Sensor-Ureterorenoscope 9 FR WL 600 MM (short RIWO D-URS), Adapter For Controller 5525
K201897 · Richard Wolf Medical Instruments Corporation · FGB · Jan 15, 2021 · Gastroenterology, Urology
Device Facts
Record ID
K201897
Device Name
Sensor-Ureterorenoscope 9 FR WL 600 MM (short RIWO D-URS), Adapter For Controller 5525
Applicant
Richard Wolf Medical Instruments Corporation
Product Code
FGB · Gastroenterology, Urology
Decision Date
Jan 15, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Sensor-Ureterorenoscopes are used within the scope of therapeutic interventions. The products are active, sterile and for single-use. They are used for visualizing body cavities via natural passages and for the insertion of auxiliary instruments through the working channel as well as the supply of irrigation fluid. The Sensor-Ureterorenoscope is used in Urology for examination, diagnostics and/or therapy of the upper urinary tract. User These products are exclusively intended for use by specialized medical personnel and must only be used by medically qualified and adequately instructed persons. ADAPTER FOR CONTROLLER 5525 The products are used for operating a sensor endoscope in conjunction with an ENDOCAM Logic 5525 Camera Controller, for the mechanical connection to the sensor endoscope as well as for controlling the signals generated. User This product is designed exclusively for use by specialized medical personnel and may only be applied by medically qualified and adequately trained persons.
Device Story
Sensor-Ureterorenoscope (RIWO D-URS) is a single-use, sterile, semi-rigid endoscope with an actively deflectable tip (+/-300°); features two working channels for laser fibers and auxiliary instruments (e.g., stone-extraction baskets); includes irrigation/drainage channels. Used in urology for upper urinary tract visualization, diagnostics, and therapy. Operated by specialized medical personnel in professional healthcare facilities. Connects to ENDOCAM Logic 5525 Camera Controller via a reusable Adapter for Controller 5525. Adapter provides mechanical connection and signal control. Physician views real-time images on a connected controller display to guide instrument placement and therapeutic procedures. Enables minimally invasive access to urinary tract, facilitating stone management and other interventions.
Clinical Evidence
No clinical or animal studies were performed. Substantial equivalence is supported by non-clinical bench testing, including sterilization validation (EN ISO 11135:2014), packaging integrity (ASTM D4169, F1980, F1886, F2096, F1929, F88), biocompatibility (ISO 10993-1), electrical safety/EMC (IEC 60601-1, 60601-1-6, 60601-1-2, 60601-2-18), and a usability study.
Indicated for urological examination, diagnostics, and/or therapy of the upper urinary tract via natural passages in patients requiring therapeutic intervention.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
Predicate Devices
Flexible Sensor Ureterorenoscope COBRA VISION EF (COBRA) (K183188)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 15, 2021
Richard Wolf Medical Instruments Corporation Michael Loiterman US Head of Regulatory - OA/OC 353 Corporate Woods Parkway Vernon Hills, IL 60061
Re: K201897
> Trade/Device Name: Sensor-Ureterorenoscope 9 FR WL 600 MM (short RIWO D-URS), Adapter for Controller 5525 Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FGB Dated: December 17, 2020 Received: December 18, 2020
Dear Michael Loiterman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mark J. Antonino, M.S. Acting Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
### 510(k) Number (if known) K201897
### Device Name SENSOR-URETERORENOSCOPE 9 FR WL 600 MM ADAPTER FOR CONTROLLER 5525
### Indications for Use (Describe) SENSOR-URETERORENOSCOPE 9 FR WL 600 MM
Sensor-Ureterorenoscopes are used within the scope of therapeutic interventions. The products are active, sterile and for single-use. They are used for visualizing body cavities via natural passages and for the insertion of auxiliary instruments through the working channel as well as the supply of irrigation fluid.
The Sensor-Ureterorenoscope is used in Urology for examination, diagnostics and/or therapy of the upper urinary tract.
User
These products are exclusively intended for use by specialized medical personnel and must only be used by medically qualified and adequately instructed persons.
### ADAPTER FOR CONTROLLER 5525
The products are used for operating a sensor endoscope in conjunction with an ENDOCAM Logic 5525 Camera Controller, for the mechanical connection to the sensor endoscope as well as for controlling the signals generated.
User
This product is designed exclusively for use by specialized medical personnel and may only be applied by medically qualified and adequately trained persons.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a logo with the text "K201897 Page 1 of 8" at the top. Below that is the text "RICHARD WOLF" in a stylized font, with the word "WOLF" in red. To the right of the text is a logo of a dog with a medical instrument inserted into it.
