K020301 · Vista Medical Technologies, Inc. · FET · Feb 12, 2002 · Gastroenterology, Urology
Device Facts
Record ID
K020301
Device Name
MODIFICATION TO VISTA STEREOSCOPE SYSTEM
Applicant
Vista Medical Technologies, Inc.
Product Code
FET · Gastroenterology, Urology
Decision Date
Feb 12, 2002
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.1500
Device Class
Class 2
Indications for Use
The Vista StereoScope System is intended for the use in endoscopic procedures and all types o video assisted procedures, including general endoscopic and laparoscopic, thoracic, anterior an posterior spinal and as an aid in visualization of cardiac structures
Device Story
Vista StereoScope System provides visualization of body cavities, organs, or canals via natural or manmade orifices. System components include Vista Stereoscope Camera Head, Vista Stereo Endoscope, and 3D Camera Control Unit (CCU). Device integrates with commercially available light sources and video monitors/overhead displays. Used by clinicians in endoscopic and video-assisted procedures (e.g., laparoscopic, thoracic, spinal, cardiac). Modified coupler allows sterilization of camera head assembly. Provides 3D visualization to assist surgeons during minimally invasive procedures.
Clinical Evidence
No clinical data provided; substantial equivalence based on design and performance testing.
Technological Characteristics
System includes stereoscope camera head, stereo endoscope, and 3D camera control unit. Features a modified coupler for sterilization. Operates as an endoscopic visualization system using external light sources and video displays. No specific materials or software algorithm details provided.
Indications for Use
Indicated for use in endoscopic and video-assisted procedures, including general endoscopic, laparoscopic, thoracic, anterior and posterior spinal procedures, and as an aid in visualization of cardiac structures. No specific age or gender restrictions provided.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services USA. The logo features a stylized image of an eagle with its wings spread, surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement. The eagle is composed of three curved lines, and the text is in a bold, sans-serif font.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Vista Medical Technologies Mr. Graham A.L. Baillie Manager, Quality Assurance and Service 134 Flanders Road Westborough. MA 01581
JUL 2 7 2015
Re: K020301 Trade/Device Name: Vista Stereoscope System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FET. GCJ Dated (Date on orig SE ltr): January 28, 2002 Received (Date on orig SE Itr): January 29, 2002
Dear Mr. Baillie,
This letter corrects our substantially equivalent letter of February 12, 2002.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{1}------------------------------------------------
Page 2 -
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
z/www.fda.gov/MedicalDevices/ResoucesforIndustry/default.htm
Sincerely yours.
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive. Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### 11) Indications for Use Statement
Statement of Intended Use
510(k) Number (if Known):
K020301
Device Name:
Vista Stereoscope System
Indications For Use:
The Vista StereoScope System is intended for the use in endoscopic procedures and all types o video assisted procedures, including general endoscopic and laparoscopic, thoracic, anterior an posterior spinal and as an aid in visualization of cardiac structures
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE I NEEDED)
### Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use _____________________________________________________________________________________________________________________________________________________________
Over-The-Counter Use (Optional Format 1-2-96)
Muriam C. Provost
Division of OM
OR
(Division Sign-Off) Division of General, Restorative and Neurological Devices
510(k) Number K020361
{3}------------------------------------------------
K020301
# FEB 1 2 2002
Image /page/3/Picture/2 description: The image shows a logo for Avista Medical Technologies. The logo consists of a stylized letter "A" with a triangle on the left side and the rest of the letters forming the word "AVISTA". Below the word "AVISTA" is the phrase "Medical Technologies". The logo is in black and white.
#### 9) 510(k) Summary
510(k) Summary for Vista Stereoscope System
A. Sponsor
Vista Medical Technologies 134 Flanders Road Westborough, MA 01581
#### B. Contact Name
Graham A. L. Baillie Manager, Quality Assurance and Service Vista Medical Technologies Phone: (508) 366-3668 ext. 8279 Facsimile: (508) 366-8858
#### C. Device Name
Vista Stereoscope System
#### D. Predicate Device(s)
Vista Stereoscope System (K990635); Vista Single Chip Video Camera (K971373)
#### E. Device Description
The Vista StereoScope System is a device used to allow observation in body cavities, organs, or canals through manmade or natural orifices. It is designed for use in all types of endoscopic and endoscopic assisted procedures. The system is supplied as a Vista Stereoscope Camera Head, Vista Stereo Endoscope and a 3 D Camera control unit (CCU). The device is designed to work with commercially available light sources and video monitor overhead mounted displays. The modified coupler enables the camera head assembly to be sterilized.
#### F. Intended Use
The Vista StereoScope System is intended for the use in endoscopic procedures and all types of video assisted procedures, including general endoscopic and laparoscopic, thoracic, anterior and posterior spinal and as an aid in visualization of cardiac structures
#### G. Substantial Equivalence
The proposed modified Vista Stereoscope System is substantially equivalent to the currently legally marketed Vista Stereoscope System (K990635) in terms of intended use, operating principle, basic design, and shelf life. Testing demonstrates that the modifications proposed herein do not adversely effect safety and effectiveness.
Vista Modical Tochnologics
134 Flanders Road Westborough, MA D1581
Ph 508 366-8882 Fx 508 366-8858
13