ENDOLIGHT LED 1.1 ENDOLIGHTLED 1.2

K121724 · Richard Wolf Medical Instruments Corp. · FCW · Sep 20, 2012 · Gastroenterology, Urology

Device Facts

Record IDK121724
Device NameENDOLIGHT LED 1.1 ENDOLIGHTLED 1.2
ApplicantRichard Wolf Medical Instruments Corp.
Product CodeFCW · Gastroenterology, Urology
Decision DateSep 20, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The ENDOLIGHT LED Light Source Projector is intended to be used to illuminate the surgical site during minimally invasive surgical procedures by producing light that is transmitted from the light source through fiber optic cable and a scope.

Device Story

The ENDOLIGHT LED Light Source is a fiber optic illumination system for minimally invasive surgery. It utilizes a high-power, solid-state white LED to generate visible light, which is transmitted to the surgical site via an optical cable and endoscope. The device consists of the LED, a cooling fan, a light port, and a protective housing. It is operated by specialized medical personnel in a clinical setting. The device features manual brightness controls. The 1.2 model includes a light cable detection feature that automatically shuts off the LED when the cable is detached to prevent overheating, and a 'Safe Start' function that manages brightness settings following power interruptions. The output provides consistent illumination for the surgical field, assisting the surgeon in visualizing the anatomy during procedures.

Clinical Evidence

No clinical tests performed. Bench testing and design verification were conducted to demonstrate safety and effectiveness, including compliance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-18 standards.

Technological Characteristics

Solid-state white LED light source; fiber optic transmission; manual brightness control; cooling fan; light cable detection (model 1.2); 'Safe Start' power management (model 1.2). Conforms to IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-18.