#### 5 510(k) Summary
#### l Submitter
Richard Wolf Medical Instruments Corporation 353 Corporate Woods Parkway Vernon Hills, IL 60061
Phone:(847) 913 1113 Fax: (847) 913 0924
Contact Person, Title: Michael Loiterman, US Head of Regulatory - QA/QC Date Prepared: December 16, 2020
## Legal Manufacturer
Richard Wolf GmbH Pforzheimer Straße 32 75438 Knittlingen
#### ll Device
| Device name | Common name | Commercial name | Model Number(s) | Classification name | Regulation Number and Name | Product Code | Device Class |
|-----------------------------------------------|--------------|-----------------|------------------------|-------------------------------------------------------|---------------------------------------------|--------------|--------------|
| SENSOR-<br>URETERORENOSCOPE<br>9 FR WL 600 MM | Ureteroscope | RIWO D-URS | 473572075<br>473572076 | Ureteroscope<br>and<br>Accessories,<br>Flexible/rigid | 876.1500<br>Endoscope<br>and<br>accessories | FGB | II |
| ADAPTER FOR<br>CONTROLLER 5525 | --- | --- | 5525410 | | | | |
#### ��� Predicate Device
Name of Predicate Device: Flexible Sensor Ureterorenoscope COBRA VISION EF (COBRA) (73561076) 510(k) Number: K183188 Regulatory Class: II Product Code: FGB / FGA / ODC Manufacturer: Richard Wolf GmbH
The predicate has not been subject to a design-related recall. No reference devices were used in this submission.
{4}------------------------------------------------
#### Subject Device Description 5.1
#### 5.1.1 Device Identification
| Type number | Description | Package<br>unit |
|-------------|----------------------------------------|-----------------|
| 473572075 | SENSOR-URETERORENOSCOPE 9 FR WL 600 MM | 1 |
| 473572076 | SENSOR-URETERORENOSCOPE 9 FR WL 600 MM | 3 |
| 5525410 | ADAPTER FOR CONTROLLER 5525 | 1 |
#### Device characteristics 5.1.2
### Software
The RIWO D-URS and ADAPTER FOR CONTROLLER 5525 do not contain software.
## Materials with patient contact
The RIWO D-URS includes components with direct and indirect patient contact.
The ADAPTER FOR CONTROLLER 5525 has no components with direct or indirect patient contact.
## Single use / reusable
The RIWO D-URS is a single use device.
The ADAPTER FOR CONTROLLER 5525 is reusable and does requires cleaning and reprocessing during its use-life, i.e. users are required to clean and disinfect the device before every application and before returning for repairs. Methods of cleaning and reprocessing are detailed in the Instruction for Use.
### Delivered sterile / non-sterile
The RIWO D-URS is delivered sterile. The ADAPTER FOR CONTROLLER 5525 is delivered non-sterile.
# Sterilization Method
The RIWO D-URS is sterilized with ethylene oxide. The ADAPTER FOR CONTROLLER 5525 is not intended to be sterilized.
#### 5.1.3 Environment of Use
The RIWO D-URS and ADAPTER FOR CONTROLLER 5525 are intended to be used in professional health care facility environment.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the text 'K201897' at the top, followed by 'Page 3 of 8'. Below this is the name 'RICHARD WOLF' in a stylized font, with the word 'WOLF' in a bold, red, sans-serif font. To the right of the name is a logo featuring a dog-like figure standing next to a vertical line with a hook at the top.
#### 5.1.4 Brief Written Description of the Device
The Sensor-Ureterorenoscope features a semi-rigid instrument sheath and a flexible and actively deflectable tip. It is equipped with two separate working channels for applications with laser fibers and/or flexible auxiliary instruments. The two working channels have different diameters, allowing the insertion of the laser fiber via working channel (2) and a stone-extraction basket or flexible auxiliary instruments via working channel (1)
Irrigation is effected via working channel (1) and is connected via the luer connector marked in blue on the housing.
A third channel integrated in the instrument sheath allows the return flow via distal holes on the instrument head. The drain is effected via the lower luer connector marked in gray.
The flexible deflectable tip can be controlled in both directions (upward and downward) +/-300° via the control wheel on the handle. Thanks to the high torsional stiffness of the semi-rigid instrument sheath the sheath movements are transmitted directly in radial and axial directions. The working channel (2) permits the insertion of laser fibers up to a diameter of 272 µm.
For visualization the Sensor-Ureterorenoscope is connected to an endoscopy camera controller. The Sensor-Ureterorenoscope is sterile and for single-use.
Image /page/5/Figure/8 description: This image shows a medical device with several labeled parts. The labels indicate the "Objective Lens", "Working channel (1)", "Working channel (2)", and "Light exit". The device appears to be an endoscope or similar instrument used for medical imaging and procedures.
