# Flexible Video Cysto-Urethroscope (C-view) (K202957)

_KARL STORZ Endoscopy-America, Inc. · FAJ · Oct 29, 2020 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FAJ/K202957

## Device Facts

- **Applicant:** KARL STORZ Endoscopy-America, Inc.
- **Product Code:** [FAJ](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FAJ.md)
- **Decision Date:** Oct 29, 2020
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Flexible Video Cysto-Urethroscope C-VIEW is used to provide visualization and operative access during diagnostic and therapeutic endoscopic procedures of urinary tract including the urethra, bladder, ureters, and kidneys.

## Device Story

Flexible video cysto-urethroscope (C-VIEW) provides visualization and operative access in urinary tract; used for diagnostic and therapeutic procedures. Device requires connection to compatible Camera Control Units (CCUs) (C-MAC or C-HUB II) to function; scope captures video signals via internal CMOS imager and LED light source; CCUs decode and display video for clinician. Used in clinical settings by physicians. Output allows real-time visualization of urinary anatomy to guide operative instruments through a 2.3 mm working channel. Benefits include minimally invasive access to urethra, bladder, ureters, and kidneys for diagnosis and treatment.

## Clinical Evidence

No clinical data. Substantial equivalence established through non-clinical bench testing, including electrical safety (IEC 60601-1, IEC 60601-2-18), EMC (IEC 60601-1-2), biocompatibility (ISO 10993), and optical performance testing (color reproduction, illumination uniformity, spatial resolution, depth of field, latency, distortion, SNR, and sensitivity).

## Technological Characteristics

Flexible endoscope with 5.2 mm shaft diameter, 37 cm length, and 2.3 mm working channel. Features CMOS imager, internal LED light source, and 210° up/140° down deflection. Connectivity via proprietary video signal to C-MAC or C-HUB II CCUs. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, IEC 62471, ISO 10993, and ISO 8600.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Flexible HD Cysto-Urethroscope ([K191357](/device/K191357.md))
- Flexible CMOS-Video-Cysto-Urethroscope ([K131364](/device/K131364.md))

## Reference Devices

- C-MAC ([K182186](/device/K182186.md))
- C-HUB II ([K182186](/device/K182186.md))
- 11272VHU ([K182723](/device/K182723.md))
- 11272VU ([K131364](/device/K131364.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

October 29, 2020

KARL STORZ Endoscopy-America, Inc Winkie Wong Regulatory Affairs Manager 2151 E. Grand Avenue EI Segundo, CA 90245

Re: K202957 Trade/Device Name: Flexible Video Cysto-Urethroscope (C-view) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FAJ, FBO Dated: September 29, 2020 Received: September 30, 2020

Dear Winkie Wong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Mark Antonino Acting Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

K202957

Device Name Flexible Video Cysto-Urethroscope (C-view)

Indications for Use (Describe)

The Flexible Video Cysto-Urethroscope C-VIEW is used to provide visualization and operative access during diagnostic and therapeutic endoscopic procedures of urinary tract including the urethra, bladder, ureters, and kidneys.

| Type of Use (Select one or both, as applicable)                                                 |
|-------------------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label>             |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Karl Storz Endoskope. The logo is in blue and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the text "KARL STORZ - ENDOSKOPE" in a smaller font.

