FUJINON DOUBLE BALLOON ENTEROSCOPY SYSTEM

K040048 · Fujinon, Inc. · FDA · Jun 7, 2004 · Gastroenterology, Urology

Device Facts

Record IDK040048
Device NameFUJINON DOUBLE BALLOON ENTEROSCOPY SYSTEM
ApplicantFujinon, Inc.
Product CodeFDA · Gastroenterology, Urology
Decision DateJun 7, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The device is intended for the Optical visualization of the upper gastrointestinal tract. This includes the esophagus, stomach, duodenum, and small bowel. It is intended for the observation, diagnosis, and endoscopic treatment.

Device Story

Fujinon Double Balloon Enteroscopy System facilitates visualization and treatment of the upper GI tract, specifically the small intestine. System components include a G5 endoscope (EN-450P5/20), over-tube (TS-12140), balloons (BS-1), and a balloon pump controller (PB-10). The device operates by using balloons and an over-tube to ensure complete positioning of the enteroscope within the small intestine, allowing the tip to reach the area of diagnosis. It connects to a Series 400 processor and various peripherals (light source, monitor, video printer, electrosurgical instruments). Used in clinical settings by physicians, the system provides real-time optical visualization to the clinician via a monitor. This enables direct observation of the GI tract, facilitating diagnostic procedures and therapeutic interventions. The balloon-assisted mechanism allows for deeper access into the small intestine compared to standard enteroscopy, potentially improving diagnostic yield and therapeutic access for patients with small bowel pathology.

Clinical Evidence

Bench testing only. The device underwent electrical safety, thermal, EMC, and biocompatibility testing. No clinical data was provided.

Technological Characteristics

Endoscope: Interline color CCD chip, 411,988 pixels, 120-degree FOV, 8.5mm diameter. Materials: Natural rubber (balloon), polyvinylpyrrolidone hydromer coating (overtube). Energy: 120V/230V power supply for pump. Connectivity: Compatible with Fujinon Series 400 processor. Sterilization: Compatible with standard endoscopic reprocessing methods. G5 upgrade features external/removable check valves and updated cleaning adaptors (CA-500).

