DIGITRAPPER DELTA, MODEL 9043G0201 AND POLYGRAM '98 PH EXTENSION MODEL 9043S0111
Applicant
Medtronic Functional Diagnostics A/S
Product Code
FFX · Gastroenterology, Urology
Decision Date
Feb 4, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1725
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The pH System is intended to record, store, view and analyze esophageal and gastric pH data to diagnose reflux disorders. The pH System can also be used to locate the position of the proximal Lower EsopHageal Sphincter (LES) manometrically, to assist in the accurate placement of the pH catheter.
Device Story
pH Testing System consists of Digitrapper Delta recording device and Polygram '98 pH Extension software on a general-purpose PC. Technician intubates patient with pH catheter; device records physiological data during normal daily activity. Data transferred via IrDA 1.1 to PC for analysis. Software marks refluxes; calculates metrics including DeMeester score and Boix-Ochoa scoring method. Physician reviews hardcopy reports to assist in reflux disorder diagnosis. System also performs manometric LES localization to guide catheter placement. Benefits include improved diagnostic capability for reflux disorders through automated data analysis and enhanced LES positioning.
Clinical Evidence
Bench testing only. Verification results confirm the enhanced system performs as the predicate system. No clinical data provided.
Technological Characteristics
System includes Digitrapper Delta recorder and PC-based software. Features 12-bit resolution, adjustable gain, 4Mb Flash RAM, and IrDA 1.1 connectivity. Powered by 2 x 1.5V AA alkaline batteries. Complies with IEC 60601-1, UL 2601-1, CAN/CSA-22.2 No 601.1-M90, JIST 1001, and AS/NZS 3200-1-0. Catheters are biocompatible; accessories supplied non-sterile.
Indications for Use
Indicated for pediatric to adult patients requiring esophageal and gastric pH data acquisition and analysis for the diagnosis of reflux disorders, and for manometric localization of the proximal Lower Esophageal Sphincter (LES) to assist in pH catheter placement.
Regulatory Classification
Identification
A gastrointestinal motility monitoring system is a device used to measure peristalic activity or pressure in the stomach or esophagus by means of a probe with transducers that is introduced through the mouth into the gastrointestinal tract. The device may include signal conditioning, amplifying, and recording equipment. This generic type of device includes the esophageal motility monitor and tube, the gastrointestinal motility (electrical) system, and certain accessories, such as a pressure transducer, amplifier, and external recorder.
{0}------------------------------------------------
## 510(k) Notification pH Testing System
4 1998 FEB
# 510(k) SUMMARY
as required per 807.92(c)
## 1. Submitters Name, Address:
Medtronic Synectics AB Renstiernas gata 12 S-116 28 STOCKHOLM Tel: (45 ) 44 57 90 00 Fax: (45 ) 44 57 90 10 Contact person for this submission: Ann-Christine Jönsson Date submission was prepared: 18th May, 1998
## 2. Trade Name, Common Name and Classification Name:
A. Trade Name: pH Testing System
B. Common Name, Classification Name, Class and Regulation Number:
| Common Name | Classification<br>Number | Class | Regulation Number |
|--------------------------------------|--------------------------|-------|-------------------|
| Digitrapper Delta | 78 FFX | II | 21 CFR 876.1725 |
| Polygram '98, pH Testing Application | 78 FFX | II | 21 CFR 876.1725 |
| pH catheters | 78 FFT | I | 21 CFR 876.1400 |
## 3. Predicate Device Identification:
The functionality and intended use of the pH Testing System is equivalent to Medtronic Synectics Digitrapper MK III (K913387) with Polygram for Windows, Base Module (K946322) and EsopHogram reflux Analysis (K961346).
{1}------------------------------------------------
#### 4. Device Description:
The pH Testing System consists of a general purpose PC running the pH Application software namned EsopHogram '98, a recording device namned Digitrapper Delta and catheters appropriate for the study to be performed. The illustration below shows the various components.
Image /page/1/Figure/3 description: The image shows a pH system with several components. A host PC with platform and pH extension software is connected via an IrDA 1.1 connection to a Delta Device with embedded software. Two catheters are connected to the Delta Device. The image provides a schematic representation of the pH system setup.
