The ACCLARENT® SE Inflation Device is an instrument intended to inflate, deflate and monitor pressure in balloon catheters used in sinus procedures.
Device Story
Manual inflation device used during sinus balloon catheter procedures; enables clinician to inflate/deflate balloon and monitor pressure. Device consists of high-pressure syringe barrel, piston with locking mechanism, ergonomic handle, pressure gauge, and high-pressure tubing. Clinician operates device manually to deliver fluid to balloon catheter; pressure gauge provides visual feedback on inflation pressure. Used in clinical settings by physicians; output informs clinician of balloon inflation status and pressure levels, facilitating safe and effective sinus dilation. Benefits patient by enabling controlled, monitored sinus opening.
Clinical Evidence
Bench testing only. Evaluated dimensional attributes, cycle fatigue, and reliability after three years of simulated accelerated aging (ASTM F 1980-07). Biocompatibility testing performed per EN ISO 10993-1:2009 (intracutaneous reactivity, sensitization, cytotoxicity, systemic toxicity) with passing results. Sterilization validated per AAMI/ANSI/ISO 11135-1:2007; ethylene oxide residuals met ISO 10993-7:2008. No clinical data required.
Technological Characteristics
Manual, rigid inflation device. Features: 7cc syringe capacity, pressure gauge (atm), ergonomic piston handle, high-pressure tubing. Biocompatibility per ISO 10993 series. Sterilization: Ethylene oxide (AAMI/ANSI/ISO 11135-1:2007). Shelf life: 3 years (ASTM F 1980-07). No software or electronic components.
Indications for Use
Indicated for use as an instrument to inflate, deflate, and monitor pressure in balloon catheters during sinus procedures. No specific patient population age or gender restrictions provided.
Regulatory Classification
Identification
An ear, nose, and throat manual surgical instrument is one of a variety of devices intended for use in surgical procedures to examine or treat the bronchus, esophagus, trachea, larynx, pharynx, nasal and paranasal sinus, or ear. This generic type of device includes the esophageal dilator; tracheal bistour (a long, narrow surgical knife); tracheal dilator; tracheal hook; laryngeal injection set; laryngeal knife; laryngeal saw; laryngeal trocar; laryngectomy tube; adenoid curette; adenotome; metal tongue depressor; mouth gag; oral screw; salpingeal curette; tonsillectome; tonsil guillotine; tonsil screw; tonsil snare; tonsil suction tube; tonsil suturing hook; antom reforator; ethmoid curette; frontal sinus-rasp; nasal curette; nasal rasp; nasal rongeur; nasal saw; nasal scissors; nasal snare; sinus irrigator; sinus trephine; ear curette; ear excavator; ear rasp; ear scissor, ear snare; ear spoon; ear suction tube; malleous ripper; mastoid gauge; microsurgical ear chisel; myringotomy tube inserter; ossici holding clamp; sacculotomy tack inserter; vein press; wire ear loop; microrule; mirror; mobilizer; ear, nose, and throat punch; ear, nose and throat knife; and ear, nose, and throat trocar.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and features the department's name around the perimeter. In the center of the seal is a stylized representation of three human profiles facing right, with flowing lines connecting them. The profiles are arranged in a stacked formation, creating a sense of unity and collaboration.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 7, 2015
Acclarent, Inc. Mr. James Patrick Garvey II Sr. Manager, Regulatory Affairs 1525-B O'Brien Drive Menlo Park, CA 94025
Re: K150172
> Trade/Device Name: Acclarent SE Inflation Device Regulation Number: 21 CFR 874.4420 Regulation Name: Ear, Nose, And Throat Manual Surgical Instrument Regulatory Class: Class I Product Code: LRC Dated: March 12, 2015 Received: March 13, 2015
Dear Mr. Garvey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150172
Device Name ACCLARENT SE Inflation Device
Indications for Use (Describe)
The ACCLARENT® SE Inflation Device is an instrument intended to inflate, deflate and monitor pressure in balloon catheters used in sinus procedures.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "Acclarent" in black font. The last few letters of the word are in purple font. There is a purple arc above the last few letters of the word.
