← Product Code [FCG](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FCG) · K133008

# BNX DELIVERY SYSTEM PRE-LOADED WITH 19, 22 AND 25 GAUGE NEEDLE, 19, 22 OR 25 GAUGE NEEDLES (INDIVIDUALLY STERILE-PACKED, (K133008)

_Beacon Endoscopic · FCG · Nov 20, 2013 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FCG/K133008

## Device Facts

- **Applicant:** Beacon Endoscopic
- **Product Code:** [FCG](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FCG.md)
- **Decision Date:** Nov 20, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1075
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The BNX FNA System is used to sample targeted sub-mucosal and extramural gastrointestinal lesions through the accessory channel of an ultrasound endoscope. The needle is designed with a passive (i.e., automatic) safety shielding feature to aid in the prevention of needle stick injury.

## Device Story

BNX Fine Needle Aspiration (FNA) System is a sterile, single-patient-use endoscopic ultrasound aspiration needle; used to acquire tissue samples from sub-mucosal and extramural gastrointestinal lesions. Device consists of aspiration delivery system and aspiration needle (19, 22, 25 gauge); inserted through accessory channel of ultrasound endoscope. Physician operates device; needle tip penetrates lesion while suction is applied via standard hypodermic syringe. Integrated passive needle protection shield automatically engages over distal needle tip during removal to prevent inadvertent needle sticks. Modular design allows needle exchange without disconnecting handle. Device benefits patient by enabling diagnostic tissue sampling while protecting healthcare providers from sharps injuries.

## Clinical Evidence

Bench and simulated use testing only. Simulated use study evaluated >500 cycles of the sharps injury prevention feature with zero failures. Results met FDA guidance criteria for sharps injury prevention features, demonstrating 97.5% confidence that the true failure rate is ≤0.7% and 99.5% confidence that it is ≤1.1%. Testing confirmed the safety mechanism does not impede clinical performance.

## Technological Characteristics

Modular endoscopic biopsy instrument; metallic needle with stylet; sheath and handle assembly. Features etched needle tips for ultrasonic visualization. Passive, automatic needle protection shield mechanism. Luer-lock connection for endoscope interface. Manual piston-driven needle advancement with screw-lock mechanism. Sterile, single-use.

## Regulatory Identification

A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Predicate Devices

- BNX Fine Needle Aspiration System ([K103668](/device/K103668.md))
- Personna Plus Safety Scalpel System ([K032242](/device/K032242.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### 5.0 510(k) SUMMARY Page 1 of 3

This 510(k) Summary for the BNX Fine Needle Aspiration System is being submitted in accordance with 21 CFR 807.92.

| Submitter's Name and Address: | Beacon Endoscopic Corp.<br>2000 Commonwealth Ave. Suite 110<br>Auburndale, MA 02466-2008                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:               | Annette Fagnant, Regulatory Affairs Consultant,<br>MedDRA Assistance Inc.<br>53 Kennedy Road<br>Foster, RI 02825<br>Phone: 401-392-0287<br>Fax: 401-397-6531                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Date:                         | September 23, 2013                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                               | NOV 20 2013                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Name of Medical Device:       | Device Regulation: 21 CFR 876.1075, Class II<br>Product Code: FCG<br>Common/Usual Name: Kit, Needle, Biopsy (FCG)<br>Proprietary Name: BNX Fine Needle Aspiration System<br>Classification Panel: Gastroenterology-Urology Devices Panel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Predicate Devices:            | The subject device is substantially equivalent to the:<br>- BNX Fine Needle Aspiration System (reference K103668, cleared December 30, 2010).<br>The sharps safety protection mechanism is similar in design and use as that of<br>the Personna Plus Safety Scalpel System (OMI Manufacturing Pty. Ltd,<br>K032242, cleared September 23, 2003)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Device Description:           | The BNX Fine Needle Aspiration (FNA) System is a sterile, single patient use<br>endoscopic ultrasound aspiration needle. The device consists of the BNX<br>Aspiration Delivery System and BNX Aspiration Needle which are assembled<br>before insertion through the accessory channel of an ultrasound endoscope.<br>The needle is used to acquire aspiration samples from lesions targeted using an<br>ultrasound endoscope. An aspiration sample is obtained by penetrating the<br>lesions with the needle while applying suction. The device is offered with needle<br>sizes of 19, 22 and 25 gauge. The BNX FNA System has an integrated needle<br>protection shield that automatically engages over the distal end of the needle<br>during removal to cover the needle sharp. In this manner, the needle tip is<br>covered to help protect against inadvertent needle sticks. |

Traditional 510(k) – BNX™ Fine Needle Aspiration System-Sharps Injury Prevention Feature

24

{1}------------------------------------------------

:

