← Product Code [FBN](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FBN) · K231497

# Choledochoscope System (K231497)

_Shenzhen HugeMed Medical Technical Development Co., Ltd. · FBN · Mar 6, 2024 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FBN/K231497

## Device Facts

- **Applicant:** Shenzhen HugeMed Medical Technical Development Co., Ltd.
- **Product Code:** [FBN](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FBN.md)
- **Decision Date:** Mar 6, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Intended Use

Choledochoscope System is intended to be used by physicians through percutaneous insertion to access, visualize, and perform procedures in the pancreaticobiliary system including the hepatic ducts and the common bile duct. The image processor provides illumination for the choledochoscope, and is also used to receive the signal from the endoscope, convert it into an image and display it on the examination monitor. The instrument enables delivery and use of accessories such as biopsy forceps, laser fibers, graspers and retrieval baskets at a surgical site

## Device Story

System comprises single-use flexible choledochoscope (models CH-M32, CH-M40, CH-M50, CH-M52) and reusable image processor (HUV-01). Choledochoscope features on-tip CMOS camera and LED illumination; control body with articulation controls and accessory ports. Processor receives signals from endoscope, converts to image, and displays on monitor. Used by physicians in surgical settings for pancreaticobiliary procedures. Enables delivery of accessories (biopsy forceps, laser fibers, graspers, retrieval baskets). Benefits include visualization and diagnostic/therapeutic access to hepatic and common bile ducts.

## Clinical Evidence

No clinical data. Bench testing only. Includes optical performance (ISO 8600), color reproduction, geometric distortion, resolution, depth of field, SNR, dynamic range, frame frequency, system delay, and mechanical testing (flow rate, bending, leak testing). Biocompatibility (ISO 10993), electrical safety (IEC 60601-1, 60601-2-18), and EMC (IEC 60601-1-2) testing performed.

## Technological Characteristics

Single-use flexible choledochoscope; reusable image processor. CMOS imaging; LED illumination. Sterilization: EO (SAL 10^-6). Biocompatibility: ISO 10993-5, -10, -11. Electrical/EMC: ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-18. Shelf life: 3 years (ASTM F1980-16). Packaging: ISO 11607-1/2, ASTM F1886, F88, F1929.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- WiScope® OM Endoscope System ([K210579](/device/K210579.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged in a circular pattern. To the right of the seal, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" below. The logo is simple and professional, reflecting the FDA's role in regulating food and drugs.

March 6, 2024

Shenzhen HugeMed Medical Technical Development Co., Ltd % Yang Jie Consultant Shenzhen Chonconn Medical Device Consulting Co., Ltd. Room 504, Block C, No. 1029 Nanhai Avenue Nanshan District Shenzhen, Guangdong 518067 China

Re: K231497

Trade/Device Name: Choledochoscope System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: FBN, FET Dated: February 5, 2024 Received: February 5, 2024

Dear Yang Jie:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801: medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

# Shanil P. Haugen -S

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

510(k) Number (if known) K231497

Device Name Choledochoscope System

Choledochoscope System is intended to be used by physicians through percutaneous insertion to access, visualize, and perform procedures in the pancreaticobiliary system including the hepatic ducts and the common bile duct.

The image processor provides illumination for the choledochoscope, and is also used to receive the signal from the endoscope, convert it into an image and display it on the examination monitor.

The instrument enables delivery and use of accessories such as biopsy forceps, laser fibers, graspers and retrieval baskets at a surgical site

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

2

{4}------------------------------------------------

# 510(K) Summary

Prepared in accordance with the requirements of 21 CFR Part 807.92

### Prepared Date: 2023/05/05

### 1. Submission sponsor

Name: Shenzhen HugeMed Medical Technical Development Co., Ltd. Address: 415, 416-1,516-1, Building 2, No. 1, Mawu Road, Baoan Community, Yuanshan Street, Longgang District, Shenzhen, Guangdong, 518115, China Contact person: Cathy Title: Regulatory engineer E-mail: cathy.shi@hugemed.cn Tel: +86-755-22275833

#### 2. Submission correspondent

Name: Shenzhen Chonconn Medical Device Consulting Co., Ltd. Address: Room 504, Block C, No. 1029 Nanhai Avenue, Nanshan District, Shenzhen, Guangdong, P.R. China Contact person: Yang Jie E-mail: yangjie(@chonconn.com Tel: +86-755 33941160

| Trade/Device Name | Choledochoscope System                                                                |
|-------------------|---------------------------------------------------------------------------------------|
| Model             | single-use Choledochoscope: CH-M50, CH-M52, CH-M40, CH-M32<br>image processor: HUV-01 |
| Common Name       | Choledochoscope System                                                                |
| Regulatory Class  | Class II                                                                              |
| Classification    | 21CFR 876.1500 / Choledochoscope And<br>Accessories, Flexible/Rigid / FBN, FET        |
| Submission type   | Traditional 510(K)                                                                    |

#### 3. Subject Device Information

#### Predicate Device 4.

510(k) number: K210579 Product name: WiScope® OM Endoscope System Submitter: OTU Medical Inc.

