EG-3800T, VIDEO GASTROSCOPE

K962460 · Pentax Precision Instrument Corp. · FDS · Sep 16, 1996 · Gastroenterology, Urology

Device Facts

Record IDK962460
Device NameEG-3800T, VIDEO GASTROSCOPE
ApplicantPentax Precision Instrument Corp.
Product CodeFDS · Gastroenterology, Urology
Decision DateSep 16, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The EG-3800T, Video Gastroscope, is intended to provide optical visualization (via a video monitor) of, and therapeutic access to, the Upper Gastrointestinal Tract. The Upper Gastrointestinal Tract includes, but is not limited to, the organs; tissues; and subsystems: Esophagus, Stomach, Duodenum, and Small Bowel. The instrument is introduced per orally when indications consistent with the requirement for procedure(s) are observed in Adult and Pediatric patient populations.

Device Story

EG-3800T is a video gastroscope for upper GI tract visualization and therapy. Input: optical images captured via CCD sensor; illumination provided by 300W xenon lamp in external video processor. Operation: light transmitted via fiber bundles; CCD collects image data synchronized with RGB color wheel strobes; video processor compiles full-color frames. Output: real-time video displayed on monitor. Used in clinical settings by physicians. Features flexible insertion tube, angulation controls, air/water delivery, suction, and two working channels for biopsy forceps. Benefits: enables minimally invasive diagnostic and therapeutic access to GI organs.

Clinical Evidence

No clinical data; substantial equivalence based on bench testing and technical comparison.

Technological Characteristics

Flexible video gastroscope with CCD imaging sensor and fiber optic illumination. Connects to external video processor (300W xenon lamp, RGB color wheel). Features two working channels, angulation controls, air/water/suction ports. System requires software-controlled video processor for image compilation and display.

Indications for Use

Indicated for adult and pediatric patients requiring optical visualization and therapeutic access to the upper gastrointestinal tract, including the esophagus, stomach, duodenum, and small bowel.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{1} AUG-07-96 WED 11:54 PENTAX SESC FAX NO. 305 430 8505 P. 01 DOCUMENT NUMBER: KS034.B7 DATE PREPARED: 08-07-96 PREPARED BY: PAUL SILVA PAGE 1 OF 1 510(K) SUMMARY DEVICE: EG-3800T, VIDEO GASTROSCOPE 1. SUBMITTER INFORMATION: Pentax Precision Instrument Corporation 30 Ramland Road Orangeburg, NY, 10962 TEL: (914)-365-0700 SEP 16 1996 K962460 2. NAME OF DEVICE: Trade Name: EG-3800T, Video Gastroscope Classification Name: Gastroscope, Gastro- Urology (78FDS), 876.1500 3. PREDICATED DEVICE(S) INFORMATION: 1. FG-36TA, Fiber Gastroscope Pentax K833924 2. EC-3800TL, Video Colonoscope Pentax K951579 3. EPM-3300, Video Processor Pentax K934918 4. DEVICE DESCRIPTION: The EG-3800T, Video Gastroscope, must be used with a Video Processor (a software controlled device). The endoscope has a flexible insertion tube, a control body, and umbilicus. The umbilicus provides connection to the video processor. The control body includes controls for up/ down/ left/ right angulation, air/water delivery, suction, two (2) accessory inlet ports, and a suction channel selector. The device contains light carrying bundles, to illuminate the body cavity, and a charge couple device (CCD) to collect image data. The instrument contains two (2) working channels through which biopsy devices, or other devices, may be introduced (the instrument is supplied with two biopsy forceps). The Video Processor contains a 300 watt xenon lamp which provides white light that is filtered, via a Red, Green, Blue color wheel, and is focused at the connected video endoscope lightguide prong. The endoscope light carrying bundles present the color strobes to the body cavity and the CCD collects image data for each strobe of color. The video processor stores the CCD information until all three color strobes are completed and a full color frame is compiled. Image data and other screen display information are formatted and presented to the video outputs of the video processor for display. 5. INTENDED USE: The EG-3800T, Video Gastroscope, is intended to provide optical visualization (via a video monitor) of, and therapeutic access to, the Upper Gastrointestinal Tract. The Upper Gastrointestinal Tract includes, but is not limited to, the organs; tissues; and subsystems: Esophagus, Stomach, Duodenum, and Small Bowel. The instrument is introduced per orally when indications consistent with the requirement for procedure(s) are observed in Adult and Pediatric patient populations. 6. COMPARISON TO PREDICATED DEVICE(S): The submission for substantial equivalence included EG-3800T literature including specifications, the identification of standard set components, and identification of optional accessories, comparison tables were provides to illustrate the comparisons to the predicated devices. The submission for substantial equivalence was not based on an assessment of clinical performance data.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...