Dragonfly™ Pancreaticobiliary Scope; Dragonfly™ Digital Controller; Dragonfly™ Instrument Channel Caps

K233752 · CenterPoint Systems, LLC · FBN · May 2, 2024 · Gastroenterology, Urology

Device Facts

Record IDK233752
Device NameDragonfly™ Pancreaticobiliary Scope; Dragonfly™ Digital Controller; Dragonfly™ Instrument Channel Caps
ApplicantCenterPoint Systems, LLC
Product CodeFBN · Gastroenterology, Urology
Decision DateMay 2, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Dragonfly Pancreaticobiliary Scope is intended to provide direct visualization and to guide both optical and accessory devices for diagnostic and therapeutic applications during endoscopic procedures in the pancreaticobiliary system including the hepatic ducts. The Dragonfly Digital Controller is intended to provide illumination power and receive, process and output images from the Dragonfly Pancreaticobiliary Scope for diagnostic applications during endoscopic procedures in the pancreaticobiliary system including the hepatic ducts. The Dragonfly Instrument Channel Caps are used to cover the biopsy/suction channel inlet of an Olympus®, Fujifilm® or Pentax® duodenoscope. The Instrument Channel Caps are intended for exclusive use with the Dragonfly Pancreaticobiliary Scope to provide access for passage, while maintaining leakage of biomaterial from the biopsy port throughout the gastrointestinal endoscopic procedure.

Device Story

System consists of single-use Dragonfly Pancreaticobiliary Scope and reusable Dragonfly Digital Controller. Scope features camera, illumination elements, handle, and catheter for per-oral cholangiopancreatoscopy. Controller provides illumination power, receives/processes video signals from scope, and outputs to external monitor. Instrument Channel Caps cover duodenoscope biopsy/suction channel inlets to maintain insufflation and minimize biomaterial leakage during accessory passage. Used in clinical settings by physicians; system provides direct visualization to guide diagnostic/therapeutic accessories. Output allows real-time visualization of pancreaticobiliary anatomy, aiding clinical decision-making during procedures. Benefits include improved access and visualization for pancreaticobiliary interventions.

Clinical Evidence

Bench testing only. No clinical data presented. Performance testing included biocompatibility (ISO 10993-1), sterilization (ISO 11135), packaging (ISO 11607-1), electrical/EMC safety (IEC 60601 series), simulated use, dimensional verification, deflection, and comprehensive optical performance testing (resolution, depth of field, SNR, dynamic range, color performance).

Technological Characteristics

Materials: PEBAX, PTFE, Stainless Steel, Silicone. Sensing: CMOS image sensor. Illumination: LED. Connectivity: DVI video output to external monitor. Sterilization: Ethylene Oxide (SAL 10^-6). Form factor: Single-use catheter (3.6mm distal OD, 5.1F working channel) and reusable controller. Electrical: Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18.

