K964131 · Akos Biomedical, Inc. · FAM · Jun 27, 1997 · Gastroenterology, Urology
Device Facts
Record ID
K964131
Device Name
ENDOSCOPE
Applicant
Akos Biomedical, Inc.
Product Code
FAM · Gastroenterology, Urology
Decision Date
Jun 27, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The endoscope system is used to examine body cavities, hollow organs and canals, specifically the lower GI tract (colon) only, and, using additional accessories, to perform various diagnostic and therapeutic procedures.
Device Story
Endoscope system for lower GI tract (colon) examination; includes control head, disposable assembly, imaging system, and pneumatically driven air/fluid system. Disposable assembly encompasses all inner passage channels to prevent contamination; eliminates need for mechanical valves in path. Lavage function provides water bolus through working channel to flush debris. Operated by clinicians in clinical settings. Provides visualization of colon; assists in diagnostic/therapeutic procedures. Benefits include reduced cross-contamination risk via single-use disposable channels.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Endoscope with control head, disposable assembly, imaging system, and pneumatic air/fluid system. Materials comply with USP classification of plastics and ISO 10993-1 biocompatibility standards. Sterilization via E-Beam radiation (SAL 10^-6) per ISO 11137. Complies with IEC 601-1, IEC 601-2-18, and electromagnetic compatibility standards (IEC 801, 1000-4).
Indications for Use
Indicated for examination of the lower GI tract (colon) and performance of diagnostic/therapeutic procedures in patients requiring colonoscopy.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
Predicate Devices
Olympus EVS-Endoscopic Video Image & Data Sys. (K853585)
{0}
964131
akos JUN 27 1997 P1n3
akos biomedical, inc.
6450 Lusk Blvd., Suite 109
San Diego, CA 92121
510(k) SUMMARY
February 10, 1997
voice: 619.450.6767
fax: 619.450.9249
e-mail: akos@cts.com
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
Submitter’s Name and Address: akos biomedical, inc.
Contact Person: Lacyne Avery
Trade or Proprietary Name: To Be Determined
Common Name: Endoscope
Classification Name: Endoscope
Device Class: Class II, within the Division of Gastroenterology-Urology Devices,
Regulation Number: 21CFR Part 876.1500 Procode: 78KOG
Indications for Use: The endoscope system is used to examine body cavities, hollow organs and canals, specifically the lower GI tract (colon) only, and, using additional accessories, to perform various diagnostic and therapeutic procedures.
Device Description: The endoscope system includes the endoscope with control head, a disposable assembly, an imaging system, and a fluid and air system. The endoscope system features: 1) a smooth control head containing a minimal number of outside angles and joint crevices, 2) a single use disposable system encompassing all of the inner passage channels that start at the manifold, continue through the endoscope cavity, and seal onto the outside of the rigid distal tip, 3) a pneumatically driven air and fluid system eliminating the need for mechanical valves in the contamination path, and 4) a lavage function that provides a bolus of water through the working channel to loosen and flush debris from the field of view as the insertion tube advances through the colon.
**Predicate Device 510(k) Information:**
| Company Name | 510 (k) No. | 510(k) Description per CDRH Electronic Docket |
| --- | --- | --- |
| Olympus Corporation | K853585 | Olympus EVS-Endoscopic Video Image & Data Sys. |
| Pentax Precision Inst. | K951576 | FS-34P2, Fiber Sigmoidoscope |
| | K951577 | ES-3800, Video Sigmoidoscope |
| | K951579 | EC-3800TL, Video Colonoscope |
| | K951574 | EC-3800L, Video Colonoscope |
| Vision Sciences, Inc. | K932843 | Flexible Video Sigmoidoscope System w/Disp. Sheath |
| | K933247 | Protective Sheath - Flexible Nasopharyngo-Larygnoscope |
| | K921690 | Flexible Fiberoptic Sigmoidoscope System |
| | K921244 | Disposable Protective Sheath |
| Fujinon, Inc. | K944759 | 200 Series Gastro-Intestinal Video System |
| | K933906 | Choledochoscope/Endoscope Modification |
| Welch Allyn, Inc. | K954704 | VS-200 Video Sigmoidoscope |
| | K950429 | Video Sigmoidoscope |
| | K801247 | CCD Endoscope |
Pg 1 of 3
2
{1}
K964131
P2013
Sterilization: The Disposable Assembly will be marketed as “Sterile” - “Contents Are Sterile If Package Is Unopened and Undamaged”, it will NOT be labeled as “pyrogen free.” Sterilization with a SAL of 10-6 is to be performed by E-Beam radiation at TITAN Scan Systems, San Diego, California, using “Method 1” as defined in the ANSI/AAMI/ISO 11137-1994 “Sterilization of health care products - Requirements for validation and routine control - Radiation sterilization.”
