CHF-Y0005

K113120 · Olympus Medical Systems Corp. · FBN · Sep 25, 2012 · Gastroenterology, Urology

Device Facts

Record IDK113120
Device NameCHF-Y0005
ApplicantOlympus Medical Systems Corp.
Product CodeFBN · Gastroenterology, Urology
Decision DateSep 25, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment. This instrument is intended to be used without a DUODENOSCOPE. The CHOLEDOCHO VIDEOSCOPE CHF-Y0005 is indicated for endoscopy and endoscopic surgery within the biliary tract and the duodenum.

Device Story

Flexible video endoscope for biliary tract and duodenal endoscopy/surgery. Inputs: optical images captured via distal end. Transformation: transmits video signals to external Olympus video system center/monitor for visualization. Features: two bending sections for navigation; gas feeding function (via CO2 regulation unit) to aid insertion; Narrow Band Imaging (NBI) for tissue observation; compatibility with electrosurgical units/accessories. Used in clinical settings by physicians. Output: real-time endoscopic video feed. Clinical benefit: enables visualization and minimally invasive surgical access to biliary/duodenal anatomy; NBI enhances mucosal/vascular visualization; gas feeding improves procedural access.

Clinical Evidence

Bench testing only. Verification and performance testing conducted per IEC 60601-1, 60601-1-1, 60601-1-2, and 60601-2-18. Mechanical and functional performance evaluated against established in-house acceptance criteria based on ISO 14971:2007 risk analysis.

Technological Characteristics

Flexible video endoscope; two bending sections; gas feeding capability; Narrow Band Imaging (NBI) observation mode. Compatible with electrosurgical units. Standards: IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-2-18, ISO 14971, ASTM E1837-96, ANSI/AAMI/ISO 11135-1, ISO 10993-7. Sterilization: Ethylene oxide (implied by ISO 11135-1).

