This instrument has been designed to be used with the OLYMPUS EVIS Video System Center, EVIS Universal Light Source, Documentation Equipment, Video Monitor, Endo-Therapy Accessories (such as a Forceps) and other ancillary equipment for endoscopy and endoscopic surgery within the small intestine.
Device Story
Small intestinal videoscope system; includes XSIF-1TQ140A scope, XB01-681-26 stylet, and ST-S1 splinting tube. Scope uses CCD imaging sensor to capture visual data of small intestine; transmits video to external EVIS Video System Center and monitor for physician visualization. Variable stiffness mechanism allows manual adjustment of insertion tube flexibility to facilitate navigation through gastrointestinal tract. Stylet increases insertion part stiffness; splinting tube straightens path from esophagus to duodenum. Used in clinical settings by physicians for diagnostic and therapeutic endoscopic procedures. Output provides real-time visualization for clinical decision-making, enabling diagnosis and surgical intervention within small intestine.
Clinical Evidence
No clinical data; bench testing only. Device evaluated for safety and effectiveness via comparison to predicate devices and compliance with voluntary safety standards (IEC 60601 series).
Technological Characteristics
CCD-based imaging system; variable stiffness mechanical insertion tube; patient-contacting materials identical to previously cleared devices (ISO 10993-1 compliant). Connectivity via proprietary Olympus EVIS system interface. Complies with IEC 60601-1, IEC 60601-1-1, and IEC 60601-2-18.
Indications for Use
Indicated for endoscopy and endoscopic surgery within the small intestine.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
Predicate Devices
SIF-B Small Intestinal Fiberscope (Preamendment Device)
SIF-SW Sonde-type Small Intestinal Fiberscope (K904800)
PR-2B (Cannula with Stylet) for EVIS-200 System (K931154)
ST-C6/C8 Splinting Tube for EVIS-140 Series Scopes (K954451)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the bird figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
OLYMPUS OPT-ELECTRONICS CO., LTD. % Ms. Tina Steffanie-Oak Senior R.A. Analyst Olympus America, Inc. Two Corporate Center Drive Melville, NY 11747-3157
JUL 2 7 2015
Re: K031256
Trade/Device Name: XSIF-1TQ140A Small Intestinal Videoscope XB01-681-26 Stylet ST-S1 Splinting Tube
Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FDA Dated (Date on orig SE ltr): April 18, 2003 Received (Date on orig SE Itr): April 24, 2003
Dear Ms. Steffanie-Oak,
This letter corrects our substantially equivalent letter of July 23, 2003.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{1}------------------------------------------------
Page 2 -
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm_for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# 510(k) Number(if known): XSIE-1TQ140A SMALL INTESTINAL VIDEOSCOPE Device Name: XBO1-681-26 STYLET ST-S1 SPLINTING TUBE
Indications for Use:
#### XSIF-1TQ140A SMALL INTESTINAL VIDEOSCOPE
This instrument has been designed to be used with the OLYMPUS EVIS Video System Center, EVIS Universal Light Source, Documentation Equipment, Video Monitor, Endo-Therapy Accessories (such as a Forceps) and other ancillary equipment for endoscopy and endoscopic surgery within the small intestine.
#### XBO1-681-26 STYLET
This product has been designed to increase stiffness of insertion part of Olympus small intestinal videoscope XSIF-1TQ140A by inserting to the endoscope biopsy channel.
#### ST-S1 SPLINTING TUBE
This product has been designed to be used with the XSIF-1TQ140A for endoscopy within the small intestinal tract. This product is intended to facilitate insertion of the scope by straightening from the esophagus through stomach to descending part of duodenum.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | <div style="display:inline-block;"> <span style="text-decoration: overline;">✓</span> </div> |
|----------------------|----------------------------------------------------------------------------------------------|
| (Per 21 CFR 801.109) | |
OR
| (Division Sign-Off) | |
|---------------------------------------------------------------|---------|
| Division of Reproductive, Abdominal, and Radiological Devices | |
| 510(k) Number | K031256 |
| Over-The-Counter Use | |
|--------------------------|--|
| (Optional Format 1-2-86) | |
{3}------------------------------------------------
# SMDA 510(k) SUMMARY
K03/256
pg/183
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21CFR, Part 807, Subpart E, Section 807.92.