The general process for use of the system occurs in two parts. In the first part, physiological data is collected by the Delta. This is done by a technician intubating a patient with the appropriate catheter(s), connect the catheter(s) to a Digitrapper Delta, turning on the recording, and allowing the patient to proceed with their normal life. At the prescribed time, the patient returns to the technician who removes the equipment from the patient.
Following that, the technician transfers the recorded data in the Digitrapper Delta to the PC where the pH Extension software is used to analyze it. This analysis includes marking refluxes in the data traces and calculating a variety a measurements such as DeMeester score and the Boix-Ochoa scoring method. Hardcopies can be printed of the raw data as well as tailored reports.
Results of the analysis are then used by a doctor to help make a diagnosis
#### 5. Intended Use:
The pH System is intended to record, store, view and analyze esophageal and gastric pH data to diagnose reflux disorders. The pH System can also be used to locate the position of the proximal Lower EsopHageal Sphincter (LES) manometrically, to assist in the accurate placement of the pH catheter.
{2}------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
# 6. Table of Device Similarities and differences to predicate device
| Manufacturer | Medtronic Synectics AB | Medtronic Synectics AB | |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number | Predicate devices<br>MK III<br>- K 913387<br>Polygram software, base<br>module<br>- K 946322<br>pH Analysis Module<br>- K 961346 | Modified Device<br>Polyram 98' pH Testing<br>System, i.e<br>Digitrapper Delta &<br>Polygram '98,<br>pH Testing Application<br>- K 981733 | - |
| General: | Predicate devics:<br>- all | Modified Device<br>- pH System | Explanation of the differences<br>compared<br>to the Predicate devices |
| Intended Use /<br>Indication of Use | Esophageal and gastric pH data<br>acquisition and analysis to diagnose<br>reflux disorders. | Same | -- |
| Intended<br>Populations | Pediatric to Adults | Same | -- |
| Sterilization | Accessories are not supplied sterile,<br>manufacturer label the accessories<br>with cleaning instructions. | Same | -- |
| Biocompatibility | Catheters are the only part that<br>comes into contact with the<br>patients. | Same | -- |
| Technical<br>Features: | Predicate devices<br>- MK III | Modified Device<br>- Digitrapper Delta | Explanation of the differences<br>compared<br>to the Predicate devices |
| Number of<br>Channels | 4 pH, 1 p | Same | -- |
| Sampling rate | 1/4, 1/8, 1/16 Hz | Programmable through 1/16Hz<br>up to 8 Hz | This new range captures all the<br>old range while also allowing the<br>possibility to be flexible enough<br>to cope with possible future<br>demands for higher sampling<br>rates. |
| Memory Size | 96 kB | 4Mb Flash Data RAM | Larger memory size to<br>incorporate possible future<br>enhancements. |
| Event markers | 4 | 3 | Customer requirement |
| Built in memory<br>back up | 14 days | Now has Flash ram (no backup<br>required). Clock has 14 days<br>backup. | Flash RAM is non-volatile<br>therefor requires no 'backup'.<br>However the clock will keep the<br>time for convenience sake. (you<br>don't have to reset the time). |
| Power supply | 9 V alkaline battery | 2 x 1.5 AA alkaline | More available batteries |
| Measuring range | 0-9 pH | Same | -- |
| Recording time | 24, 48 & 96 hours | 24 Hours | Customer requirement with only<br>24 hours. |
| Current<br>consumption | 4 mA | 20mA | The new device runs at 3V so the<br>current increases to meet the<br>power demands. |
| Communication | 8 bit parallel Synectics Liberty or<br>Combi Interface, Serial interface to<br>computer or printer | IrDA Communication<br>(Infra Red) | This is a standard device that is<br>present on all Notebooks. The IR<br>Dongle can also be plugged into |
{3}------------------------------------------------
# 510(k) Notification pH Testing System
:
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
-----
| Resolution | 0.04 pH | Better than 0.01pH | not require a removal of the computer cover to install.<br>The MKIII has an 8 bit converter and fixed gain. The new one has 12bit resolution and adjustable gain. | | | | |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|------|----|----|
| Display | 32 character LCD LES identification display | 119 by 73 pixel graphic display. | Enhances user interface | | | | |
| Clock | Built-in real time clock | Same | -- | | | | |
| Dimension | 4.8 x 2.6 x 1.0 in | 115 x 140 x 50 | -- | | | | |
| Weight | 10.6 oz (300g) | Approx. the same | -- | | | | |
| On-line monitoring | Via own Screen | Same | -- | | | | |