## 510(k) SUMMARY Page 2 of 3

| Indication For<br>Use:            | The BNX FNA System is used to sample targeted sub-mucosal and extramural<br>gastrointestinal lesions through the accessory channel of an ultrasound<br>endoscope. The needle is designed with a passive (i.e., automatic) safety<br>shielding feature to aid in the prevention of needle stick injury.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>Characteristics: | The proposed BNX FNA System has similar intended use and identical<br>technological characteristics compared with the predicate BNX FNA System<br>cleared in K103668. There have been no changes made to the design of the<br>BNX FNA System to allow for the expanded labeling claim.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                   | The referenced BNX FNA device incorporates a long stiff metallic needle with<br>stylet housed in a sheath with handle assembly. The handle is screwed onto<br>the luer-lock connection of the endoscope. The needle is manipulated by a<br>handle piston which is locked and unlocked by means of a screw to avoid<br>advancement of the needle during introduction and withdrawal of the biopsy<br>assembly. The tips of the Aspiration Needles are etched for enhanced<br>ultrasonic needle visualization. Tissue samples are acquired into the lumen of<br>the needle via applied suction using a standard hypodermic syringe. The BNX<br>FNA System is modular in design, i.e., the sheath and handle assembly are<br>incorporated in a Delivery System as a separate component from the<br>Aspiration Needle/stylet assembly. The modular design facilitates exchange of<br>any size aspiration needle as the needle can be removed from the scope<br>without requiring that the handle be disconnected. Additionally, the BNX FNA<br>System has an integrated needle protection shield as a sharps injury<br>protection feature that automatically engages over the distal end of the needle<br>during removal to cover the needle sharp. In this manner, the needle tip is<br>covered to help protect against inadvertent needle sticks. If a second sample<br>acquisition cycle is required, the needle protector shield is captured within the<br>handle housing to be actuated during the next needle removal cycle. |
| Performance<br>Data:              | Bench and simulated use testing were performed demonstrating that the<br>subject device is substantially equivalent to the predicate devices for the<br>proposed intended use. As per CDRH's Guidance for Industry and FDA:<br>Medical Devices with Sharps Injury Prevention Features, simulated use<br>studies were conducted on the subject device evaluating greater than 500<br>cycles of product use with zero failures for the sharps injury prevention feature.<br>Results of this evaluation met FDA's minimum standard for a "97.5%<br>confidence that the true failure rate was no higher than 0.7% and 99.5%<br>confidence that it is no higher than 1.1% " as per the FDA Guidance and<br>documented that the BNX FNA System's passive sharps safety prevention<br>mechanism did not impede or adversely affect the intended clinical<br>performance of the device and provided protection against unintended sharps<br>injury through disposal.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |

Traditional 510(k) - BNX™ Fine Needle Aspiration System-Sharps Injury Prevention Feature

25

{2}------------------------------------------------

#### 510(k) SUMMARY

### Page 3 of 3

Beacon Endoscopic Corp. has demonstrated that the proposed BNX FNA Conclusion: System is substantially equivalent to the predicate BNX FNA System and has demonstrated that the passive needle protection sheath provides sharps safety protection supporting expansion of the device's indication to include that the needle is designed with a passive (i.e., automatic) safety shielding feature to aid in the prevention of needle stick injury.

{3}------------------------------------------------

Food and Drue Administration 10903 New Hampshire Avenue Document Control Center - WO66-Ghos Silver Spring, MD 20993-0002

November 20, 2013

Beacon Endoscopic Corporation. % Annette M. Fagnant Regulatory Affairs Consultant MedDRA Assistance, Inc. 53 Kennedy Road Foster, RI 02825

Re: K133008

> Trade/Device Name: BNX Fine Needle Aspiration System Regulation Number: 21 CFR§ 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: II Product Code: FCG Dated: September 23, 2013 Received: September 25, 2013

Dear Annette M. Fagnant,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, merclore, mains of the Act include requirements for annual registration, listing of general controls providering practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be four of the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean i least be advised that I Dris lesantes over device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{4}------------------------------------------------

Page 2 - Annette M. Fagnant

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/About/DA/CentersOffices/CDRH/CDRHOffices/ncm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Herbert P. Lerner -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive. Gastro-Renal. and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K133008

#### 4.0 Indication for Use Statement

510(k) Number (if known):

Device Name: BNX Fine Needle Aspiration System

Indications for Use:

The BNX FNA System is used to sample targeted sub-mucosal and extramural gastrointestinal lesions through the accessory channel of an ultrasound endoscope. The needle is designed with a passive (i.e., automatic) safety shielding feature to aid in the prevention of needle stick injury.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

# Herbert P. Lerner -S 2013.11.20 16:46:03

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FCG/K133008](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FCG/K133008)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
