# Disposable Endoscopic Injection Needles (K252021)

_Changzhou New Med Micro-Medtech Co., Ltd. · FBK · Dec 2, 2025 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FBK/K252021

## Device Facts

- **Applicant:** Changzhou New Med Micro-Medtech Co., Ltd.
- **Product Code:** [FBK](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FBK.md)
- **Decision Date:** Dec 2, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

This product is intended to be used in conjunction with an endoscope for the submucosal injection in the digestive tract.

## Device Story

Disposable Endoscopic Injection Needle; used with endoscope for submucosal injection in digestive tract. Device components: handle assembly, inner/outer tubes, introducer, needle tip. Operated by trained doctors/technicians in hospital settings. Device delivers therapeutic agents to submucosal tissue; facilitates endoscopic procedures. Benefits: enables targeted delivery of fluids/medications to GI tract tissue.

## Clinical Evidence

No clinical data. Bench testing only. Conformance demonstrated with ISO 10993 (biocompatibility), ASTM F1980 (shelf life), ASTM F88/F1929/F1886 (packaging integrity), and USP <85> (endotoxins).

## Technological Characteristics

Materials: Needle tip/introducer (06Cr19Ni10), inner/outer tubes (PTFE), handle (ABS). Dimensions: Working lengths 1200-2300mm; needle sizes 21G-25G; needle lengths 3-6mm. Sterilization: Ethylene Oxide (SAL 10^-6). Standards: ISO 10993, ASTM F1980, ASTM F88/88M, ASTM F1929, ASTM F1886, USP <85>.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Injection Needle ([K213914](/device/K213914.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

December 2, 2025

Changzhou New Med Micro-Medtech Co., Ltd.
% Kyra Kang
Official Correspondent
Landlink Healthcare Technology (Shanghai) Co., Ltd.
Room1308, Baohua International Plaza,
West Guangzhong Road 555, Jingan District
Shanghai, China

Re: K252021
Trade/Device Name: Disposable Endoscopic Injection Needles
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope And Accessories
Regulatory Class: Class II
Product Code: FBK
Dated: October 27, 2025
Received: October 27, 2025

Dear Kyra Kang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K252021 - Kyra Kang
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K252021 - Kyra Kang
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Shanil P. Haugen -S

Shanil P. Haugen, Ph.D.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity and Transplant Devices
OHT3: Office of GastroRenal, ObGyn,
General Hospital and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Submission Number (if known)
K252021

Device Name
Disposable Endoscopic Injection Needles

Indications for Use (Describe)
This product is intended to be used in conjunction with an endoscope for the submucosal injection in the digestive tract.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

&gt; "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}

K252021
Page 1 of 4

510(k) Summary

I. Submitter
Changzhou New Med Micro-Medtech Co., Ltd.
No. 99 of Longxi Avenue, Zhulin Town, Jintan District, Changzhou City, Jiangsu Province, China
Contact person: Ashe Chen
Position: manager
Tel.: 86-18352577510
E-mail: ashe.chen@newmed-medical.com
Preparation date: March 18, 2025

Submission Correspondent
Ms. Kyra Kang
Landlink Healthcare Technology (Shanghai) Co., Ltd.
E-mail: kyra.kang@landlink-health.com

II. Proposed Device
Device Trade Name: Disposable Endoscopic Injection Needles
Common name: Endoscope and accessories
Regulation Number: 21 CFR 876.1500
Regulatory Class: Class II
Product code: FBK
Review Panel: Gastroenterology/Urology

III. Predicate Devices
510(k) Number: K213914
Trade name: Injection Needle
Common name: Endoscope and accessories
Classification: Class II
Product Code: FBK
Manufacturer: Jiangsu Vedkang Medical Science and Technology Co., Ltd.

Page 1 of 4

{5}

K252021
Page 2 of 4

## IV. Device description

The Disposable Endoscopic Injection Needle is intended to be used in conjunction with an endoscope for the submucosal injection in the digestive tract.