Indications for Use

Indicated for patients undergoing diagnostic and therapeutic endoscopic procedures in the pancreaticobiliary system, including hepatic ducts, requiring direct visualization and guidance of optical/accessory devices.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. May 2, 2024 CenterPoint Systems LLC Conner Johnson Regulatory/Ouality Associate 3338 Parkway Blvd West Valley City, Utah 84119 Re: K233752 Trade/Device Name: Dragonfly™ Pancreaticobiliary Scope; Dragonfly™ Digital Controller; DragonflyTM Instrument Channel Caps Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: FBN, NTN, FET, ODC Dated: April 2, 2024 Received: April 2, 2024 Dear Conner Johnson: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {1}------------------------------------------------ (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2}------------------------------------------------ Sincerely, # Shanil P. Haugen -S Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use ### 510(k) Number (if known) K233752 ### Device Name Dragonfly™ pancreaticobiliary scope; Dragonfly™ digital controller; Dragonfly™ instrument channel caps ### Indications for Use (Describe) The Dragonfly Pancreaticobiliary Scope is intended to provide direct visualization and to guide both optical and accessory devices for diagnostic and therapeutic applications during endoscopic procedures in the pancreaticobiliary system including the hepatic ducts. The Dragonfly Digital Controller is intended to provide illumination power and receive, process and output images from the Dragonfly Pancreaticobiliary Scope for diagnostic applications during endoscopic procedures in the pancreaticobiliary system including the hepatic ducts. The Dragonfly Instrument Channel Caps are used to cover the biopsy/suction channel inlet of an Olympus®, Fujifilm® or Pentax® duodenoscope. The Instrument Channel Caps are intended for exclusive use with the Dragonfly Pancreaticobiliary Scope to provide access for passage, while maintaining leakage of biomaterial from the biopsy port throughout the gastrointestinal endoscopic procedure. Type of Use (Select one or both, as applicable) | <span> <b> ☑ </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> | |--------------------------------------------------------------------------------------------------------| | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ### 510(K) SUMMARY | 1 | SUBMITTER | | |---|-----------------|-------------------------------------------| | | Name: | CenterPoint Systems LLC | | | Address: | 3338 Parkway Blvd<br>West Valley City, UT | | | Phone: | 877-848-0828 | | | Contact Person: | Conner Johnson | | | Date Prepared: | December 21, 2023 | #### 2 DEVICE | Name of Device: | Dragonfly Pancreaticobiliary Scope System | |-----------------------|-------------------------------------------| | Common or Usual Name: | Pancreaticobiliary Scope | | Classification Name: | Endoscope and accessories | | Regulatory Class: | Class II per 21 CFR 876.1500 | | Product Code: | FBN, NTN, FET | | Name of Device: | Dragonfly Instrument Channel Caps | |-----------------------|-----------------------------------| | Common or Usual Name: | Instrument Caps | | Classification Name: | Endoscope and accessories | | Regulatory Class: | Class II per 21 CFR 876.1500 | | Product Code: | ODC | #### 3 PREDICATE DEVICES Predicate Name and 510(k) Number: SpyGlass Visualization System, K181439 This predicate has not been subject to a design-related recall. No reference predicates were used in this submission. Predicate Name and 510(k) Number: BioShield biopsy valve EUS – Linear, K202583 This predicate has not been subject to a design-related recall. No reference predicates were used in this submission. ## CONFIDENTIAL {5}------------------------------------------------ 510(k) Summary - Pancreaticobiliary Scope: Traditional 510(k) Premarket Notification Submission DragonflyTM Pancreaticobiliary Scope System with Instrument Channel Caps #### 4 DEVICE DESCRIPTION ## Dragonfly Pancreaticobiliary Scope and Digital Controller The Dragonfly Pancreaticobiliary Scope is a sterile, single-use endoscope used for singleoperator per-oral 2holoangiopancreatoscopy. The Dragonfly Pancreaticobiliary System is comprised of two components: The Dragonfly Pancreaticobiliary Scope and the Dragonfly Digital Controller. The Dragonfly Pancreaticobiliary Scope System is intended to provide direct visualization and to guide both optical and accessory devices for diagnostic and therapeutic applications during endoscopic procedures in the pancreaticobiliary system including the hepatic ducts. The Dragonfly Pancreaticobiliary Scope is a single-use catheter which provides direct visualization with a camera and illumination, and enables access and delivery of accessories to targeted pancreaticobiliary anatomy. The Dragonfly Pancreaticobiliary Scope comprises of a handle, a catheter, imaging and illumination elements, and a video connection mechanism. The Dragonfly Digital Controller is a reusable accessory which controls illumination and