Wi-Fi body scale, model HS5 is indicated to measure body weight, estimate body fat, body water percentage, body muscle mass, bones mass, visceral fat rating and daily calorie intake (DCI) using BIA (bioelectrical impedance analysis). This product is for use by generally healthy adults,,who are not ill, feverish, have a chronic or acute disease, or a condition that affect the level of hydration such as pregnancy.
Device Story
Device is a body analysis scale using Bioelectrical Impedance Analysis (BIA). User stands barefoot on four electrodes; device applies 500 uA current at 50 kHz. Internal software calculates body composition metrics (fat, water, muscle, bone, visceral fat, DCI) using impedance data combined with user-inputted height, weight, age, and gender. Designed for home use by healthy adults. Device connects via Wi-Fi to mobile devices (iPod Touch, iPhone, iPad) for data display. Output provides users with body composition metrics to monitor health status.
Clinical Evidence
Clinical testing was performed to validate the effectiveness and accuracy of the device. All test results were satisfactory. Bench testing for electrical, mechanical, and environmental safety was conducted per EN/IEC 60601-1 and EN/IEC 60601-1-2.
Technological Characteristics
Impedance plethysmograph; BIA sensing principle using 500 uA at 50 kHz. Four-electrode configuration. Connectivity via Wi-Fi to mobile platforms (iPod Touch, iPhone, iPad). Complies with EN/IEC 60601-1 and EN/IEC 60601-1-2 standards.
Indications for Use
Indicated for generally healthy adults to measure body weight and estimate body composition (fat, water, muscle, bone, visceral fat, DCI) via BIA. Contraindicated for individuals who are ill, feverish, have chronic/acute disease, or conditions affecting hydration (e.g., pregnancy).
Regulatory Classification
Identification
An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
# 510(k) Summary Jun 2 8 2012
K121527
PAGE 1 OF 4
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of SMDA and 21 CFR 807.92.
### 1.0 submitter's information
Name: Shenzhen Healthcare Electronic Technology Co., Ltd. Address: 2/F, Block B, Tsinghua Information Termi North Zone Of Hi-Tech Industrial Park,Na Shenzhen, CHINA 518057 Phone number: 86-755-27179235 Fax number: 86-755-27179236 Contact: Mrs. SHELLY SHI Date of Application: 05/10/2012
#### 2.0 Device information
| Trade name: | Wi-Fi body scale, model HS5 |
|----------------------|------------------------------------------------|
| Common name: | Body Analysis scale/ Electronic Body Fat Scale |
| Classification name: | Impedance plethysmograph |
#### 3.0 Classification
Production code: MNW- Body Fat Analyzer. Regulation number: 870.2770 Classification: II Panel: Cardiovascular
### 4.0 Predict device information
Shenzhen Healthcare Electronic Technology Co., Ltd Manufacturer: Body Analysis Scale, Model BG 17 Device: 510(k) number: K110928
#### 5.0 Device description
The patient steps on the scale device, where four electrodes are located. The patient must step on the electrodes with bare feet, with normal moisture. Through harmless current stimulation of 500 uA, at 50 kHz, the Wi-Fi body scale calculates the body fat percentage. This calculation is done via the Bioelectrical Impedance Method. The current is passed through the body and the impedance of the body
{1}------------------------------------------------
determines the body fat. The calculation is based upon electrical impedance, height, weight, age, and gender. The calculation is performed via internal software, which uses the variables programmed in by the user. Wi-Fi body scale, model HS5 can be used with an iPod Touch, iPhone or iPad.
#### 6.0 Intended use
Wi-Fi body scale, model HS5 is indicated to measure body weight, estimate body fat, body water percentage, body muscle mass, bones mass, visceral fat rating and daily calorie intake (DCI) using BIA (bioelectrical impedance analysis). This product is for use by generally healthy adults,,who are not ill, feverish, have a chronic or acute disease, or a condition that affect the level of hydration such as pregnancy.
