K152109 · W.O.M. World of Medicine GmbH · FET · Oct 13, 2015 · Gastroenterology, Urology
Device Facts
Record ID
K152109
Device Name
HD-Camera HDC1000
Applicant
W.O.M. World of Medicine GmbH
Product Code
FET · Gastroenterology, Urology
Decision Date
Oct 13, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Indications for Use
The HD-Camera HDC1000 is intended to be used during diagnostic and operative endoscopic procedures to provide illumination and visualization of body cavities, hollow organs and canals and to allow for image recording and documentation. The HD-Camera HDC1000 is indicated for use with compatible rigid i flexible endoscopes and other video equipment and endoscopic accessories.
Device Story
HD-Camera HDC1000 is a video-management system for endoscopic procedures; comprises camera control unit (CCU), coupler, and camera head with integral cable. System integrates LED light source for surgical site illumination. Optical images from endoscopes/microscopes are converted to electronic signals via CCD sensor. CCU processes signals for display on monitors and storage on SDHC cards. Used in clinical settings by physicians; CCU remains outside sterile field; camera head/cable are reusable and sterilizable via STERRAD or sterile covers. Provides real-time visualization and documentation; assists clinicians in diagnostic and operative decision-making.
Clinical Evidence
Bench testing only. Comparative testing against predicate included signal-to-noise ratio (SNR), resolution, latency, shading correction accuracy, white balance accuracy, and maximum light source intensity. Electrical safety and EMC testing performed per IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-18. Sterilization validation performed per AAMI/ANSI/ISO 14937 and AAMI TIR12.
Technological Characteristics
1-Chip endoscopic video system. Components: CCU, optical coupler, camera head with CCD sensor. Integrated LED light source. Connectivity: HDMI signal outputs, SDHC card storage. Sterilization: STERRAD for camera head/cable; CCU non-sterilizable. Software: GUI-based, menu-driven settings. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, IEC 62304, AAMI/ANSI/ISO 14937.
Indications for Use
Indicated for use during diagnostic and operative endoscopic procedures requiring illumination and visualization of body cavities, hollow organs, and canals. Compatible with rigid or flexible endoscopes and endoscopic accessories.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 13, 2015
W.O.M. World Of Medicine GmbH % Susanne Raab Regulatory Consultant Susan Raab 1480 Cambridge Street Cambridge, MA 02139
Re: K152109
> Trade/Device Name: HD-Camera HDC1000 Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FET, NTN Dated: July 24, 2015 Received: July 29, 2015
Dear Susanne Raab,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Herbert P. Lerner -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152109
Device Name HD-Camera HDC1000
Indications for Use (Describe)
The HD-Camera HDC1000 is intended to be used during diagnostic and operative endoscopic procedures to provide illumination and visualization of body cavities, hollow organs and to allow for image recording and documentation. The HD-Camera HDC1000 is indicated for use with compatible rigid or flexible endoscopes and other video equipment and endoscopic accessories.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span></span> New construction/Modification | <span></span> Emergency/Standby Use |
|---------------------------------------------|-------------------------------------|
|---------------------------------------------|-------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 SubpartC)
#### CONTINUE ON A SEPARATE PAGE IFNEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for WOM, which is a stylized circle with a red triangle at the bottom. Below the logo is the text "WOM" in a sans-serif font. Below that is the text "HD-Camera HDC1000" and "510(k) Premarket Notification". The text is in a smaller font than the logo.
