← Product Code [FBK](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FBK) · K022484

# COOK INJECTION NEEDLE (K022484)

_Cook Urological, Inc. · FBK · Oct 25, 2002 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FBK/K022484

## Device Facts

- **Applicant:** Cook Urological, Inc.
- **Product Code:** [FBK](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FBK.md)
- **Decision Date:** Oct 25, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Cook Injection Needles are used to deliver a variety of injectable materials into tissues during laparoscopic, hysteroscopic, cystoscopic transurethral procedures and open surgical procedures. The type of material to be injected will be dependent on the nature of the procedure. The needle is intended to be used with legally marketed drugs and devices.

## Device Story

Cook Injection Needles consist of a polyurethane catheter with a stainless steel needle tip and luer lock hub. Catheter sizes range from 3.7 French to 9 French; lengths from 15cm to 65cm; needle gauges from 16G to 25G. Used by physicians during endoscopic or open surgical procedures to deliver injectable materials into tissues. Device functions as a conduit for clinician-administered substances. Benefits include facilitating targeted delivery of therapeutic or diagnostic agents during minimally invasive or open surgery.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on design, material, and intended use similarities to predicate devices.

## Technological Characteristics

Polyurethane catheter; stainless steel needle tip; luer lock hub. Catheter sizes: 3.7F to 9F; lengths: 15cm to 65cm. Needle gauges: 16G to 25G. Manual operation; no energy source or software.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Advanced UroScience Injection Needles ([K982890](/device/K982890.md))
- Martech Endoscopic Needle ([K960519](/device/K960519.md))
- Genyx Medical Injection Needles ([K990996](/device/K990996.md))
- Tunis Transurethral Injection/Aspiration System Probe Devices ([K983765](/device/K983765.md))
- Vance Cystoscopic Injection Needle ([K812057](/device/K812057.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K 022484

510(k) Premarket Notification Cook Iniection Needles Cook Urological, Inc.

510(k) SUMMARY I.

Submitted By:

Debbie Schmitt Cook Urological 1100 West Morgan Street Spencer, Indiana 47460 (812) 829-4891

Date: July 26, 2002

Device

Cook Injection Needles Trade Name: Proposed Classification Name: Class II Needle, Endoscopic 78 FBK Specifically, the Cook Injection Needles are similar to The Predicate Devices: Advanced UroScience Injection Needles (K#982890), The Martech Endoscopic Needle (K#960519), The Genyx Medical Injection Needles (K#990996), The Tunis Transurethral Injection/Aspiration System Probe Devices (K#983765) and The Vance Cystoscopic Injection Needle (K#812057). The Cook Injection Needle will consist of a Polyurethane Device Description: Catheter with a Stainless Steel Needle Tip and a luer lock hub. The Catheter sizes range from 3.7 French to 9 French and will be 15cm to 65cm's in length. The Stainless Steel Needle portion will be 16 Gage though 25 Gage. These Devices will be manufactured according to specified Substantial Equivalence: controls and a Ouality Assurance Program. The devices will undergo packaging similar to the devices currently marketed and distributed by Cook Urological. Being similar with respect to indications for use, materials and physical construction to predicate devices, these devices meet the requirements for section 510(k) substantial equivalence.

7

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle with three human profiles within the eagle's body.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

## OCT 2 5 2002

Ms. Debbie Schmitt Regulatory Affairs Manager Cook® Urological 1100 W. Morgan Street SPENCER IN 46460

Re: K022484

Trade/Device Name: Cook Injection Needles Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: 78 FBK Dated: July 26, 2002 Received: July 29, 2002

Dear Ms. Schmitt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy Chrogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Page _________________________________________________________________________________________________________________________________________________________________________ of

510(k) Number (if known): 人 0 22 48 4

Device Name: Cook Injection Needles

Indications For Use: The Cook Injection Needles are used to deliver a variety of injectable materials into tissues during laparoscopic, hysteroscopic, cystoscopic transurethral procedures and open surgical procedures. The type of material to be injected will be dependent on the nature of the procedure. The needle is intended to be used with legally marketed drugs and devices.

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use L/

OR

Over-the-Counter______________________________________________________________________________________________________________________________________________________________

David A. Depson

(Division Sign-Off Division of Reproductiv and Radiological Devic 510(k) Number

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FBK/K022484](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FBK/K022484)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
