← Product Code [FBK](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FBK) · K232200

# Disposable Injection Needles (K232200)

_Ningbo Xinwell Medical Technology Co., Ltd. · FBK · Apr 12, 2024 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FBK/K232200

## Device Facts

- **Applicant:** Ningbo Xinwell Medical Technology Co., Ltd.
- **Product Code:** [FBK](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FBK.md)
- **Decision Date:** Apr 12, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The device is intended to be used with an endoscope to perform endoscopic vascular or submucosal injection in the GI tract.

## Device Story

Disposable injection needle used with endoscope for vascular or submucosal injection in GI tract. Device consists of needle tube, inner/outer tubes, protective sleeve, booster tube, handle, limitation ring, buckle, and conical connector. Operator inserts device through endoscopic channel to target lesion; pushes conical connector to extend needle tube from outer tube; inserts needle into tissue; connects syringe to Luer interface for fluid injection. After injection, operator retracts needle into outer tube and withdraws device. Used in hospital settings by trained clinicians. Provides targeted delivery of therapeutic agents to GI tissue, facilitating endoscopic procedures.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing including biocompatibility (cytotoxicity, irritation, sensitization, systemic toxicity, hemocompatibility) and performance testing (tensile strength, fracture force, patency, injection residue, needle retraction, and connector standards).

## Technological Characteristics

Materials: 06Cr19Ni10 stainless steel (needle/booster tubes), PP (inner/outer tubes), ABS (handle/connector). Sensing/actuation: Mechanical manual extension/retraction. Energy: None (manual). Dimensions: Outer tube 1.8mm/2.3mm; working length 1200-2300mm; needle 21G-25G. Connectivity: None. Sterilization: Ethylene oxide. Standards: ISO 10993, ISO 20695, ISO 9626, ISO 13402, ISO 80369-7, ISO 7864, ASTM F756, F2382, F2888.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Injection Needle ([K213914](/device/K213914.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

April 12, 2024

Ningbo Xinwell Medical Technology Co., Ltd Lei Xi Registration Manager No.188 Binjiang Road, Cixi High-tech Industrial Development Zone Cixi, Zhejiang 315301 China

Re: K232200

Trade/Device Name: Disposable Injection Needles Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FBK Dated: March 12, 2024 Received: March 12, 2024

Dear Lei Xi:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Sivakami Venkatachalam -S

for

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

Submission Number (if known)

K232200

Device Name

### Disposable Injection Needles

Indications for Use (Describe)

The device is intended to be used with an endoscope to perform endoscopic vascular or submucosal injection in the GI tract.

Type of Use (Select one or both, as applicable)

< | Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

#### 1. SUBMITTER

Ningbo Xinwell Medical Technology Co., LTD

No.188 Binjiang Road, Cixi High-tech industrial Development Zone, Cixi City, 315301 Zhejiang,P.R.China

Contact Person: Lei Xi

Title: RA manager

Tel: +86574-63026388

Email: xilei@xinhaigroup.com

Summary Prepared: 2024-04-04

#### 2. DEVICE

Name of device: Disposable Injection Needles Regulation Number: 21 CFR 876.1500 Classification name: Endoscope and accessories Common name:Endoscopic injection needle Regulatory Class: Class II Product Code: FBK

#### 3. PREDICATE DEVICE

The predicate device is Injection Needle(K213914).

#### 4. DEVICE DESCRIPTION

The device is intended to be used with an endoscope to perform endoscopic vascular or submucosal injection in the GI tract.The Disposable Injection Needles are consist of needle tube, inner tube, outer tube, protective sleeve, booster tube, handle, limitation ring, buckle and conical connector. The needle tube and booster tube are made of 06Cr19Ni10. The inner tube are made of PP.The handle, limitation ring, buckle and connector are made of ABS. The diameter of the outer tube is available in two specifications, respectively are 1.8mm and 2.3mm. There are five specifications for the working length of the outer tube, respectively are 1200mm,1800mm,2000mm and 2300mm. There are three specifications for outer diameters of needle tube, respectively are 21G, 23G and 25G. Five specifications for lengths of needle tube, respectively are 4mm, 5mm, 7mm and 8mm.

The device is sterilized by ethylene oxide. The shelf life is 3 years.

The inner, middle and transportation packaging ways respectively are sealed pouch, box and carton.The corresponding product quantities is 1 pc per bag, 10 pcs per box and 5 boxes per carton.

