Disposable Coaxial Biopsy Needle is intended for use with biopsy devices cannula during soft tissue core biopsy procedures. The device is not intended for use in bone.
Device Story
Disposable coaxial biopsy needle; consists of cannula, stylet, and housing; available in various gauges (11G-19G) and lengths. Used as a guide channel for biopsy needles to access soft tissue (breast, kidney, liver, lung); allows multiple samples from a single puncture site; reduces skin trauma. Operated by clinicians under medical imaging guidance (excluding MRI). Device provides a stable path for biopsy instruments; facilitates accurate tissue acquisition. Single-use; sterile; stainless steel construction.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via penetration force, depth projection, mechanical durability, and ultrasound visibility testing. Biocompatibility confirmed per ISO 10993.
Technological Characteristics
Stainless steel (06Cr19Ni10) construction; conforms to ISO 9626. Coaxial introducer design with blunt/sharp stylet options. Gauges: 11G, 13G, 15G, 17G, 19G. Lengths: 70mm-178mm. Sterilization: Ethylene Oxide (SAL 10^-6). Packaging: ISO 11607-1 compliant. Standalone mechanical device.
Indications for Use
Indicated for soft tissue core biopsy procedures in patients requiring tissue sampling; not for use in bone.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA acronym is in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
Suzhou Leapmed Healthcare Corporation % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.1801, No.161, East Lujiazui Rd., Pudong Shanghai, Shanghai 200120 China
# Re: K212822
Trade/Device Name: Disposable Coaxial Biopsy Needle Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology Biopsy Instrument Regulatory Class: Class II Product Code: FCG Dated: May 27, 2022 Received: June 6, 2022
## Dear Boyle Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm = identifies = combination = product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical
July 6, 2022
{1}------------------------------------------------
device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and -CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-DICE website assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212822
Device Name Disposable Coaxial Biopsy Needle
Indications for Use (Describe)
Disposable Coaxial Biopsy Needle is intended for use with biopsy devices cannula during soft tissue core biopsy procedures. The device is not intended for use in bone.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K212822
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.
# 1.0 Submitter's Information
| Name: | Suzhou Leapmed Healthcare Corporation |
|----------|-----------------------------------------------------------------------------------------------------------------------------------|
| Address: | Wuzhong Science Park, NO.2 & NO.3 Building, 38# North<br>Guandu Road, Wuzhong District, Suzhou City, 215100 Jiangsu,<br>P.R.China |
| Tel: | (+86) 512 66554498-8043 |
| Contact: | Yu Zhu |
## Designated Submission Correspondent
| Contact: | Mr. Boyle Wang |
|----------------|------------------------------------------------------------------------|
| Name: | Shanghai Truthful Information Technology Co., Ltd. |
| Address: | Room 1801, No. 161 East Lujiazui Rd., Pudong Shanghai,<br>200120 China |
| Tel:<br>Email: | +86-21-50313932<br>Info@truthful.com.cn |
Date of Preparation: Jul.4,2022
# 2.0 Device Information
| Trade name: | Disposable Coaxial Biopsy Needle |
|----------------------|--------------------------------------------|
| Common name: | Biopsy Needle |
| Classification name: | Gastroenterology-urology biopsy instrument |
| Model(s): | 11G、13G、15G、17G、19G |
| Production code: | FCG |
| Regulation number: | 21 CFR 876.1075 |
| Classification: | Class II |
| Panel: | Gastroenterology/Urology |
# 3.0 Predicate Device Information
| Manufacturer: | M.D.L. S.r.l. |
|----------------|--------------------------------------|
| Device: | MDL INTRO Coaxial Introducer Needles |
| 510(k) number: | K160316 |
{4}------------------------------------------------
## 4.0 Device Description
The disposable coaxial biopsy needle consists of canular housing, canula, marker, stylet with sharping, stylet without sharping, stylet housing and protecting sheath. It is available in several gauge sizes and lengths.
The disposable coaxial biopsy needle (hereinafter referred to as coaxial biopsy needle) is divided into 11G、13G、15G、17G、19G in accordance with the outer diameter of canula and 70mm、78mm、100mm、130mm、138mm、170mm、170mm and 178mm in accordance with the length of canula.
The materials used for construction of disposable coaxial biopsy needle are typical for this type of medical device. The only material in direct patient contact is the stainless steel 06Cr19Ni10.
The disposable coaxial biopsy needle can be used with the matching biopsy needle under the guidance of medical imaging, as a channel for other biopsy needles to enter the body and obtain biopsy samples through negative pressure or cutting for clinical diagnosis or treatment. In clinical practice, doctors sometimes repeatedly puncture the target site for biopsy. At this time, the coaxial needle is configured, which provides a channel that the device can take multiple samples at the target site. That avoids repeated puncture and injury on the patient's skin surface.
As the device is single use device, which is individually packaged sterile devices. The packaging is compatible with the product's EO sterilization method. The sterilization validation confirms the packaging is qualified bacterial film to maintain the sterilization condition of the device.
## 5.0 Indication for Use Statement
Disposable Coaxial Biopsy Needle is intended for use with biopsy devices cannula during soft tissue core biopsy procedures. The device is not intended for use in bone.
## 6.0 Summary of Non-Clinical Testing
Summary of non-clinical and performance testing- Bench testing was performed to evaluate the performance and functionality of the subject device against requirement specification. The subject device has been subjected to compliance testing according to, by FDA, recognized consensus standards ISO 9626, ISO 7864, ISO 10993-7, ISO 10993-1, ISO 11607-1. Results from testing performed confirms that the design requirement specification and user needs have been met. The subject device is confirmed to be safe and effective for the intended use.