This device consists of the handle assembly, inner and outer tubes, introducer, and needle tip assembly.

The proposed devices are sterilized by Ethylene Oxide Gas to achieve a SAL of 10-6 and supplied sterility maintenance package which could maintain the sterility of the device during the shelf life of 3 years.

## V. Indication for use

This product is intended to be used in conjunction with an endoscope for the submucosal injection in the digestive tract.

## VI. Comparison of technological characteristics with the predicate devices

Table 1 General Comparison of Disposable Endoscopic Injection Needles

|  Characteristics | Proposed device | Predicate device (K213914) | Discussion  |
| --- | --- | --- | --- |
|  Product | Disposable Endoscopic Injection Needles | Injection Needle | /  |
|  Product Code | FBK | FBK | Same  |
|  Regulation Number | 21 CFR 876.1500 | 21 CFR 876.1500 | Same  |
|  Class | Class II | Class II | Same  |
|  Indication for Use | This product is intended to be used in conjunction with an endoscope for the submucosal injection in the digestive tract. | The device is intended to be used with an endoscope to perform endoscopic vascular or submucosal injection in the GI tract. | Similar  |
|  Configuration | handle assembly, inner and outer tubes, introducer, and needle tip | needle, connector tube, guiding head, inner tube, outer tube, protective sleeve, front handle, injection handle, front handle cover, boosting tube | Similar  |
|  Environment of use | Hospital | Hospital | Same  |
|  Intended users | The device must be used by trained doctors or technicians | The device must be used by trained doctors or technicians | Same  |
|  Single Use | Single Use | Single Use | Same  |
|  Label/Labeling | Complied with 21 CFR part 801 | Complied with 21 CFR part 801 | Same  |
|  Working Length | 1200mm,1600mm, 1800mm, 2300mm | 1200mm,1800mm,2300mm, 3000mm | Similar  |

Page 2 of 4

{6}

K252021
Page 3 of 4

|  Minimum working channel | 2.0mm, 2.8mm | 2.0mm, 2.8mm | Same  |
| --- | --- | --- | --- |
|  Needle Size | 21G, 23G, 25G | 19G, 20G, 21G, 22G, 23G, 25G | Same (covered by Predicate device)  |
|  Needle Length | 3mm, 4mm, 5mm, 6mm | 3mm, 4mm, 5mm, 6mm, 8mm | Same (covered by Predicate device)  |
|  Patient contact material | Needle tip:06Cr19Ni10
Introducer:06Cr19Ni10
Inner tube:PTFE
Outer Tube: PTFE
Handle Assembly: ABS | Needle: S30400
Guiding Head: S30300
Inner Tube: PTFE or PP
Outer Tube: PTFE or PP
Injection Handle: ABS
Boosting Tube: S30400 | Similar  |
|  Biocompatibility | Comply with the ISO 10993 Standards | Comply with the ISO 10993 Standards | Same  |
|  Sterilization Method | EO Sterilized | EO Sterilized | Same  |
|  SAL | 10-6 | 10-6 | Same  |
|  Endotoxin Limit | 20 EU per device | 20 EU per device | Same  |

VII. Non-Clinical Testing

The device described in this summary was tested and demonstrated to be in conformance with the following standards:

ISO 10993-4:2017
ISO 10993-5:2009
ISO 10993-7:2008
ISO 10993-10:2021
ISO 10993-11:2017
ISO 10993-23:2021
ASTM F1980-16
ASTM F88/88M-15
ASTM F1929-15
ASTM F1886-16
USP &lt;85&gt; Bacterial Endotoxins Test

Page 3 of 4

{7}

K252021
Page 4 of 4

## VIII. Clinical Testing

No clinical study is included in this submission.

## IX. Conclusion

The proposed device has the same indication for use and has similar design features and technological characteristic as the predicate device. Performance testing data demonstrates that the proposed device is safety and effectiveness as the predicated device. Accordingly, the proposed device is substantially equivalent to the predicate device.

Page 4 of 4

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FBK/K252021](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FBK/K252021)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).