5. PRINCIPLE OF OPERATION

{5}------------------------------------------------

The Disposable Injection Needles achieve its intended use by combining them with the flexible channel of the endoscope. Firstly, the Disposable Injection Needles passed through the endoscopic channel to the channel port, and the lesion was determined through the endoscopic visual field. After the lesion was determined, the needle tube of the Disposable Injection Needles was extended out of the outer tube by pushing the conical connector at the end of the handle, and then inserted into the target tissue. And the Luer interface at the end of the handle is connected to the syringe for fluid injection; After the injection is completed, pull the conical connector at the handle to retrieve the needle tube into the outer tube, and then withdraw the Disposable Injection Needles from the endoscopic channel.

- 6. Available model

| Specification and Model | Model Code | Outer tube diameter (mm) | Working length (mm) | Outer diameter of needle tube (G) | Length of needle tube (mm) |
|-------------------------|------------|--------------------------|---------------------|-----------------------------------|----------------------------|
| XY-ZSZ-C-18×1200×21×4   | C          | 1.8                      | 1200                | 21                                | 4                          |
| XY-ZSZ-C-18×1200×21×5   | C          | 1.8                      | 1200                | 21                                | 5                          |
| XY-ZSZ-C-18×1200×21×6   | C          | 1.8                      | 1200                | 21                                | 6                          |
| XY-ZSZ-C-18×1200×21×7   | C          | 1.8                      | 1200                | 21                                | 7                          |
| XY-ZSZ-C-18×1200×21×8   | C          | 1.8                      | 1200                | 21                                | 8                          |
| XY-ZSZ-C-18×1200×23×4   | C          | 1.8                      | 1200                | 23                                | 4                          |
| XY-ZSZ-C-18×1200×23×5   | C          | 1.8                      | 1200                | 23                                | 5                          |
| XY-ZSZ-C-18×1200×23×6   | C          | 1.8                      | 1200                | 23                                | 6                          |
| XY-ZSZ-C-18×1200×23×7   | C          | 1.8                      | 1200                | 23                                | 7                          |
| XY-ZSZ-C-18×1200×23×8   | C          | 1.8                      | 1200                | 23                                | 8                          |
| XY-ZSZ-C-18×1200×25×4   | C          | 1.8                      | 1200                | 25                                | 4                          |
| XY-ZSZ-C-18×1200×25×5   | C          | 1.8                      | 1200                | 25                                | 5                          |
| XY-ZSZ-C-18×1200×25×6   | C          | 1.8                      | 1200                | 25                                | 6                          |
| XY-ZSZ-C-18×1200×25×7   | C          | 1.8                      | 1200                | 25                                | 7                          |
| XY-ZSZ-C-18×1200×25×8   | C          | 1.8                      | 1200                | 25                                | 8                          |
| XY-ZSZ-C-18×1600×21×4   | C          | 1.8                      | 1600                | 21                                | 4                          |
| XY-ZSZ-C-18×1600×21×5   | C          | 1.8                      | 1600                | 21                                | 5                          |
| XY-ZSZ-C-18×1600×21×6   | C          | 1.8                      | 1600                | 21                                | 6                          |
| XY-ZSZ-C-18×1600×21×7   | C          | 1.8                      | 1600                | 21                                | 7                          |
| XY-ZSZ-C-18×1600×21×8   | C          | 1.8                      | 1600                | 21                                | 8                          |
| XY-ZSZ-C-18×1600×23×4   | C          | 1.8                      | 1600                | 23                                | 4                          |
| XY-ZSZ-C-18×1600×23×5   | C          | 1.8                      | 1600                | 23                                | 5                          |
| XY-ZSZ-C-18×1600×23×6   | C          | 1.8                      | 1600                | 23                                | 6                          |
| XY-ZSZ-C-18×1600×23×7   | C          | 1.8                      | 1600                | 23                                | 7                          |
| XY-ZSZ-C-18×1600×23×8   | C          | 1.8                      | 1600                | 23                                | 8                          |
| XY-ZSZ-C-18×1600×25×4   | C          | 1.8                      | 1600                | 25                                | 4                          |
| XY-ZSZ-C-18×1600×25×5   | C          | 1.8                      | 1600                | 25                                | 5                          |
| XY-ZSZ-C-18×1600×25×6   | C          | 1.8                      | 1600                | 25                                | 6                          |
| XY-ZSZ-C-18×1600×25×7   | C          | 1.8                      | 1600                | 25                                | 7                          |
| XY-ZSZ-C-18×1600×25×8   | C          | 1.8                      | 1600                | 25                                | 8                          |
| XY-ZSZ-C-18×1800×21×4   | C          | 1.8                      | 1800                | 21                                | 4                          |
| XY-ZSZ-C-18×1800×21×5   | C          | 1.8                      | 1800                | 21                                | 5                          |
