As a guiding needle in obtaining core biopsy samples from soft tissue such as liver, kidney, spleen, lymph nodes and various soft tissue lesions.
Device Story
Gangi-SoftGuard and Gangi-HydroGuard Coaxial Needles are sterile, single-use, manually operated introducer needles for soft tissue biopsy. The device consists of a cannula, a trocar stylet, and a spring-loaded blunt-tip stylet. During insertion, the blunt tip shields the sharp cannula tip to prevent trauma. Upon encountering resistance, the blunt tip retracts, exposing the sharp cannula for tissue penetration. Once resistance decreases, the blunt tip automatically springs back to shield the cannula. The Gangi-HydroGuard variant includes a hollow blunt stylet with a distal hole and a Luer hub, allowing fluid injection without removing the stylet. Used in clinical settings under imaging guidance (excluding MRI) by trained physicians. The device facilitates biopsy procedures by providing a protected path for biopsy instruments to reach targeted lesions.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified through mechanical testing (penetration force, spring force), biocompatibility (ISO 10993), and compliance with ISO 7864 (sterile needles), ISO 9626 (tubing), and ISO 594 (Luer fittings).
Technological Characteristics
Manually operated coaxial introducer needle. Materials: Stainless steel AISI 304 and AISI 302. Features spring-loaded blunt-tip stylet mechanism for non-traumatic insertion. Luer lock fittings per ISO 594. Sterile (Ethylene Oxide). Single-use. No electronic components or software.
Indications for Use
Indicated for use as a guiding needle in obtaining core biopsy samples from soft tissue, including liver, kidney, spleen, lymph nodes, and various soft tissue lesions.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 18, 2018
AprioMed AB Katrin Svensson Director Quality & Regulatory Affairs Virdings Allé 28 Uppsala, 75450 Sweden
Re: K181756 Trade/Device Name: Gangi-SoftGuard Coaxial Needle Gangi-HydroGuard Coaxial Needle Regulation Number: 21 CFR§ 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: II Product Code: FCG Dated: November 12, 2018 Received: November 16, 2018
Dear Katrin Svensson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Daniel G. Walter Jr -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K181756
Device Name Gangi-SoftGuard Coaxial Needle Gangi-HydroGuard Coaxial Needle
Indications for Use (Describe)
As a guiding needle in obtaining core biopsy samples from soft tissue such as liver, kidney, spleen, lymph nodes and various soft tissue lesions.
| <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;">□</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Type of Use (Select one or both, as applicable)
{3}------------------------------------------------
#### 5 510(K) SUMMARY
As Required by 21 CFR 807.92(c) 510(k) Summary
#### Submitter/510(k) Holder 5.1
AprioMed AB Virdings Allé 28 Uppsala, SWEDEN 754 50
Contact Person: Katrin Svensson, Director Quality & Regulatory Affairs Telephone: +46 (0) 73- 345 14 40
Date Prepared: December 12, 2018
#### 5.2 Device
| Proprietary Name: | Gangi-SoftGuard Coaxial Needle<br>Gangi-HydroGuard Coaxial Needle |
|----------------------|-------------------------------------------------------------------|
| Common/Usual Names: | Biopsy Needle |
| Regulation number: | 21 CFR 876.1075 |
| Classification Name: | Gastroenterology-Urology Biopsy Instrument |
| Class: | II |
| Product Code: | FCG |
| 510(k) number: | K181756 |
#### Predicate Devices and Reference Device identification 5.3
### Predicate device:
| Proprietary Name: | Bard TruGuide Disposable Coaxial Biopsy Needle |
|----------------------|------------------------------------------------|
| Common/Usual Names: | Biopsy Needle |
| Regulation number: | 21 CFR 876.1075 |
| Classification Name: | Gastroenterology-Urology Biopsy Instrument |
| Class: | II |
| 510(k) number: | K936194 |
| Product Code: | FCG |
### Reference device:
| Proprietary Name: | Surgineedle™ Pneumoperitoneum Needle |
|----------------------|--------------------------------------|
| Common/Usual Names: | Pneumoperitoneum Needle |
| Regulation number: | 21 CFR 876.1500 |
| Classification Name: | Endoscope and accessories |
| Class: | II |
| 510(k) number: | K863330 |
| Product Code: | FHO |
{4}------------------------------------------------
#### 5.4 Device Description
Gangi-SoftGuard Coaxial Needle together with Gangi-HydroGuard Coaxial Needle forms the Gangi Coaxial Needle family. Gangi-SoftGuard Coaxial Needle and Gangi-HydroGuard Coaxial Needle are manually operated, sterile, single use coaxial guiding needles with a spring loaded blunt-tip stylet, to be used as a guiding needle for biopsy instruments in procedures in soft tissue. Gangi-SoftGuard Coaxial Needle and Gangi-HydroGuard Coaxial Needle has the same intended use.
