CLOCKMINI, MODEL DF301; CLOCKNAVI, MODEL DF311

K023942 · Yamato Corp. · MNW · Feb 24, 2003 · Cardiovascular

Device Facts

Record IDK023942
Device NameCLOCKMINI, MODEL DF301; CLOCKNAVI, MODEL DF311
ApplicantYamato Corp.
Product CodeMNW · Cardiovascular
Decision DateFeb 24, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2770
Device ClassClass 2
AttributesPediatric

Indications for Use

The Yamato ClockMINI and ClockNAVI Body Fat Monitors are intended for use in the estimation of body fat percent in users between the ages of 10 and 80 years old.

Device Story

Over-the-counter body composition monitors (ClockMINI DF301, ClockNAVI DF311) use Bioelectrical Impedance Analysis (BIA) to estimate body fat percentage. Device passes 50kHz/500µA current through user's hands, arms, and upper torso; measures electrical resistance. Algorithm processes resistance data to calculate body fat percent; classifies results into categories (low, average, high, very high). ClockNAVI also accepts user-inputted height, weight, age, and gender to calculate Body Mass Index (BMI) and obesity level. Used by consumers in home settings. Output displayed on device screen; informs user of body composition status for personal health monitoring.

Clinical Evidence

Bench testing compared Yamato devices against underwater weighing (UWW) in 149 subjects (51 males, 98 females). Results showed high similarity; average difference was 0.1% for males and -0.2% for females, with standard deviations between 3.2% and 3.3%.

Technological Characteristics

Battery-powered (2 AAA); BIA sensing principle; 50kHz/500µA current; measures electrical resistance; calculates body fat % and BMI; standalone operation.

Indications for Use

Indicated for estimation of body fat percent in users aged 10 to 80 years.

Regulatory Classification

Identification

An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.

Special Controls

*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ # 9. 510(k) Summary FEB 2 4 2003 ### 510(k) SUMMARY YAMATO ClockMINI and ClockNAVI BODY FAT MONITORS #### SUBMITTER INFORMATION Company name / address: Enscicon Corporation 518 17th Street Suite 1350 Denver, CO 80202 Contact name / numbers: Clay Anselmo Phone: 800-341-4255 Fax: 303-832-6700 Date summary prepared: October 22, 2002 Body Fat Analyzer #### DEVICE IDENTIFICATION Trade names: Yamato ClockMINI Body Fat Monitor, DF301 Yamato ClockNAVI Body Fat Monitor, DF311 Common name: Classification name: Impedance Plethysmograph Omron HBF-306 Body Fat Analyzer ### PREDICATE DEVICE Trade name: 510(k) number: K011652 ## DEVICE DESCRIPTION The Yamato ClockMINI Body Fat Monitor -- DF301 (ClockMINI) is an over-the-counter body composition analyzer and clock that uses Bioelectrical Impedance Analysis (BIA) to estimate body fat percent by measuring the electric resistance sensed when a 50kHz/500 µ A current is passed through the body. From this measurement the device uses an algorithm to display an estimate of body fat percent. In addition, the ClockMINI classifies the body fat percent into one of four categories: low, average, high, and very high. The Yamato ClockNAVI Body Fat Monitor - DF311 (ClockNAVI) is an over-the-counter body composition analyzer and clock that uses Bioelectrical Impedance Analysis (BIA) to measure body fat percent by measuring the electric resistance sensed when a 50kHz/500 µ A current is passed through the body. From this measurement the device uses an algorithm to display an estimate of body fat percent and a classification of obesity level. The device also calculates Body Mass Index (BMI) from data input by the user and displays an obesity level classification based on the calculated BMI. {2}------------------------------------------------ # INDICATIONS FOR USE The Yamato ClockMINI and ClockNAVI Body Fat Monitors are intended for use in the estimation of body fat percent in users between the ages of 10 and 80 years old. ## TECHNOLOGICAL CHARACTERISTICS COMPARISON Both of the Yamato devices and the Omron device are battery powered (2 "AAA" batteries) and pass a small electrical current (~500µA, 50kHz) through the hands, arms, and upper torso of the user for the purpose of calculating the bio-electrical impedance of the user. This method is commonly referred to as bio-electrical impedance analysis (BIA.) The primary indication for use for the Yamato devices and the Omron device is the estimation of body fat percent. The intended user group for all of the devices for normal subjects is 10 to 80 vears. (The Omron device has an "athlete" mode that is not present in the Yamato devices.) The technological characteristics for all the devices are equivalent including the method for estimating body fat percent and the method for calculating body mass index (BMI). The following primary characteristics of the two Yamato devices are equivalent to the Omron device. - Method for estimating body fat percent (BIA) - . Calculation for Body Mass Index (BMI) - Current and Frequency specifications - Route of current through the body . - Data input (height, weight, age, gender) . - Output parameters (BMI (ClockNAVI only), Body Fat%) . ## PERFORMANCE DATA To show acceptable performance of the Yamato ClockMINI and ClockNAVI devices for estimating body fat percentage, the devices were compared in a Yamato study against the under water weighing (UWW) method. Fifty-one males and 98 females had their body fat percentages estimated by both methods. The two methods provided very similar results overall. In both the male and female populations, the average difference between the Yamato device and the UWW method was extremely small (0.1% m, and -0.2% f) and the standard deviation of the difference was between 3.2% and 3.3%. When considering the variability associated with both methods, including measurement of residual volume in the lung, unpredictability of body water, and the estimation of intestinal tract volume, these numbers are quite small. ## CONCLUSION The Yamato ClockMINI and ClockNAVI Body Fat Monitors are substantially equivalent to the Omron HBF-306 Body Fat Analyzer. {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Yamato Corporation c/o Mr. Clay Anselmo Submission Correspondent Enscicon Corporation 518 17th Street Suite 1350 DENVER CO 80202 FEB 2 4 2003 Re: K023942 Trade/Device Name: Yamato ClockMINI Body Fat Monitor, DF301; and Yamato ClockNAVI Body Fat Monitor, DF311 Regulation Number: 21 CFR §870.2770 Regulation Name: Impedance plethysmograph Regulatory Class: II Product Code: 74 MNW Dated: November 18, 2002 Received: November 26, 2002 Dear Mr. Anselmo: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter: | 8xx.1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely vours. Nancy C. brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure Page 2 {5}------------------------------------------------ Page ___ of ___ | 510(k) Number (if known): | K023942 | |---------------------------|------------------------------------------------------| | Device Name: | Yamato ClockMINI (DF301)<br>Yamato ClockNAVI (DF311) | Indications for Use: The Yamato ClockMINI and ClockNAVI Body Fat Monitors are intended for use in the estimation of body fat percent in users between the ages of 10 and 80 years old. ### PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) David A. Leppanen (Division Sign-Off) Division of Reproductive, A and Radiological Devices 510(k) Number Over-the-Counter Use. . (Optional Format 3-10-98)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...