| XY-ZSZ-C-18×1800×21×6   | C          | 1.8                      | 1800                | 21                                | 6                          |
| XY-ZSZ-C-18×1800×21×7   | C          | 1.8                      | 1800                | 21                                | 7                          |
| XY-ZSZ-C-18×1800×21×8   | C          | 1.8                      | 1800                | 21                                | 8                          |
| XY-ZSZ-C-18×1800×23×4   | C          | 1.8                      | 1800                | 23                                | 4                          |
| XY-ZSZ-C-18×1800×23×5   | C          | 1.8                      | 1800                | 23                                | 5                          |
| XY-ZSZ-C-18×1800×23×6   | C          | 1.8                      | 1800                | 23                                | 6                          |
| XY-ZSZ-C-18×1800×23×7   | C          | 1.8                      | 1800                | 23                                | 7                          |
| XY-ZSZ-C-18×1800×23×8   | C          | 1.8                      | 1800                | 23                                | 8                          |
| XY-ZSZ-C-18×1800×25×4   | C          | 1.8                      | 1800                | 25                                | 4                          |
| XY-ZSZ-C-18×1800×25×5   | C          | 1.8                      | 1800                | 25                                | 5                          |
| XY-ZSZ-C-18×1800×25×6   | C          | 1.8                      | 1800                | 25                                | 6                          |
| XY-ZSZ-C-18×1800×25×7   | C          | 1.8                      | 1800                | 25                                | 7                          |
| XY-ZSZ-C-18×1800×25×8   | C          | 1.8                      | 1800                | 25                                | 8                          |
| XY-ZSZ-C-18×2000×21×4   | C          | 1.8                      | 2000                | 21                                | 4                          |
| XY-ZSZ-C-18×2000×21×5   | C          | 1.8                      | 2000                | 21                                | 5                          |
| XY-ZSZ-C-18×2000×21×6   | C          | 1.8                      | 2000                | 21                                | 6                          |
| XY-ZSZ-C-18×2000×21×7   | C          | 1.8                      | 2000                | 21                                | 7                          |
| XY-ZSZ-C-18×2000×21×8   | C          | 1.8                      | 2000                | 21                                | 8                          |
| XY-ZSZ-C-18×2000×23×4   | C          | 1.8                      | 2000                | 23                                | 4                          |
| XY-ZSZ-C-18×2000×23×5   | C          | 1.8                      | 2000                | 23                                | 5                          |
| XY-ZSZ-C-18×2000×23×6   | C          | 1.8                      | 2000                | 23                                | 6                          |
| XY-ZSZ-C-18×2000×23×7   | C          | 1.8                      | 2000                | 23                                | 7                          |
| XY-ZSZ-C-18×2000×23×8   | C          | 1.8                      | 2000                | 23                                | 8                          |
| XY-ZSZ-C-18×2000×25×4   | C          | 1.8                      | 2000                | 25                                | 4                          |
| XY-ZSZ-C-18×2000×25×5   | C          | 1.8                      | 2000                | 25                                | 5                          |
| XY-ZSZ-C-18×2000×25×6   | C          | 1.8                      | 2000                | 25                                | 6                          |
| XY-ZSZ-C-18×2000×25×7   | C          | 1.8                      | 2000                | 25                                | 7                          |
| XY-ZSZ-C-18×2000×25×8   | C          | 1.8                      | 2000                | 25                                | 8                          |
| XY-ZSZ-C-18×2300×21×4   | C          | 1.8                      | 2300                | 21                                | 4                          |
| XY-ZSZ-C-18×2300×21×5   | C          | 1.8                      | 2300                | 21                                | 5                          |
| XY-ZSZ-C-18×2300×21×6   | C          | 1.8                      | 2300                | 21                                | 6                          |
| XY-ZSZ-C-18×2300×21×7   | C          | 1.8                      | 2300                | 21                                | 7                          |
| XY-ZSZ-C-18×2300×21×8   | C          | 1.8                      | 2300                | 21                                | 8                          |
| XY-ZSZ-C-18×2300×23×4   | C          | 1.8                      | 2300                | 23                                | 4                          |
| XY-ZSZ-C-18×2300×23×5   | C          | 1.8                      | 2300                | 23                                | 5                          |
| XY-ZSZ-C-18×2300×23×6   | C          | 1.8                      | 2300                | 23                                | 6                          |
| XY-ZSZ-C-18×2300×23×7   | C          | 1.8                      | 2300                | 23                                | 7                          |
| XY-ZSZ-C-18×2300×23×8   | C          | 1.8                      | 2300                | 23                                | 8                          |
| XY-ZSZ-C-18×2300×25×4   | C          | 1.8                      | 2300                | 25                                | 4                          |
| XY-ZSZ-C-18×2300×25×5   | C          | 1.8                      | 2300                | 25                                | 5                          |
