← Product Code [FBK](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FBK) · K982890

# ADVANCED UROSCIENCE INJECTION NEEDLE (K982890)

_Advanced Uroscience, Inc. · FBK · Oct 26, 1998 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FBK/K982890

## Device Facts

- **Applicant:** Advanced Uroscience, Inc.
- **Product Code:** [FBK](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FBK.md)
- **Decision Date:** Oct 26, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Advanced UroScience Injection needle is an accessory for currently marketed endoscopes to allow delivery of injectable materials into tissues during an endoscopic procedure.

## Device Story

Device is an endoscopic injection needle; consists of stainless steel needle attached to plastic luer lock hub. Used as accessory for endoscopes; allows delivery of injectable materials (collagen, sclerosing agents, anesthetics, saline, contrast media) into tissues. Operated by clinicians during endoscopic procedures (cystoscopic, esophagoscopic, gastroscopic, laryngoscopic, colonoscopic). Needle lengths up to 200 cm; gauges 14-25. Provided sterile; single-use only. Standard syringe attaches to luer lock hub to facilitate injection. Benefits patient by enabling targeted delivery of therapeutic or diagnostic agents during minimally invasive endoscopic interventions.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Stainless steel needle; plastic luer lock hub. Needle lengths up to 200 cm; gauges 14-25. Manual operation via standard syringe. Sterile, single-use device.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Single Wall Introducer Needle, Ocean Medical Products ([K843719](/device/K843719.md))
- Martech Endoscopic Injection Needle, Martech Medical Products, Inc ([K960519](/device/K960519.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# OCT 2 6 1998

| 16. 510(k) SUMMARY  | K 982890                                                                                                                                              | Pg 1 of 3 |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| General Information |                                                                                                                                                       |           |
| Date Prepared       | August 14, 1998                                                                                                                                       |           |
| Classification      | Class II                                                                                                                                              |           |
| Trade Name          | Advanced UroScience Injection Needle                                                                                                                  |           |
| Common Name         | Endoscopic Needle                                                                                                                                     |           |
| Submitter           | Advanced UroScience<br>1290 Hammond Road<br>St. Paul, MN 55110<br>651-653-8512                                                                        |           |
| Contact             | Karen E. Peterson<br>Vice President of Regulatory, Clinical, and Quality Affairs                                                                      |           |
| Predicate Device    | Single Wall Introducer Needle,<br>Ocean Medical Products (K843719)<br>Martech Endoscopic Injection Needle,<br>Martech Medical Products, Inc (K960519) |           |

{1}------------------------------------------------

K982896 Pg 2 of 3

### Device Description

The Advanced UroScience Injection Needle consists of a stainless steel needle attached to a plastic luer lock hub. Needle length ranges up to 200 cm to accommodate the length of the endoscope channel and gauge ranges from 14 to 25. The needle length and gauge will be identified on the label. The luer lock hub, which is molded onto the needle, is designed to accommodate a standard syringe.

The Advanced UroScience Injection Needle is intended for use as an accessory for currently marketed endoscopes and provides delivery of injectable materials during an endoscopic procedure. The type of material to be injected is dependent on the nature of the procedure, but may include delivery of collagen during cystoscopic procedures, sclerosing agents during esophagoscopic and gastroscopic procedures, local anesthetics during cystoscopic or laryngoscopic procedures, or saline or contrast media during colonoscopic procedures.

The Advanced UroScience Injection Needle is provided sterile and is intended for single use only.

## Indication

The Advanced UroScience Injection Needle is an accessory for currently marketed endoscopes to allow delivery of injectable materials into tissues during an endoscopic procedure.

#### Technological Characteristics

The Advanced UroScience Injection Needle consists of a stainless steel needle and luer lock connector hub where a standard syringe can be attached for injection of materials through the lumen of the needle into tissue. Multiple needle lengths are available to accommodate the length of the endoscope channel.

{2}------------------------------------------------

$$\nwarrow \emptyset \lhd \lhd \emptyset^{\lozenge \hzenge \hzenge \hzenge \hzenge}$$

# Summary

In summary Advanced UroScience believes the above listed predicate devices and the Advanced UroScience Injection Needle are substantially equivalent based on design, materials, methods of fabrication and indications for use.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES" are arranged in a circle around the eagle. The eagle is depicted in black, and the text is also in black.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 26 1998

Ms. Karen E. Peterson Vice President of Regulatory, Clinical, and Quality Affairs Advanced UroScience, Inc. 1290 Hammond Road Saint Paul, Minnesota 55110

Re: K982890

Advanced UroScience Injection Needle Dated: August 14, 1998 Received: August 17, 1998 Regulatory Class: II 21 CFR 876.1500/Procode: 78 FBK

Dear Ms. Peterson:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/odrh/dsmaldsmamain.html".

Sincerely yours

Lillian Yin, Ph.D.

Director, Division of Reproductive, Abdominal, Ear, Nose and Throat and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K982890 510(k) Number (if known)

Advanced UroScience Injection Needle Device Name

Indications for Use

The Advanced UroScience Injection needle is an accessory for currently marketed endoscopes to allow delivery of injectable materials into tissues during an endoscopic procedure.

(PLEASE DO NOT WRITE BELOW THIS LINE—CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109

OR

Over the Counter Use **__**

(Optimal Format 1-2-96)

(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices

| <strong>Labels</strong> | <strong>Values</strong> |
|-------------------------|-------------------------|
| 510(k) Number           | K982890                 |

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FBK/K982890](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FBK/K982890)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
