K191472 · Ushare Medical, Inc. · FCG · Jan 29, 2020 · Gastroenterology, Urology
Device Facts
Record ID
K191472
Device Name
Biopsy Needle
Applicant
Ushare Medical, Inc.
Product Code
FCG · Gastroenterology, Urology
Decision Date
Jan 29, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1075
Device Class
Class 2
Indications for Use
The Biopsy Needle is intended for use in obtaining core biopsy samples from soft tissue such as kidney, liver, prostrate, spleen, lymph nodes, and various soft tissue masses. Not intended for use in bone. The Biopsy Needle is also indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality. The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
Device Story
Biopsy Needle; single-use, sterile, disposable device. Input: soft tissue (kidney, liver, prostate, spleen, lymph nodes, breast). Operation: clinician-operated, single-hand automatic activation; guillotine coring mechanism; provides single or two-stage (sequential) automatic advancement for fixed sample length. Output: core tissue sample for histologic evaluation. Used in clinical settings; features include cannula centimeter marks, echogenic radiographic visibility, color-coded hubs, and adjustable depth stops. Benefits: facilitates accurate tissue sampling for diagnosis of abnormalities; allows partial or complete removal of imaged breast abnormalities for examination.
Clinical Evidence
Bench testing only. Performance verified via comparison to predicate device, including firing force, penetration force, biopsy sample quality, and ultrasound visibility. Biocompatibility testing per ISO 10993 series and sterilization validation per ISO 11135/11607 performed.
Technological Characteristics
Stainless steel needle tubing (ISO 9626). Single-use, EO sterilized (ISO 11135). Features: guillotine coring, single-hand automatic activation, echogenic tip, depth stops. Biocompatibility per ISO 10993-1, -4, -5, -10, -11. Dimensions: 14G-20G, lengths 100-200mm.
Indications for Use
Indicated for obtaining core biopsy samples from soft tissue (kidney, liver, prostate, spleen, lymph nodes, masses) and diagnostic sampling of breast abnormalities. Not for use in bone.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Predicate Devices
Achieve Programmable Automatic Biopsy Systems (K141552)
Submission Summary (Full Text)
{1}------------------------------------------------
January 29, 2020
Ushare Medical Inc. % Raymond Luo Technical Manager Shanghai Sungo Management Consulting Co., Ltd. 13th F, 1500# Century Avenue Shanghai, China 200122
Re: K191472
Trade/Device Name: Biopsy Needle Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology-Urology Biopsy Instrument Regulatory Class: Class II Product Code: KNW Dated: December 2, 2019 Received: December 2, 2019
Dear Raymond Luo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191472
Device Name Biospy Needle
#### Indications for Use (Describe)
The Biopsy Needle is intended for use in obtaining core biopsy samples from soft tissue such as kidney, liver, prostrate, spleen, lymph nodes, and various soft tissue masses. Not intended for use in bone.
The Biopsy Needle is also indicated to provide breast tissue sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality.
The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
| Type of Use (Select one or both, as applicable) | <span> <span style="text-decoration: underline;">Prescription Use (Part 21 CFR 801 Subpart D)</span> </span> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) Summary
# Date of preparation: 2020-1-28
#### 1. Applicant:
Company Name: Ushare Medical Inc. Company Address: 445 An Ji Zhong Road, San Zao Town, Zhuhai, Guangdong, P. R. China Contact Person Name: Amy Wang (Ms.) Title: Quality Manager Tel: 0086-756-7516888 Mail: qm(@usharemedical.com
# Official Contact Person Information
Company Name: Shanghai Sungo Management Consulting Company Limited Contact Person Name: Raymond Luo (Mr.) Title: Technical Manager Tel: 0086-21-68828050 Mail: fda.sungo@gmail.com
# 2. Current Device:
The proprietary name of the new device: Biopsy Needle The generic name of the device: Instrument, Biopsy Classification regulation: 21 CFR 876.1075 Classification: Class II. Regulation Medical Specialty: Gastroenterology/Urology Product code: KNW
# 3. Predicate device:
K number: K141552 Company: CareFusion Address: 75 North Fairway Drive, Vernon Hills, IL 60061 USA Predicate Device: Achieve Programmable Automatic Biopsy Systems
# 4. Intended use of the device:
The Biopsy Needle is intended for use in obtaining core biopsy samples from soft tissue such as kidney, liver, prostrate, spleen, lymph nodes, and various soft tissue masses. Not intended for use in bone.