| XY-ZSZ-C-18×2300×25×6   | C          | 1.8                      | 2300                | 25                                | 6                          |
| XY-ZSZ-C-18×2300×25×7   | C          | 1.8                      | 2300                | 25                                | 7                          |
| XY-ZSZ-C-18×2300×25×8   | C          | 1.8                      | 2300                | 25                                | 8                          |
| XY-ZSZ-C-23×1200×21×4   | C          | 2.3                      | 1200                | 21                                | 4                          |
| XY-ZSZ-C-23×1200×21×5   | C          | 2.3                      | 1200                | 21                                | 5                          |
| XY-ZSZ-C-23×1200×21×6   | C          | 2.3                      | 1200                | 21                                | 6                          |
| XY-ZSZ-C-23×1200×21×7   | C          | 2.3                      | 1200                | 21                                | 7                          |
| XY-ZSZ-C-23×1200×21×8   | C          | 2.3                      | 1200                | 21                                | 8                          |
| XY-ZSZ-C-23×1200×23×4   | C          | 2.3                      | 1200                | 23                                | 4                          |
| XY-ZSZ-C-23×1200×23×5   | C          | 2.3                      | 1200                | 23                                | 5                          |
| XY-ZSZ-C-23×1200×23×6   | C          | 2.3                      | 1200                | 23                                | 6                          |
| XY-ZSZ-C-23×1200×23×7   | C          | 2.3                      | 1200                | 23                                | 7                          |
| XY-ZSZ-C-23×1200×23×8   | C          | 2.3                      | 1200                | 23                                | 8                          |
| XY-ZSZ-C-23×1200×25×4   | C          | 2.3                      | 1200                | 25                                | 4                          |
| XY-ZSZ-C-23×1200×25×5   | C          | 2.3                      | 1200                | 25                                | 5                          |
| XY-ZSZ-C-23×1200×25×6   | C          | 2.3                      | 1200                | 25                                | 6                          |
| XY-ZSZ-C-23×1200×25×7   | C          | 2.3                      | 1200                | 25                                | 7                          |
| XY-ZSZ-C-23×1200×25×8   | C          | 2.3                      | 1200                | 25                                | 8                          |
| XY-ZSZ-C-23×1600×21×4   | C          | 2.3                      | 1600                | 21                                | 4                          |
| XY-ZSZ-C-23×1600×21×5   | C          | 2.3                      | 1600                | 21                                | 5                          |
| XY-ZSZ-C-23×1600×21×6   | C          | 2.3                      | 1600                | 21                                | 6                          |
| XY-ZSZ-C-23×1600×21×7   | C          | 2.3                      | 1600                | 21                                | 7                          |
| XY-ZSZ-C-23×1600×21×8   | C          | 2.3                      | 1600                | 21                                | 8                          |
| XY-ZSZ-C-23×1600×23×4   | C          | 2.3                      | 1600                | 23                                | 4                          |
| XY-ZSZ-C-23×1600×23×5   | C          | 2.3                      | 1600                | 23                                | 5                          |
| XY-ZSZ-C-23×1600×23×6   | C          | 2.3                      | 1600                | 23                                | 6                          |
| XY-ZSZ-C-23×1600×23×7   | C          | 2.3                      | 1600                | 23                                | 7                          |
| XY-ZSZ-C-23×1600×23×8   | C          | 2.3                      | 1600                | 23                                | 8                          |
| XY-ZSZ-C-23×1600×25×4   | C          | 2.3                      | 1600                | 25                                | 4                          |
| XY-ZSZ-C-23×1600×25×5   | C          | 2.3                      | 1600                | 25                                | 5                          |
| XY-ZSZ-C-23×1600×25×6   | C          | 2.3                      | 1600                | 25                                | 6                          |
| XY-ZSZ-C-23×1600×25×7   | C          | 2.3                      | 1600                | 25                                | 7                          |
| XY-ZSZ-C-23×1600×25×8   | C          | 2.3                      | 1600                | 25                                | 8                          |
| XY-ZSZ-C-23×1800×21×4   | C          | 2.3                      | 1800                | 21                                | 4                          |
| XY-ZSZ-C-23×1800×21×5   | C          | 2.3                      | 1800                | 21                                | 5                          |
| XY-ZSZ-C-23×1800×21×6   | C          | 2.3                      | 1800                | 21                                | 6                          |
| XY-ZSZ-C-23×1800×21×7   | C          | 2.3                      | 1800                | 21                                | 7                          |
| XY-ZSZ-C-23×1800×21×8   | C          | 2.3                      | 1800                | 21                                | 8                          |
| XY-ZSZ-C-23×1800×23×4   | C          | 2.3                      | 1800                | 23                                | 4                          |
