The THD Anopress device must be used exclusively to assess the average sphincter tone due to the pressure exerted by the muscles in the anal canal on the specially designed THD Press Probe. THD Anopress may only be used by appropriately trained medical staff.
Device Story
THD Anopress is a handheld, portable anorectal manometry device used at the bedside by trained medical staff to assess anal sphincter tone in adult patients. The system consists of a main unit with a keyboard and display, and a non-sterile, disposable polyurethane THD Press Probe. The probe is connected to the unit via a Luer-Lock connector; an internal pump inflates the probe with air to 150 mmHg. Upon insertion into the anal canal, the sphincter muscles compress the membrane. The device detects pressure changes in real-time, which are visualized on the unit's display. Optionally, data can be transmitted via Bluetooth to a PC running dedicated Anopress software for storage. The device provides quantitative measurements of resting and squeeze pressures, endurance, strain, and various reflexes, assisting clinicians in evaluating anorectal function and informing clinical decision-making regarding patient treatment.
Clinical Evidence
No clinical data. Bench testing only. Performance verified through biocompatibility (ISO 10993-1, -5, -10), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and software validation (IEC 62304). Functional accuracy for strain, squeeze, and resting pressure was verified in simulated use conditions and compared to the predicate device.
Technological Characteristics
Handheld manometry system; air-charged polyurethane balloon probe; internal pump inflation; wireless Bluetooth connectivity to PC; real-time pressure monitoring. Complies with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), and ISO 10993 (biocompatibility). Software developed per IEC 62304.
Indications for Use
Indicated for adult patients requiring assessment of anal sphincter tone, including resting pressure, squeeze pressure, endurance, strain pressure, squeeze/rest rate, anal contractile reflex, and cough reflex.
Regulatory Classification
Identification
A gastrointestinal motility monitoring system is a device used to measure peristalic activity or pressure in the stomach or esophagus by means of a probe with transducers that is introduced through the mouth into the gastrointestinal tract. The device may include signal conditioning, amplifying, and recording equipment. This generic type of device includes the esophageal motility monitor and tube, the gastrointestinal motility (electrical) system, and certain accessories, such as a pressure transducer, amplifier, and external recorder.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other, resembling a bird in flight.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 14, 2017
THD S.p.A. % Maurizio Pantaleoni CEO of Isemed (THD's Consultant) ISEMED S.R.L. Via A. Altobelli Bonetti 3/A Imola, BO 40026 Italy
Re: K161785
> Trade/Device Name: THD ANOPRESS Regulation Number: 21 CFR§ 876.1725 Regulation Name: Gastrointestinal Motility Monitoring System Regulatory Class: II Product Code: KLA Dated: February 3, 2017 Received: February 8, 2017
Dear Maurizio Pantaleoni:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely,
Joyce M. Whang -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
Device Name THD ANOPRESS
Indications for Use (Describe)
The THD Anopress device must be used exclusively to assess the average sphincter tone due to the pressure exerted by the muscles in the anal on the specially designed THD Press Probe. THD Anopress may only be used by appropriately trained medical staff.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
_ | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary of the THD ANOPRESS
This 510(k) Summary is being submitted as required by21 CFR 807.92.
### 1. General Information
| Submitter: | THD SPA<br>Via dell'Industria 1, Correggio (RE), 42015, Italy<br>Tel. +39 0522 634311<br>Fax +39 0522634371 |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Consultant/ Contact: | Maurizio Pantaleoni<br>ISEMED srl<br>Via Altobelli Bonetti 3/a<br>Imola, BO40026ITALY<br>Mob. +393484435155<br>Tel. +390542683803<br>Fax+390542698456<br>Email: regulatory@isemed.eu |
Summary Prepared Date: 07/03/2017
#### 2. Names & Classification
Trade Name: THD ANOPRESS Common Name: Anorectal Manometry System Classification name: Gastrointestinal motility monitoring system Class II Regulation Number and Product Code: 21 CFR 876.1725; KLA
#### 3. Predicate Devices
The THD ANOPRESS is substantially equivalent to the following device:
| Applicant | Device name | 510(k) Number |
|-----------|-------------|---------------|
| Medspira | Mcompass | K120088 |
{4}------------------------------------------------
#### 4. Device Description
THD Anopress is a handheld, portable anorectal manometry device for measuring anal pressure, and especially the average sphincter tone, at the bedside on adult patients population.
The subject device consists of two main components:
- THD Anopress: This is the main unit with a keyboard for the selection of input command for pressure measurement and display for the visualization of detected pressure parameters in real time;
- THD Press Probe: This is a non-sterile disposable probe to be connected via the dedicated Luer-Lock connector to the THD Anopress. The probe is charged with air by the pump contained into the THD Anopress main unit and then is introduced into the anal canal of patient for the detection of pressure.