#### 5. Device Description

The Choledochoscope System consists of a single-use choledochoscope to provide illumination and intuitive images in endoscopic surgery of the pancreatic ductal system and 

{5}------------------------------------------------

serves as a guide to diagnosis and management with accessories and an image processor for clinical image processing.

The single-use choledochoscope is a sterile single-use flexible choledochoscope and has the following 4 models:

| Models | Description of similarities and difference                       |
|--------|------------------------------------------------------------------|
| CHM32  | The four models are the same except for the difference in the    |
| CH-M40 | characteristics of the insertion tube, including diameter of the |
| CH-M50 | insertion tube, maximum insertion portion width and minimum      |
| CH-M52 | insertion channel width.                                         |

The image processor is a reusable monitor.

The single-use choledochoscope is comprised of a control body with articulation controls and accessory access ports, and a flexible insertion tube with an on-tip camera module and LED lighting source. The image processor processes the images from the choledochoscope and outputs video signals to a display.

#### 6. Intended use & Indication for use

Choledochoscope System is intended to be used by physicians through percutaneous insertion to access, visualize, and perform procedures in the pancreaticobiliary system including the hepatic ducts and the common bile duct.

The image processor provides illumination for the choledochoscope, and is also used to receive the signal from the endoscope, convert it into an image and display it on the examination monitor.

The instrument enables delivery and use of accessories such as biopsy forceps, laser fibers, graspers and retrieval baskets at a surgical site.