Coaxial Needle and Gangi-SoftGuard Gangi-HydroGuard Coaxial Needle are gastroenterological, manually operated, sterile, single use coaxial introducer needles. Both Gangi-SoftGuard Coaxial Needle and Gangi-HydroGuard Coaxial Needle consists of three parts: a needle (cannula) provided with a trocar stylet, and a spring-loaded blunt tip stylet. The difference from Gangi-SoftGuard Coaxial Needle is that Gangi-HydroGuard Coaxial Needle has a Luer hub located on the blunt stylet housing and the blunt stylet is hollow and has a distal tip hole. This enables injection of fluid without the need to remove the blunt stylet.
#### Technological Characteristics ર્સ્ટ
Gangi-SoftGuard Coaxial Needle and Gangi-HydroGuard Coaxial Needle are delivered with the trocar stylet inserted in the needle (cannula). When the trocar is exchanged to the blunt stylet, the blunt tip protrudes out from the distal end of the needle (cannula) which makes the coaxial needle tip non-traumatic. shielding the surroundings from the sharp needle (cannula tip).
The position of the coaxial needle is monitored using imaging technique. If resistance or hard surface is met when using the device, the blunt tip retracts into the needle (cannula), exposing the sharp needle (cannula) tip and enabling cutting through tissue. The retraction of the blunt tip can be prevented by pressing the proximal part of the stylet (covered by a plastic cap) or the proximal Luer hub on the housing.
After cutting through the tissue with the sharp needle (cannula) and the resistance is reduced, the hollow blunt stylet tip springs back i.e., the non-traumatic tip will once again shield the surrounding tissue from the sharp needle (cannula tip).
When the targeted lesion has been reached, Gangi-SoftGuard Coaxial Needle and Gangi-HydroGuard Coaxial Needle enables biopsy to be performed per normal hospital routines.
The materials used for construction of the needles are typical for this type of medical device. The only material in direct patient contact is stainless steel AISI 304 and stainless steel AISI 302. The needles are designed in accordance with ISO 7864 - Sterile hypodermic needles for single use.
{5}------------------------------------------------
#### 5.6 Indications for Use
Gangi-SoftGuard Coaxial Needle is intended as a guiding needle in obtaining core biopsy samples from soft tissue such as liver, kidney, spleen, lymph nodes and various soft tissue lesions.
Gangi-HydroGuard Coaxial Needle is intended as a guiding needle in obtaining core biopsy samples from soft tissue such as liver, kidney, spleen, lymph nodes and various soft tissue lesions.
#### Substantial Equivalence 5.7
AprioMed AB has determined that Gangi-SoftGuard Coaxial Needle and Gangi-HydroGuard Coaxial Needle are substantially equivalent to the predicate device based on the Indications for Use, compliance with internationally recognized design and performance standards, material specifications, use environments, and performance. The differences between subject devices and the predicate device used for the same Indications for Use do not raise new issues of safety and effectiveness.