receives, processes, and outputs video signals from the Dragonfly Pancreaticobiliary Scope. The system does not treat or diagnosis conditions. ## Dragonfly Instrument Channel Caps The Dragonfly Instrument Channel Caps, which are included in package with the Dragonfly Pancreaticobiliary Scope, are used to cover the opening to the biopsy/suction channel inlet of a Olympus®, Fujifilm® or Pentax® duodenoscope. The Instrument Channel Caps are intended for exclusive use with the Dragonfly Pancreaticobiliary Scope to provide access for passage, while maintaining insufflation and minimizing leakage of biomaterial from the biopsy port throughout the gastrointestinal endoscopic procedure. {6}------------------------------------------------ 510(k) Summary - Pancreaticobiliary Scope: Traditional 510(k) Premarket Notification Submission DragonflyTM Pancreaticobiliary Scope System with Instrument Channel Caps #### 5 INDICATIONS FOR USE ## Dragonfly Pancreaticobiliary Scope and Digital Controller The Dragonfly Pancreaticobiliary Scope is intended to provide direct visualization and to guide both optical and accessory devices for diagnostic and therapeutic applications during endoscopic procedures in the pancreaticobiliary system including the hepatic ducts. The Dragonfly Digital Controller is intended to provide illumination power and receive, process and output images from the Dragonfly Pancreaticobiliary Scope for diagnostic and therapeutic applications during endoscopic procedures in the pancreaticobiliary system including the hepatic ducts. ## Dragonfly Instrument Channel Caps The Dragonfly Instrument Channel Caps are used to cover the opening to the biopsy/suction channel inlet of a Olympus®, Fujifilm® or Pentax® duodenoscope. The Instrument Channel Caps are intended for exclusive use with the Dragonfly Pancreaticobiliary Scope to provide access for passage, while maintaining insufflation and minimizing leakage of biomaterial from the biopsy port throughout the gastrointestinal endoscopic procedure. {7}------------------------------------------------ 510(k) Summary - Pancreaticobiliary Scope: Traditional 510(k) Premarket Notification Submission DragonflyTM Pancreaticobiliary Scope System with Instrument Channel Caps #### 6 COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE -DRAGONFLY PANCREATICOBILIARY SCOPE AND DIGITAL CONTROLER The proposed Dragonfly Pancreaticobiliary Scope System and the predicate SpyGlass Visualization System (K181439) are similar in indications for use, intended use, and technological characteristics, and principles of operation. The differences between the Proposed Device and the Predicate Device are minor and raise no different questions of safety and effectiveness, thus it was concluded that the Proposed Device is substantially equivalent to the Predicate Device. In accordance with 21CFR807.92(a)(6) a summary of how the technological characteristics of the Proposed Device compares to the Predicate Device is provided below. | Feature | Dragonfly Pancreaticobiliary<br>Scope (proposed device) | Primary Predicate: SpyScope<br>DS Access & Delivery<br>Catheter(K181439) | Same/Differe<br>nt between<br>Proposed &<br>Predicate | |------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------| | Intended<br>Use/Indications for<br>Use (Scope) | The Dragonfly Pancreaticobiliary<br>Scope System is intended to<br>provide direct visualization and to<br>guide both optical and accessory<br>devices for diagnostic and<br>therapeutic applications during<br>endoscopic procedures in the<br>pancreaticobiliary system including<br>the hepatic ducts. | The SpyGlass Visualization<br>System is intended to provide<br>direct visualization and to guide<br>both optical and accessory devices<br>for diagnostic and therapeutic<br>applications during endoscopic<br>procedures in the<br>pancreaticobiliary system<br>including the hepatic ducts. | Same | | Device Class | II | II | Same | | Product Code | FBN, NTN, FET | FBN, NTN, KQM | Same | | Regulation number | 21 CFR 876.1500 | 21 CFR 876.1500 | Same | | Duration of use | Single-use, Transient | Single-use, Transient | Same | | Insertion Method | Through a duodenoscope | Through a duodenoscope | Same | | Sterilization | Ethylene Oxide | Ethylene Oxide | Same | | Prescription<br>Device | Yes | Yes | Same | | Working Channel<br>Inner Diameter | 5.1F | 3.6F | Different<br>(Substantially<br>Equivalent) | | Working Length | 139cm | 214cm | Different<br>(Substantially<br>Equivalent) | | Type of image<br>sensor | CMOS | CMOS | Same | | Illumination Type | LED | LED | Same | | Field of View | 120 degrees | 120 degrees | Same | | Outer Shaft<br>Diameter<br>(Proximal) | 3.6mm | 3.6mm | Same | ### Table 1: Dragonfly Pancreaticobiliary Scope Comparison to Predicate Device {8}------------------------------------------------ 510(k) Summary – Pancreaticobiliary Scope: Traditional 510(k) Premarket Notification Submission Dragonfly™ Pancreaticobiliary Scope System with Instrument Channel Caps | Feature | Dragonfly Pancreaticobiliary<br>Scope (proposed device) | Primary Predicate: SpyScope<br>DS Access & Delivery<br>Catheter(K181439) | Same/Differe<br>nt between<br>Proposed &<br>Predicate | |--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------| | Outer Shaft<br>Diameter (Distal) | 3.6mm | 3.5mm | Different<br>(Substantially<br>Equivalent) | | Articulation | Yes, using deflection knobs on<br>handle | Yes, using deflection knobs on<br>handle | Same | | Materials/<br>Biocompatibility | Standard medical device materials,<br>including PEBAX, PTFE, Stainless<br>Steel<br>As described in the 510(k) the<br>materials above were tested during<br>biocompatibility testing. The<br>biocompatibility tests demonstrates<br>that there are no adverse<br>biocompatibility risks associated<br>with use of this material. All test<br>results met the requirements of ISO<br>10993-1. | Standard medical device materials.<br>The biocompatibility tests<br>demonstrates that there are no<br>adverse biocompatibility risks<br>associated with use of this<br>material. All test results met the<br>requirements of ISO 10993-1. | Same | | Features | Handle, Deflection knobs,<br>Umbilical Cable, Accessory<br>Working Channel, Irrigation Ports, | Handle, Articulation Lever,<br>Umbilical Cable, Accessory<br>Access Port, Irrigation Port | Same | | Video Processing<br>Unit | Separate VPU which connects to an<br>existing monitor. | Separate VPU which connects to<br>an existing monitor | Same | | Direction of View | 0 degrees | 0 degrees | Same | | Minimum<br>Duodenoscope<br>Working Channel | 4.2mm | 4.2mm | Same | | EMC Safety<br>Testing | Complies with applicable clauses of<br>IEC 60601 | Complies with applicable clauses<br>of IEC 60601 | Same | | | | Table 2: Dragonfly Digital Controller Comparison to Predicate Device | | | |--|--|----------------------------------------------------------------------|--|--| | Feature | Dragonfly Digital Controller | Primary Predicate: SpyGlass DS Digital Controller (K181439) | Same/Different between Proposed & Predicate | |-----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------| | Intended Use/Indications for Use (Controller) | The Dragonfly Digital Controller is intended to provide illumination power and receive, process and output images from the Dragonfly Pancreaticobiliary Scope for diagnostic and therapeutic applications during endoscopic procedures in the pancreaticobiliary system including the hepatic ducts. | The SpyGlass DS Digital Controller is intended to provide illumination and receive, process, and output images from the SpyScope DS Access and Delivery Catheter or SpyScope DS II Access and Delivery Catheter for diagnostic and therapeutic applications during endoscopic procedures in the pancreaticobiliary system including the hepatic ducts. | Different (Substantially Equivalent) | {9}------------------------------------------------ 510(k) Summary – Pancreaticobiliary Scope: Traditional 510(k) Premarket Notification Submission Dragonfly™ Pancreaticobiliary Scope System with Instrument Channel Caps | Feature | Dragonfly Digital Controller | Primary Predicate: SpyGlass DS<br>Digital Controller (K181439) | Same/Differe<br>nt between<br>Proposed &<br>Predicate | |-----------------------------|--------------------------------------------------|----------------------------------------------------------------|-------------------------------------------------------| | Device Class | II | II | Same | | Product Code | FBN | FBN | Same | | Regulation number | 21 CFR 876.1500 | 21 CFR 876.1500 | Same | | Video Outputs | DVI | S-video, VGA or DVI | Same | | White Balancing | Automatic | Automatic | Same | | Brightness Control | Yes | Yes | Same | | Separate Monitor | Yes | Yes | Same | | Energy<br>used/Power source | Yes | Yes | Same | | Reusability | Yes | Yes | Same | | EMC Safety<br>Testing | Complies with applicable clauses of<br>IEC-60601 | Complies with applicable clauses<br>of IEC-60601 | Same | {10}------------------------------------------------ ### 7 COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE -DRAGONFLY INSTRUMENT CHANNEL CAPS The proposed Dragonfly Instrument Channel Caps and the predicate BioShield biopsy valve EUS - Linear (K202583) are similar in indications for use, intended use, and technological characteristics, and principles of operation. The differences between the Proposed Device and the Predicate Device are minor and raise no different questions of safety and effectiveness, thus it was concluded that the Proposed Device is substantially equivalent to the Predicate Device. In accordance with 21CFR807.92(a)(6) a summary of how the technological characteristics of the Proposed Device compares to the Predicate Device is provided below. | Feature | Dragonfly Instrument Caps<br>(proposed device) | Primary Predicate: BioShield<br>biopsy valve EUS - Linear<br>(K202583) | Same/Different<br>between<br>Proposed &<br>Predicate | |------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------| | Intended Use | The Dragonfly Instrument Channel<br>Caps are intended to cover the<br>opening to the biopsy/suction<br>channel of flexible<br>echoendoscopes. It