" in blue. At the top of the image are the words "Page 2 of 5" and the code "K231323". mm or greater. The device is indicated to ensure complete positioning of an endoscope in the small intestine, by either oral or anal insertion, and assist with optical visualization, diagnosis, and endoscopic treatment. The device is designed to be used with a Balloon Endoscopy System (BES) which includes a multi-channel endoscope, an endoscopy tower (light source, display monitor, recording equipment, etc.) and a balloon inflation unit (equipment that inflates or deflates by a push button control, inflation is to a stated maximum pressure and the pressure is maintained during active use when inflated). ## INDICATIONS FOR USE The Ancora-SB is an accessory to an endoscope. The Ancora-SB is intended for use with any standard endoscope that has an outer diameter of 9.0 - 9.4 mm and a working length of 1680 mm or greater. The Ancora-SB is intended for use with any standard flexible endoscopy balloon inflation unit with a set pressure of 5.4 kPa (+2.6 kPa, -1.8 kPa). The device is indicated to ensure complete positioning of an endoscope in the small intestine by either oral or anal insertion, and assist with optical visualization, diagnosis, and endoscopic treatment. ## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS Aspero Medical, Inc. believes that the Ancora-SB is substantially equivalent to the predicate device based on the information summarized here: The subject device has the same intended use, i.e., it is a single use tube intended to ensure complete positioning of an endoscope in the small intestine, by either oral or anal insertion, and assist with optical visualization, diagnosis, and endoscopic treatment, as the splinting tube ST-SB1 cleared as an accessory in K071254. The subject device has similar technological characteristics (same operating principle, similar design and dimensions, similar materials). as the splinting tube ST-SB1 cleared as an accessory in K071254. Differences in technological characteristics includes the following: - The silicone balloon of the subject device features a proprietary textured surface . designed to provide stability and advancement of the endoscope. Smooth, round balloons currently on the market, such as the one on the splinting tube ST-SB1, are prone to slippage against the mucosa, reducing the time efficiency of small bowel enteroscopy procedures. - . The Ancora-SB is provided non-sterile. This is similar to the reference device cleared in K221452. These technological characteristics have undergone testing to ensure the device is as safe and effective as the predicates. Table below provides a summary of the technological characteristics of the subject device compared to the predicate and reference devices. {6}------------------------------------------------ Image /page/5/Picture/1 description: The image contains the logo for Aspero Medical. The logo consists of a series of vertical bars of varying heights and shades of gray, followed by the word "aspero" in a large, bold, black font. Below "aspero" is the word "MEDICAL" in a smaller, blue font. | | Subject Device | Primary Predicate Device | Reference Device | |----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Ancora-SB | Splinting Tube ST-SB1 | DiLumen C1, EZ1 and Tool | | | | (Accessory to Small Intestinal | Mount | | | | Videoscope System) | | | | | K071254 | K221452 | | Product<br>Codes /<br>Regulation<br>Number | FDA / 21 CFR 876.1500<br>FED / 21 CFR 876.1500 | Same | FDF / 21 CFR 876.1500 | | Intended<br>Use | Single use tube intended to<br>ensure complete positioning of<br>an endoscope in the small<br>intestine, by either oral or anal<br>insertion, and assist with<br>optical visualization, diagnosis,<br>and endoscopic treatment. | Same | Single use tube intended to<br>ensure complete positioning of<br>an endoscope in the large<br>intestine, and assist with<br>optical visualization, diagnosis,<br>and endoscopic treatment. | | Device<br>Component<br>s | Single Use Splinting Tube Air supply tube assembly Luer adaptor | Small Intestinal Videoscope Single Use Splinting Tube<br>(ST-SB1) Balloon Control Unit<br>(OBCU) Accessories | Sleeve Manual inflation bulb | | Balloon<br>Outer<br>Diameter | 38 mm | 40 mm | 60 mm | | Balloon<br>Injection<br>Volume | 43 mL | Same¹ | Unknown | | Insertion<br>Tube<br>Maximum<br>Insertion<br>Width | 15.9 mm | Same | Unknown | | Insertion<br>Tube Inner<br>Diameter | 11 mm | Same | Unknown | | Insertion<br>Tube<br>Working<br>Length | 1,320 mm | Same | 1,300 mm | | Insertion<br>Tube Total<br>Length | 1,400 mm | Same | Unknown | | Air Flow<br>Tube<br>Length | 3,700 mm | Same | Unknown | | Materials | Hydrophilic-coated silicone<br>Radiopaque material on distal | Same | Hydrophilic-coated extruded<br>polyurethane | | | Subject Device | Primary Predicate Device | Reference Device | | | Ancora-SB | Splinting Tube ST-SB1<br>(Accessory to Small Intestinal<br>Videoscope System) | DiLumen C1, EZ1 and Tool<br>Mount | | | | K071254 | K221452 | | | end | | Low durometer polyurethane<br>(balloon)<br>Radiopaque material on distal<br>end | | Balloon | Single textured balloon on<br>distal end of splinting tube | Single smooth balloon on distal<br>end of splinting tube | Single smooth balloon on distal<br>end of splinting tube | | Balloon<br>Inflation<br>Methods | Standard flexible endoscopy<br>balloon inflation unit with a set<br>pressure of 5.4 (+2.6/-1.8) kPa. | Olympus OBCU delivering set<br>inflation pressure of 5.4 (+2.6/- 0) kPa | Integrated manual inflation<br>bulb | | Proximal<br>End<br>Connections | • Endoscope insertion port<br>• Fluid/flush port<br>• Air/inflation port | Same | Same | | Sterilization<br>Method | Non-sterile | EO | Same | | Compatible<br>Endoscopes | Any standard endoscope with a<br>distal tip outer diameter of 9.0<br>- 9.4 mm and a working length<br>of 1680 mm or greater | Olympus SIF Type Q180, Q260,<br>each having outer diameter of<br>9.2 mm and working length of<br>2000 mm. | Any standard endoscope with a<br>distal tip outer diameter of<br>12.5 – 14.3 mm and a working<br>length of 1680 mm or greater | {7}------------------------------------------------ ^Volume of air needed to reach the outside diameter specification (40 mm) of the balloon in a commercial splinting tube ST-SB1, as measured by Aspero Medical. ## SUMMARY OF NON-CLINICAL TESTING The following tests were performed to demonstrate safety based on current industry standards: - Biocompatibility testing per ISO 10993-1, -5, -10, and -23 ● - Distribution testing per ASTM D4169-22 - Shelf-life testing per ASTM F1980-21 - Bench performance testing including dimension inspection, packaging inspection, . components and features inspection, balloon compatibility and reliability, overtube friction force and reliability, balloon inflation time, balloon anchor force, and joint strength - GLP animal study . The results of these tests indicate that the Ancora-SB is substantially equivalent to the predicate device. #### CONCLUSION Based on the testing performed, it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate device. The similar indications for {8}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for Aspero Medical. The logo has the word "aspero" in bold black letters, with the word "MEDICAL" in blue letters below it. To the left of the word "aspero" are three vertical bars of varying heights. The text "Page 5 of 5" and "K231323" are in the upper right corner of the image. use, technological characteristics, and performance characteristics for the proposed Ancora-SB are assessed to be substantially equivalent to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...