K152109
# 510(k) SUMMARY OF SAFETY & EFFECTIVENESS
## 1) General Information:
| Submitter: | W.O.M. WORLD OF MEDICINE GmbH |
|--------------------------|-------------------------------------|
| | Salzufer 8 |
| | 10587 Berlin |
| | Germany |
| | Phone: +49 30 39981 594 |
| | Fax: +49 30 39981 593 |
| FDA Registration Number: | 3001556604 |
| Contact Person: | Susanne Raab |
| | Regulatory Consultant |
| | 1480 Cambridge Street |
| | Cambridge, MA 02139 |
| | Phone: (617) 547-0628 |
| | Fax:<br>(617) 520-2401 |
| | e-mail: sbraab@comcast.net |
| Date Prepared: | July 24, 2015 |
| 2) Proposed Device: | |
| Trade Name: | HD-Camera HDC1000 |
| Common Name: | Endoscopic Video Imaging |
| | System/Component, Gastroenterology- |
| | Urology |
| Classification Name: | Endoscope and accessories |
| Regulation Number: | 21 C.F.R. 876.1500 |
| Regulatory Class: | II |
| Product Code: | FET. NTN |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the WOM logo, which is a stylized symbol above the letters "WOM". Below the logo is the text "HD-Camera HDC1000" followed by "510(k) Premarket Notification". The logo consists of a circle with a design inside, and a small triangle at the bottom of the design.
## 3) Predicate Devices:
| Trade Name: | OVS1 Video System |
|----------------------|------------------------------------------|
| 510(k) Number: | K024291 |
| Classification Name: | Laparoscope, General and Plastic Surgery |
| Regulation Number: | 21 C.F.R. 888.1100 |
| Regulatory Class: | II |
| Product Code: | GCJ, FCW |
### 4) Device Description:
The HD-Camera HDC1000 is a video-management system that consists of a camera, an integrated light source and a capture module for storing and compressing video and still images. It is used in conjunction with an endoscope or microscope, a light cable and other endoscopic accessories to illuminate surgical sites and to view the surgical sites on video monitors during diagnostic and operative endoscopic procedures. The camera of the proposed device consists of a camera control unit (CCU), a coupler, and a camera head with an integral cable that connects to the camera control unit (CCU). The optical image is transferred from the surgical site to the camera head by compatible rigid and flexible endoscopes that are attached to the camera head using an optical coupler. The camera head including camera cable and the coupler are reusable devices and must be sterilized using the STERRAD ® advanced sterilization process or may be used with a sterile cover or sleeve. The camera control unit (CCU) is not intended to enter the sterile field, and cannot be sterilized.
## 5) Intended Use:
The HD-Camera HDC1000 is intended to be used during diagnostic and operative endoscopic procedures to provide illumination and visualization of
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for WOM, which is a circle with a stylized figure inside and a red triangle at the bottom. Below the logo is the text "WOM" in a larger font. Below that is the text "HD-Camera HDC100" and "510(k) Premarket Notification".
body cavities, hollow organs and canals and to allow for image recording and documentation. The HD-Camera HDC1000 is indicated for use with compatible rigid i flexible endoscopes and other video equipment and endoscopic accessories.
### 6) Comparison of Technological Characteristics:
The proposed device HD-Camera HDC1000 and the predicate device OVS1 Video System incorporate the same basic design. Specifically, both the HD-Camera HDC1000 and the predicate device are 1-Chip cameras that consist of three main components: a camera head, various optical couplers, and a camera control unit (CCU). In addition, the HD-Camera HDC 1000 and the predicate device are designed with an integrated LED light source.
Both the HD-Camera HDC1000 and the predicate device receive optical images from a variety of scopes that are attached to the camera head via a coupler or direct attachment and use a CCD image sensor to convert the optical image into an electronic signal. The camera head of the HD-Camera HDC1000 and the predicate device incorporate a button keypad to control the system itself and any other connected video equipment.
The camera control unit (CCU) of the proposed device and the predicate device are software controlled and are equipped with a graphical user interface (GUI). The GUI of both the HD-Camera HDC1000 and the predicate device is designed with a menu key that allows the user to select and change various preprogrammed settings and to make changes in the assignment of the buttons on the keypad of the camera head.
The differences in the technological characteristics of both the proposed device HDC1000 and Predicate Device are minor and do not raise new questions of
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for WOM, which is a circle with a stylized figure inside and a red triangle at the bottom. Below the logo is the text "WOM" in a larger font. Below that is the text "HD-Camera HDC100" and "510(k) Premarket Notification".
safety and effectiveness. The differences between the HDC1000 and the Predicate Device are the following:
- Unlike the camera head unit of the Predicate Device, the camera head unit . of the proposed device can be sterilized.
