ACMI M4 TELESCOPES

K031758 · Acmi Corporation · FBP · Sep 2, 2003 · Gastroenterology, Urology

Device Facts

Record IDK031758
Device NameACMI M4 TELESCOPES
ApplicantAcmi Corporation
Product CodeFBP · Gastroenterology, Urology
Decision DateSep 2, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

ACMI M4 Series Telescopes are intended for use in patients requiring endoscopic evaluation or for surgical procedures of the urological or gynecological systems.

Device Story

ACMI M4 Series Telescopes are rigid, autoclavable optical instruments (4mm diameter, 31cm length) used for visualization during diagnostic and surgical procedures in urology and gynecology. The device is inserted through endoscopic instruments such as cystourethroscopes, hysteroscopes, or resectoscopes. Available in four viewing angles (0°, 12°, 30°, 70°), the telescope transmits optical images to assist the physician in viewing anatomy. Operated by physicians in clinical settings, the device provides direct visualization to guide therapeutic or diagnostic interventions, potentially improving procedural accuracy and patient outcomes.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Rigid, autoclavable optical telescope; 4mm diameter, 31cm length; four viewing angles (0°, 12°, 30°, 70°). Designed for use with endoscopic accessories (cystourethroscopes, hysteroscopes, resectoscopes).

Indications for Use

Indicated for patients requiring endoscopic evaluation or surgical procedures of the urological or gynecological systems.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K031758 510(k) Summary of Safety and Effectiveness ACMI Corporation ACMI® M4™ Telescope (Page 1 of 2) #### General Information Manufacturer: ACMI Corporation 136 Turnpike Rd. Southborough, MA. 01772-2104 FDA Establishment Registration: Contact Person: 2020483 Gabriel J. Muraca, Jr. Senior Regulatory Affairs Specialist Date Prepared: June 3, 2003 #### Device Description Trade Name: ACMI® M4™ Series Telescopes ACMI M4 Series Telescopes are optical devices that are used with endoscopic accessories in diagnostic and surgical procedures of the urological and gynecological systems. Classification Names: 21CFR 876.1500 - Endoscope and accessories and 21 CFR 884.1690 - Hysteroscope and accessories Generic/Common Name: Telescopic system used along with endoscopic accessories in diagnostic and surgical procedures of the urological and gynecological systems. #### Predicate Devices ACMI Resectoscope System and accessories K890328/B ## Intended Uses ACMI M4 Series Telescopes are intended for use in patients requiring endoscopic evaluation or for surgical procedures of the urological or gynecological systems. {2}------------------------------------------------ K031758 510(k) Summary of Safety and Effectiveness ACMI Corporation ACMI® M4™ Telescope (Page 2 of 2) ## Product Description ACMI M4 Telescopes are optical devices that are used along with endoscopic accessories for diagnostic and therapeutic procedures in urology and gynecology. They are designed to be inserted through an endoscopic device such as an ACMI cystourethroscope, hysteroscope, or resectoscope. M4 Telescopes are (4mm diameter, 31 cm long) rigid autoclavable optical instruments available in four directions of view: M4-0A, direct (0°), M4-12A, operative (12°), M4-30A, foroblique (30°), and M4-70A, lateral (70°), to assist the physician in visualization of the anatomy during diagnostic and surgical procedures. ## Technological Characteristics and Substantial Equivalence The M4 Series Telescopes are substantially equivalent to features incorporated in the M2 and M3 telescopes, as described in K890328/B, the ACMI Resectoscope System and accessories. The M2 and M3 telescopes are legally marketed predicate devices. The M4 Series designs utilize similar telescope technology (design, materials, and performance specifications) as in the predicate devices and, therefore, are substantially equivalent to the ACMI M2 and M3 series telescopes. ## Summary of Safety and Effectiveness This 510(k) proposes several modifications in materials, design, and the manufacturing process for the M4 Series telescopes. The indications for use and principles of operation remain the same as in the predicate devices. The proposed modifications for the M4 Series Telescopes, as described in this submission, are substantially equivalent to the predicate devices. The proposed modification in materials, design, and processing are not substantial changes or modifications, and do not significantly affect the safety or efficacy of the devices. In summary, ACMI believes the ACMI M4 Series telescopes are substantially equivalent to the predicate devices, because they have the same intended use and similar technological characteristics. {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three lines representing its wings. The words "Department of Health & Human Services" are arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP - 2 2003 Mr. Gabriel J. Muraca, Jr., RAC Senior Regulatory Affairs Specialist ACMI Corporation 136 Turnpike Road SOUTHBOROUGH MA 01772-2104 Re: K031758 Trade/Device Name: ACMI® M4TM Telescopes Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: 78 FBP Dated: June 3, 2003 Received: June 6, 2003 Dear Mr. Muraca: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adultcration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter: | 8xx. 1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive. Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Device Name: ACMI® M4™ Telescope 510(k) Number: Ko31758 # Indications for Use: ACMI® M4™ Series Telescopes are intended for use in patients requiring endoscopic evaluation or for surgical procedures of the urological or gynecological systems. ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use: __ X __ OR Over-the-Counter Use: ________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801.109) Nancy Brodon (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number K031758
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...