7.1 Sterilization and shelf life disposable coaxial biopsy needle is delivered sterile and have successfully been tested according to ISO 11607- 1. The label shelf life is 3 years.
{5}------------------------------------------------
7.2 Biocompatibility testing of disposable coaxial biopsy needle has successfully been tested for cytotoxicity, sensitization, intracutaneously irritation, acute systemic toxicity and material medicated pyrogenicity. The test results verify that the biocompatibility criteria given in ISO 10993 are fulfilled. Disposable coaxial biopsy needle is non-toxic and biocompatible.
7.3 Performance testing - Bench: The performance of disposable coaxial biopsy needle has been verified. Tests as described in table 1 have been completed.
| Test Item | Description |
|---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Depth projection | To confirm that subject device will not extend over the stylet tip (over-throw) during use. Conformity has been demonstrated. |
| Penetration force | To confirm that the penetration force of the subject device is equivalent to predicate device. Conformity has been demonstrated. |
| Mechanical durability | To confirm that subject device withstands the forces applied to the subject device during normal use. Conformity has been demonstrated. |
| Ultrasound Visibility | To confirm that the invasive part of subject device is visible in the area guided by the puncture frame or within the ultrasound range. |
| Qualification metal tubing/needle component | The stainless-steel tubing fulfills the requirement in ISO 9626 Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods. Conformity has been demonstrated |
# Table 1: Performance testing summary - Bench
# 7.0 Summary of Clinical Testing
No clinical study is included in this submission.
## 8.0 Technological Characteristic Comparison Table
| Item | Subject Device | Predicate Device |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) No. | K212822 | K160316 |
| Product Code | FCG | FCG |
| Regulation No. | 21 CFR 876.1075 | 21 CFR 876.1075 |
| Class | II | II |
| Intended Use | Disposable Coaxial Biopsy Needle is intended for use with biopsy devices cannula during soft tissue | Coaxial Introducer Needle is intended for use with biopsy devices cannula during soft tissue |
| | | |
| | core biopsy procedures. The<br>device is not intended for use in<br>bone. | core biopsy procedures. The<br>device is not intended for use in<br>bone.<br>The extent of histological<br>abnormality cannot be reliably<br>determined from it mammographic<br>appearance. Therefore the extent<br>of removal of the imaged evidence<br>of an abnormality does not predict<br>the extent of removal of a<br>histological abnormality (e.g.<br>malignancy). When the sampled<br>abnormality is not histologically<br>benign, it is essential that the<br>tissue margins be examined for<br>completeness of removal using<br>standard surgical procedures. |
| Anatomical<br>sites | Specimens from soft tissue such<br>as breast, kidney, liver, lung and<br>various soft tissue masses | Specimens from soft tissue such<br>as breast, kidney, liver, lung and<br>various soft tissue masses. |
| Device type | Coaxial introducer needle with a<br>blunt tip stylet | Coaxial introducer needle with a<br>blunt tip stylet |
| Visualization<br>technique | Conventional imaging guidance<br>equipment excluding MRI | Conventional imaging guidance<br>equipment excluding MRI |
| Needle material | Stainless Steel.<br>Only Stainless steel is in direct<br>surgical contact with all soft<br>tissues of the patient. | Stainless Steel |
| Needle<br>diameter<br>(Gauge) | 11G,13G,15G,17G,19G | 13G,15G,17G,19G |
| Needle length<br>(mm) | 70、78、100、108、130、138、<br>170、178 | 90,132,170,217 |
| Cannula and<br>Stylet | The cannula is designed with an<br>outer cutting cannula having a<br>sharpened tip and an inner stylet<br>with sharping or without sharping. | The cannula is designed with an<br>outer cutting cannula having a<br>sharpened tip and an inner stylet<br>with sharping or without sharping. |
| Sterile | Ethylene Oxide, SAL: 10-6 | Ethylene Oxide |
| Shelf Life | 3 years | 5 years |
| Single Use | Single Use | Single Use |
| Labeling | Conform with 21 CFR 801 | Conform with 21 CFR 801 |
| Performance<br>Comparison<br>testing | Comparative test of penetration force, stiffness, resistance to breakage, bonding strength and depth projection on the current subject device and the predicate device has been performed. | |
| Biocompatibility | Conform with ISO10993-1<br>(ISO10993-4, ISO10993-5,<br>ISO10993-10, ISO10993-11) | Conform with ISO 10993 standards |
#### Table 2- Comparison of Technology Characteristics
{6}------------------------------------------------
{7}------------------------------------------------
510(k) Summary
The technological characteristics of the subject device are identical to those of predicate device. The subject device has the same basic design as the predicate device. The comparison between the subject and predicate devices is based on the following:
- Same intended use
- · Same indications for use
- · Similar material types that meet ISO 10993 biocompatibility requirements
- · Same sterilization methods
- · Same fundamental technology
The Disposable Coaxial Biopsy Needle designs display minor differences between the subject device and the predicate devices for needle length. The Max Needle length of the current device are bigger than those of the predicate device, but the comparative performance testing results per FDA recognized standards ISO 9626 and ISO 7864 on the subject device and the predicate device shown there is no significant risk raised by the difference.
# 9.0 Conclusion
The conclusions drawn from the comparison and analysis above demonstrate that the proposed device is as safe, as effective, and performs as well as the legally marketed predicated device in K160316 and raises no new questions of safety or effectiveness. The differences between both devices are insignificant in terms of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.