THD Anopress may be used in conjunction with a PC for data storage (not provided with THD Anopress). For this purpose, the PC requires installation of a dedicated program (Anopress SW) and the insertion of the Bluetooth Dongle accessory into the USB port. The Bluetooth dongle as well as Anopress SW are optionally provided by THD SpA
THD PressProbe membrane has a diameter varying from 14,4 mm (empty membrane) to 16,6 mm (membrane inflated to 150 mmHg).
THD Press Probe is inflated to 150 mmHg, then it is inserted in the anal so that the sphincter exerts a compression on the membrane until 390mmHg.
#### 5. Indications for Use
The THD Anopress device must be used exclusively to assess the average sphincter tone due to the pressure exerted by the muscles in the anal canal on the specially designed THD Press Probe. THD Anopress may only be used by appropriately trained medical staff.
#### 6. Comparison of technological characteristics with the predicate device
| Characteristics | THD AnoPress | Mcompass<br>Anorectal Manometry System<br>(K120088) |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| Composition | • manometer<br>• Inflating system: pump<br>• Software on manometer<br>• Software on PC(optional)<br>• catheter (probe) with one sensitive<br>balloon | • manometer<br>• inflating system: syringe<br>• Software<br>• catheter (probe) with 5 sensitive<br>balloons |
| Pressure Data<br>Transmission | Wireless ; in real time | Wireless ; in real time |
| Mechanism of action | Air-charged probe for measuring anal<br>pressure | Air-charged probe for measuring local<br>anorectal pressure |
{5}------------------------------------------------
#### COMPANY NAME 510(K) PREMARKET NOTIFICATION
| Characteristics | THD AnoPress | Mcompass<br>Anorectal Manometry System<br>(K120088) |
|--------------------------------------|-----------------------------------------|-----------------------------------------------------|
| Detected parameters | Resting Pressure | Resting Pressure |
| | Squeeze Pressure | Squeeze Pressure |
| | Endurance | Endurance |
| | Strain Pressure | Strain Pressure |
| | Squeeze/Rest Rate | Squeeze/Rest Rate |
| | Real Time Pressure Measurement | Real Time Pressure Measurement |
| | Anal Contractile Reflex | Anal Contractile Reflex |
| | Cough Reflex | Cough Reflex |
| Component in<br>contact with patient | Disposable Catheter (sensitive balloon) | Disposable Catether (sensitive balloon) |
| Materials patient<br>contacting | Poly-urethane | Urethane |
| Sterility | Non sterile | Non sterile |
## 7. Performance data
# 1) Non clinical tests performed on the subject device:
| Non clinical Performance test Summary-THD AnoPress | | |
|------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Characteristics | Reference standards | Results |
| Biocompatibility | ISO 10993-1 ISO 10993-5 ISO 10993-10 | Obtained results demonstrate compliance<br>to the standards |
| Electrical safety | • IEC 60601-1 | Obtained results demonstrate compliance<br>to the standard |
| EMC | • IEC 60601-1-2 | Obtained results demonstrate compliance<br>to the standard |
| Software Verification<br>and Validation | • IEC 62304 | Obtained results demonstrate compliance<br>to the standard |
| Functional<br>Performance and<br>accuracy verification<br>in simulated use<br>conditions | • Verification of functional<br>performances ad accuracy of the THD<br>Anopress at :<br>Strain pressure squeeze pressure resting pressure using different THD AnoPress and<br>different THDPress-Probe | The performances and the accuracy<br>obtained are comparable with the<br>predicate device for all the products tested |
{6}------------------------------------------------
| Non clinical Performance test Summary-THD Press Probe | | |
|-------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristics | Test Performed | Results |
| Mechanical<br>Performance Tests<br>(new and aged<br>devices | • Verification of geometric<br>specifications (length and diameter)<br>• Verification of membrane<br>expanded/unexpanded diameter<br>• Verification of membrane and tubing<br>leak testing<br>• Verification of membrane expanded /<br>unexpanded diameter after repeated<br>use<br>• Verification of membrane and tubing<br>leak testing after repeated use<br>• Verification of membrane maximum<br>volume<br>• Verification of functional performance<br>of the THD AnoPress (strain, squeeze,<br>resting) with not aged/aged Press<br>Probes.<br>• Verification of the membrane burst<br>pressure<br>• Verification of tensile testing of the<br>joints and bonded components of the<br>tubing<br>• Verification of the peeling of the joints<br>and bonded components of the<br>membrane | No differences in results between aged and<br>not aged samples.<br>When applicable the relevant<br>performances of THD AnoPress and THD<br>PressProbe are comparable with the<br>performances of the predicate device. |
### 2) Clinical tests performed on the subject device:
N/A: No clinical tests were performed on the subject device.
#### 8. Conclusions
In light of evidence summarized above and based on classification, intended use, technological characteristics, and performance data, the subject device is substantially equivalent to the predicate device K120088.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.