# A. GENERAL INFORMATION
| 1. Applicant: | OLYMPUS OPT-ELECTRONICS CO., LTD. |
|---------------------------------|-----------------------------------------------------------------------------------------------|
| | (New Company Name: AIZU OLYMPUS CO., LTD) |
| Address: | 500 Aza Muranishi Ooaza Niidera, Monden-machi<br>Aizuwakamatsu-shi Fukushima, Japan, 965-8520 |
| Establishment Registration No.: | 9610595 |
| | |
| 2. Submission Correspondent: | Tina Steffanie-Oak |
| Title: | Senior R.A. Analyst |
| | OLYMPUS AMERICA INC. |
| Address: | Two Corporate Center Drive, Melville, NY 11747-3157 |
| Telephone: | 631-844-5477 |
| Facsimile: | 631-844-5416 |
| E-mail address: | Tina.Steffanie-Oak@olympus.com |
| Establishment Registration No.: | 2429304 |
#### 3. Initial Impoter: Olympus America Inc. Two Corporate Center Drive, Melville, NY 11747-3157 Address: Establishment Registration No .: 2429304
### B. DEVICE IDENTIFICATION
- 1. Common/Usual Name SMALL INTESTINAL VIDEOSCOPE
- 2. Device Name SIF-1TQ140A SMALL INTESTINAL VIDEOSCOPE
## 3. Classification Name
| ---------------------<br>I a suall in a g a g a mich armines a<br>Jumbe | The first and the contract of the country of the country of the college of the<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br> | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------|--|
| 1 - 1 - 1 - 1 -<br>.<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | LELL HILL HOND RESERVED FOLL & ARREN COLMANIAL BAL<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>- 10 -<br>ADSCOLE<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br><br>The first and the many of the many of the management of consisted and program and the program the first and | o vini<br>Company of Children<br>1 | |
{4}------------------------------------------------
K 031,256
pg 2 of 3
# C. IDENTIFICATION OF LEGALLY MARKETED DEVICES WHICH WE CLAIM SUBSTANTIAL EQUIVALENCE
The following listed devices are considered as predicate devices in consideration of their characteristics, and the following table shows their regulatory histories.
| Model | 510(k)# | Manufacturer | Class | Product<br>Code |
|----------------------------------------------------------|------------------------|---------------------------|-------|-----------------|
| SIF-B<br>Small Intestinal Fiberscope | Preamendment<br>Device | Olympus Optical Co., Ltd. | II | 78-KOG |
| SIF-SW Sonde-type Small<br>Intestinal Fiberscope | #K904800 | Olympus Optical Co., Ltd. | II | 78-KOG |
| EVIS EXERA<br>Colonovideoendoscopes | #K001241 | Olympus Optical Co., Ltd. | II | 78-KOG |
| PR-2B(Cannula with Stylet)<br>for EVIS-200 System | #K931154 | Olympus Optical Co., Ltd. | II | 77-EOQ |
| ST-C6/C8 Splinting Tube<br>for EVIS-140 Series<br>Scopes | #K954451 | Olympus Optical Co., Ltd. | II | 78-FDS |
#### D. DEVICE DESCRIPTION
#### 1. Summarv
The subject device, the XSIF-1TQ140 is basically identical to the preamendment device, the SIF-B except that the mechanical structure of variable stiffness has been added and the imaging system has been changed from fiber to CCD.
While the mechanical structure of variable stiffness (Refer to Attachment 2, "Flexibility Adjustment Function") and the CCD imaging system for use in the small intestine are considered new technologies, these features are identical to those of another predicate device, the EVIS EXERA Colonovideoendoscopes (#K001241). Any new insertion method and techniques in consequence of the added mechanical structure of variable stiffness are not considered. A variable stiffness enteroscope enhances insertion depth compared to the conventional enteroscope with or without overtube. The mechanical structure of variable stiffness facilitates insertion of this scope into the small intestine (Refer to Appendix-3, "Clinical Literature").
By changing the imaging systems from fiber to CCD, this scope shows more excellent images than the SIF-B. Since the optical system of the subject device is equivalent to that of the predicate device, the CF-Q160AL (#K001241), it does not affect the safety and efficacy for diagnoses and treatments.
In conclusion, the subject device is substantially equivalent to the predicate devices. A comparison table of the subject device and predicate devices is found in Attachment 1.
#### 2. Design
XSIF-1TQ140A has been designed, manufactured and tested in compliance with voluntary safety standards. It meets the requirement of IEC60601-1, IEC60601-1-1and IEC60601-2-18.
{5}------------------------------------------------
Ko 3 1256 pg 3 of 3
#### 3. Materials
All the patient contacting materials used in this endoscope and ancillary equipment are identical materials that have been cleared in the past 510(k) submissions. And all materials have been confirmed with ISO 10993-1.
#### 4. Intended Use of the device
This instrument has been designed to be used with the OLYMPUS EVIS Video System Center, EVIS Universal Light Source, Documentation Equipment, Video Monitor, Endo-Therapy Accessories (such as a Forceps) and other ancillary equipment for endoscopy and endoscopic surgery within the small intestine.
# 5. Summary including conclusion drawn form Non-clinical Tests
When compared to the preamendment/predicate device, XSIF-1TQ140A does not incorporate any significant changes in the intended use, method of operation, material, or designed that could affect the safety effectiveness. Therefore the clinical data is not necessary for its evaluation of safety and efficacy.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.