a. Endoscopes: The Henke-Sass, Wolf Cystoscopes/Hysteroscopes are indicated to provide the user with the means for endoscopic diagnostic and therapeutic surgical procedures. Examples for the use of the devices include the visualization and manipulation of the anatomy, as the surgeon deems appropriate. The Henke-Sass, Wolf Cystoscopes/Hysteroscopes are intended to be used in general urological and gynecological surgery through a minimally invasive approach by utilizing natural orifices to access the surgical site. b. Urology Instruments: The Henke-Sass, Wolf Urology Instruments in combination with endoscopes are intended to provide the user with the means for endoscopic visualization during examination, diagnosis and therapy of the lower urinary tract in conjunction with endoscopic accessories, or as the surgeon deems appropriate. The instruments are intended for use in urological procedures through the minimally invasive approach, by utilizing natural orifices to access the surgical site.
Device Story
Device consists of rigid cystoscopes, hysteroscopes, and associated sheaths/bridges; used for endoscopic visualization, diagnosis, and therapeutic procedures in urology and gynecology. Surgeon inserts device through natural orifices to access surgical site; provides optical visualization of anatomy. Device is non-sterile; requires manual/automated cleaning and steam sterilization before use. Operates with external light sources (Xenon or LED). Output is direct visualization for the clinician to perform procedures like biopsy or manipulation. Benefits include minimally invasive access to internal anatomy.
Clinical Evidence
Bench testing only. Verification testing evaluated mechanical, optical, and illumination performance. Cleaning and steam sterilization validated to 10^-6 sterility assurance level. Conformance to ISO 8600 series, IEC 60601-2-18, and ISO 14971 risk management standards.
Technological Characteristics
Rigid endoscopes; working length 269-302mm; outer diameter 2.7-4.0mm; deflection 0°-70°. Compatible with Xenon/LED light sources. Materials validated for manual/automated cleaning and steam sterilization. Standards: ISO 8600-1, 3, 4, 5, 6; IEC 60601-2-18; ISO 14971.
Indications for Use
Indicated for patients undergoing endoscopic diagnostic and therapeutic surgical procedures in general urology and gynecology, specifically for visualization and manipulation of anatomy via natural orifices.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
Predicate Devices
Stryker Urology and Gynecology Hardware System (K040390)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 7, 2017
Henke-Sass, Wolf GmbH Anna Reifschneider, RAC Regulatory Affairs Manager Keltenstrasse 1 Tuttlingen, 78532 De Germany
Re: K171336
> Trade/Device Name: Cystoscope 4.0mm autoclavable 0° WL 302mm; Hysteroscope 4.0mm autoclavable 0° WL 302mm; Cysto-Urethroscope Sheath 19 Fr .; Bridge for 17-25 Fr. Cysto-Urethroscope Sheath Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FAJ, HIH Dated: October 27, 2017 Received: October 27, 2017
Dear Anna Reifschneider:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
## 510(k) Number (if known) K171336
#### Device Name
Cystoscope 4.0mm autoclavable 0° WL 302mm; Hysteroscope 4.0mm; Cysto-Urethroscope Sheath 19 Fr .; Bridge for 17-25 Fr. Cysto-Urethroscope Sheath
#### Indications for Use (Describe)
Indications for Use
#### a. Endoscopes
The Henke-Sass, Wolf Cystoscopes are indicated to provide the user with the means for endoscopic diagnostic and therapeutic surgical procedures. Examples for the devices include the visualization and manipulation of the anatomy, as the surgeon deems appropriate. The Henke-Sass, Wolf Cystoscopes are intended to be used in general urological surgery through a minimally invasive approach by utilizing natural orifices to access the surgical site.