### 7.0 Summary comparing technological characteristics with predicate device
| | Wi-Fi body scale, model HS5 | Body Analysis Scale, Model BG 17(K110928) | Same or not? |
|--------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Intended use | Wi-Fi body scale, model HS5 is indicated to measure body weight, estimate body fat, body water percentage, body muscle mass, bones mass, visceral fat rating and daily calorie intake (DCI) using BIA (bioelectrical impedance analysis). This product is for use by generally healthy adults,,who are not ill, feverish, have a chronic or acute disease, or a condition that affect the level of hydration such as pregnancy. | Body Analysis Scale, Model BG 17 is indicated to measure body weight, estimate body fat, body water percentage, body muscle mass, bones mass, visceral fat rating and daily calorie intake (DCI) using BIA (bioelectrical impedance analysis). This product is for use by generally healthy adults,,who are not ill, feverish, have a chronic or acute disease, or a condition that affect the level of hydration | Same |
{2}------------------------------------------------
:
1. 1. 1. 1. 1.
:
:
:
:
| | | such as pregnancy. | |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| Product<br>Description | Wi-Fi body scale, model<br>HS5 is indicated to<br>measure body weight,<br>estimate body fat,<br>body water<br>percentage, body<br>muscle mass, bones<br>mass, visceral fat<br>rating and daily calorie<br>intake (DCI) using BIA<br>(bioelectrical<br>impedance analysis). | Body Analysis Scale,<br>Model BG 17 is<br>indicated to measure<br>body weight,<br>estimate body fat,<br>body water<br>percentage, body<br>muscle mass, bones<br>mass, visceral fat<br>rating and daily<br>calorie intake (DCI)<br>using<br>BIA<br>(bioelectrical<br>impedance analysis). | Same |
| Analytical<br>method | BIA<br>(bioelectrical<br>impedance analysis) | BIA (bioelectrical<br>impedance analysis) | Same |
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
:
.
:
and the comments of the comments of
{3}------------------------------------------------
#### 8.0 Performance summary
Electrical, mechanical, environmental safety and performance testing according to standard EN/IEC 60601-1 and EN/IEC 60601-1-2. Clinical testing was used to validate the effectiveness and accuracy of the device. All test results were satisfactory.
### 9.0 Comparison to the predict device and the conclusion
Our device Wi-Fi body scale, model HS5 is substantially equivalent to the ody Analysis Scale, Model BG 17 whose 510(k) number is K110928.
The two devices are very similar in the intended use, the design principle, the material, the energy source and the applicable standards. Wi-Fi body scale, model HS5 can be used with an iPod Touch, iPhone or iPad.
However, the test in this submission provides demonstration that these small differences do not raise any new questions of safety and effectiveness.
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or other bird in flight. The symbol is composed of three stylized lines that suggest the bird's head, body, and tail.
#### Public Health Service
### JUN 28 2012
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Shenzhen Healthcare Electronic Technology Co., Ltd. % Ms. Yang Wang Andon Health Co., Ltd. No 3. Jinping Road Ya'an Street, Nankai District TIANJIN 300190 CHINA
Re: K121527
> Trade/Device Name: Wi-Fi body scale, model HS5 Regulation Number: 21 CFR& 870.2770 Regulation Name: Impedance plethysmograph Regulatory Class: II Product Code: MNW Dated: May 23, 2012 . Received: May 30, 2012
#### Dear Ms. Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{5}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Benjamin K. twho
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Statement of Indications for Use
510(k) Number :
121527
Device name:
Wi-Fi body scale, model HS5
### Indications for use:
Wi-Fi body scale, model HS5 is indicated to measure body weight, estimate body fat, body water percentage, body muscle mass, bones mass, visceral fat rating and daily calorie intake (DCI) using BIA (bioelectrical impedance analysis). This product is for use by generally healthy adults,,who are not ill, feverish, have a chronic or acute disease, or a condition that affect the level of hydration such as pregnancy.
Prescription use Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR YES (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-COUNTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Huber Steur
ductive, Gastro-Renal, and
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.