A substantial equivalence comparison table of subject devices, Gangi-SoftGuard Coaxial Needle and Gangi-HydroGuard Coaxial Needle and the predicate device, Bard TruGuide Disposable Coaxial Biopsy Needle (K936194) is provided in Table 5-1.
{6}------------------------------------------------
| Table 5-1: Substantial equivalence table | | |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Parameter | Subject Devices | Predicate Device |
| Proprietary Name | Gangi-SoftGuard Coaxial Needle<br>Gangi-HydroGuard Coaxial Needle | Bard TruGuide Disposable Coaxial Biopsy Needle |
| 510(k) | This 510(k) application | K936194 |
| Device Classification<br>Name | Biopsy needle | Biopsy needle |
| Product Code | FCG | FCG |
| Regulation<br>description | Gastroenterology-Urology Biopsy<br>Instruments | Gastroenterology-Urology Biopsy<br>Instruments |
| Regulation Number | 21 CFR 876.1075 | 21 CFR 876.1075 |
| Intended Use/<br>Indications for Use | As a guiding needle in obtaining core<br>biopsy samples from soft tissue such as<br>liver, kidney, spleen, lymph nodes and<br>various soft tissue lesions. | The Coaxial biopsy needle guide is<br>intended for use as a guiding needle in<br>obtaining core biopsy samples from soft<br>tissue such as liver, kidney, spleen, lymph<br>nodes and various soft tissue lesions. |
| Device type | Coaxial introducer needle with a spring<br>loaded blunt tip stylet/hollow blunt tip stylet | Coaxial introducer needle with a blunt tip<br>stylet |
| Visualization<br>technique | Conventional imaging guidance equipment<br>excluding MRI | Conventional imaging guidance equipment<br>excluding MRI |
| Single Use | Yes | Yes |
| Sterile | Yes, Ethylene Oxide | Yes, Ethylene Oxide |
| Needle material | Stainless Steel | Stainless Steel |
| Needle Hub | Luer | Luer |
| Needle size | Gangi-SoftGuard Coaxial Needle:<br>12G-17G<br>7,1-17,1 cm<br>Gangi-HydroGuard Coaxial Needle:<br>15G-17G<br>12,1-17,1 cm | 11G-19G<br>7.0-17.8 cm |
| Needle penetration<br>depth | 17.1 cm free length | 17.8 cm free length |
| Anatomical sites | Liver, kidney, spleen, lymph nodes and<br>various soft tissue lesions | Liver, kidney, spleen lymph nodes and<br>various soft tissue lesions |
Table 5-1: Substantial equivalence table
{7}------------------------------------------------
#### 5.7.1 Intended Use/Indication for Use
AprioMed has determined that the intended use of Gangi-HydroGuard and Gangi-SoftGuard coaxial biopsy needle are substantially equivalent to the predicate device.
#### 5.7.2 Materials
AprioMed has determined that the Gangi-SoftGuard Coaxial Needle and Gangi-HydroGuard Coaxial Needle are substantially equivalent to the predicate device based on materials. Gangi-SoftGuard Coaxial Needle and Gangi-HydroGuard Coaxial Needle and the predicate is constructed with medical grade stainless steel. Gangi Coaxial Needle material also fulfill the, by FDA, recognized consensus standard, ISO 9626 Stainless Steel Needle Tubing for the Manufacture of Medical Devices.
#### 5.7.3 Use Environment
Gangi-SoftGuard Coaxial Needle and Gangi-HydroGuard Coaxial Needle is intended to be used in a standard clinical setting for percutaneous imaging-guided biopsy sampling excluding in Magnetic Resonance Imaging (MRI) equipment. The users are physicians trained in percutaneous image-guided biopsy sampling of soft tissue. AprioMed has determined that Gangi-SoftGuard Coaxial Needle and Gangi-HydroGuard Coaxial Needle are substantially equivalent to the predicate device based on the use environment.
