Clear-Jet Injection Catheter

K181690 · Finemedix Co., Ltd. · FBK · Dec 21, 2018 · Gastroenterology, Urology

Device Facts

Record IDK181690
Device NameClear-Jet Injection Catheter
ApplicantFinemedix Co., Ltd.
Product CodeFBK · Gastroenterology, Urology
Decision DateDec 21, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Clear-Jet Injection Catheter is to be used in conjunction with an endoscopic injections, such as the treatment of esophageal and gastric varies and for submucosal dye marking in the GI tract.

Device Story

Clear-Jet Injection Catheter is a manual endoscopic accessory used to deliver therapeutic solutions into tissue. The device consists of a stainless-steel needle, inner tube, catheter tube, handle, and slider mechanism. A standard syringe attaches to the slider to inject fluids through the needle lumen. The device is inserted through the working channel of an endoscope by a clinician (e.g., gastroenterologist) during endoscopic procedures. By enabling precise submucosal injection, the device facilitates the treatment of esophageal and gastric varices and assists in tissue marking. It is supplied sterile for single-patient use and is not intended for reuse or re-sterilization.

Clinical Evidence

Bench testing only. Testing included biocompatibility (ISO 10993-1, -5, -10, -11), physical/chemical safety (pH, heavy metals, non-volatile residue), performance testing (appearance, dimensions), EO sterilization validation (ISO 11737), and shelf-life testing (ASTM F1980).

Technological Characteristics

Materials: Stainless-steel needle, thermoplastic-PTFE polymer catheter tube. Dimensions: 21G/23G needles; 4mm/6mm lengths; 1800mm/2300mm working lengths; 2.35mm outer sheath diameter. Manual actuation via syringe/slider. Supplied sterile (EO).