| XY-ZSZ-C-23×1800×23×5   | C          | 2.3                      | 1800                | 23                                | 5                          |
| XY-ZSZ-C-23×1800×23×6   | C          | 2.3                      | 1800                | 23                                | 6                          |
| XY-ZSZ-C-23×1800×23×7   | C          | 2.3                      | 1800                | 23                                | 7                          |
| XY-ZSZ-C-23×1800×23×8   | C          | 2.3                      | 1800                | 23                                | 8                          |
| XY-ZSZ-C-23×1800×25×4   | C          | 2.3                      | 1800                | 25                                | 4                          |
| XY-ZSZ-C-23×1800×25×5   | C          | 2.3                      | 1800                | 25                                | 5                          |
| XY-ZSZ-C-23×1800×25×6   | C          | 2.3                      | 1800                | 25                                | 6                          |
| XY-ZSZ-C-23×1800×25×7   | C          | 2.3                      | 1800                | 25                                | 7                          |
| XY-ZSZ-C-23×1800×25×8   | C          | 2.3                      | 1800                | 25                                | 8                          |
| XY-ZSZ-C-23×2000×21×4   | C          | 2.3                      | 2000                | 21                                | 4                          |
| XY-ZSZ-C-23×2000×21×5   | C          | 2.3                      | 2000                | 21                                | 5                          |
| XY-ZSZ-C-23×2000×21×6   | C          | 2.3                      | 2000                | 21                                | 6                          |
| XY-ZSZ-C-23×2000×21×7   | C          | 2.3                      | 2000                | 21                                | 7                          |
| XY-ZSZ-C-23×2000×21×8   | C          | 2.3                      | 2000                | 21                                | 8                          |
| XY-ZSZ-C-23×2000×23×4   | C          | 2.3                      | 2000                | 23                                | 4                          |
| XY-ZSZ-C-23×2000×23×5   | C          | 2.3                      | 2000                | 23                                | 5                          |
| XY-ZSZ-C-23×2000×23×6   | C          | 2.3                      | 2000                | 23                                | 6                          |
| XY-ZSZ-C-23×2000×23×7   | C          | 2.3                      | 2000                | 23                                | 7                          |
| XY-ZSZ-C-23×2000×23×8   | C          | 2.3                      | 2000                | 23                                | 8                          |
| XY-ZSZ-C-23×2000×25×4   | C          | 2.3                      | 2000                | 25                                | 4                          |
| XY-ZSZ-C-23×2000×25×5   | C          | 2.3                      | 2000                | 25                                | 5                          |
| XY-ZSZ-C-23×2000×25×6   | C          | 2.3                      | 2000                | 25                                | 6                          |
| XY-ZSZ-C-23×2000×25×7   | C          | 2.3                      | 2000                | 25                                | 7                          |
| XY-ZSZ-C-23×2000×25×8   | C          | 2.3                      | 2000                | 25                                | 8                          |
| XY-ZSZ-C-23×2300×21×4   | C          | 2.3                      | 2300                | 21                                | 4                          |
|                         |            |                          |                     |                                   |                            |
| XY-ZSZ-C-23×2300×21×5   | C          | 2.3                      | 2300                | 21                                | 5                          |
| XY-ZSZ-C-23×2300×21×6   | C          | 2.3                      | 2300                | 21                                | 6                          |
| XY-ZSZ-C-23×2300×21×7   | C          | 2.3                      | 2300                | 21                                | 7                          |
| XY-ZSZ-C-23×2300×21×8   | C          | 2.3                      | 2300                | 21                                | 8                          |
| XY-ZSZ-C-23×2300×23×4   | C          | 2.3                      | 2300                | 23                                | 4                          |
| XY-ZSZ-C-23×2300×23×5   | C          | 2.3                      | 2300                | 23                                | 5                          |
| XY-ZSZ-C-23×2300×23×6   | C          | 2.3                      | 2300                | 23                                | 6                          |
| XY-ZSZ-C-23×2300×23×7   | C          | 2.3                      | 2300                | 23                                | 7                          |
| XY-ZSZ-C-23×2300×23×8   | C          | 2.3                      | 2300                | 23                                | 8                          |
| XY-ZSZ-C-23×2300×25×4   | C          | 2.3                      | 2300                | 25                                | 4                          |
| XY-ZSZ-C-23×2300×25×5   | C          | 2.3                      | 2300                | 25                                | 5                          |
| XY-ZSZ-C-23×2300×25×6   | C          | 2.3                      | 2300                | 25                                | 6                          |
| XY-ZSZ-C-23×2300×25×7   | C          | 2.3                      | 2300                | 25                                | 7                          |
| XY-ZSZ-C-23×2300×25×8   | C          | 2.3                      | 2300                | 25                                | 8                          |