The Biopsy Needle is also indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality.
The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically
{5}------------------------------------------------
benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
# 5. Device Description:
The sketch of the product structure and the dimension is shown in the figure below.
The Biopsy Needle are used to remove, by cutting, a specimen of tissue for microscopic evaluation. The organs in which the device may be used include but are not limited to breast, kidney, liver, prostate, spleen and lymph nodes plus various soft tissue masses. As the device is single use device, which is individually packaged sterile devices. In the package, there is a whole device with the structure shown in the picture above without any accessories. The packaging is compatible with the product's EO sterilization method. The sterilization validation confirms the packaging is qualified bacterial film to maintain the sterilization condition of the device.
# 6. Current Device and Predicate Device Technical Characteristics
The proposed Biopsy Needle is substantially equivalent to the predicate Achieve Programmable Automatic Biopsy Systems with regards to claims, design, technology, and intended use. Refer to the Side by Side Comparison Table below.
#### 6.1 Device Format
The subject disposable biopsy devices share format as the predicate devices for sterile, single use, and EO compatible packaging. Results from device testing indicate that subject devices are non-pyrogenic.
{6}------------------------------------------------
| Predicate | |
|-----------|----------------------------------------------------------------------------------------|
| Device | |
| | A Fully Automatic<br>D-Stylet Only<br>ACHIEVE™<br>Programmable Automatic Biopsy System |
# 6.2 Technical Characteristics
| Device | Current Device | Predicate Device |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Ushare Medical Inc. K191472 | CareFusion K141552 |
| Model Name | MBN Series Biopsy Needle | Achieve Programmable Automatic Biopsy Systems |
| Classification | Class II Device, KNW | Class II Device, KNW, |
| Intend use | The Biopsy Needle is intended for use in obtaining core biopsy samples from soft tissue such as kidney, liver, prostrate, spleen, lymph nodes, and various soft tissue masses. Not intended for use in bone.<br>The Biopsy Needle is also indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality.<br>The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures. | The Achieve Programmable Automatic Biopsy Systems is intended for use in obtaining core biopsy samples from soft tissue such as kidney, liver, prostrate, spleen, lymph nodes, and various soft tissue masses. Not intended for use in bone.<br>The Achieve Programmable Automatic Biopsy Systems is also indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality.<br>The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures. |
| Designated<br>metric size | 14G, 16G, 18G, 20G | 12G, 14G, 16G, 18G, 20G |
| Length L(mm) | 100mm, 150mm, 200mm | 60mm, 90mm, 110mm, 150mm, 200mm, 250mm |
| Slot size | 10mm, 15mm, 20mm | 20mm |
| Cannula and<br>Stylet | The cannula is designed with an outer cutting cannula having a sharpened tip and an inner | The cannula is designed with an outer cutting cannula having a sharpened tip and an inner |
{7}------------------------------------------------
| | stylet with sample slot. | stylet with sample slot. |
|-----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Needle<br>Advancement<br>and Tissue<br>Access | Biopsy Needle with guillotine coring and the<br>predicate device with guillotine coring provide<br>the clinician with the same single or two-stage<br>(sequential) automatic advancement for fixed<br>sample length for tissue penetration and cutting. | Same as current device |
| Mechanism of<br>Action | single-hand automatic activation | Same as current device |
| Usability and<br>Convenience | Provide design features that facilitate clinician<br>use during biopsies: cannula centimeter marks,<br>echogenic radiographic visibility, color coded<br>needle hubs, and adjustable depth stops. | Same as current device |
| Standard | Biocompatibility: ISO10993-1 (ISO10993-4,<br>ISO10993-5, ISO10993-10, ISO10993-11)<br>Sterilization: ISO11135, ISO11138, ASTM<br>F1980, ISO11737-1, ISO10993-7, ISO11607<br>Performance: ISO9626 | Biocompatibility: ISO10993-1<br>Sterilization: ISO11135, ISO11138, ASTM<br>F1980, ISO11737-1, ISO10993-7, ISO11607<br>Performance: ISO9626 |
| Comparison<br>testing | Pressing parts, Pressing parts firing force,<br>Cannula firing force, Safety switch, Sampling<br>switch, Scale marks firmness, Total heavy metal<br>content, Scale mark identification, Sampling<br>structure, Sampling method, Penetration force,<br>Biopsy Sample Testing, Ultrasound Visibility<br>Testing | Pressing parts, Pressing parts firing force,<br>Cannula firing force, Safety switch, Sampling<br>switch, Scale marks firmness, Total heavy metal<br>content, Scale mark identification, Sampling<br>structure, Sampling method, Penetration force,<br>Biopsy Sample Testing, Ultrasound Visibility<br>Testing |
The technological characteristics of the subject device are identical to those of predicate device. The subject device has the same basic design as the predicate device. The comparison between the subject and predicate devices is based on the following:
- Same intended use
- Same indications for use
- · Similar material types that meet ISO 10993 biocompatibility requirements
- · Same sterilization methods
- · Same fundamental technology/principal of operation/user interface
The disposable biopsy device needle designs display minor differences between the subject device and the predicate devices for gauge and needle length. The Max and Min size of the current device were covered by the predicate device. There is no significant risk raised by the difference.