## Safety Performance Specification / Standards (Applicable Sections):
FCC 47 CFR Parts 0 to 19 (Part 18 is the applicable section)
UL 94 Tests for Flammability of Plastic Materials for Parts in Devices and Appliances
UL 1572 High Intensity Discharge Lighting Fixtures
UL 2601-1 Medical Electrical Equipment, Part 1: General Requirements for Safety
CSA C22.2 No. 0.4-M1982 Bonding and Grounding of Electrical Equipment
CSA C22.2 No. 601.1 Medical Electrical Equipment, General Requirements for Safety
G95-1 FDA Blue Book Memo - Required Biocompatibility Training and Toxicology Profiles for Evaluation of Medical Devices
U.S.P. United States Pharmacopeia - Table 1. Classification of Plastics
ISO10993-1R Biological Evaluation of Medical Devices, Part 1: Evaluation and Testing
ISO 11137 Sterilization of Health Care Products, Requirements for Validation and Routine Control, Radiation Sterilization.
IEC 320-1 Appliance Couplers for Household and Similar General Purposes
IEC 601-1 Medical Electrical Equipment, Part 1: General Requirements
IEC 601-1-1 Medical Electrical Equipment, Part 1: General Requirements for Safety
Collateral Standard: Safety Requirements for Medical Electrical Systems
IEC 601-1-2 Medical Electrical Equipment, Part 1: General Requirements for Safety
Collateral Standard: Electromagnetic Compatibility - Requirements and Tests
IEC 601-2-18 Medical Electrical Equipment, Part 2: Particular Requirements for the Safety of Endoscopic Equipment
IEC 801-1 Electromagnetic Compatibility for Industrial-Process Measurement and Control Equipment, Part 1: General Introduction
IEC 801-5 Electromagnetic Compatibility for Industrial-Process Measurement and Control Equipment, Part 5: Surge Immunity Requirements
IEC 1000-4-2 Electromagnetic Compatibility (EMC), Part 4: Testing and Measurement Techniques, Section 2: Electrostatic Discharge Immunity Test Basic EMC Publication
IEC 1000-4-3 Electromagnetic Compatibility (EMC), Part 4: Testing and Measurement Techniques, Section 3: Radiated, Radio-Frequency, Electromagnetic Field Immunity Test
IEC 1000-4-4 Electromagnetic Compatibility (EMC)
EN 50082-1 Generic Immunity Standard for a Light Industrial Environment
EN 55011 Limits and Methods of Measurement of Electromagnetic Disturbance Characteristics of Industrial, Scientific and Medical (ISM) Radio Frequency Equipment CISPR 11
EN 60950 Safety of Information Technology Equipment Including Electrical Business Equipment
4169-94 ASTM Standard Practice for Performance Testing of Shipping Containers and Systems
P2013
3
{2}
K964131
P373
TABLE OF COMPARISONS TO PREDICATE DEVICES
| | akos (TBD) | akos (TBD) | Olympus CF-100S | Olympus CF-130S | Olympus E-GIF | Olympus E-CF | Pentax FS-34P2 | Pentax ES-3801 | Pentax EC-3801S2 | Pentax EC-3400M |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Viewing Direction | Forward | Forward | Forward | Forward | Forward | Forward | | | | |
| Optical Field of View | 135degrees | 125degrees | 140degrees | 140degrees | 113degrees | 133degrees | 120degrees | 120degrees | 120degrees | 120degrees |
| Optical Depth of Field | 5-100mm | 5-100mm | 5-100mm | 5-100mm | 8-85mm | 7-80mm | 3-100mm | 5-100mm | 5-100mm | 5-100mm |
| Rigid Tip Diameter | 13.7mm | 14.2mm | 15.4mm | 13.2mm | 11.4mm | 15.4mm | 11.5mm | 13.4mm | 13.4mm | 13.0mm |
| Bending Tip Deflection | 180up/dn 180r/left | 180up/dn 180r/left | 180up/dn 180r/left | 180up/dn 180r/left | 210up/90dn&100r/left | 180up/dn 180r/left | 180up/dn 180r/left | 180up/dn 180r/left | 180up/dn 180r/left | 180up/dn 180r/left |
| Bending OD | 13.4mm | 14mm | | | 11.2mm | 13.8mm | | | | |
| Insertion Tube Dia. | 13.2mm | 13.9mm | 13.3mm | 13.3mm | 11.4mm | 13.8mm | 11.5mm | 12.8mm | 12.8mm | 11.5mm |