Indications for Use

Indicated for patients requiring endoscopic injections, including treatment of esophageal and gastric varices and submucosal dye marking in the GI tract.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue font. December 21, 2018 Finemedix Co. Ltd. % April Lee Consultant Withus Group Inc. 106 Superior Irvine, CA 92620 Re: K181690 > Trade/Device Name: Clear-Jet Injection Catheter Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FBK Dated: November 13, 2018 Received: November 20, 2018 Dear April Lee: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {2}------------------------------------------------ You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Daniel G. Walter Jr -S for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K181690 Device Name Clear-Jet Injection Catheter Indications for Use (Describe) The Clear-Jet Injection Catheter is to be used in conjunction with an endoscopic injections, such as the treatment of esophageal and gastric varies and for submucosal dye marking in the GI tract. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for FINEMEDIX. The logo is in blue and has an orange circle above the letter 'E'. The text is in all caps and is a sans-serif font. ## FINEMEDIX CO., LTD. 60, Maeyeo-ro, Dong-gu, Daegu, Republic of Korea (41065) #### 510(k) Summary #### Submitter FINEMEDIX CO., LTD. Hee Kyung Kwon 60, Maeyeo-ro, Dong-gu, Daegu 41065 South Korea Email: hkyung1230@finemedix.com Tel. +82-53-741-8388 Fax. +82-53-741-8168 #### Official Correspondent Withus Group Inc April Lee 106 Superior. Irvine, CA 92620 USA Email: withus6664@gmail.com Phone: 1-909-274-9971 Fax: 1-909-460-8122 #### Device Information - Trade Name: Clear-Jet Injection Catheter ● - Common Name: Endoscopic Injection Needle ● - Classification Name: Endoscopic injection needle, gastroenterology-urology ● - . Product Code: FBK - Panel: Gastroenterology/Urology - Regulation Number: 21 CFR 876.1500 Endoscope and Accessories - Device Class: Class II - . Date Prepared: 06/20/2018 #### Predicate Devices: The subject device is substantially equivalent to the following predicate devices: - K 150434, Injection Needle manufactured by Micro-Tech (Nanjing) Co., Ltd. #### Indication for Use: The Clear-Jet Injection Catheter is to be used in conjunction with an endoscope to perform endoscopic injections, such as the treatment of esophageal and gastric varies and for submucosal dye marking in the GI tract. #### Device Description: Clear-Jet Injection Catheter consists of a stainless-steel needle attached to inner tube, an Injector Head attached to Catheter tube, and Slider where a standard syringe can be attached for injection of solutions through the lumen of the needle into tissue. It is available in various sizes and working lengths. The needle sizes are 21 gauge (outer diameter 0.8mm) and 23 gauge (outer diameter 0.6mm), and the needle lengths are 4.0mm and 6.0mm. The needle wall thicknesses are 0.1mm, 0.05mm, and the Catheter Tube length (=Working length) includes 1,800mm, 2,300mm. There are two types of Clear-Jet Injection Catheter such as Head type and Non-head type has the Injector Head made of stainless steel alloys attached to the Catheter Tube, Non -head type does not have the Injector Head made of stainless steel alloys. {5}------------------------------------------------ T Image /page/4/Picture/1 description: The image shows the word "FINEMEDIX" in a bold, sans-serif font. The letters are a dark blue color. Above the letters "ME" is a small orange circle. This device passes through the working channel of endoscope, and the average contact time with mucosa of the human digestive tract is less than 1 hour. This device is supplied sterile for single-patient use and shall be not reused or re-sterilized. #### Summary of Technological Characteristics: | | | Subject Device | Primary Predicate | Substantial<br>Equivalence | |-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------| | Company | | Finemedix Co., Ltd. | Micro-Tech Co., Ltd. | - | | Device Name | | Clear-Jet Injection Catheter | Injection Needle | - | | 510(k) Number | | NA | K150434 | - | | Device<br>Classification Name | | Endoscopic injection needle,<br>gastroenterology-urology | Endoscopic injection needle,<br>gastroenterology-urology | me | | Product Code | | FBK | FBK | Same | | Regulation Number | | 876.1500 | 876.1500 | Same | | Indications for Use | | The Clear-Jet Injection<br>Catheter is to be used in<br>conjunction with an<br>endoscope to perform<br>endoscopic injections, such as<br>the treatment of esophageal<br>submucosal dye marking in<br>the GI tract | The Injection Needle is to be<br>used in conjunction with an<br>endoscope to perform<br>endoscopic injections, such as<br>the treatment of esophageal<br>Sa<br>and gastric varies, and for<br>and gastric varies and for<br>submucosal dye marking in<br>the GI tract. | me | | Material | Needle<br>Tube | Stainless-Steel<br>Thermoplastic-PTFE Polymer | Stainless-Steel<br>Thermoplastic-PTFE Polymer | Same<br>Same | | Components | | Needle, Inner Tube, Catheter<br>Tube, Handle, Slider | Needle, Innersheath, outer<br>shell, outer sheath and luer<br>Simi | lar | | Gauge size | | 21,23 Gauge | 19,22,25 Gauge | Within the range | | Supplied Sterile | | Yes | Yes | Same | | Outer Sheath<br>Diameter | 2.35mm | 2.3mm | Similar | | | Working Length | 1800, 2300 mm | 1800, 2000, 2300 mm | Within the range | | | Needle Length | 4, 6 mm | 4,5,6 mm | Within the range | | | Packaging | Single-use EO Sterilized<br>Pouch with one device per<br>pouch | Single-use EO Sterilized<br>Pouch with one device per<br>pouch | Same | | | Single Use | Yes | Yes | Same | | | Similarities | Both subject and predicate device are substantially equivalent in the indications for use,<br>Material, Components, Design, sterility, outer sheath diameter, working length, needle length,<br>and packaging as charted. | | | | | Differences | The differences between the subject and predicate devices are gauge size and outer sheath<br>diameter. The subject device has 21 and 23 gauge and the predicate device has 19,22,25 gauge.<br>But the subject device gauges are contained within the range of the predicate's gauges, so it<br>does not affect the safety and effectiveness.<br>The other difference is outer sheath diameter. The outer sheath diameter of the subject device is<br>2.35mm and the outer sheath diameter of the predicate device is 2.3mm. But this size difference<br>doesn't affect the safety and effectiveness during surgery.<br>Any differences in technology characteristics are accompanied by information that<br>demonstrated the device is substantially equivalent as the predicate and do not raise different<br>questions of safety and effectiveness than the predicate. | | | | {6}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the word "FINEMEDIX" in a bold, sans-serif font. The letters are a dark blue color. Above the "E" in "FINEMEDIX" is a small orange circle. ## FINEMEDIX CO., LTD. 60, Maeyeo-ro, Dong-gu, Daegu, Republic of Korea (41065) {7}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the logo for FINEMEDIX. The logo is in purple and has an orange circle above the letters 'ME'. The text is bold and sans-serif. #### Non-clinical testing data: The subject device was tested to evaluate its substantial equivalence according to the following standards. - Biocompatibility testing according to ISO 10993-1:2009, ISO 10993-5:2009, ISO 10993-10:2010 . and ISO 10993-11:2006 - . Performance testing such as appearance and dimension - Physical and Chemical Safety Testing such as color and transparent, pH, KMnO4 Consumption, . Non-Volatile residue, ultraviolet absorption, heavy metal as lead, EO Sterilization residuals test, and sterility test - EO Sterilization Testing according to ISO 11737-1:2006 and ISO 11737-2:2009 - Shelf Life Testing according to ASTM F1980 ● The biocompatibility evaluation for Clear-Jet Injection Catheter was conducted in accordance with ISO 10993-1: 2009 "Biological Evaluation of Medical Devices –Part 1: Evaluation and Testing within a Risk Management Process" and FDA's biocompatibility guidance, G95-1 Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" (May 1, 1995). The following tests were completed: Cytotoxicity, Sensitization, Intracutaneous reactivity and Acute Systemic Toxicity. Performance testing such as appearance and dimension and extractable testing such as color and transparent, pH, KMnO4 Consumption, Non-Volatile residue, ultraviolet absorption, and heavy metal as lead were performed as per Finemedix's design control system. The non-clinical testing results demonstrate that the subject device is substantially equivalent to the predicate device. #### Conclusion: Clear-Jet Injection Catheter constitutes a substantially equivalent medical device, meeting all the declared requirements of its intended use. This system has the same intended use and fundamental scientific technology as its predicate device. Therefore, Clear-Jet Injection Catheter and its predicate are substantially equivalent.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...