The Duet-with Anorectal Manometry Suite is intended to measure and record the functioning of the patient's urinary tract and anorectal function*. The measurements comprise recording of pressure, volume, flow and EMG. The system includes transducers, devices, tubing, catheters and electrodes. * expanded indication for use due to the Anorectal Manometry Suite.
Device Story
Anorectal Manometry Suite is a software-enabled diagnostic tool for use with the Duet/Duet MultiP system. It records pressure, volume, flow, and EMG signals from the anorectal region. The system utilizes transducers, tubing, catheters, and electrodes to capture physiological data. It performs predefined tests: Sphincter Profile, Resting Pressure, Rectal Balloon test, Balloon Expulsion Test, Squeeze Test, and Recto-Anal Inhibitory Test. Operated in clinical settings, the software requires a license to enable functionality. It processes raw signal data to calculate values such as maximums, minimums, and durations. Results are displayed as signal tracings and reports for clinician review to assist in diagnosing anorectal function. The device does not perform automated interpretations or comparisons against normal values.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Software-enabled diagnostic suite for Duet/Duet MultiP hardware. Records pressure, volume, flow, and EMG. Connectivity via Windows 95/98 PC environment. Uses MS Access (ODBCII driver) for database management. Complies with EN 60601-1:1990 (and amendments A1, A11, A12, A13). Accessories are non-sterile.
Indications for Use
Indicated for pediatric to adult patients requiring diagnostic measurement and recording of urinary tract and anorectal function, including pressure, volume, flow, and EMG signals.
Regulatory Classification
Identification
A gastrointestinal motility monitoring system is a device used to measure peristalic activity or pressure in the stomach or esophagus by means of a probe with transducers that is introduced through the mouth into the gastrointestinal tract. The device may include signal conditioning, amplifying, and recording equipment. This generic type of device includes the esophageal motility monitor and tube, the gastrointestinal motility (electrical) system, and certain accessories, such as a pressure transducer, amplifier, and external recorder.
{0}------------------------------------------------
# 510(k) Notification 23S20 Anorectal Manometry Suite for Duet / Duet MultiP
# 510(k) SUMMARY
as required per 807.92(c) 動画 2017 06:00 【 10:00 【 10:00 【 10:00 【 10:00 【 10:00 】 10:00 【 10:00 】 10:00 【 1
# 1. Submitters Name, Address:
Medtronic Functional Diagnostics A/S Tonsbakken 16-18 DK-2740 SKOVLUNDE Tel: + 45 44 57 95 02 Fax: + 45 44 57 90 10 Contact person for this submission: Ann-Christine Jönsson Date submission was prepared: 19th April, 1999
# 2. Trade Name, Common Name and Classification Name:
Anorectal Manometry Suite for Duet / Duet MultiP A. Trade Name:
## B. Common Name, Classification Name, Class and Regulation Number:
| Common Name | Classification<br>Number | Class | Regulation Number |
|---------------------------|--------------------------|-------|-------------------|
| Anorectal Manometry suite | 78 FFX | II | 21 CFR 876.1725 |
## 3. Predicate Device Identification:
The scientific technology and the functionality and intended use of the Anorectal Manometry Suit for Duet is equivalent to Medtronic Synectics Anorectal Manomentry Analysis Module (K972439).
## 1 (4)
#### COMPANY CONFIDENTIAL
Medtronic Functional Diagnostics A/S
Tonsbakken 16-18 MIT 2740 CUATIT TINING
{1}------------------------------------------------
#### 4. Device Description:
The Anorectal Manometry Suit is together with Duet/Duet MultiP intended to record a patients anorectal function compromising EMG and pressure recordings. The whole system includes transducers, devices, tubing, catheters and electrodes.
The Duet / Duet MultiP is an apparatus capable of registration of pressure signals over a long time. The measurements and analysis have originally been developed for urodynamics tests, but since anorectal tests also are measuring pressure signals over a period of time, it is suitable to use the same technique for these types of tests, i.e. anorectal tests.
The device is used in a dedicated clinical setup. The following sets of test types are predefined tests for measuring in the rectum:
- Sphincter Profile,
- . Resting Pressure,
- Rectal Balloon test,
- 비 Balloon Expulsion Test,
- 피 Squeeze Test
- l Recto-Anal Inhibitory Test.
The tests are operated just like the urodynamic tests. A license number is required to enable the software for the Anorectal Manometry Suite.
#### 5. Intended Use:
The Duet-with anorectal manometry suite is intended as a diagnostic tool to measure and record the functioning of the patient's urinary tract and anorectal function *. The measurements comprise recording of pressure, volume, flow and EMG.
* expanded indication for use due to the Anorectal Manometry Suit.