| Insert. Working Length | 700mm | 1750mm | 820mm | 840mm | 1025mm | I:1330 / L:1680mm | 700mm | 700mm | 1150mm | 1300mm |
| Biopsy Channel ID | 3.2mm | 3.8mm | 3.2mm | 3.2mm | 2.8mm | 3.2mm | 3.5mm | | | |
| Total Length | 1020mm | 2070mm | 920mm | 935mm | | | 1015mm | 1020mm | 1470mm | 1620mm |
| Op. Env. Amb. Temp | 12-35deg C | 12-35deg C | 10-40deg C | 10-40deg C | | | | | | |
| Op. Env. R Humidity | 15-85 % | 15-85 % | 30-85 % | 30-85 % | | | | | | |
| Atmospheric Pressure | 70-106kPa | 70-106kPa | 700-1080mbar | 70-106kPa | | | | | | |
| Disposable Channels | Yes | Yes | No | No | No | No | No | No | No | No |
| | Vision Sci. VS2000 | Vision Sci. S-F100 | Fujinon COL-ST2/LT2 | Fujinon SIG-ET2 | Fujinon COL-MP/LP2 | Fujinon SIG-QP | Fujinon EC7-MR/LR2 | Fujinon ES7-ER2 | Fujinon EC7-MT/LT2 | Welch Allyn FX-100 |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Viewing Direction | Forward | Forward | Forward | Forward | Forward | Forward | Forward | Forward | Forward | Forward |
| Optical Field of View | 120degrees | 120degrees | 110degrees | 110degrees | 105degrees | 105degrees | 125degrees | 125degrees | 125degrees | 100degrees |
| Optical Depth of Field | | 3-100mm | 4-120mm observation | 4-120mm observ. | 4-120mm observ. | 3-100mm observ. | 5-100mm observ. | 5-100mm observ. | 5-100mm observ. | 3-100mm focal |
| Rigid Tip Diameter | | 13.8mm | 14.0mm | 14.0mm | 11.5mm | 11.8mm | 13.0mm | 13.0mm | 14.3mm | |
| Bending Tip Deflection | 180up/dn 180r/left | 180up/dn 180r/left | 180up/dn 180r/left | 180up/dn 180r/left | 180up/dn 180r/left | 180up/dn 180r/left | 180up/dn 180r/left | 180up/dn 180r/left | 180up/dn 180r/left | |
| Bending OD | | | 14.0mm | 14.0mm | 11.5mm | 11.8mm | 13.0mm | 13.0mm | 14.0mm | |
| Insertion Tube Dia. | 12.8mm | 12.8mm | | | | | | | | 13.5mm |
| Insert. Working Length | 65cm | 65cm | 1010-1735mm | 765mm | 1475-1735 | 650mm | 1520-1770mm | 790mm | 1520-1770mm | 65cm |
| Biopsy Channel ID | 3.2mm | 3.2mm | 4.3mm | 4.3mm | 3.2mm | 3.2mm | 3.2mm | 3.2mm | 4.3mm | 3.2mm |
| Total Length | | | 1265-1990mm | 1020mm | 1730-1990mm | 980mm | 1850-2100mm | 1120mm | 1850-2100mm | |
| Op. Env. Amb. Temp | | | | | | | | | | |
| Op. Env. R Humidity | | | | | | | | | | |
| Atmospheric Pressure | | | | | | | | | | |
| Disposable Channels | Yes | Yes | No | No | No | No | No | No | No | No |
Pg 3 of 3
{3}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
Ms. Lacyne Y. Avery
Director, Quality Assurance
and Regulatory Affairs
Akos Biomedical, Inc.
6450 Lusk Boulevard, Suite 109
San Diego, California 92121
Re: K964131
Video Endoscope System
Dated: April 11, 1997
Received: April 14, 1997
Regulatory class: II
21 CFR §876.1500/Product code: 78 FAM and FDF
Dear Ms. Avery:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4616. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Lillian Yin, Ph.D.
Director, Division of Reproductive,
Abdominal, Ear, Nose and Throat,
and Radiological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{4}
03/20/97 FRI 18:25 FAX 619 450 4338 akos biomedical, inc. 028
Page 1 of 1
K964131
510(k) Number (if known): ________________________
Video Endoscope System
Device Name: ________________________
Indications For Use:
The akos endoscope system is used to examine body cavities, hollow organs and canals, specifically the lower GI tract (colon) only, and, using additional accessories, to perform various diagnostic and therapeutic procedures.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number K964131
Prescription Use ☑
(Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐
(Optional Format 1-2-96)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.