K171956 · Changzhou Detain Medical Devices Co., Ltd. · FCG · Dec 12, 2017 · Gastroenterology, Urology
Device Facts
Record ID
K171956
Device Name
Sterile Endoscope Biopsy Sampling Needle
Applicant
Changzhou Detain Medical Devices Co., Ltd.
Product Code
FCG · Gastroenterology, Urology
Decision Date
Dec 12, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1075
Device Class
Class 2
Indications for Use
The Sterile endoscope biopsy sampling needle is used to sample targeted submucosal and extramural lesions within or adjacent to the tracheobronchial tree or gastrointestinal tract through the accessory channel of an ultrasound endoscope.
Device Story
Sterile endoscope biopsy sampling needle (Model DT-EN-W7) used for soft tissue sampling. Device consists of needle head, PEEK outer tube, and handle. Operated by physician via ultrasound endoscope accessory channel. Procedure involves advancing needle to target site, applying firm pressure to puncture tissue, connecting external end to negative pressure aspirator, retracting needle head into tube, and withdrawing device. Sample retrieved from stylet notch. Benefits include targeted tissue acquisition for diagnostic purposes.
Clinical Evidence
Bench testing and animal studies (porcine model) performed. Bench tests included smoothness, leakproofness, joint strength, puncture force, stiffness, resistance to breakage, and chemical/pH analysis. Biocompatibility testing (ISO 10993-1) included cytotoxicity, sensitization, irritation, systemic toxicity, and material-mediated pyrogenicity. Animal testing confirmed performance equivalence to predicate.
Technological Characteristics
Materials: PEEK outer tube. Dimensions: 18-22 gauge, 0-80 mm length. Performance specs: Puncture force ≤ 8 N; joint strength ≥ 15 N; stiffness (15 mm span, 10 N load, deflection ≤ 0.50 mm). Biocompatibility per ISO 10993-1. Sterilization: Sterile. Shelf life: 2 years per ASTM F1980-07.
Indications for Use
Indicated for sampling targeted submucosal and extramural lesions within or adjacent to the tracheobronchial tree or gastrointestinal tract via ultrasound endoscope accessory channel.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Predicate Devices
Echotip® Ultra Endobronchial High Definition Ultrasound Needle, Echotip Procore® Endobronchial High Definition Ultrasound Biopsy Needle (K160229)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 12, 2017
Changzhou Detain Medical Devices Co., Ltd. % Ethan Liu RA Specialist Shanghai Thinkwell Consulting Co., Ltd. Xinling Rd, 211/6F Shanghai, 201100 China
Re: K171956
Trade/Device Name: Sterile Endoscope Biopsy Sampling Needle Regulation Number: 21 CFR§ 876.1075 Regulation Name: Gastroenterology-Urology-Biopsy Instrument Regulatory Class: II Product Code: FCG Dated: October 26, 2017 Received: October 30, 2017
Dear Ethan Liu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/6 description: The image shows the name "Benjamin R. Fisher -S" in a large, sans-serif font. The text is black and appears to be centered on a white background. The letters are clear and easy to read, with consistent spacing between them.
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K171956
Device Name Sterile Endoscope Biopsy Sampling Needle
Indications for Use (Describe)
The Sterile endoscope biopsy sampling need to sample targeted submucosal and extramural lesions within or adjacent to the tracheobronchial tree or gastrointestinal tract through the accessory channel of an ultrasound endoscope.
| Type of Use (Select one or both, as applicable) | <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5: 510(k) Summary
# 510(k) Number K
# 5.1 Date Prepared
December 4, 2017
#### 5.2 Submitter
| Company: | Changzhou Detian Medical Devices Co., Ltd. |
|------------|------------------------------------------------------------------------|
| Address: | No. 47, North Qingyang Rd., Tianning District, Changzhou City, Jiangsu |
| | Province, R. R. China |
| Telephone: | +86-21-60409623 |
| Fax: | +86-21-60732022 |
| Website: | www.jsczdt.com |
| | |
# 5.3 Contact/Consult
| Company: | Shanghai Thinkwell consulting Co., Ltd. |
|------------|-----------------------------------------|
| Website: | www.thinkwellmed.com |
| Contact: | Ethan Liu |
| Telephone: | +86-15216699240 |
| Email: | xtdeepwater@126.com |
#### 5.4 Device
| Trade Name: | Sterile Endoscope Biopsy Sampling Needle |
|-----------------------|--------------------------------------------|
| Model: | DT-EN-W7 |
| Common Name: | Biopsy Needle |
| Regulation Number: | 21 CFR §876.1075 |
| Classification Name: | Gastroenterology-urology biopsy instrument |
| Product Code: | FCG |
| Classification Panel: | Gastroenterology/Urology |
| Regulatory Class: | Class II |
#### 5.5 Predicate Device
Echotip® Ultra Endobronchial High Definition Ultrasound Needle, Echotip Procore® Endobronchial High Definition Ultrasound Biopsy Needle (Cook Ireland Ltd.) K160229
#### 5.6 Device Description
Sterile endoscope biopsy sampling needle, Model: DT-EN-W7 is mainly composed of needle head, PEEK outer tube assembly, and handle. Sizes are available in 18 - 22 gauge needles. The length of the needles range is between 0 and 80 mm. The needle is used by advancing the entire device to the site of soft tissue sampling through endoscopic channel aperture. The needle is advanced with gentle but firm pressure. Once the needle is in position, stab hard into the tissue and connect the external end of needle tube with aspirator in negative pressure. Retreat the needle head into tube, and get the needle out of the endoscope through the
{4}------------------------------------------------
Page 2. of 3
channel. The sample can then be expelled from the stylet notch.