Indications for Use

Indicated for optical visualization, observation, diagnosis, and endoscopic treatment of the upper gastrointestinal tract, including the esophagus, stomach, duodenum, and small bowel in patients requiring GI endoscopy.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K 040048 pg ' of 4 # 510 (k) Summary JUN - 7 2004 Date Prepared [21 CFR 807.92(a)(1)] January 9, 2004 # Submitter's Information [21 CFR 807.92(a)(1)] Joseph M. Azary C/o Fujinon Inc. 543 Long Hill Avenue Shelton, CT. 06484 Azary Technologies has received authorization to submit this 510(k) on behalf of the sponsor Fujinon Inc. 10 High Point Drive, Wayne, NJ 07470, Establishment Registration# 2431293. # Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)] The device trade names are: Fujinon Double Balloon Enteroscopy System Common Name: Enteroscope Classification: Class II, 21 CFR 876.1500, FDA #### Predicate Device [21 CFR 807.92(a)(3)] - Fujinon Enteroscope EN-410WM K993704 ● The subject device have the same indications for use, material composition, viewing direction, image size, bending, reprocessing/sterilization method, and working dimensions as the predicate. The subject device uses the same processor and peripherals as the predicate device. The main differences between the subject device and predicate device are as follows: - Minor differences with observation range, field of view, diameter, and length. . - Subject device includes an overtube, balloons, and a balloon pump. ● - The subject device includes the G5 upgrade, which is characterized by the following minor differ-. ences: - The L-Port has been eliminated. The L-port functioned as a lens wash port . Doctors had o the option to take a syringe to inject a fluid to use it as a high pressure wash for the lens. This function was eliminated because demand was low and it was rarely used by the surgeons. - The J-Port was repositioned. The J-Port is used as a jost water wash port. The J-Port was 0 repositioned based on doctor feedback. The port was moved from the bottom part to the top (end) of the scope. There was also a desire to eligainate check valves to facilitate reprocessing and cleaning, as well as prevent clogging. {1}------------------------------------------------ K040.48 pg 2 of 4 - A G5 forceps inlet port was modified. The new port is smaller and comes with a rubber O cap. The smaller port and rubber cap help increase suction and reduce leakage. - The jet wash line check valve was removed. Internal check valves were removed to elimio nate the potential for clogging and to facilitate cleaning, disinfection, and sterilization. The valves are now external and removable. - The suction and air/water cylinders and valves were upgraded. They were updated to ac-0 commodate the new valves. The function of the valves is the same. - Addition of the FOV, which is the rubber forceps inlet valve cover. This helps create a wa-O tertight seal when the endoscope is used. - Upgrade to CA-500 cleaning adaptor. The cleaning adaptor allows the scope to be con-0 nected to tubes for cleaning. #### Description of the Device [21 CFR 807.92(a)(4)} The small intestine is one of the most difficult organs to access in gastro-intestinal (GI) tract endoscopy. The Fujinon double balloon enteroscopy system facilitates diagnosis and treatment of the upper GI tract including the small intestine. The Fujinon double balloon enteroscopy system utilized balloons and over-tube to ensure complete positioning of the enteroscope in the small intestine. The tip of the scope can be smoothly inserted to reach the area of diagnosis. The Fujinon double balloon enteroscopy system includes the following: - G5 endoscope / enteroscope (EN-450P5/20) . - Over-Tube (TS-12140) . - Balloon (BS-1) . - Balloon Pump Controller (PB-10) . The subject device is used with the Fujinon Series 400 processor, the same processor used with the predicate device. The Fujinon double balloon enteroscopy system is used in conjunction with other peripherals specified in the Operation Manual such as: - Light Source . - . Processor - Cart . - Data Keyboard - Foot Switch . - Monitor . - Video Printer . - Camera and Hard Copy Unit . - VCR . - ElectroSurgical Tastruments . {2}------------------------------------------------ \$\qquad K^{v}4\quad 0048\quad B_{y}\quad ^{3}g\quad 4\$ # Specifications Chart: | | EN-450P5/20 | | |---------------------------|--------------|--| | Viewing Direction | Forward | | | Observation Range | 5~100mm | | | Field of view | 120 | | | Image size | 130% (super) | | | Distal end diameter | 8.5mm | | | Flexible portion diameter | 8.5mm | | | Bending capability | See below | | | Up | 180 | | | Down | 180 | | | Left | 160 | | | Right | 160 | | | Forceps channel diameter | 2.2mm | | | Working length | 2000mm | | | Total length | 2300mm | | | Lens diameter | 2.1mm | |----------------------------|----------------------| | # of pixels | 411,988 pixels | | Pixels per square mm | 83, 892 | | Pixel size (H x V) | .0032 x .003725 | | Active area of CCD (H x V) | 2.458 x 1.840 | | Type of CCD chip | Interline color chip | | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>.<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>-----------------------------------------------------------------------------------------------------------<br>NAL R BLOGARIA<br>Door | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>Book Book Child Child Child Children Children Children Children Children Children Children Children Children Children Children Children Children Children Children Children Ch<br>A-4-4-4-4-4<br>Section of the country of the closed on the controlled to the contribution of the comments of the comments of the comments of the comments of the comments of the comments of<br>A more. I cost and consisted com accession of<br>FEE A<br>REER CRIMINATION COLUMN AN<br><br><br><br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br><br>-----------------------------------<br>Access the contributed to the contributed to the contributed to the contributed to the contributed to the contributed to the contributed to the contributed to the contributed<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>and the collection of the desires the first the first the first the first the<br>Acres of the Council A children and consider and any and<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>Advertising the<br>.<br><br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | | |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--| | SELL CLEAN I LAND OR BALL BOOK OF<br>and the most of the property of the continued any and changes and the considered and the contribution of the concesses<br>CAR COLOR COLLECT OF COLLECTION<br>Company Controller Comments of<br>Company of the call of the collection | | | | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>Company of Canada Cases of Children Company Company Company Company Comments of Children Comments of Children<br>.<br>.<br>AND LI & A R ANNUA MOND SHOULD STATUS CONSULER | | | | <br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br><br><br>AND LA BREAD LE LANDRED<br>-------------------------------------------------------------------------------------------------------------- | <br>FIR IN IN MINNERALING A<br><br>.