Single Use Aspiration Needle NA-U401SX-4025/NA-U401SX-4025N
K180668 · Olympus Medical Systems Corp. · FCG · Nov 15, 2018 · Gastroenterology, Urology
Device Facts
Record ID
K180668
Device Name
Single Use Aspiration Needle NA-U401SX-4025/NA-U401SX-4025N
Applicant
Olympus Medical Systems Corp.
Product Code
FCG · Gastroenterology, Urology
Decision Date
Nov 15, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1075
Device Class
Class 2
Indications for Use
This instrument has been designed to be used with ultrasound endoscopes for ultrasound guided fine needle aspiration(FNA) of submucosal and extramural lesions of the gastrointestinal tract.
Device Story
Single Use Aspiration Needle (ViziShot 2) facilitates ultrasound-guided fine needle aspiration (FNA) of GI tract lesions. Device comprises handle, insertion section (sheath, needle, stylet), and optional medallion syringe. Operated by physician during endoscopic procedure; needle tube extends from sheath to puncture target tissue; syringe attached to handle port provides aspiration. Needle features dimpled surface for echo enhancement under ultrasound visualization. Device enables tissue/cell collection for diagnostic purposes; benefits patient by providing minimally invasive biopsy capability. Used in clinical settings with compatible ultrasound endoscopes.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified through bench testing including insertion, flexibility, piercing, ultrasound visibility, needle extraction/retraction, aspiration, withdrawal, handle locking force, and depth limitation. Sterilization and shelf-life validated per ASTM F1980-16 and ISO 11135:2014. Biocompatibility evaluated per ISO 10993-1:2009.
Technological Characteristics
Single-use aspiration needle with dimpled echo-enhanced needle tube. Materials evaluated per ISO 10993-1. Sterilization per ISO 11135:2014. Shelf-life per ASTM F1980-16. Mechanical operation via handle-actuated needle slider. Compatible with ultrasound endoscopes for visualization. No software or electronic components.
Indications for Use
Indicated for ultrasound-guided fine needle aspiration (FNA) of submucosal and extramural gastrointestinal tract lesions in patients requiring tissue or cell sampling via ultrasound endoscope.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Predicate Devices
NA-U401SX-4022 (Single Use Aspiration Needle NA-U401SX) (K160098)
Reference Devices
ECHO-HD-25-EBUS-O (Echotip Ultra Endobronchial High Definition Ultrasound Biopsy Needle) (K160229)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 15, 2018
Olympus Medical Systems Corp. % Sheri L. Musgnung Manager, Regulatory Affairs Olympus Corporation of the Americas 3500 Corporate Parkway P.O. Box 610 Center Valley, PA 18034-0610
Re: K180668
> Trade/Device Name: Single Use Aspiration Needle NA-U401SX-4025/NA-U401SX-4025N Regulation Number: 21 CFR§ 876.1075 Regulation Name: Gastroenterology-Urology Biopsy Instrument Regulatory Class: II Product Code: FCG Dated: October 11, 2018 Received: October 15, 2018
Dear Sheri L. Musgnung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Glenn B. Bell -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K180668
Device Name
Single Use Aspiration Needle NA-U401SX-4025/NA-U401SX-4025N
Indications for Use (Describe)
This instrument has been designed to be used with ultrasound endoscopes for ultrasound guided fine needle aspiration(FNA) of submucosal and extramural lesions of the gastrointestinal tract.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are thick and sans-serif, giving the word a strong and modern appearance. A thin yellow line is visible beneath the word, adding a subtle accent to the logo.
October 10, 2018
# 5. 510(k) Summary
# 5.1 GENERAL INFORMATION
- 510(k) Submitter: OLYMPUS MEDICAL SYSTEMS CORP. 2951 Ishikawa-cho, Hachioji-shi, Tokyo, Japan 192-8507
- Sheri L. Musgnung ■ Contact Person: Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610, USA Phone: 484-896-3147 Fax: 484-896-7128 Email: sheri.musgnung(@olympus.com
# 5.2 DEVICE IDENTIFICATION
| ■ Device Name | Single Use Aspiration Needle |
|------------------------|--------------------------------------------|
| ■ Model Name | NA-U401SX-4025, NA-U401SX-4025N |
| ■ Common Name | Aspiration Needle |
| ■ Regulation Number | 21 CFR 876.1075 |
| ■ Regulation Name | Gastroenterology-urology biopsy instrument |
| ■ Regulatory Class | II |
| ■ Product Code | FCG |
| ■ Classification Panel | Gastroenterology/Urology |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The word is positioned above a thick, horizontal, yellow line. The font is sans-serif and the letters are closely spaced, giving the word a strong, unified appearance. The overall design is simple and clean, with a focus on the brand name.
K180668
# 5.3 PREDICATE DEVICE
#### 1) Predicate device
Table 12-1 Primary predicate device for NA-U401SX-4025/NA-U401SX-4025N
| Device name | 510(k) Submitter | 510(k) No. |
|---------------------------------------------------------------|----------------------------------|------------|
| NA-U401SX-4022<br>(Single Use Aspiration Needle<br>NA-U401SX) | OLYMPUS MEDICAL<br>SYSTEMS CORP. | K160098 |
#### 2) Reference device
#### Table 12-2 Reference device for NA-U401SX-4025/NA-U401SX-4025N
| Device name | 510(k) Submitter | 510(k) No. |
|------------------------------------------------------------------------------------------------|-------------------|------------|
| ECHO-HD-25-EBUS-O<br>(Echotip Ultra Endobronchial High<br>Definition Ultrasound Biopsy Needle) | Cook Ireland Ltd. | K160229 |
#### 5.4 DEVICE DESCRIPTION
#### 1) General Description of the subject device
The Single Use Aspiration Needle NA-U401SX-4025/NA-U401SX-4025N (aka ViziShot 2) is used in conjunction with Olympus ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA).