provides access<br>for endoscopic device passage and<br>exchange, helps maintain<br>insufflation, minimizes leakage of<br>biomaterial from the biopsy port<br>throughout the endoscopic<br>procedure. | The BioShield biopsy valve EUS<br>– Linear (K202583) is intended to<br>cover the opening to the<br>biopsy/suction channel of flexible<br>echoendoscopes. It provides<br>access for endoscopic device<br>passage and exchange, helps<br>maintain insufflation, minimizes<br>leakage of biomaterial from the<br>biopsy port throughout the<br>endoscopic procedure. | Same | | Indications for<br>Use | The Dragonfly Instrument Channel<br>Caps are used to cover the opening<br>to the biopsy/suction channel inlet<br>of a Olympus®, Fujifilm® or<br>Pentax® duodenoscope. The<br>Instrument Channel Caps are<br>intended for exclusive use with the<br>Dragonfly Pancreaticobiliary Scope<br>to provide access for passage,<br>while maintaining insufflation and<br>minimizing leakage of biomaterial<br>from the biopsy port throughout the<br>gastrointestinal endoscopic<br>procedure. | The single use BioShield(@ biopsy<br>valve EUS is used to cover the<br>opening to the biopsy/suction<br>channel of flexible<br>echoendoscopes. It provides<br>access for endoscopic device<br>passage and exchange, helps<br>maintain insufflation, minimizes<br>leakage of biomaterial from the<br>biopsy port throughout the<br>endoscopic procedure and<br>provides access for irrigation. | Different<br>(Substantially<br>Equivalent) | | Device Class | II | II | Same | {11}------------------------------------------------ 510(k) Summary – Pancreaticobiliary Scope: Traditional 510(k) Premarket Notification Submission Dragonfly™ Pancreaticobiliary Scope System with Instrument Channel Caps | Feature | Dragonfly Instrument Caps<br>(proposed device) | Primary Predicate: BioShield<br>biopsy valve EUS – Linear<br>(K202583) | Same/Different<br>between<br>Proposed &<br>Predicate | |----------------------------------|---------------------------------------------------|------------------------------------------------------------------------|------------------------------------------------------| | Regulation<br>number | 21 CFR 876.1500 | 21 CFR 876.1500 | Same | | Duration of use | Single-use, Transient | Single-use, Transient | Same | | Construction | Cap | Valve Body, Valve Cap | Different<br>(Substantially<br>Equivalent) | | Sterilization<br>Method | Ethylene Oxide | Ethylene Oxide | Same | | Sterilization<br>Assurance Level | 10-6 | 10-6 | Same | | Prescription<br>Device | Yes | Yes | Same | | Device<br>Dimensions | Length: 0.615"<br>Width: 0.615"<br>Height: 0.480" | Length: 0.52"<br>Width: 0.52"<br>Height: 0.77" | Different<br>(Substantially<br>Equivalent) | | Target Population | GI Endoscopic procedures | GI Endoscopic procedures | Same | | Energy<br>used/delivered | None | None | Same | | Method of<br>application | Manual application | Manual application | Same | | Compatible<br>endoscopes | Olympus, Pentax, FujiFlim<br>duodenoscopes | Linear echoendoscopes | Different<br>(Substantially<br>equivalent) | | Materials/<br>Biocompatibility | Silicone | Thermoplastic elastomer | Same | {12}------------------------------------------------ 510(k) Summary - Pancreaticobiliary Scope: Traditional 510(k) Premarket Notification Submission DragonflyTM Pancreaticobiliary Scope System with Instrument Channel Caps #### 8 PERFORMANCE DATA All necessary performance testing has been conducted on the Dragonfly Pancreaticobiliary Scope System and Instrument Channel Caps to assure substantial equivalence to the predicate devices and to demonstrate the device performs as intended. All testing was performed on test units representative of finished devices. The device passed the following tests: - . Biocompatibility testing per FDA Final Guidance Document, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" (2023) - . Sterilization validation per ANSI/AAMI/ISO 11135 - Packaging validation per ● ANSI/AAMI/ISO 11607-1 - Electrical Safety and electromagnetic (EMC) Testing per applicable requirements of IEC 60601-1, IEC 60601-1-2, & IEC 60601-2-18. - Simulated use testing, including use with ancillary devices - . Scope and controller Dimensional verification, including distal OD, working length, working channel length and ID - . Deflection verifications - . Image testing, including illumination verification, Field of View and direction measurement, camera function, camera light ingress and glare - HDMI compatibility ● - Image functionality and gain function ● - Sheath compatibility ● - Leak test - . Tensile tests - Optics testing, including Resolution, . Depth of Field, Field of View, Geometric distortion, Signal-to-Noise Ratio, Dynamic Range, Image intensity uniformity, color performance #### 9 CONCLUSION Upon reviewing the information provided in this submission and comparing the intended use, principle of operation and overall technological characteristics, the is substantially equivalent to existing legally marketed devices.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...