<br>AN LAND CANDREAL CONNECT COLLEGION<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>A has and a comments a che all can and a comments of the first and<br>Company of Concession Company Company Company Company Company Company Company Company Company Company Company Company Company Company Company Company Company Company Company | | # Over-Tube Specifications | .<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>1<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>Specificant program and a che consisted and a set a superior contract<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>.<br><br>ARE E ME RE LEBERT<br>----------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | LE MILE II DAME<br>AND AND A COLLEGION FOR<br>- Status Support For Children Comments a<br>.<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>.<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>Company of the county of the comments of the promote on<br><br><br> | |----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | CO CALL COLLECTION<br>1<br>Carl Property Concession Come<br>Can Beach and Child A company a<br>.<br>.<br>.<br><br>CON EL LAND - NA AN AND AND AND AND | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br><br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br><br>.<br>.<br><br><br>. | Balloon Pump Controller with Remote Switch PB-10 Specifications | Set Pressure of Balloon | 3.6 kpa - 7.6 kpa (5.6 kpa +/- 2 kpa) | |---------------------------|---------------------------------------| | Maximum flow rate of pump | 170 ml/10 sec | | Power Supply | 120V / 230V | | Dimensions | 300 (W) x 200(H) x 300(D) mm | | Weight | 8.7 kg | The balloon is composed of natural rubber, which has been subjected to and passed biocompatibility I his banding vertube is composed of polyvinylpyrolidone hydromer coating, which has been subjected to and passed biocompatibility testing. > Fujinon Inc. Double Balloon Enteroscopy 510(k) Page 5-3 {3}------------------------------------------------ K cttco48 Pg 4 of 4 ### Intended Use [21 CFR 807.92(a)(5)] The device is intended for the Optical visualization of the upper gastrointestinal tract. This includes the esophagus, stomach, duodenum, and small bowel. It is intended for the observation, diagnosis, and endoscopic treatment. # Technological Characteristics [21 CFR 807.92(a)(6)] Fujinon, Inc. believes that the subject device is substantially equivalent to the predicate device. The subject device have the same indications for use, material composition, image size, bending, reprocessing/sterilization method, and working dimensions as the predicate. The subject device uses the same processor and peripherals as the predicate device. The main technological difference is the balloon and overtube as a way to facilitate diagnosis and treatment of the upper GI tract including the small intestine, as well as the minor changes associated with the G5 upgrade. #### Performance Data [21 CFR 807.92(b)(1)} The subject device has been subjected to and passed electrical safety, thermal, and EMC testing requirements. The materials in the endoscope are identical to the materials used in the predicate device. The balloon and overtube materials have been subjected to and passed biocompatibility testing. ### Conclusion [21 CFR 807.92(b)(3)] We believe the changes are minor and conclude that the subject devices are as safe and effective as the predicate devices. > Fujinon Inc. Double Balloon Enteroscopy 510(k) Page 5-4 {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle with three horizontal lines above it, representing the department's commitment to health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle symbol. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN - 7 2004 Fujinon, Inc. c/o Mr. Joseph M. Azary Azary Technologies LLC 543 Long Hill Avenue SHELTON CT 06484 Re: K040048 Trade/Device Náme: Fujinon Double Balloon Enteroscopy System Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Product Code: 78 FDA Regulation Number: 21 CFR §876.5980 Regulation Name: Gastrointestinal tube and accessories Product Code: 78 KNT Regulatory Class: II Dated: May 5, 2004 Received: May 7, 2004 Dear Mr. Azary: We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becamined the device is substantially equivalent (for the indications ferenced above and nave acterimes ally marketed predicate devices marketed in interstate for use slated in the enolosure) is regars and ment date of the Medical Device Amendments, or to commerce prior to way 20, 1976, the exaccidance with the provisions of the Federal Food, Drug, devices that have occh receired in asse equire approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, therefore, mance are as not include requirements for annual registration, listing of general controls provisions of the 1100 ling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), II your device is classified (see above) into entire regulations affecting your device can be it may be subject to additional controller Life 21, Parts 800 to 898. In addition, FDA may round in the Code of Possients concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean i Tease oc advised that FDF i robation a vour device complies with other requirements of the Act that I DA has made a decommandin an administered by other Federal agencies. You must of any I ederal Statutes and regulations, but not limited to: registration and listing comply with an the rice I requirements, in 201); good manufacturing practice requirements as set (21 CFR Part 807), labemg (21 CFR Part 820); and if applicable, the electronic forth in the quality systems (QS) . og (sections 531-542 of the Act); 21 CFR 1000-1050. {5}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will allow you to begin marteenig your ential equivalence of your device to a legally premarket notification. The PDA inding of backantal epical on thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice tor your device on one of the following numbers, based on the regulation number at the top of the letter: | 8xx.1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Additionally, 101 questions on the promote of the regulation entitled, "Misbranding Office of Compliance at (301) 594-4639. Othice of Compliance at (501) 594-4057. Fab.) production of Super Bear the Division of Same by reference to premarket notification (21 OF Act may be obtained from the Division of Small information on your responsibilities under the Act may be sitted from the 1800) 6 information on your responsionalities unec tales tales tales toll-free number (800) 638-2041 or Manufacturers, International and Consumer Assistance at its toll-fremation htt Manufacturers, International and Colladmor i assistement of the main.html. Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ 5 10(k) Number (if known): // 04 00 48 Device Name: Fujinon Inc. Double Balloon Enteroscopy System Indications For Use: The device is intended for the Optical visualization of the upper gastrointestinal tract. This includes the esophagus, stomach, duodenum, and small bowel. It is intended for the observation, diagnosis, and endoscopic treatment. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH Office of Device Evaluation (ODE) OR Prescription Use (Per 21 CFR 801.109) Allin (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number. Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (Optional Format 1-2-96) Fujinon Inc. Double Balloon Enteroscopy 510(k) Page 4
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...