- The button keypad of the camera head of the Predicate Device is designed . with three buttons. The proposed device HDC1000 is designed with two camera head buttons.
- . The camera head of the Predicate device incorporates a digital zoom feature whereas the proposed device HDC1000 is to be used with a zoom lens.
- The chip pixels of the camera head of the devices differ slightly. The . camera head of the proposed device incorporates a chip with 1280 x 720 pixels versus 1920 x 1080 pixels of the predicate device.
- The proposed device HDC1000 has a frame rate of 30fps and the ● Predicate Device of 60fps.
- . The CCU of the Predicate Device contains a touchscreen with a graphical interface including display. The HDC1000 is designed with a plastic foil keypad with a graphical interface and adjustments are only displayed on the monitor.
- The CCU of the Predicate Device offers fiberscope and ENT in addition . to the above mentioned pre-programmed procedure profiles.
- . The CCU of the proposed device incorporates two HDMI signal outputs and the Predicate device one DVI (HDMI), one 3G-SDI, one USB 2.0 and one Ethernet signal output.
- The resolution of the video output of the CCU of the HDC1000 is 1280 x . 720 pixels versus 1920 x 1080 pixels of the Predicate device.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for WOM, which is a circle with a design inside and a small triangle at the bottom. Below the logo, the text "WOM" is written in a bold font. The text "HD-Camera HDC1000 510(k) Premarket Notification" is written below the logo in a smaller font.
- The CCU of the HDC1000 is designed to be used with a SDHC card as a ● storage media and that of the Predicate Device with a USB flash thumb.
# 7) Performance Data:
Electrical safety and electromagnetic compatibility testing was performed by independent laboratories in accordance with the following standards:
- · IEC 60601-1:2005 Medical Electrical Equipment Part 1: General Requirements for Basic Safety and Essential Performance;
• IEC60601-1-2:2007 - Medical Electrical Equipment – Part 1-2: General Requirements for Basic Safety and Essential Performance – Collateral Standard: Electromagnetic Compatibility - Requirements and Tests;
· IEC60601-2-18:2009 - Medical Electrical Equipment - Part 2-18: Particular Requirements for the Basic Safety and Essential Performance of Endoscopic Equipment.
Test results demonstrate that the proposed device conforms to the above standards.
The software was developed, tested, and verified in accordance with the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and in accordance with the following standard:
· IEC62304:2006 - Medical Device Software - Software Life Cycle Processes.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for WOM, which is a circle with a design inside and a triangle at the bottom. Below the logo is the text "WOM" in a larger font. Below that is the text "HD-Camera HDC1000" and "510(k) Premarket Notification".
The cleaning instructions provided in the labeling for the camera head and cable were validated. Sterilization validation testing was performed in accordance with the below standard:
· AAMI/ANSI/ISO 14937:2009 - Sterilization of Health Care Products -General Requirements for Characterization of a Sterilizing Agent and the Development, Validation and Routine Control of a Sterilization Process for Medical Devices;
· AAMI TIR12:2010 - Designing, Testing and Labeling reusable medical devices for reprocessing in health care facilities: A Guide for Medical Device Manufacturers.
Design verification testing of the HD-Camera HDC1000 demonstrates that the device performs as intended and that the performance does not raise new questions of safety and effectiveness.
Finally, comparative bench testing was performed to demonstrate that the performance of the HD-Camera HDC1000 is substantially equivalent to that of the predicate device OVS1 Video System. The comparative bench test included signal to noise ratio (SNR), resolution, latency, shading correction accuracy, white balance accuracy and maximum light source intensity.
## 8) Conclusions:
Based on the same intended use, the same basic technological characteristics and performance testing, the HD-Camera HDC1000 is substantially equivalent to the predicate device OVS1 Video System. The differences between the
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for WOM, which is a circle with a stylized figure inside and a small triangle at the bottom. Below the logo is the text "WOM" in a larger font. Below that is the text "HD-Camera HDC1000" and "510(k) Premarket Notification".
proposed device and the predicate device do not raise new questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.