Figure 1: Image of the SENSOR-URETERORENOSCOPE 9 FR WL 600 MM with detailed display of the tip
The ADAPTER FOR CONTROLLER 5525 is an accessory to the RIWO D-URS, which serves as an interface to connect the RIWO D-URS to an ENDOCAM Logic 5525 Camera Controller (not part of this submission).
#### 5.1.5 Materials of Use
The RIWO D-URS consists of stainless steel, glass, polymer, glue and instrument oil. The materials can come into direct and indirect body contact with the patient for a maximal contact duration of 3 hours.
The ADAPTER FOR CONTROLLER 5525 consists of stainless steel, polymer, composite, copper and gold. The materials do not have any body contact with the patient.
#### 5.1.6 Key Performance Characteristics
The RIWO D-URS provides a deflection angle of 300° upwards and downwards, two working channels, a working length of 600 mm and a 160000 pixel CMOS sensor for imaging.
The ADAPTER FOR CONTROLLER 5525 provides electromechanical connectivity with a camera controller.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the text "K201897 Page 4 of 8" in the upper left corner. Below that is the text "RICHARD WOLF" in a stylized font. To the right of the text is a logo of a dog with a nail being hammered into its head.
#### Indications for Use 5.2
#### 5.2.1 Statement
# SENSOR-URETERORENOSCOPE 9 FR WL 600 MM
Sensor-Ureterorenoscopes are used within the scope of therapeutic and diagnostic interventions. The products are active, sterile and for single-use. They are used for visualizing body cavities via natural passages and for the insertion of auxiliary instruments through the working channel as well as the supply of irrigation fluid.
The Sensor-Ureterorenoscope is used in Urology for examination, diagnostics and/or therapy of the upper urinary tract.
User
These products are exclusively intended for use by specialized medical personnel and must only be used by medically qualified and adequately instructed persons.
## ADAPTER FOR CONTROLLER 5525
The products are used for operating a sensor endoscope in conjunction with an ENDOCAM Logic 5525 Camera Controller, for the mechanical connection to the sensor endoscope as well as for controlling the signals generated.
User
This product is designed exclusively for use by specialized medical personnel and may only be applied by medically qualified and adequately trained persons.
#### 5.2.2 Explanation on differences to predicate device
In comparison to the predicate device, the subject device does have only the urological indication of use and restricts usage of the device to natural passages.
However, the full indication of use for the subject device is covered by the predicate device and reducing the scope does not lead to new questions regarding safety and effectiveness.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the text "K201897 Page 5 of 8" in the upper left corner. Below this text is the name "RICHARD WOLF" in a stylized font, with "RICHARD" in black and "WOLF" in red. To the right of the name is a logo featuring a dog-like figure with a tool or instrument near its head.
#### Comparison of Technological Characteristics with the Predicate Device 5.3
#### 5.3.1 Overview
| Technolgical<br>characteristic | Submitted device<br>RIWO D-URS | Predicate device |
|------------------------------------|--------------------------------|---------------------|
| Software | NA | NA |
| Materials with<br>patient contact | Direct and indirect | Direct and indirect |
| Single use /<br>reusable | Single use | Reusable |
| Delivered sterile /<br>non-sterile | Delivered sterile | Non-sterile |
| Sterilization<br>method | Ethylene oxide | Hydrogen peroxide |
#### 5.3.2 Discussion
There are some differences between the subject device and the predicate device that had to be verified or validated before claiming substantial equivalence.
Main difference:
- l. The RIWO D-URS is a single use and sterile delivered device whereas the predicate device can be reprocessed which lead to changes in material, sterilization process, storage, etc.
Minor differences:
- i. The RIWO D-URS requires the ADAPTER FOR CONTROLLER 5525 to establish the connection with an ENDOCAM Logic 5525 Camera Controller whereas the predicate device can be connected directly.
- ii. The RIWO D-URS has a different handle design which requires different operation of the instrument.
- The RIWO D-URS is designed with small deviations in technical characteristics such as iii. sheath diameter, working channel diameter, working length, deflection range, LED positioning and pixel size
- l. For the main difference, several testing activities of the subject device were necessary to prove safety and effectiveness. This includes
- Storage and packaging validation of the sterile packaging
- Validation of the sterilization process ●
- Biocompatibility testing due to the new materials
- Performance testing due to new materials
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the text "K201897 Page 6 of 8 RICHARD WOLF" in black font. Below the text is the word "WOLF" in large red font. To the right of the word "WOLF" is a logo of a dog with a nail in its mouth.
- i. The first minor difference was evaluated by bench testing and electromagnetic compatibility testing to prove the compatibility of the devices. Every functional performance test was done using the ADAPTER FOR CONTROLLER 5525 as otherwise no image would have been visible.
- ii. The second difference was evaluated by a usability study which proved that the subject device can be operated as safe and effective as the predicate device.