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

### 7. Indication for use

The device is intended to be used with an endoscope to perform endoscopic vascular or submucosal injection in the GI tract.

- 8. Comparison of technological characteristics with the predicate device
The Disposable Injection Needles are compared with the predicate device Injection Needles(K213914).The results are shown below in the Technological Characteristics Comparison Table:

| ITEM                           | Proposed Device                                                                                                                                          | Predicate Device<br>K213914                                                                                                                                           | Remark     |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Product                        | Disposable Injection<br>Needles                                                                                                                          | Injection Needle                                                                                                                                                      | /          |
| Product Code                   | FBK                                                                                                                                                      | FBK                                                                                                                                                                   | Same       |
| Regulation Number              | 21 CFR 876.1500                                                                                                                                          | 21 CFR 876.1500                                                                                                                                                       | Same       |
| Class                          | Class II                                                                                                                                                 | Class II                                                                                                                                                              | Same       |
| Indication for Use             | The device is intended<br>to be used with an<br>endoscope to perform<br>endoscopic vascular or<br>submucosal injection in<br>the GI tract.               | The device is intended<br>to be used with an<br>endoscope to perform<br>endoscopic vascular or<br>submucosal injection in<br>the GI tract.                            | Same       |
|                                |                                                                                                                                                          |                                                                                                                                                                       |            |
| Configuration                  | Needle tube, Inner tube,<br>Outer tube, Protective<br>sleeve, Handle,<br>Limitation ring, Buckle,<br>Conical connector,<br>Booster tube.                 | needle, connector tube,<br>guiding head, inner<br>tube, outer tube,<br>protective sleeve, front<br>handle, injection<br>handle, front handle<br>cover, boosting tube. | Difference |
| Environment of use             | Hospital                                                                                                                                                 | Hospital                                                                                                                                                              | Same       |
| Intended users                 | The device must be<br>used by trained doctors<br>or technicians.                                                                                         | The device must be<br>used by trained doctors<br>or technicians.                                                                                                      | Same       |
| Single Use                     | Single Use                                                                                                                                               | Single Use                                                                                                                                                            | Same       |
| Label/Labeling                 | Complied with 21 CFR<br>part 801                                                                                                                         | Complied with 21 CFR<br>part 801                                                                                                                                      | Same       |
| Specifications                 |                                                                                                                                                          |                                                                                                                                                                       |            |
| Working Length                 | 1200mm,1600mm,<br>1800mm,2000mm<br>2300mm                                                                                                                | 1200mm, 1800mm,<br>2300mm, 3000mm                                                                                                                                     | Difference |
| Minimum working<br>channel     | 2.0mm, 2.8mm                                                                                                                                             | 2.0mm, 2.8mm                                                                                                                                                          | Same       |
| Needle Size                    | 21G, 23G, 25G                                                                                                                                            | 19G, 20G, 21G, 22G,<br>23G, 25G                                                                                                                                       | Difference |
| Needle Length                  | 4mm, 5mm, 6mm,<br>7mm,                                                                                                                                   | 3mm, 4mm, 5mm,<br>6mm, 8mm                                                                                                                                            | Difference |
| Patient<br>contact<br>material | Needle tube :Stainless<br>steel (06Cr19Ni10)<br>Inner tube :PP<br>Outer tube:PP<br>Booster tube:Stainless<br>steel (06Cr19Ni10)<br>Conical connector:ABS | Needle: S30400<br>Guiding Head: S30300<br>Inner Tube: PTFE or PP<br>Outer Tube: PTFE or PP<br>Injection Handle: ABS<br>Boosting Tube: S30400                          | Difference |
| Biocompatibility               |                                                                                                                                                          |                                                                                                                                                                       |            |
| Cytotoxicity                   | No cytotoxicity                                                                                                                                          | No cytotoxicity                                                                                                                                                       | Same       |
| Irritation                     | No intracutaneous<br>reactivity                                                                                                                          | No intracutaneous<br>reactivity                                                                                                                                       | Same       |
| Sensitization                  | No skin sensitization                                                                                                                                    | No skin sensitization                                                                                                                                                 | Same       |
| Systemic Toxicity              | No systemic toxicity                                                                                                                                     | No systemic toxicity                                                                                                                                                  | Same       |
| Pyrogen                        | No Pyrogen                                                                                                                                               | No Pyrogen                                                                                                                                                            | Same       |