Acclarent SE Inflation Device
Special 510(k) Premarket Notification -
| Sponsor/Submitter: | Acclarent, Inc.<br>1525-B O'Brien Drive<br>Menlo Park, California 94025 |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | James Patrick Garvey II<br>Sr. Manager, Regulatory Affairs<br>Phone: (650) 687-4807<br>Fax: (650) 687-4847 |
| Date of Submission: | March 12, 2015 |
| Device Trade Name: | ACCLARENT SE Inflation Device |
| Common Name: | Balloon Inflation Device |
| Device Classification: | Class I |
| Regulation Number: | 21 CFR 874.4420 |
| Classification Name: | Ear, nose, and throat manual surgical instrument |
| Product Code: | LRC |
| Predicate Devices: | Acclarent Balloon Inflation Device (K052198) |
| Device Description: | The ACCLARENT SE Inflation Device is an instrument intended to inflate,<br>deflate and monitor pressure in balloon catheters used during sinus procedures. |
| | When used in a Balloon Sinuplasty dilation procedure, the inflation device<br>enables the user to inflate the sinus balloon catheter, monitor the pressure<br>within the balloon, and deflate the balloon. |
| | The device has a pressure gauge that indicates the atmospheric pressure at<br>which the attached balloon is inflated, a high pressure syringe barrel that<br>indicates the amount of fluid in the chamber, a mechanism to lock/unlock the<br>piston, an ergonomic piston handle, and high pressure tubing used to attach the<br>inflation device to the balloon catheters. The ACCLARENT SE Inflation<br>Device should only be used with Acclarent Sinus Balloon Catheters with<br>balloon sizes less than or equal to 7mmx24mm. Consult the individual<br>Acclarent sinus balloon catheter carton labels for balloon size information. |
| Indications for Use: | The Acclarent® SE Inflation Device is an instrument intended to inflate, deflate<br>and monitor pressure in balloon catheters used sinus procedures. |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the word "Acclarent" in a bold, sans-serif font. The word is black, except for the tail of the "t", which is purple and extends into a curved shape above the word. The logo appears to be for the company Acclarent.
## Acclarent SE Inflation Device
### Technological Characteristics:
| Attribute | Predicate Device<br>Acclarent Balloon<br>Inflation Device | Subject Device<br>ACCLARENT SE<br>Inflation Device |
|---------------------------------------------|-----------------------------------------------------------------------------|----------------------------------------------------|
| 510(k) number | K052198 | This Application |
| Model Number | BID30 | SE Inflation Device |
| Rigidity | Rigid | Same |
| Pressure Gauge | Values Displayed in<br>atmospheres (atm) &<br>Pounds per Square in<br>(PSI) | Values Displayed in<br>atmospheres (atm) only |
| Length<br>(Plunger not extended) | 8.1" | 8.8" |
| Length<br>(Plunger extended) | 11.3" | 11.1" |
| Device Diameter (at<br>hand position) | 1.0 inch | 1.3 inches |
| Flexible reinforced<br>high pressure tubing | Yes | Same |
| Syringe capacity | 20cc | 7cc |
| Luminescent dial | Yes | Yes |
#### Performance Data:
Bench testing met all acceptance criteria for attributes such as dimensional attributes, cycle fatigue, and reliability following three years of simulated accelerated aging.
Biocompatibility testing was performed for the ACCLARENT SE Inflation Device in accordance with EN ISO 10993-1: 2009. Biocompatibility testing for the SE Inflation Device included:
| Test | Test Standard | Results |
|------------------------------|-------------------|---------|
| Intracutaneous<br>Reactivity | ISO 10993-10:2013 | Pass |
| Sensitization | ISO 10993-10:2013 | Pass |
| In Vitro Cytotoxicity | ISO 10993-5:2009 | Pass |
| Systemic Toxicity | ISO 10993-11:2009 | Pass |
Biocompatibility data indicates that the likelihood of a toxic biologic effect from the SE Inflation Device is negligible. Additionally, the ACCLARENT SE Inflation Device is determined to have met the requirements of EN ISO 10993-1 and FDA General Program Memorandum #G95-1, and is therefore considered safe from a biocompatibility perspective.
The sterilization process was validated per AAMI/ANSI/ISO 11135-1: 2007 and demonstrated a sterility assurance level of 100. Testing of ethylene oxide residuals met ISO 10993-7:2008 requirements.
The subject device is not tested nor labeled as "non-pyrogenic". Packaging shelf life was established at three years via accelerated aging per ASTM F 1980-07. Clinical data was not necessary to establish the safety and efficacy
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "Acclarent" in black font. There is a purple curved line above the word that starts at the end of the word and extends to the right. The font is sans-serif and the word is slightly italicized.
Acclarent SE Inflation Device
Special 510(k) Premarket Notification
of the ACCLARENT SE Inflation Device. The performance data demonstrates that the subject device performs as intended. The performance data demonstrate that the device performs as intended.
Summary of Substantial Equivalence:
The ACCLARENT SE Inflation Device is substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.