| Features                                                              | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Comparison |
|-----------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Trade name                                                            | Choledochoscope<br>System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | WiScope®OM Endoscope<br>System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | /          |
| 510(K) Submitter                                                      | Shenzhen<br>HugeMed<br>Medical<br>Technical<br>Development Co.,<br>Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | OTU Medical Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | /          |
| K number                                                              | K231497                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | K210579                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | /          |
| Classification<br>Regulation                                          | 21CRF 876.1500                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 21CRF 876.1500                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same       |
| Features                                                              | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Comparison |
| Classification<br>and Code                                            | Class II,<br>FBN, FET                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Class II,<br>FGB, FBN                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Different  |
| Common name                                                           | Choledochoscope<br>and Accessories,<br>Flexible/rigid                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Choledochoscope and<br>Accessories, Flexible/rigid                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same       |
| Indication for use                                                    | Choledochoscope<br>System is<br>intended to be<br>used by<br>physicians<br>through<br>percutaneous<br>insertion to<br>access, visualize,<br>and perform<br>procedures in the<br>pancreaticobiliary<br>system including<br>the hepatic ducts<br>and the common<br>bile duct.<br>The image<br>processor<br>provides<br>illumination for<br>the<br>choledochoscope,<br>and is also used to<br>receive the signal<br>from the<br>endoscope,<br>convert it into an<br>image and display<br>it on the<br>examination<br>monitor.<br>The instrument<br>enables delivery<br>and use of | WiScope® OM Endoscope<br>System is intended to be<br>used by physicians to access,<br>visualize, and perform<br>procedures in the urinary<br>tract and the kidney.<br>WiScope® OM Endoscope<br>System is also intended to be<br>used by physicians through<br>percutaneous insertion to<br>access, visualize, and<br>perform procedures in the<br>pancreaticobiliary system<br>including the hepatic ducts<br>and the common bile duct.<br>The instrument enables<br>delivery and use of<br>accessories such as biopsy<br>forceps, laser fibers, graspers<br>and retrieval baskets at a<br>surgical site. | Different  |
| Features                                                              | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Comparison |
|                                                                       | accessories such as biopsy forceps, laser fibers, graspers and retrieval baskets at a surgical site.                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |            |
| Item                                                                  | Choledochoscope System<br>- single-use<br>Choledochoscope<br>: CH-M50, CH-M52, CH-M40, CH-M32                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | WiScope OM Endoscope System<br>- WiScope Single-Use<br>Digital Flexible<br>Ureteroscope.<br>Choledochoscope                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |            |
| Choledochoscope                                                       | Single-Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Single-Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same       |
| Digital video<br>technology                                           | CMOS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | CMOS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same       |
| Illumination                                                          | LED                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | LED                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same       |
| Field of view<br>(degree)                                             | 120°±15%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 100°                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Different  |
| Working length<br>(mm)                                                | 605                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 670                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Different  |
| Outer Shaft<br>Diameter                                               | CH-M52:<br>$Φ$ 5.2±10%mm<br>CH-M50:<br>$Φ$ 5.0±10%mm<br>CH-M40:<br>$Φ$ 4.4±10%mm<br>CH-M32:<br>$Φ$ 3.2±10% mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 2.8mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Different  |
| Working Channel<br>Diameter                                           | CH-M52: $Φ$ ≥2.10 mm<br>CH-M50:<br>$Φ$ 2.8±10%mm<br>CH-M40: $Φ$ ≥1.1 mm<br>CH-M32:<br>$Φ$ 1.2±10%mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 1.2mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Different  |
| Up/Down                                                               | 210° upward and                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | UP: 275°                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Different  |
| Features                                                              | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Comparison |
| Deflection                                                            | 210° downward                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | DOWN: 275°                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same       |
| Direction of<br>view (degree)                                         | 0°±10°                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 0°                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same       |
| Camera Head<br>Configurable                                           | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Different  |
| Sterilization                                                         | EO<br>SAL: 10-6                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | EO<br>SAL: 10-6                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same       |
| Packaging                                                             | Choledochoscope<br>is packaged in a<br>tray which is<br>sealed by sterile<br>barrier.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Ureteroscope/Choledochoscope<br>is packaged in a tray<br>which is sealed by sterile<br>barrier.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same       |
| Label and<br>Labeling                                                 | Meet FDA's<br>Requirements                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Meet FDA's Requirements                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same       |
| Cytotoxicity                                                          | Comply with ISO<br>10993-5,<br>No cytotoxicity effect                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Comply with ISO 10993-5,<br>No cytotoxicity effect                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same       |
| Irritation                                                            | Comply with ISO<br>10993-10,<br>not an irritant                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Comply with ISO 10993-10,<br>not an irritant                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same       |
| Sensitization                                                         | Comply with ISO<br>10993-10,<br>not a sensitizer.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Comply with ISO 10993-10,<br>not a sensitizer.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same       |
| Acute systemic<br>toxicity &<br>Material-<br>Mediated<br>Pyrogenicity | Comply with ISO<br>10993-11,<br>no acute systemic<br>toxicity, no<br>temperature rise                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Comply with ISO 10993-11,<br>no acute systemic toxicity,<br>no temperature rise                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same       |
| Item                                                                  | Choledochoscope<br>System -Image<br>Processor HUV-<br>01                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | WiScope OM Endoscope System<br>-WiScope Image System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |            |
| Output Formats                                                        | USB/DVI/S-<br>Video/CVBS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | USB/AV/HDMI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Different  |
| Brightness<br>Control                                                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same       |
| White Balance                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same       |
| Image/Video<br>Capture                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Different  |
| Features                                                              | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Comparison |
| Separate monitor                                                      | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same       |
| Reusability                                                           | Reusable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Reusable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same       |
| Electrical                                                            | Comply with                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Comply with                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same       |
| Performance                                                           | Comply with<br>ANSI AAMI<br>ES60601-1<br>IEC60601-1-2<br>IEC 60601-2-18                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Comply with<br>ANSI AAMI ES60601-1<br>IEC60601-1-2<br>IEC 60601-2-18                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same       |

#### Comparison to the Predicate Device 7.

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

#### Performance Data 8.

The following performance data were provided in support of the substantial equivalence determination.

## Biocompatibility testing

Biocompatibility of the single-use choledochoscope was evaluated in accordance with the FDA guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA.

### Sterilization and shelf life testing

The Single-use choledochoscope is provided sterile and its shelf-life is 3 years. Sterilization Process has been validated accordance with ISO 11135:2014.

EO/ECH residual test was performed according to ISO 10993-7:2008.

The shelf life is determined based on optical testing and product performance testing after accelerated aging test according to ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices.

Package validation was conducted according to ISO 11607-1:2019 and ISO 11607-2:2019, and ASTM F1886/F1886M-16, ASTM F88/F88M-15, ASTM F 1929-15.

## Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the Choledochoscope System. The system complies with the IEC 60601-1 and IEC60601-2-18 for safety and the IEC 60601-1-2 for EMC.

### Software Verification and Validation Testing

Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern.

{10}------------------------------------------------

### Bench performance testing

The following bench tests were performed:

1. Optical performance testing according to ISO 8600 series.

2. Color performance (color reproduction), geometric distortion, optical performance (resolution, depth of field and image intensity uniformity), SNR and dynamic range, image frame frequency and system delay test compared with the predicate device.

3. Mechanical testing including Working Channel Flow Rate, bending and Insertion Portion Leak testing.

#### 9. Clinical study

No clinical study is included in this submission

### 10. Conclusion

Performance testing and compliance with voluntary standards demonstrate that the proposed subject device is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FBN/K231497](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FBN/K231497)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