## 510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KSEA's knowledge.

| Applicant:                            | KARL STORZ Endoscopy-America, Inc<br>2151 E. Grand Avenue<br>El Segundo, CA 90245                                                                                                                                                                                                                                     |                                                                             |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| Contact:                              | Winkie Wong<br>Regulatory Affairs Manager<br>310-658-3427 (phone)                                                                                                                                                                                                                                                     |                                                                             |
| Date of<br>Preparation:               | October 20th, 2020                                                                                                                                                                                                                                                                                                    |                                                                             |
| Type of 510(k)<br>Submission:         | Special                                                                                                                                                                                                                                                                                                               |                                                                             |
| Device<br>Identification:             | Trade Name:                                                                                                                                                                                                                                                                                                           | Flexible Video Cysto-Urethroscope (C-view)                                  |
|                                       | Common Name:                                                                                                                                                                                                                                                                                                          | Cystoscope and accessories, Flexible/Rigid (FAJ)<br>Cystourethroscope (FBO) |
| Product Code:                         | FAJ, FBO                                                                                                                                                                                                                                                                                                              |                                                                             |
| Classification<br>Number and<br>Name: | 21 CFR 876.1500 (Endoscope and Accessories)                                                                                                                                                                                                                                                                           |                                                                             |
| Predicate<br>Device(s):               | Flexible HD Cysto-Urethroscope (K191357) – Primary<br>Flexible CMOS-Video-Cysto-Urethroscope (K131364) – Secondary<br>**The above predicate devices have not been subject to any recall**                                                                                                                             |                                                                             |
| Device<br>Description:                | The Flexible Video Cysto-Urethroscope C-VIEW (Part Number: 11272VUE) is<br>intended to be used with multiple compatible CCUs: C-MAC (8403ZX, cleared via<br>K182186) and C-HUB II (20290301, cleared via K182186). Identical to the<br>predicates, the scope cannot be operated on its own because it is a videoscope |                                                                             |