## b. Urology Instruments
The Henke-Sass, Wolf Urology Instruments in combination with endoscopes are intended to provide the user with the means for endoscopic visualization during examination, diagnosis and therapy of the lower urinary tract in conjunction with endoscopic accessories, or as the surgeon deems appropriate. The intended for use in urological procedures through the minimally invasive approach, by utilizing natural orifices to access the surgical site.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
{3}------------------------------------------------
# 510(k) Summary
510(k) Number: 171336
# 5.1 Sponsor Information
| Date Prepared: | December 7th, 2017 |
|---------------------------|-------------------------------------------------------------------------|
| Company Name and Address: | Henke-Sass, Wolf GmbH<br>Keltenstrasse 1<br>78532 Tuttlingen<br>Germany |
| Contact Person: | Anna Reifschneider, RAC<br>Regulatory Affairs Manager |
# 5.2 Device Information
| Device Type: | Cystoscopes, Hysteroscopes and Accessories |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation Description: | Endoscope and Accessories |
| Review Panel: | Gastroenterology/Urology, Obstetrics/Gynecology |
| Regulation Number: | 21 CFR 876.1500 |
| Product Code: | FAJ, HIH |
| Device Class: | II |
| Device Name: | Cystoscope 4.0mm autoclavable 0° WL 302mm;<br>Hysteroscope 4.0mm autoclavable 0° WL 302mm; Cysto-<br>Urethroscope Sheath 19 Fr.; Bridge for 17-25 Fr. Cysto-<br>Urethroscope Sheath |
# 5.3 Predicate Device
The legally marketed device to which substantial equivalence is being claimed is:
| 510(k) Number: | K040390 |
|--------------------|------------------------------------------------|
| Applicant: | Stryker Endoscopy |
| Device Name: | Stryker Urology and Gynecology Hardware System |
| Regulation Number: | 21 CFR 884.1690 |
| Product Code: | HIH, FAJ, FAS, and KQT |
| Device Class: | II |
{4}------------------------------------------------
# 5.4 Device Description
The Henke-Sass, Wolf Cystoscopes, Hysteroscopes and Accessories are composed of cystoscopes, hysteroscopes and cystoscopic accessories, which provide the user with the means for endoscopic diagnostic and therapeutic procedures. Examples of use of the product include the visualization and manipulation of anatomy, biopsy, and as the surgeon deems appropriate.
# 5.5 Indications for Use
## a. Endoscopes:
The Henke-Sass, Wolf Cystoscopes/Hysteroscopes are indicated to provide the user with the means for endoscopic diagnostic and therapeutic surgical procedures. Examples for the use of the devices include the visualization and manipulation of the anatomy, as the surgeon deems appropriate. The Henke-Sass, Wolf Cystoscopes/Hysteroscopes are intended to be used in general urological and gynecological surgery through a minimally invasive approach by utilizing natural orifices to access the surgical site.
## b. Urology Instruments:
The Henke-Sass, Wolf Urology Instruments in combination with endoscopes are intended to provide the user with the means for endoscopic visualization during examination, diagnosis and therapy of the lower urinary tract in conjunction with endoscopic accessories, or as the surgeon deems appropriate. The instruments are intended for use in urological procedures through the minimally invasive approach, by utilizing natural orifices to access the surgical site.
## 5.6 Summary of Substantial Equivalence
The Henke-Sass, Wolf Cystoscopes, Hysteroscopes and Accessories are substantially equivalent to the existing 510(k) cleared devices under the predicate Stryker Urology and Gynecology Hardware System (K040390). The indications for use of the Henke-Sass, Wolf Cystoscopes, Hysteroscopes and Accessories are substantially equivalent to the predicate device. The Henke-Sass, Wolf Cystoscopes, Hysteroscopes and Accessories do not incorporate any special technology or characteristics when compared to its predicate device.
## 5.7 Non-Clinical Performance Data
Bench testing was conducted to verify the performance of the Henke-Sass, Wolf Cystoscopes, Hysteroscopes, and Accessories. Design verification testing was conducted to evaluate the mechanical, optical, and illumination performance.
{5}------------------------------------------------
The cleaning instructions were also validated. The Henke-Sass, Wolf Cystoscopes, Hysteroscopes, and Accessories has been validated for manual and automated cleaning as well as steam sterilization to provide a sterility assurance of 106. The devices are provided non-sterile. The risk analysis was carried out in accordance with established in-house acceptance criteria based on ISO 14971.