- iii. The third difference was evaluated by bench testing, proving that the deviations have no negative impact on the safety and effectiveness of the device and that it can perform within its specifications.
The following subsections provide an overview of the testing activities and methods, for more detail please check the applicable section.
#### 5.4 Clinical Performance Testing
No clinical testing was performed with RIWO D-URS or ADAPTER FOR CONTROLLER 5525
No animal studies were performed with RIWO D-URS or ADAPTER FOR CONTROLLER 5525
#### Non-Clinical Performance Testing 5.5
#### 5.5.1 Sterilization
The ethylene oxide sterilization process for the RIWO D-URS was validated according to EN ISO 11135:2014 Annex B (half-cycle method). The sterile packaging was subjected to several tests to ensure the sterility including
- . Environmental conditioning ASTM D4169-16 and ASTM D4332-14
- Accelerated aging ASTM F1980-16 .
- Transportation Simulation ASTM D4169-16 .
- . Visual Inspection of packaging and seal integrity ASTM F1886-16
- . Packaging Challenging Testing [Bubble Test] ASTM F2096-11
- . Seal Integrity Test [Liquid Dye] ASTM F1929-15
- . Seal Strength Test ASTM F88-15
All test methods were applied to non-aged as well as accelerated aged products.
The ADAPTER FOR CONTROLLER 5525 is delivered non-sterile, therefore no validation was required.
The results proved that the RIWO D-URS is as safe as the predicate device and that the RIWO D-URS including ADAPTER FOR CONTROLLER 5525 can be deemed substantially equivalent to the predicate device.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows a logo with the text "K201897" at the top, followed by "Page 7 of 8". Below this is the name "RICHARD WOLF" in a stylized font, with "WOLF" in large red letters. To the right of the name is a logo featuring a dog-like figure with a medical instrument, possibly a scope, inserted into it.
#### Biocompatibility 5.5.2
The RIWO D-URS includes components with direct and indirect patient contact. Therefore, a biological risk assessment was performed considering the following:
- . ISO 10993-1:2018: Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process
- . Use of International Standard ISO 10993, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process". Guidance for Industry and Food and Drug Administration Staff, issued June 16, 2016
The ADAPTER FOR CONTROLLER 5525 has no components with direct or indirect patient contact, therefore biocompatibility testing according to ISO 10993-1 was not necessary.
Biological testing and chemical characterization proved that the RIWO D-URS is as safe as the predicate device and that the RIWO D-URS including ADAPTER FOR CONTROLLER 5525 can be deemed substantially equivalent to the predicate device.
#### 5.5.3 Electromagnetic Compatibility and Electrical Safety
Electromagnetic compatibility testing proved compliance with the following standards for RIWO D-URS including ADAPTER FOR CONTROLLER 5525
- IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012; Medical electrical equipment Part 1: . General requirements for basic safety and essential performance
- . IEC 60601-1-6:2010 + A1:2013 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- . IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests
- . IEC 60601-2-18:2009 Medical electrical equipment - Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment
It could be shown that regarding electromagnetic compatibility and electrical safety the devices are as safe and effective and therefore substantially equivalent to the predicate device.
#### 5.5.4 Performance and Operational testing
The efficacy and safety of Richard Wolf's RIWO D-URS including ADAPTER FOR CONTROLLER 5525 is documented by the verification and validation testing, which confirms that the product meets all the reguirements and specifications for overall design, basic safety and essential performance, and that the design inputs and specifications are met.
This includes Packaging, Transportation, Environmental conditions, photobiological safety and functionality testing, performed for the RIWO D-URS as well as the ADAPTER FOR CONTROLLER 5525. In addition, a usability study was performed, and optical performance was evaluated for the RIWO D-URS and parameters without defined acceptance criteria were compared to those of the predicate device to prove equivalence.
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the text "K201897 Page 8 of 8 RICHARD WOLF" in a bold, sans-serif font. The word "WOLF" is in a larger font size and is red. To the right of the text is a logo featuring a dog-like figure with a medical instrument.
It could be shown that the RIWO D-URS including the ADAPTER FOR CONTROLLER 5525 functions as safe as effective as the predicate device and therefore can be deemed substantially equivalent.
#### 5.6 Conclusions
The non-clinical data support the safety of the device and the verification and validation demonstrate that the RIWO D-URS and ADAPTER FOR CONTROLLER 5525 should perform as intended in the specified use conditions. The data demonstrate that the RIWO D_URS and ADAPTER FOR CONTROLLER 5525 perform comparably to the predicate device that is currently marketed for an equivalent intended use.
Therefore, the RIWO D-URS with ADAPTER FOR CONTROLLER 5525 has been deemed to be substantially equivalent to the legally marketed device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.