#### 5.7 Indication for Use
The Sterile endoscope biopsy sampling needle is used to sample targeted submucosal and extramural lesions within or adjacent to the tracheobronchial tree or gastrointestinal tract through the accessory channel of an ultrasound endoscope.
#### 5.8 Technological characteristics
- Smoothness and leakproofness: Needle head can move freely and smoothly in the outer tube. The tip should stay in outer tube after retracement. The pull rod can be fixed apace when operate aspirator and turn it. No air leakage except the air passageway.
- Joint strength: The connect of outer tube and handle ≥15N, connection of needle head and handle ≥15N.
- Puncture force: Needle tip should be sharp with maximum puncture force ≤ 8 N.
- Stiffness and resistance to breakage: Sample needle should have good stiffness with span 15 mm ±0.1 mm, load 10 N ±0.1 N, deflection value ≤ 0.50 mm. Sample needle should have good resistance to breakage with span 17.5 mm ± 0.1 mm, and shall not be broken in the same plane at 20 degrees by two-way repeated bending for 20 times.
- Reducing substance: Compared with the same batch blank contrast solution of the same volume, the KMnO4 solution consumption volume of the test liquid is less than 6.5 mL.
- Total content of heavy metal: The color of test solution is not darker than that of p (Pb2+) = 1 ug/mL standard contrast liquid.
- pH value: Test liquid compared with the same batch blank liquid, the difference of pH value should be ≤ 1.5.
- Decay Resistance: test results satisfy with request of Class 5.4b after needle head erosion test.
- Compatibility with Endoscope: The sterile endoscope biopsy sampling needles shall go through the channel of endoscope smoothly.
#### 5.9 Substantial Equivalency Summary
By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device, or has the same intended use and different technological characteristics, and it can be demonstrated that the device is as safe and effective as the predicate device, and that the new device does not raise different questions regarding its safety and effectiveness as compared to the predicate device.
It has been shown in this 510(k) submission that the differences between the Sterile Endoscope Biopsy Sampling Needle and the predicate device do not raise any questions regarding its safety and effectiveness. Sterile Endoscope Biopsy Sampling Needle, as designed and manufactured, is as safe and effective as the predicate.
# 5.10 Non-clinical Testing 5.10.1 Performance Data Summary
{5}------------------------------------------------
Non-clinical comparative performance bench testing was successfully completed to establish substantial equivalence between the proposed Sterile Endoscope Biopsy Sampling Needle and its predicate device Echotip® Ultra Endobronchial High Definition Ultrasound Needle, Echotip Procore® Endobronchial High Definition Ultrasound Biopsy Needle. The following tests were conducted on the Sterile Endoscope Biopsy Sampling Needle:
- 1. Smoothness and leakproofness
- 2. Joint strength
- 3. Puncture force
- Stiffness and resistance to breakage 4.
- న. Reducing substance
- 6. Total content of heavy metal
- 7. PH value.
- 8. Decay Resistance
- Compatibility with Endoscope 9.
# 5.10.2 Sterility Testing and Shelf Life Summary
The 2 years shelf life of the proposed Sterile Endoscope Biopsy Sampling Needle has been established in accordance with ASTM F1980-07:2011Standard Guide for Accelerated Aging of Sterile Barrie System for Medical Device. The sterility testing has been conducted in accordance with ISO 11737-2:2009 Sterilization of Medical Devices - Microbiological Methods - Part 2: Test of Sterility Performed in the Definition, Validation and Maintenance of a Sterilization Process after shelf life study and proved package can keep the product sterile during the shelf life after product sterilization process.
#### 5.10.3 Biocompatibility Testing Summary
The proposed Sterile Endoscope Biopsy Sampling Needle was evaluated biocompatibility in accordance with ISO 10993-1:2009 Evaluation and Testing. The following tests were performed with acceptable results on the patient contacting portions of the Sterile Endoscope Biopsy Sampling Needle: Cytotoxicity, Sensitization, Irritation, Systemic Toxicity and material-mediated Pyrogenicity test.
#### 5.10.4 Animal Testing Summary
Animal testing has been conducted by using Sterile Endoscope Biopsy Sampling Needle and Echotip Ultra/Procore Endobronchial High Definition Ultrasound Needle in pigs. And the results demonstrate that the proposed device has the substantial equivalent performance on biopsy sampling with the predicate device.
#### 5.11 Conclusion
The Sterile Endoscope Biopsy Sampling Needle is as safe and effective as the predicate device. It has the same or similar intended use, indications for use, technological characteristics, and principles of operation as those of the predicate device. The minor differences between the Sterile Endoscope Biopsy Sampling Needle and its predicate device raise no new issues of safety or effectiveness. Thus, the Sterile Endoscope Biopsy Sampling Needle is substantially equivalent to its predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.