| Features: | Predicate devices<br>- Polygram software, base module &<br>- pH Analysis Module, EsopHogram pH Reflux Analysis | Modified Device<br>- Polygram '98, pH Testing Application | Explanation of the differences compared to the Predicate devices | | | | |
| Signals to analyze | pH | Same | -- | | | | |
| User commands | Menu selections, keyboard combinations, screen "buttons" | Same | -- | | | | |
| Calculated parameters | 1. Maximum, Minimum<br>2. Duration of period<br>3. Number of acid refluxes<br>4. Number of long acid refluxes<br>5. Longest acid reflux<br>6. Total time pH below 4<br>7. Fraction time pH below 4<br>8. Symptom index<br>9. Symptom Association Probability<br>10. Number of alkaline shifts<br>11. Number of long alkaline shifts<br>12. Longest alkaline shift<br>13. Total tome pH above 8<br>14. Fraction time pH above 8 | Same<br>Not Available<br>Not Available | --<br>Excluded in the first version<br>Excluded in the first version | | | | |
| Scoring, Normals | 1. DeMeester & Johnson (adult)<br>2. Boix-Ochoa (pediatric)<br>3. Infant normals percentile graph (ESPGAN normals) | Same<br>Not Available<br>Not Available | --<br>Excluded in the first version<br>Excluded in the first version | | | | |
| LES identification | Bar Graph | Displays an actual manometric tracing on screen. | Enhanced LES identification. | | | | |
| Reports | Signal tracings and reports. Optionally selections only. | Same | -- | | | | |
| Patient database | Relational database with logical patient- recording relations | Same | -- | | | | |
| Additional data | User definable additional patient/recording parameters | Same | -- | | | | |
| User help system | Online help system with descriptions of procedures | Same | -- | | | | |
| Signal review method | Time - tracing based | Same | -- | | | | |
| Recording control | Real time monitoring of signals | Same | -- | | | | |
| Recording configuration | Recording<br>configuration | A template is used for each type of recording. User definable. Once used, not possible to change, | A template is used for each type of<br>recording. User definable. Once<br>used, not possible to change,<br>ensuring recording integrity | Same | Same | -- | -- |
{4}------------------------------------------------
- 7. Assessment of non-clinical performance data for equivalence:
Verifications results shows that the enhanced system performs as its predicate system.
## 8. Assessment of clinical performance data for equivalence: Fields tests are on going.
- 9. Biocompatability: Not applicable .
10. Sterilization: Not applicable
## 11. Standards and Guidances:
The pH System complies to the following standard:
- IEC 60601-1, Second edition, 1988 with Amendment 1, 1991 and Amendment 2, 1995. (EN 60601-1:1990 and Amendments A1, A11, A12 and A13)
- UL 2601-1, 1994
- CAN/CSA-22.2 No 601.1-M90
- JIST 1001
- AS/NZS 3200-1-0
{5}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
4 1999 FEB
> Ann-Christine Jönsson Regulatory Affairs Medtronic Functional Diagnostics AB 16-18 Tonsbakken DK-2740 Skovlunde DENMARK
Re: K981733
Digitrapper Delta, Model 9043G0201 and Polygram '98 pH Extension Dated: November 4, 1998 Received: November 9, 1998 Regulatory Class: II 21 CFR 876.1725/Procode: 78 FFX
Dear Ms. Jönsson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce pror to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act, You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
M. Daniel C. Stahl, M.D.
Capt. Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radioloqical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
:
510(k) Notification pH System
# Indication for Use Statement
Page 1 of 1 _
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name:__________________________________________________________________________________________________________________________________________________________________
Indications for Use:
The pH System is intended to record, store, view and analyze esophageal and gastric pH data to diagnose reflux disorders. The pH System can also be used to locate the position of the proximal Lower Esophageal Sphincter (LES) manometrically, to assist in the accurate placement of the pH catheter.
MRI Compatibility Statement:
The pH System is not compatible for use in a MRI magnetic field.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use (Per 21 CFR 801.109) | <div> <br/> </div> | OR | Over-The-Counter Use (Optional Format 1-2-96) |
|---------------------------------------|--------------------|----|-----------------------------------------------|
|---------------------------------------|--------------------|----|-----------------------------------------------|
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
| 510(k) Number | K981733/500 |
|---------------|-------------|
|---------------|-------------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.