The malleable penile prosthesis is indicated for implantation into the corpora cavernosa of the penis in men who are diagnosed with erectile dysfunction. The prosthesis is implanted to provide adequate penile rigidity for sexual intercourse. The prosthesis is designed for the treatment of organic erectile dysfunction in men due to: Pelvic fracture, Spinal cord injury or disease, Prostatectomy, Multiple sclerosis, Diabetes mellitus, Arteriosclerosis, Hypertensive vascular disease, Priapism, Peyronie's disease, Selectively for psychogenic impotence.
Device Story
Rigi10 Malleable Penile Prosthesis is a sterile, single-use surgical implant for treating erectile dysfunction. Device consists of two cylinders implanted into corpora cavernosa; each cylinder features a malleable stainless-steel core, titanium end caps, PTFE tube cover, and silicone outer surface. Urologists or surgeons perform implantation. Device provides manual rigidity for intercourse and allows concealment under clothing. Extenders are provided to adjust length to patient anatomy. No electronic components or software; purely mechanical implant.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing and biocompatibility evaluation. Bench tests included simulated life/fatigue, positioning/concealability, buckling, and tip extender joint strength. Biocompatibility testing followed ISO 10993-1, including cytotoxicity, sensitization, reactivity, systemic toxicity, pyrogen, endotoxin, implantation, chemical characterization, and hemolysis tests. All tests met established acceptance criteria.
Technological Characteristics
Implantable penile prosthesis; materials include silicone elastomer, stainless-steel core, titanium caps, and PTFE cover. Dimensions: 23-25 cm length, 9-13 mm diameter. Accessories include 0.5 cm and 1.0 cm extenders. Sterilization via ethylene oxide (EtO). Mechanical operation; no energy source or software.
Indications for Use
Indicated for men with organic or psychogenic erectile dysfunction requiring penile rigidity for intercourse, including those with pelvic fracture, spinal cord injury/disease, prostatectomy, multiple sclerosis, diabetes, arteriosclerosis, vascular disease, priapism, or Peyronie's disease.
Regulatory Classification
Identification
A penile rigidity implant is a device that consists of a pair of semi-rigid rods implanted in the corpora cavernosa of the penis to provide rigidity. It is intended to be used in men diagnosed as having erectile dysfunction.
Special Controls
The special control for this device is the FDA guidance entitled “Guidance for the Content of Premarket Notifications for Penile Rigidity Implants.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font, also in blue.
April 15, 2019
Rigicon, Inc. Ahmet Melih Luleci Regulatory Representative 2805 Veterans Highway, Suite 13 Ronkonkoma, NY 11779
Re: K181673
> Trade/Device Name: RIGI 10 MALLEABLE PENILE PROSTHESIS Regulation Number: 21 CFR§ 876.3630 Regulation Name: Penile Rigidity Implant Regulatory Class: II Product Code: FAE Dated: March 13, 2019 Received: March 13, 2019
Dear Ahmet Melih Luleci:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Glenn B. Bell -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K181673
Device Name RIGI 10 MALLEABLE PENILE PROSTHESIS
Indications for Use (Describe)
The malleable penile prosthesis is indicated for implantation into the corpora cavernosa of the penis in men who are diagnosed with erectile dysfunction. The prosthesis is implanted to provide adequate penile rigidity for sexual intercourse. The prosthesis is designed for the treatment of organic erectile dysfunction in men due to:
Pelvic fracture, Spinal cord injury or disease, Prostatectomy,Multiple sclerosis, Diabetes
mellitus, Arteriosclerosis, Hypertensive vascular disease, Priapism, Peyronie's disease, Selectively for psychogenic impotence.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
{3}------------------------------------------------
Rigicon Inc. Malleable Penile Prosthesis Abbreviated 510(k)
Image /page/3/Picture/1 description: The image shows the text "K181673 Page 1 of 4 9". The number 9 is in a larger font than the other text. The background of the image is a blue and white abstract design.
## Section 5 - 510(k) Summary
#### 1) SUBMITTER
Rigicon, Inc. 2805 Veterans Highway Suite 13, Ronkonkoma, NY 11779 Phone: 631-676-3376
Contact Person : Ahmet Melih Luleci Rigicon, Inc. Requlatory Representative regulatory@rigicon.com
: March 08, 2019
Date Prepared
#### 2) DEVICE
| Name of Device | : Rigi10™ Malleable Penile Prosthesis |
|----------------------|---------------------------------------|
| Common or Usual Name | : Penile Prosthesis |
| Classification Name | : Penile Rigidity Implant |
| Regulatory Class | : Class II |
| Regulation | : 21 CFR 876.3630 |
| Product Code | : FAE |
## 3) PREDICATE DEVICE
Coloplast (formerly Mentor Corp.) Genesis Penile Prosthesis, K040959 This predicate has not been subject to a design-related recall.
No reference devices were used in this submission.
4) REASON FOR 510(K) SUBMISSION
Rigi10 Malleable Penile Prosthesis is a new model.
#### 5) DEVICE DESCRIPTION
Rigi10 malleable penile prosthesis is a sterile, non-pyrogenic, single-use surgical implant.
The device core consists of a malleable stainless-steel core, with titanium caps on both ends, and a PTFE (Polytetrafluoroethylene) tube cover. The entire outside surface of the cylinder is made out of silicone. The malleable stainlesssteel core enables the prosthesis to be moved into an erect position when desired and also enables the concealment of the
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the text "K181673 Page 2 of 4" at the top. The number 10 is in the middle of the image. The image also contains a blue and white graphic design.
prosthesis under clothing. All components consist of implantable, biocompatible materials.