#### 5.7.4 Size, stylet size and penetration depth
AprioMed has determined that the sizes, stylet size and penetration depth of Gangi-SoftGuard Coaxial Needle and Gangi-HydroGuard Coaxial Needle is substantially equivalent to the predicate device.
#### 5.7.5 Tip design
AprioMed has determined that the tip design of Gangi-SoftGuard Coaxial Needle and Gangi-HydroGuard Coaxial Needle is substantially equivalent to the predicate device.
#### 5.7.6 Penetration force
AprioMed has determined that Gangi-SoftGuard Coaxial Needle and Gangi-HydroGuard Coaxial Needle are substantially equivalent to the predicate device with respect to needle penetration.
#### 5.7.7 Safety Features
AprioMed believes that Gangi-SoftGuard Coaxial Needle and Gangi-HydroGuard Coaxial Needle is substantially equivalent to the predicate device in respect to safety features. Neither Gangi-SoftGuard Coaxial Needle, Gangi-HydroGuard Coaxial Needle nor the predicate has any specific safety features.
#### Summary of non-clinical and performance testing 5.8
Testing in the form of bench testing was performed to evaluate the performance and functionality to evaluate the device against requirements specification. The devices have been subjected to compliance testing to voluntary standards e.g. ISO 7864, ISO 9626, ISO 10993-7, ISO 10993-1, ISO 11607-1.
Results from verification and validation testing demonstrates that conformance to applicable technical requirements specification and user needs have been met and substantial equivalence has been demonstrated.
{8}------------------------------------------------
#### 5.8.1 Biocompatibility testing
The material used in Gangi-SoftGuard Coaxial Needle and Gang-HydroGuard Coaxial Needle has been tested for biocompatibility in accordance with ISO 10993 for externally communicating devices in contact with tissue for 24 hours or less. The material in direct patient contact is the cannula needle and stylets made of stainless steel AISI 304 and stainless steel AISI 302. Gangi-SoftGuard Coaxial Needle and Gang-HydroGuard Coaxial Needle have successfully been tested for:
- Cytotoxicity ।
- Sensitization ।
- Intracutaneously irritation -
- Acute Systemic Toxicity -
- -Pyrogenicity
The test results verify that the biocompatibility criteria given in ISO 10993 are fulfilled. AprioMed concludes that Gangi-SoftGuard Coaxial Needle and Gangi-HydroGuard Coaxial Needle is non-toxic and biocompatible.
#### 5.8.2 Performance testing - Bench
5.8.2.1 Qualification metal tubing
The stainless-steel tubing is qualified in compliance with ISO 9626:2016 Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods, and conformity demonstrated.
#### Qualification of Needle component 5.8.2.2
The needle used for Gangi-SoftGuard Coaxial Needle and Gangi-HydroGuard Coaxial Needle are qualified in compliance with ISO7864:2016 Sterile hypodermic needles for single use -Requirements and test methods.
#### 5.8.2.3 Conical Fittings
The Luer lock fittings used for Gangi-SoftGuard Coaxial Needle and Gangi-HydroGuard Coaxial Needle are qualified in compliance with ISO 594 Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment, and conformity demonstrated.
#### 5.8.2.4 Mechanical performance
The mechanical performance testing of Gangi-SoftGuard Coaxial Needle and Gangi-HydroGuard Coaxial Needle has passed all design requirements.
#### 5.8.2.5 Verification of penetration force and spring
The penetration force and the spring force have been evaluated and passed all design requirements.
#### Summary and Conclusion 5.9
Based on the above discussion, the non-clinical testing supports the safety of the device and demonstrates that the Gangi-SoftGuard Coaxial Needle and Gangi-HydroGuard Coaxial Needle should perform as intended in the specified use conditions. The data supports the substantial equivalence determination between the Gangi-SoftGuard Coaxial Needle and Gangi-HydroGuard Coaxial Needle and the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.