Indications for Use

Indicated for illumination of surgical sites during minimally invasive surgical procedures. Intended for use by specialized, medically qualified, and adequately trained personnel.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K12 / 72 4 Revision 08-21-2012 SEP 2 0 2012 # 510(k) Summary | Submitter: | | | |---------------------------------------------------------------------|--------------------------------------------------------------------|------------------------------------| | Company / Institution name: | RICHARD WOLF MEDICAL INSTRUMENTS CORP. | | | Division name (if applicable): | N.A. | | | Street address: | 353 Corporate Woods Parkway | | | City: | Vernon Hills | | | State/Province: | Illinois | | | Country: | USA | | | Contact name: | Mr. Ron Haselhorst | | | Contact title: | Quality Assurance / Regulatory Affairs Manager | | | Date of Preparation: | June 08, 2012 | | | FDA establishment registration number: | 14 184 79 | | | Phone number (include area code): | ( 847 ) 913 1113 | | | FAX number (include area code): | ( 847 ) 913 0924 | | | ZIP / Postal Code: | 60061 | | | Parent Company: | | | | Company / Institution name: | Richard Wolf GmbH | | | FDA establishment registration number: | 96 111 02 | | | Street address: | Pforzheimer Str. 32 | | | City: | Knittlingen | | | State/Province: | Baden-Württemberg | | | Country: | Germany | | | ZIP / Postal Code: | 75438 | | | Product Information: | | | | Trade name: | 5160 ENDOLIGHT LED Light Source | | | Model numbers: | 5160XXX | | | Common name: | LED Light Source | | | Classification name: | FCW - Light Source, Fiber Optic, Routine<br>NTN - LED Light Source | | | Information on devices to which substantial equivalence is claimed: | | | | 510(k) Number | Trade or proprietary or model name | Manufacturer | | 1 K983628 | 1 Auto LP 5123 Xenon Light Projector (Product Code GCT) | 1 Richard Wolf Medical Inst. Corp. | | 2 K082813 | 2 Stryker LED Light Source (Product Code FCW) | 2 Stryker Endoscopy | | 3 K102167 | 3 LO-50 LED Light Source (Product Code FCW, NTN) | 3 Fiberoptics Technology, Inc | {1}------------------------------------------------ K12/724 Revision 08-21-2012 ### Device Description: The ENDOLIGHT LED Light Source is a fiber optic light source utilizing a single, solid state, high power, white light emitting diode (LED) to produce visible light that used to illuminate surgical sites during minimally invasive surgical procedures. The light from the ENDOLIGHT LED Light Source is transmitted through an optical cable and a scope. ENDOLIGHT LED 1.1 and 1.2 are comprised of light source and power supply cord. The units consist of aforementioned LED, cooling fan, light port, and protective housing. ENDOLIGHT LED 1.1 and 1.2 are portable and are very similar in design, format, and functionality to other LED light sources including the predicate devices. This product is exclusively intended for use by specialized medical personnel and must only be used by medically qualified and adequately trained persons. ### Intended Use: The ENDOLIGHT LED Light Source Projector is intended to illuminate the surgical site during minimally invasive surgical procedures by producing light that is transmitted from the light source through fiber optic cable and a scope. #### Technological Characteristics: The ENDOLIGHT LED Light Sources 1.1 and 1.2 are technologically similar to predicate device found in this submission in that all/some of the devices: - Use fiber optic cables and scopes to transmit the light from the source to the site of the l medical procedure - Use an identical technology ie, an LED lamp to produce light while others use an equivalent | source such as a xenon or halogen lamp. - . I Utilize similar manual controls to set the light brightness - Have similar illumination and quality of light - - Operate within an equivalent temperature range - - Can utilize a number of commonly available fiber optic cables ー - ENDOLIGHT LED 1.2 has a light cable detection feature which shuts the unit off the LED l lamp when the fiber optic cable is detached reducing the chance of unit overheating. The ENDOLIGHT LED Light Sources are technologically different to predicate devices found in this submission in that: - > ENDOLIGHT LED 1.2 utilizes a "Safe Start"™ function which after a power interruption of more than five seconds and less then thirty seconds the LED lamp is shut off but retains nominal brightness setting. After thirty seconds the LED lamp is shut off and brightness is reset to 50%. {2}------------------------------------------------ #### Performance Data: Design verification testing demonstrates that the devices function as intended, and the performance did not raise any new issues of safety and effectiveness, and that formal user training is not required. Voluntary Safety and Performance Standards: The ENDOLIGHT LED 1.1 and 1.2 conform to the following Safety Standards: - > IEC 60601-1-2 Edition 3:2007-03, Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests. (General) - > IEC 60601-1, Medical Electrical Equipment Part 1: General Requirements for Safety, 1988; Amendment 1, 1991-11, Amendment 2, 1995. (General) - IEC 60601-2-18:1996, Amendment 1 2000 Medical electrical equipment Part 2: Particular A requirements for the safety of endoscopic equipment. (Dental/ENT) Testing was completed by Independent laboratories, certifications are on file. #### Clinical Data: No clinical tests performed. #### Rational for Substantial Equivalence: The Richard Wolf ENDOLIGHT LED Light Source shares the same general indications for use, has similar function features and technological characteristics as the predicate devices, the minor difference(s) do not raise new questions for safety or effectiveness. The Richard Wolf ENDOLIGHT LED Light Sources were non-clinically tested to determine the safety and efficacy under the indications for use and meet aforementioned safety standards, same as the predicate devices. For these reasons, the Richard Wolf ENDOLIGHT LED Light Source is substantially equivalent to the existing 510(k) cleared devices sold by: Richard Wolf Medical Instruments Corporation (K983628), Stryker Endoscopy (K082813), and Fiberoptics Technology, Inc (K102167). {3}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle-like symbol with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Mr. Ron Haselhorst Quality Assurance / Regulatory Affairs Manager Richard Wolf Medical Instruments Corporation 353 Corporate Woods Parkway VERNON HILLS IL 60061 Re: K121724 Trade/Device Name: ENDOLIGHT LED Light Source Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FCW, NTN Dated: August 22, 2012 Received: August 27, 2012 Dear Mr. Haselhorst: We have reviewed your Section 5.10(k) premarket notification of intent to market the indication We have reviewed your Section 3.10(x) promation is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the enclosure) to tegans ment date of the Medical Device Amendments, or to commerce prior to May 20, 1770, the enacifican with the provisions of the Federal Food. Drug, devices that have been reclasined in accerative approval of a premarket approval application (PMA).).).). and Cosment Act (Act) that to hot require approval controls provisions of the Act. The Act. The Act. The Act. The You may, therefore, market the device, buyer to many for annual registration, listing of general controls provisions of the fiel labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability adulteration. Please note. CDNT does not overaliate international and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), ( If your device is classified (SCC above) in Exting major regulations affecting your device can be it may be subject to additional controls. Extracts of an edition, FDA may it may be subject to additional controls. Externs mayor solo to 898. In addition, FDA may found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In additio found in the Code of I eactar resguinting your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that IDA's issualite of a but device with other requirements of the Act that FDA has made a determination that your device of the Force agencies. You must that FDA has made a decommancin as administered by other Federal agencies. You must or any Pederal Statutes and regulations and not limited to: registration and listing comply with an the Act s requirements, mortains, common (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical SEP 2 0 2012 {4}------------------------------------------------ Page 2 - device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set de vice-related adverse ovents) (21 CFR Part 820); and if applicable, the electronic form in the quality by selling (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please II you desire specific advice for your do need on our more offices/ucm115809.htm for go to intp://www.idd.gov/rtladiological Health's (CDRH's) Office of Compliance. Also, please the Center for Deviocs and Radiotoground by reference to premarket notification" (21CFR Part note the regulation onlined, "informations" - " - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the I ou may ovan other general mistmational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Benjamin R. Fisher, Ph.D. Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Indication for Use K121724 510(k) Number (if known): Device Name: ENDOLIGHT LED Light Source ## Intended Use: The ENDOLIGHT LED Light Source Projector is intended to be used to illuminate the surgical site during minimally invasive surgical procedures by producing light that is transmitted from the light source through fiber optic cable and a scope. Prescription use S (Part 21 CFR 801 Subpart D) and / or Over-The Counter Use (Part 21 CFR 801 Subpart C) PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ANOTHER PAGE IF NEEDED Concurrence of CDHR office of Device Evaluation (ODE) Aoy M. Whan ductive, Gastro-R Page 1 of 1 Section / Page 4-9 of 44
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...