The subject devices consist of a handle section and insertion. The handle section is connected to the endoscope's instrument channel port via the single use adapter biopsy valve (MAJ-1414). The insertion section is composed of the sheath, needle and stylet. The needle tube is stored in the sheath and extended from the sheath to puncture the target tissue to collect specimen by moving the needle slider on the handle. A syringe is attached to the aspiration port on the handle section to aspirate the specimen. The needle is dimpled for echo enhancement.
The subject devices will be sold with or without medallion syringe manufactured by Merit Medical, which consists of VACLOK Syringe and Stopcock.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are all capitalized and evenly spaced. There is a yellow line underneath the word. The registered trademark symbol is located to the right of the letter "S".
# 2) Principle of Operation
Ultrasound emitted by ultrasound transducer of ultrasound endoscope diffusely reflects at dimples on the echo-enhanced region of needle tube. The ultrasound transducer detects the part of ultrasound which diffusely reflects and the echo-enhanced region is indicated on ultrasound image.
The needle tube and stylet are extracted from the sheath by moving the needle slider to distal side and is drawn into the sheath by moving the needle slider in the counter direction.
The subject devices are used with specified ultrasound endoscopes for aspirating tissue or cells by piercing the target area in ultrasound image and attaching the medallion syringe to the aspiration port on the handle section.
# 5.5 INDICATIONS FOR USE
This instrument has been designed to be used with ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA) of submucosal and extramural lesions of the gastrointestinal tract.
# 5.6 COMPARISON OF TECHNOLOGY CHARACTERISTICS WITH THE PREDICATE DEIVCE
The Single Use Aspiration Needles, NA-U401SX-4025/NA-U401SX-4025N have the same technological characteristics and design as the predicate device except for the following new features:
- Needle width
- Needle shape
- Needle tip shape
- Sheath composition
- New colored sheath (NA-U401SX-4025 only)
All other technological characteristics of both the subject and predicate devices are identical.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, dark blue letters. The letters are closely spaced and fill most of the upper portion of the image. Below the word "OLYMPUS" is a horizontal bar in a gradient of yellow and gold. The bar is positioned directly beneath the text, providing a visual base for the logo.
K180668
# 5.7 PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
#### 1) Sterilization/Shelf life testing
Sterilization/shelf life life testing for Single Use Aspiration Needle. NA-U401SX-4025/NA-U401SX-4025N was conducted in accordance with the FDA's Guidance for Industry and Food and Drug Administration Staff, "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile".
Accelerated aging test for the NA-U401SX-4025/NA-U401SX-4025N was conducted in accordance with ASTM F1980-16, the standard guide for accelerated aging of sterile barrier systems for medical devices. The real-time aging test for three-years will be performed to demonstrate longer stability and support the results of the accelerated aging test.
#### 2) Biocompatibility testing
Biocompatibility testing for the the Single Use Aspiration Needle, NA-U401SX-4025/NA-U401SX-4025N were conducted in accordance with the FDA's Guidance for Industry and Food and Drug Administration Staff, Use of International Standard ISO 10993-1:2009, "Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process''.
#### 3) Performance testing - Bench
Bench testing for the Single Use Aspiration Needle,
NA-U401SX-4025/NA-U401SX-4025N as listed below was conducted to ensure that the subject devices perform as intended and meet design specifications.
- Inserting into endoscope
- Flexibility of the insertion portion
- Piercing
- Ultrasound visibility
- Needle extraction and retraction
- Aspiration
- Withdrawal from endoscope
- Locking force of handle portion
- Limitation of needle depth
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are a sans-serif font and are evenly spaced. There is a yellow line underneath the word. The registered trademark symbol is in the upper right corner of the image.
# 4) Risk analysis
Risk analysis for the Single Use Aspiration Needle, NA-U401SX-4025/NA-U401SX-4025N was conducted in accordance with established in-house acceptance criteria based on ISO 14971:2007. The design verification tests and their acceptance criteria were identified and performed as a result of this risk analvsis assessment.
The following standards have been applied to the Single Use Aspiration Needle, NA-U401SX-4025/NA-U401SX-4025N.
- ISO 10993-1: 2009 - AAMI/ANSI/ISO 10993-5: 2009 - ISO 10993-10: 2010 - ISO 10993-11: 2006 - ISO 11135: 2014 - ISO 10993-7: 2008+Cor1 - ISO 11607-1: 2006+A1 - ISO 11607-2: 2006+A1 - ASTM F1980-16 - ISO 14971: 2007
# 5.8 CONCLUSIONS
Based on the indications for use, technological characteristics, performance testing and technological comparison to the predicate devices, the NA-U401SX-4025/ NA-U401SX-4025N raise no new issue of safety and effectiveness and are substantially equivalent to the predicate device in terms of safety, efficacy and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.