The subject devices demonstrate conformance with the following recognized standards and guidances:
- 510(k) submission guidance for Hysteroscopes and Gynecology Laparoscopes; ●
- IEC 60601-2-18, Sec. 201.11.1.2.2 ●
- ISO 8600-1;
- ISO 8600-3;
- ISO 8600-4; ●
- . ISO 8600-5;
- . ISO 8600-6.
Results show that the subject device has met pre-defined design and performance acceptance Results of all non-clinical testing support the safety and effectiveness of the subject criteria. devices.
{6}------------------------------------------------
# 5.8 Technological Characteristics
| | The subject device compares to the legally marketed device as follows: |
|--|------------------------------------------------------------------------|
| | |
| Device | Subject Device:<br>Henke-Sass, Wolf Cystoscopes,<br>Hysteroscopes, and Accessories | Predicate Device:<br>K040390 - Stryker Urology and Gynecology<br>Hardware System |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant | Henke-Sass, Wolf | Stryker Endoscopy |
| 510(k) number | Not yet assigned | K040390 |
| Indications for Use | a. Endoscopes:<br>The Henke-Sass, Wolf<br>Cystoscopes/Hysteroscopes are indicated to<br>provide the user with the means for endoscopic<br>diagnostic and therapeutic surgical procedures.<br>Examples for the use of the devices include the<br>visualization and manipulation of the anatomy,<br>as the surgeon deems appropriate. The Henke-<br>Sass, Wolf Cystoscopes/Hysteroscopes are<br>intended to be used in general urological and<br>gynecological surgery through a minimally<br>invasive approach by utilizing natural orifices<br>to access the surgical site. | The Stryker Urology and Gynecology Hardware<br>System is intended to provide the user with the<br>means for endoscopic diagnostic and therapeutic<br>surgical procedures. Examples of use of the<br>product include the visualization and manipulation<br>of anatomy, ablation, biopsy, incision, and<br>resection of tissue, and/or as the surgeon deems<br>appropriate. The system is intended for use in<br>general urological and gynecological surgery<br>through the minimally invasive approach, by<br>utilizing natural orifices to access the surgical site.<br>The system's use is intended for, but not limited to<br>the following types of procedures: |
| | b. Urology Instruments: | • Dilation of the urethra, and cold-slitting of<br>urethral strictures |
| | The Henke-Sass, Wolf Urology Instruments in<br>combination with endoscopes are intended to<br>provide the user with the means for endoscopic | • Trans-urethral incision and resection of the<br>prostate |
| | visualization during examination, diagnosis and<br>therapy of the lower urinary tract in | • Trans-urethral removal of bladder tumors |
| | conjunction with endoscopic accessories, or as | • Trans-cervical resection and ablation of the<br>endometrium |
| | | • Trans-cervical resection of fibroids |
| Device | Subject Device:<br>Henke-Sass, Wolf Cystoscopes,<br>Hysteroscopes, and Accessories | Predicate Device:<br>K040390 - Stryker Urology and Gynecology<br>Hardware System |
| | the surgeon deems appropriate. The<br>instruments are intended for use in urological<br>procedures through the minimally invasive<br>approach, by utilizing natural orifices to access<br>the surgical site. | |
| Regulation Number | 884.1690, 876.1500 | 884.1690 |
| Product Code | FAJ, HIH | HIH, FAJ, FAS, KQT |
| Working length (± 0.5mm) | 269mm - 302mm | 269mm - 302mm |
| Outer diameter (± 0.2mm) | 2.7mm - 4.0mm | 2.7mm - 4.0mm |
| Light Source | Xenon, LED | Xenon, LED |
| Distal tip (outer diameter) | 2.7mm - 4.0mm | 2.7mm - 4.0mm |
| Deflection (°) | 0°- 70° | 0°- 70° |
{7}------------------------------------------------
## 5.8 Conclusion
Based on the above information and all data provided in this submisson of intended uses, technological characteristics, and non-clinical performance that the subject devices are substantially equivalent to the predicate device identified in this submission.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.