{10}------------------------------------------------

{11}------------------------------------------------

| Hemolysis Properties | No Hemolysis     | No Hemolysis     | Same |  |  |
|----------------------|------------------|------------------|------|--|--|
|                      |                  |                  |      |  |  |
|                      |                  |                  |      |  |  |
| Sterilization        |                  |                  |      |  |  |
| Method               | EO Sterilized    | EO Sterilized    | Same |  |  |
| SAL                  | 10-e             | 10-6             | Same |  |  |
| Endotoxin Limit      | 20 EU per device | 20 EU per device | Same |  |  |

#### 9. Performance data

### Non-clinical performance test conclusion

### Biocompatibility

Biocompatibility testing was performed on the proposed device in accordance with the following standards:

ISO 10993-5:2009 Biological evaluation of medical devices — Part 5: Tests for in vitro cytotoxicity

ISO 10993-4:2017 Biological evaluation of medical devices — Part 4: Selection of tests for interactions with blood

ISO 10993-10:2021 Biological evaluation of medical devices — Part 10: Tests for skin sensitization

ISO 10993-11:2017 Biological evaluation of medical devices — Part 11: Tests for systemic toxicity

ISO 10993-23:2021 Biological evaluation of medical devices — Part 23: Tests for irritation

ASTM F756-17 Standard Practice for Assessment of Hemolytic Properties of Materials

ASTM F2382-18 Standard Test Method for Assessment of Circulating Blood-Contacting Medical Device Materials on Partial Thromboplastin Time (PTT)

ASTM F2888-19 Standard Practice for Platelet Leukocyte Count--An In-Vitro Measure for

Hemocompatibility Assessment of Cardiovascular Materials

### Performance bench testing

The basic safety and essential performance comparison test were evaluated based on as following standards:

ISO 20695:2020 Enteral feeding systems-Design and testing (tensile strength, fracture force).

ISO 9626:2016 Stainless steel needle tubing for the manufacture of medical devices-Requirements and test methods.

ISO 13402:1995 Surgical and dental hand instruments-Determination of resistance against autoclaving, corrosion and thermal exposure.

ISO 80369-7:2021 Small-bore connectors for liquids and gases in healthcare applications-Part 7: Connectors for intravascular or hypodermic applications.

ISO 7864:2016 Sterile hypodermic needles for single use-Requirements and test methods

Testing for patency, injection residue and needle retraction suitability was performed.

{12}------------------------------------------------

### Clinical test conclusion

No clinical study is included in this submission.

#### 10. Conclusion

The conclusions drawn from the non-clinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as the legally marketed predicate device Injection Needle (K213914).

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FBK/K232200](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FBK/K232200)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
