WELCH ALLYN XENON 300 LIGHT SOURCE

K980281 · Welch Allyn, Inc. · GCT · Feb 27, 1998 · Gastroenterology, Urology

Device Facts

Record IDK980281
Device NameWELCH ALLYN XENON 300 LIGHT SOURCE
ApplicantWelch Allyn, Inc.
Product CodeGCT · Gastroenterology, Urology
Decision DateFeb 27, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The purpose of the Welch Allyn 300 Watt Xenon Light Source is to provide lighting for medical applications requiring high intensity illumination. The intended use is to provide illumination for medical products such as existing and approved headlights and various existing and approved rigid or flexible endoscopes.

Device Story

The Welch Allyn 300 Watt Xenon Light Source provides high-intensity illumination for medical applications. It serves as a light source for existing, approved medical headlights and rigid or flexible endoscopes. The device is intended for use in clinical settings by healthcare professionals. It operates by generating high-intensity light, which is then transmitted to the surgical or diagnostic site via compatible medical instruments. The device is designed to meet electrical and mechanical safety standards to protect both the patient and the user. By providing reliable, high-intensity lighting, it facilitates visualization during medical procedures, aiding clinicians in diagnostic and surgical tasks.

Clinical Evidence

Bench testing only. Safety and effectiveness are supported by compliance with electrical safety standards (UL 2601, CSA 601-1, IEC 601-1) and electromagnetic compatibility standards (IEC 601-1-2). Performance is evaluated against ACGIH TLV equations for ultraviolet, light, and near-infrared radiation.

Technological Characteristics

300 Watt Xenon lamp; high-intensity illumination source. Designed for connectivity with rigid/flexible endoscopes and headlights. Complies with UL 2601, CSA 601-1, IEC 601-1, and IEC 601-1-2 standards. Mechanical safety features include rounded, textured, or deburred contact surfaces.

Indications for Use

Indicated for medical lighting applications requiring remotely located, high intensity illumination. Not indicated for neonate transillumination.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K980281 P192 #### "510(k) SUMMARY" FEB 2 7 1998 Summary of Safety and Effectiveness | Submitter's Name & Address: | Welch Allyn Inc. | |-----------------------------|----------------------------------------| | | 4619 Jordan Road, Box 187 | | | Skaneateles Falls, New York 13153-0187 | Contact Person & Telephone: Kathy Lowther (315) 685-2897 Date Summary Prepared: January 19, 1998 Classification Name - Endoscope and Accessories Device Name: Common/Usual Name - Xenon 300 Light Source Proprietary Name - Welch Allyn Xenon 300 Light Source - Predicate Device: Luxtec Model 9300 Xenon Light Source (A 510(k) could not be located for this device. However, the following Luxtec 510(k) numbers may be appropriate: K864385, K890716, or K864380. It is assumed Luxtec is utilizing one of the above 510(k) numbers as an equivalent device.) # Device Description, intended Use & Effectiveness: The purpose of the Welch Allyn 300 Watt Xenon Light Source is to provide lighting for medical applications requiring high intensity illumination. The intended use is to provide illumination for medical products such as existing and approved headlights and various existing and approved rigid or flexible endoscopes. The Welch Allyn 300 Watt Xenon Light Source is effective for its intended use. # Technological Characteristics: See Attachment "A" for a comparison of the features and specifications of the Welch Allyn 300 Watt Xenon Light Source to the predicate device. # Summary of Safety: The Welch Allyn 300 Watt Xenon Light Source is designed to provide electrical safety to the patient as well as the user. The system is designed to meet the following standards related to electrical safety: UL 2601, CSA 601-1, and IEC 601-1. To provide mechanical safety to the user, all contact surfaces shall be rounded, textured, or deburred where possible and if necessary. {2}------------------------------------------------ 130281 P292 ACGIH TLV's (American Conference of Governmental Industrial Hygienists Threshold Limit Values (1994 - 1995.) Ultraviolet, Light, and Near Infrared Sections) equations will be used as a basis of comparison for the Welch Allyn 300 Watt Light Source and the predicate device in their intended application. The system will also be certified to the following standards: IEC601-1-2 - Conforms with provisions of European Council Directive 93/42/EEC CE Mark concerning medical devices. #### Summary of Effectiveness: The Welch Allyn 300 Watt Xenon Light Source is effective for its intended use of providing illumination for medical products such as existing and approved headlights and various existing and approved rigid or flexible endoscopes. {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name arranged in a circular fashion around a symbol. The symbol is composed of four stylized human figures, each represented by a curved line, stacked on top of each other. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 FEB 2 7 1998 Ms. Kathy Lowther Quality Engineer Welch Allyn, Inc. 4619 Jordan Road, Box 187 Skaneateles Falls, New York 13153-0187 Re: K980281 Welch Allyn 300 Watt Xenon Light Source Dated: January 20, 1998 Received: January 26, 1998 Regulatory Class: II 21 CFR §876.1500/Product Codes: 78 GCT & FCW Dear Ms. Lowther: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drue, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begine as described in your 5100k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for question and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours. Kilian Uhr Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health {4}------------------------------------------------ 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: Welch Allyn 300 Watt Xenon Light Source #### Indications For Use: The Welch Allyn 300 Watt Xenon Light Source is indicated for medical lighting applications requiring a remotely located, high intensity illumination. The device is not indicated for neonate transillumination. Rober D. Sattler/ Concurrence of CDRH, Office of Device Evaluation (ODE) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ (Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 1980281 510(k) Number _ Prescription Use __ NEEDED) OR Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________ (Optional Format 1-2-96)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...