| Characteristic | Standard | Content |
|------------------|--------------------------|----------------------------------------------------------------------------|
| Biocompatibility | AAMI/ANSI/ISO<br>10993-1 | Biological evaluation of Medical Devices Part 1: Evaluation<br>and Testing |
| Biocompatibility | ISO 10993-4 | Biological evaluation of medical devices--Part 4: Selection of tests for |
# 7. Performance Testing
#### 7.1 Biocompatibility and Sterility
{8}------------------------------------------------
| | | interactions with blood |
|------------------|------------------------|----------------------------------------------------------------------------------------------------|
| Biocompatibility | ISO 10993-5 | Biological evaluation of medical devices - Part 5: Tests for in vitro<br>cytotoxicity |
| Biocompatibility | ISO 10993-10 | Biological evaluation of medical devices - Part 10: Tests for irritation and skin<br>sensitization |
| Biocompatibility | ISO 10993-11 | Biological evaluation of medical devices - Part 11: Tests for systemic toxicity |
| Sterilization | ISO11135 | Medical Device, Validation and Routine Control of Ethylene Oxide<br>Sterilization |
| Residuals | ISO10993-7 | Biological evaluation of Medical Devices Part 7: Ethylene<br>Oxide Sterilization Residuals |
| Sterilization | ANSI/AAMI/ISO<br>11607 | Packaging for Terminally Sterilized Medical Devices |
| Sterilization | ASTM F1980-07 | Accelerated Aging of Sterile Barrier Systems |
#### 7.2 Device Shelf-life
The subject devices in their packaging were subjected to accelerated aging to simulate a 5 year shelf life (Treatment: 60°C, 162 days, <50% RH). The aged subject devices were tested for seal Strength, Dye Penetration, Vacuum Leak and Packaging resistance Bacterial performance.
| Characteristic | Standard | Content |
|----------------|-------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance | ISO 9626:2016 | Stainless Steel Needle Tubing for the Manufacture of Medical Devices. |
| Performance | The test was<br>conducted to the<br>predicate device<br>and current<br>device to<br>compare their<br>performance. | Pressing parts, Pressing parts firing force, Cannula firing force, Safety switch,<br>Sampling switch, Scale marks firmness, Total heavy metal content, Scale mark<br>identification, Sampling structure, Sampling method, Penetration force-<br>Comparison of testing result of predicate and proposed devices to prove<br>equivalency. |
| Performance | | Biopsy Sample Testing – Comparison of samples obtained by predicate and<br>proposed devices to prove equivalency. |
| Performance | | Ultrasound Visibility Testing - Verification of the proposed device ultrasound<br>visibility to ensure safety and effectiveness |
#### 7.3 Performance Testing
#### 8. Conclusion
The analysis of the Current Biopsy Devices by intended use, indications, anatomical locations, and mechanism of action supports that the subject devices are the same as those of the predicate devices.
Needle advancement for the subject device provides a single or two-stage (sequential) automatic advancement for fixed sample length for tissue penetration and cutting same as the predicate device. Materials of construction are those commonly used in medical devices and met biocompatibility requirements for medical devices. The sterile disposable devices also met the requirements for sterility per ISO and USP standards. There are no new questions concerning the safety and effectiveness of these devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.