{4}------------------------------------------------

|                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                              | K202957<br>Page 2 of 4                       |                                               |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|----------------------------------------------|-----------------------------------------------|
| whose image data output is provided in the form of video signals and sent to the CCU<br>for decoding and display. Therefore, when the scope is connected to the compatible<br>CCUs, it becomes the Flexible Video Cysto-Urethroscope C-VIEW System, which<br>provides visualization and operative access in the urinary tract. |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                              |                                              |                                               |
| Indications for<br>use:                                                                                                                                                                                                                                                                                                        | The Flexible Video Cysto-Urethroscope C-VIEW is used to provide visualization and<br>operative access during diagnostic and therapeutic endoscopic procedures of<br>urinary tract including the urethra, bladder, ureters, and kidneys.<br>The indications for use are identical to the primary predicate device. The subject                                                                                                                                                             |                                              |                                              |                                               |
|                                                                                                                                                                                                                                                                                                                                | device includes kidneys when compared to the secondary predicate. This difference<br>does not have different intended use because the kidney is part of the urinary tract.                                                                                                                                                                                                                                                                                                                |                                              |                                              |                                               |
| Technological<br>Characteristics:                                                                                                                                                                                                                                                                                              | The subject device is a modification of the primary predicate, 11272VHU, cleared via<br>K182723 and most recently via K191357 for the addition of HLD for reprocessing. It<br>is also the next generation of the secondary predicate device, 11272VU, cleared via<br>K131364 as lower cost alternative imaging systems when compared to the primary<br>predicate.                                                                                                                         |                                              |                                              |                                               |
|                                                                                                                                                                                                                                                                                                                                | The comparison of technological characteristics is as follow:                                                                                                                                                                                                                                                                                                                                                                                                                             |                                              |                                              |                                               |
|                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Subject Device                               | Primary Predicate<br>K191357                 | Secondary Predicate<br>K131364                |
|                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Flexible Video Cysto-<br>Urethroscope System | Flexible HD Cysto-<br>Urethroscope<br>System | Flexible CMOS<br>Video Cysto<br>Uretheroscope |
|                                                                                                                                                                                                                                                                                                                                | Type of scope                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Flexible                                     | Same as the subject<br>device                | Same as the subject<br>device                 |
|                                                                                                                                                                                                                                                                                                                                | Insertion Shaft Diameter                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 5.2 mm                                       | 5.5 mm                                       | 5.5 mm                                        |
|                                                                                                                                                                                                                                                                                                                                | Insertion Shaft Length                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 37 cm                                        | Same as the subject<br>device                | Same as the subject<br>device                 |
|                                                                                                                                                                                                                                                                                                                                | Working Channel<br>Diameter                                                                                                                                                                                                                                                                                                                                                                                                                                                               | 2.3 mm                                       | 2.3 mm                                       | 2.2 mm                                        |
|                                                                                                                                                                                                                                                                                                                                | Suction Port/Channel                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Via working channel                          | Yes                                          | Via working channel                           |
|                                                                                                                                                                                                                                                                                                                                | Deflection (°)                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Up: 210°<br>Down: 140°                       | Same as the subject<br>device                | Same as the subject<br>device                 |
|                                                                                                                                                                                                                                                                                                                                | (C1) Type of Imager                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | CMOS                                         | CMOS                                         | CMOS                                          |
|                                                                                                                                                                                                                                                                                                                                | (C2) Field of View                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 100°                                         | Same as the subject<br>device                | Same as the subject<br>device                 |
|                                                                                                                                                                                                                                                                                                                                | (C3) Direction of View                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 0°                                           | Same as the subject<br>device                | Same as the subject<br>device                 |
|                                                                                                                                                                                                                                                                                                                                | (C4) Depth of Field                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 5-50 mm                                      | Same as the subject<br>device                | Same as the subject<br>device                 |
|                                                                                                                                                                                                                                                                                                                                | (C5) On-axis Resolution<br>(minimal)                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 1.8 lp/mm @ 5 mm<br>16 lp/mm @50 mm          | 2.5 lp/mm @3 mm<br>40 lp /mm@50 mm           | 1.1 lp/mm @ 5 mm<br>11 lp/mm @ 50mm           |
|                                                                                                                                                                                                                                                                                                                                | (C6) Light Source                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Internal LED                                 | Same as the subject<br>device                | Same as the subject<br>device                 |
|                                                                                                                                                                                                                                                                                                                                | (C7) Compatible CCUs                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | C-MAC (8403ZX)<br>C-HUB II (P/N20290301)     | IMAGE1 S                                     | C-MAC (8402ZX)                                |
| Non-Clinical<br>Performance<br>Data:                                                                                                                                                                                                                                                                                           | There are no performance standards or special controls developed under Section<br>514 of the FD&C Act for endoscopes. However, the subject device follows the FDA<br>recognized consensus standards and is tested according to the following standards<br>and FDA Guidance:                                                                                                                                                                                                               |                                              |                                              |                                               |
|                                                                                                                                                                                                                                                                                                                                | Electrical Safety and EMC  IEC 60601-1 IEC 60601-1-2 IEC 60601-2-18 IEC 62471 ISO 10993 ISO 8600   FDA Guidance Document  Reprocessing Medical Devices in Health Care Settings: Validation<br>Methods and Labeling   Optical Performance Testing  Color Reproduction and Color Contrast Illumination Detection Uniformity Spatial Resolution & Depth of Field Latency Distortion Signal-to-Noise Ratio (SNR) & Sensitivity                                                                |                                              |                                              |                                               |
|                                                                                                                                                                                                                                                                                                                                | Additional bench testing was performed to ensure the device met its design<br>specifications. The bench testing performed verified and validated that the Flexible<br>Video Cysto-Urethroscope (C-view) has met all its design specification and is<br>substantially equivalent to its predicate devices.                                                                                                                                                                                 |                                              |                                              |                                               |
| Substantial<br>Equivalence:                                                                                                                                                                                                                                                                                                    | The indications for use, operating principles, technological characteristics and<br>features are similar, if not identical, between that subject device and the predicate<br>devices, Flexible HD Cysto-Ureteroscope (K191357) and CMOS Flexible Video<br>Endoscope (K131364). The minor difference between the subject and predicate<br>devices that does not raise new or different questions or safety and effectiveness are<br>listed above in Technological Characteristics section. |                                              |                                              |                                               |
|                                                                                                                                                                                                                                                                                                                                | As proven by the comparisons, the differences do not raise different questions of<br>safety and effectiveness because the intended use, operating principles.                                                                                                                                                                                                                                                                                                                             |                                              |                                              |                                               |

{5}------------------------------------------------

{6}------------------------------------------------

|                                  | technological characteristics, and features are similar, if not identical. Both systems             also comply with identical standards and safety testing, where applicable.                                                                                 |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                  | Substantial equivalence on the effectiveness of the subject device is supported by the             comparison of the images and standard image quality characteristics including, but             not limited to the performance testing listed above.         |
| Clinical<br>Performance<br>Data: | Clinical performance is not required to demonstrate substantial equivalence to the             predicate devices. Non-clinical bench testing was sufficient to establish substantial             equivalence.                                                  |
| Conclusion:                      | The Flexible Video Cysto-Urethroscope C-VIEW is substantially equivalent to its             predicate device. The non-clinical bench and comparative testing demonstrate that             the device is as safe and effective as the legally marketed devices. |

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FAJ/K202957](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FAJ/K202957)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).