The device consists of two cylinders, and extenders. The cylinders are surgically implanted, one into each corpus cavernosum, to provide penile rigidity.
The prosthesis comes in two lengths. This includes 23 cm and 25 cm. The length is determined by the diameter of the prosthesis. Rigi10 Malleable Penile Prosthesis is available in five different diameter sizes; 9 mm, 10 mm, 11 mm, 12 mm, and 13 mm. Cylinders with 9 mm and 10 mm diameters are available in 23 cm length while cylinders with 11mm, 12 mm and 13 mm diameters are available in 25 cm length. There are four extenders (two pieces of 0.5 cm and two pieces of 1.0 cm) included in each product package. The extenders are accessories of the device and are available for adjusting the length of the prosthesis to match the total corporal measurement of the patient.
Patients diagnosed with erectile dysfunction are candidates for Rigi10 Malleable Penile Prosthesis implantation. The device should only be implanted by specialized physicians (e.g. urologists and surgeons)
## 6) INDICATIONS FOR USE
The malleable penile prosthesis is indicated for implantation into the corpora cavernosa of the penis in men who are diagnosed with erectile dysfunction. The prosthesis is implanted to provide adequate penile rigidity for sexual intercourse.
The prosthesis is designed for the treatment of organic erectile dysfunction in men due to:
- Pelvic fracture, ●
- Spinal cord injury or disease, ●
- Prostatectomy, ●
- Multiple sclerosis,
- Diabetes mellitus,
- Arteriosclerosis, ●
- Hypertensive vascular disease, ●
- 0 Priapism,
- Peyronie's disease, ●
- Selectively for psychogenic impotence. ●
The Indications for Use statement for the Rigi10 Malleable Penile Prosthesis is not identical to the predicate device (i.e. K040959) however, the differences do not alter the intended therapeutic use of the device nor do they affect the safety and
{5}------------------------------------------------
Image /page/5/Figure/1 description: The image shows the number 11 in white text on a blue background. The background is made up of several overlapping blue shapes, creating a layered effect. The number 11 is positioned in the upper right corner of the image.
effectiveness of the device relative to the predicate. Both the subject and predicate devices have the same intended use for the treatment of erectile dysfunction, by providing adequate penile rigidity for sexual intercourse.
# 7) SUBSTANTIAL EQUIVALENCE CLAIM
Rigi10 Malleable Penile Prosthesis is substantially equivalent to the predicate device (i.e. Coloplast (formerly Mentor Corp.) Genesis Penile Prosthesis) The devices have same intended use, indications for use and target population. The devices are mainly manufactured with similar material (i.e. silicone elastomer) with minor differences such as stainless-steel core and PTFE cover in Rigi10 while the predicate device has a silver core and a polvester cover. Subject and predicate devices are sterilized with the same sterilization methods (i.e. EtO sterilization).
Testing rationale was provided to support the equivalence of the Rigi10 Malleable Penile Prosthesis and shows that no new concerns of safety and effectiveness have been introduced with this device.
The safety and effectiveness of the Rigi10 Malleable Penile Prosthesis are adequately supported by the testing rationale, substantial equivalence information, materials information, and comparison of design characteristics provided with the submission document.
Because the intended use and the fundamental scientific technology of the predicate device are similar with the subject device a clinical study was not conducted for Rigi10 Malleable Penile Prosthesis.
## 8) PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination:
Biocompatibility tests:
- In Vitro Cytotoxicity Test ●
- Skin Sensitization Test ●
- Intracutaneous Reactivity Test ●
- Acute Systemic Toxicity Test
- Pyrogen Test ●
- Bacterial Endotoxin Test
- Implantation Tests
- Chemical Characterization Test ●
- Toxicological Risk Assessment
- Rabbit Blood Hemolysis Test ●
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows a page with the text "K181673 Page 4 of 4" at the top. The number "12" is prominently displayed in the center-right of the image. The background features a blue and white abstract design with geometric shapes.
The device met all biocompatibility test specifications. The tests were carried out in accordance with the ISO 10993-1 "Biological evaluation of medical devices" guideline.
Bench performance tests:
- Simulated Life/Fatique Testing. .
- Positioning/Concealability Testing. .
- Buckling Testing, .
- · Tip Extender Joint Strength Testing.
Sterilization by ethylene oxide has been validated for Rigi10 Malleable Penile Prosthesis.
For each bench performance test, an acceptance criterion was established prior to testing and the outcome of each test was that the acceptance criteria would be met to demonstrate statistically significant substantial equivalence to the predicate device.
The device met all performance test specifications and passed all tests.
## 9) CONCLUSION
Biocompatibility evaluation for Rigi10 malleable penile prosthesis was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. These tests confirm that Rigi10 malleable penile prosthesis is as safe as the predicate device.
Bench test evaluations for Rigi10 malleable penile prosthesis was conducted in accordance with the "Guidance for the Content of Premarket Notifications for Penile Rigidity Implants" January 16, 2000. These tests confirm that Rigi10 malleable penile prosthesis is as safe and effective as the predicate device.
The performance data support the safety of Rigi10 malleable penile prosthesis and demonstrate that Rigi10 is as safe, as effective, and performs as well as the legally marketed device predicate (i.e. Coloplast (formerly Mentor Corp.) Genesis Penile Prosthesis. K040959) that is currently marketed for the same intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.