Indications for Use

Indicated for endoscopy and endoscopic surgery within the biliary tract and the duodenum. Intended for use with Olympus video system centers, light sources, monitors, and EndoTherapy accessories. Not for use with a duodenoscope.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K113120 page 1 of 4 SEP 25 2012 # 510(k) SUMMARY ### CHOLEDOCHO VIDEOSCOPE CHF-Y0005 ### General Information 1 . । - 트 Applicant: . - 트 Official Correspondent: OLYMPUS MEDICAL SYSTEMS CORP. 2951 Ishikawa-cho, Hachioji-shi, Tokyo, Japan 192-8507 Establishment Registration No: 8010047 Stacy Abbatiello Kluesner, M.S., RAC Regulatory Affairs & Quality Assurance Olympus America Inc. 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610, USA Phone: 484-896-5405 FAX: 484-896-7128 Email: stacy.kluesner@olympus.com Aizu Olympus Co., Ltd. 500 Aza-Muranishi, Ooaza-lidera, Monden-cho; Aizuwakamatsu-shi, Fukushima, Japan 965-8520 Establishment Registration No.: 9610595 #### Device Identification 2 - i Device Trade Name: Manufacturer: - Common Name: - Regulation Number: - - Regulation Name: - I Regulatory Class: - Classification Panel: . - Product Code: E CHF-Y0005 CHOLEDOCHO VIDEOSCOPE 21 CFR 876.1500 . Endoscope and accessories ll Choledochoscope And Accessories, Flexible Endoscope, Accessories, Narrow Band Spectrum FBN, NWB {2}------------------------------------------------ #### Predicate Device Information 3 | Subject Device<br>(Part of this Submission) | Predicate Device | PD's<br>510(k)<br>No. | |------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|-----------------------| | CHOLEDOCHO VIDEOSCOPE<br>CHF-Y0005<br>(Hereinafter referred<br>to as<br>CHF-Y0005) | VIDEOSCOPE OLYMPUS XCHF TYPE T160<br>(XCHF-T160) | K082576 | | | GASTROINTESTINALVIDEOSCOPE<br>XGIF-Q140M (XGIF-Q140M) | K032177 | | | CHOLEDOCHOSCOPE OLYMPUS XCHF<br>TYPE B180Y1 (XCHF-B180Y1) | K080586 | | | EVIS EXERA II GASTROINTESTINAL<br>VIDEOSCOPE OLYMPUS GIF TYPE N180<br>(GIF-N180) | K100584 | #### Device Description ব The CHF-Y0005 is designed as a flexible video endoscope used for endoscopy and endoscopic surgery within the biliary tract and the duodenum. The CHF-Y0005 is modified from the predicate XCHF TYPE T160 as it has two bending sections, a gas feeding function when connected to an Endoscopic CO2 regulation unit, Narrow Band Imaging observation and is now compatible with specified electrosurgical units and electrosurgical accessories. #### Indications for Use 5 This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment. This instrument is intended to be used without a DUODENOSCOPE. The CHOLEDOCHO VIDEOSCOPE CHF-Y0005 is indicated for endoscopy and endoscopic surgery within the biliary tract and the duodenum. ### Comparison of Technological Characteristics 6 The CHF-Y0005 is basically identical to the predicate devices in intended use, and similar in specifications. When compared to the predicate devices the subject device has the similar technological features such as the depth of field, direction of forward view and bending section angulation. The differences include two bending sections, a gas feeding function when connected to an {3}------------------------------------------------ Endoscopic CO2 regulation unit, Narrow Band Imaging observation and is now compatible with specified electrosurgical units and electrosurgical accessories. ### Substantially Equivalent Discussion 7 The indications for use, principles of operation and fundamental technology of the CHF-Y0005 are similar to the predicate devices. The major differences from the predicate devices are as follows: ### Bending Section: When compared with predicate XCHF-T160 device (K082576), the predicate XCHF-T160 device has one bending section with four bending directions (Up/Down/Right/Left). The predicate device XGIF-Q140M (K032177) also has two bending sections. The first bending section has four bending directions (Up/Down/Right/Left) while the second bending section passively bends in 4 directions (Up/Down/Left/Right). As the CHF-Y0005 has the same degree of movement as the predicate XCHF-T160. ### Gas Feeding: The gas feeding function was added to allow for insufflations with CO2 which aids in the insertion of the choleodocho scope through the GI tract. The predicate XCHF-T160 device (K082576) does not have a gas feeding function. However, the predicate XGIF-Q140M (K032177) does have a gas feeding function. ### Narrow Band Imaging: The predicate XCHF-B180Y1 device (K080586) which has similar indications for use to the subject device, was cleared with NBI observation mode. ### Compatibility with electrosurgical units: The distal end of the subject CHF-Y0005 is equivalent to the predicate XGIF-Q140M (K032177) and therefore is compatible with the specified electrosurgical units The above functions have been confirmed that the safety and effectiveness are equivalent as compared with the predicate devices. The CHF-Y0005 is similar or identical in the method of operation, materials and design as the predicate devices. ### Summary of Non-Clinical Testing 8 The following non-clinical tests were performed. Basic safety and performance testing was performed in accordance with IEC 60601-1, 60601-1-1, 60601-1-2, 60601-2-18. In addition, verification and comparison studies were conducted to evaluate the mechanical and functional performance. {4}------------------------------------------------ The following standards were used during the design and validation of the subject devices. 1) IEC 60601-1: 1988, A1: 1991, A2: 1995 2) IEC 60601-1-1: 2000, 3) IEC 60601-1-2: 2001, A1: 2004 4) IEC 60601-2-18: 1996, A1: 2000 5) ISO 14971: 2007 6) ASTM E1837-96: 2007 7) ANSI/AAMI/ISO 11135-1: 2007 8) ISO 10993-7: 2008 The risk analysis was carried out in accordance with established in-house acceptance criteria based on ISO 14971:2007. The design verifications tests and their acceptance criteria were identified and performed as a result of their risk analysis assessment. ### Conclusion 9 When compared to the predicate device, the subject CHF-Y0005 device does not incorporate . any significant changes in intended use, method of operation, material, or design that could affect the safety or effectiveness of the device. Therefore, the CHF-Y0005 is substantially equivalent to the identified predicate devices. {5}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" is arranged in a circular fashion around the symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 OLYMPUS MEDICAL SYSTEMS CORP. % Ms. Sheri Musgnung Associate Manager, Regulatory Affairs Olympus America Inc. 3500 Corporate Parkway, P.O. Box 610 CENTER VALLEY PA 18034-0610 SEP 25 2012 Re: K113120 Trade/Device Name: CHOLEDOCHO VIDEOSCOPE CHF-Y0005 Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FBN, NWB Dated: September 14, 2012 Received: September 17, 2012 Dear Ms. Musgnung: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be . found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {6}------------------------------------------------ device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set device-related adverse evones) (21 CFR Part 820); and if applicable, the electronic forth in the quality systems (QS) regardines (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please s and the many of the comments of any connect off and complised on the II you desire specific advice for your ac neves contracts offices/ucm115809.html for go to inter.har.gov/rtbound Drives Health's (CDRH's) Office of Compliance. Also, please the Center for Devices and Radioregiven on premarket notification' (21CFR Path note the regulation entitled, "Misoranding of references of promises 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the I ou may obtain other general mornational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address (600) 050-2011 of (2011 of (201) industry/default.html Sincerely yours, Benjamin K. Evans Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ ## Indications for Use 510(k) Number (if known): Device Name: CHOLEDOCHO VIDEOSCOPE CHF-Y0005 Indications For Use: This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment. This instrument is intended to be used without DUODENOSCOPE. The CHOLEDOCHO VIDEOSCOPE CHF-Y0005 is indicated for endoscopy and endoscopic surgery within the biliary tract and the duodenum. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use_ (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Aogui mrz Page 1 of ﺎ ﻟﻢ Castro-Rena
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...