# 2 (4) COMPANY CONFIDENTIAL
Medtronic Functional Diagnostics A/S
Tonsbakken 16-18 DK-2740 SKOVI I'NDU
{2}------------------------------------------------
# 6. Table of Device Similarities and differences to predicate device
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
| Manufacturer | Medtronic Synectics AB | Medtronic Functional<br>Diagnostics | - |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number | Predicate Device | Modified Device | |
| | Anorectal Manometry<br>Analysis Module for Polygraf<br>HR - K 972439 | Anorectal manometry<br>Suite for Duet/Duet<br>MultiP - K tbd | |
| General: | Predicate devices: | Modified Device | Explanation of the<br>differences compared<br>to the Predicate device |
| Intended Use /<br>Indication of Use | Analyse pressure data recorded<br>from the lower gastrointestinal<br>tract. | Record pressure data from the<br>lower gastrointestinal tract. | The Modified Device does<br>only record pressure and<br>EMG, no comparisons are<br>made, only actual results or<br>calculation heroff is made. |
| Intended Populations | Pediatric to Adults | Same | --- |
| Sterilization | Accessories are not supplied<br>sterile, manufacturer label the<br>accessories with cleaning<br>instructions. | Same | --- |
| Biocompatibility | Sensors are the only part that<br>comes into contact with the<br>patients. | Same | --- |
| Features: | Predicate devices | Modified Device | Explanation of the<br>differences compared<br>to the Predicate device |
| Signal to measure | Pressure, EMG | Same | |
| Analyzing signals | Pressure data is analyzed in terms<br>of physiological properties,<br>comparison with normal values,<br>etc. | No analysis or interparetation<br>is made - only calculations<br>(max, min, duration, etc values<br>of tracings. The following tests<br>are specified:<br>Sphincter Profile<br>Resting Pressure<br>Rectal Balloon Test<br>Squeeze Test<br>Balloon Expulsion Test<br>Recto-Anal Inhibitory Reflex | Modified Device presents<br>recorded signals and results<br>only. It does not contain<br>normal values and no<br>comparison is made. |
| Data displayed | Raw signal data | Same | --- |
| User commands<br>Software<br>Environment | Menu selections<br>Dos and Windows | Same<br>Windows | ---<br>Modified Device runs under<br>Windows 95 / 98 |
| Reports | Signal tracings and reports. | Same | --- |
| Patient database | Database | Relational database with<br>logical patient- recording<br>relations | Modified Device uses MS<br>Access (ODBCII Driver) |
| Additional patient<br>data | Comment field | Same | --- |
| User help system | Separate command description | Same | --- |
| Signal review<br>method | Time - tracing based | Same | --- |
| Recording control | Real time monitoring of signals | Same | --- |
| Recording configur. | User definable configuration | Same | --- |
#### 3 (4)
# COMPANY CONFIDENTIAL
Medtronic Functional Diagnostics A/S
Tonsbakken 16-18 DK_2740 SKAVI UNITE ...
{3}------------------------------------------------
#### 7. Assessment of non-clinical performance data for equivalence:
Verifications results shows that the enhanced system performs as its predicate system.
# 8. Assessment of clinical performance data for equivalence:
Clinical evaluation not performed.
#### 9. Biocompatability:
This new anorectal suit is an extension of the software and there is no contact with the patient. The only part of this system that comes into contact with the body are the accessories and they are already in commercial distribution on the US market.
#### 10. Sterilization:
Not applicable
## 11. Standards and Guidances:
The Duet / Duet MultiP complies to the following standard:
- EN 60601-1:1990 and Amendments A1, A11, A12 and A13
The Anorectal manometry suite is a pure software enabling and doesn't affect the hardware.
#### COMPANY CONFIDENTIAL
Medtronic Functional Diagnostics A/S
Tonsbakken 16-18 DK-2740 SKOVI IINDR
Tel: + 45 44 57 90 00 Nor. + 45 14 57 00 10
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and head. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 15 1999
Ms. Ann-Christine Jönsson Regulatory Affairs Medtronic Functional Diagnostics A/S Tonsbakken 16-18 DK-2740 Skovlunde DENMARK
Re: K991389 Anorectal Manometry Suite for Duet/Duet MultiP Dated: April 19, 1999 Received: April 21, 1999 Requiatory Class: II 21 CFR §876.1725/Procode: 78 FFX
Dear Ms. Jönsson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requirin regulatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compiance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Notification 23S20 Anorectal Manometry Suite for Duet /Duet MultiP
# Indication for Use Statement
Page 1 of 1 _ Market Land (1982) - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1991
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Anorectal Manometry Suit
Indications for Use:
The Duet-with Anorectal Manometry Suite is intended to measure and record the functioning of the patient's urinary tract and anorectal function*. The measurements comprise recording of pressure, volume, flow and EMG. The system includes transducers, devices, tubing, catheters and electrodes.
* expanded indication for use due to the Anorectal Manometry Suite.
## MRI Compatibility Statement:
The Duet with Anorectal Manometry Suite is not compatible for use in a MRI magnetic field.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|--------------------------------------------------------------------|---------|
| Division of Reproductive, Abdominal, ENT, and Radiological Devices | |
| 510(k) Number | K99/389 |
| Prescription Use | <div> </div> |
|----------------------|--------------|
| (Per 21 CFR 801.109) | |
OR
| Over-The-Counter Use | |
|----------------------|--|
|----------------------|--|
(Optional Format 1-2-96)COMPANY CONFIDENTIAL
Medtronic Functional Diagnostics A/S
Tonsbakken 16-18 1787 Tel: + 45 44 57 90 00
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.