The ViziShot 2 FLEX has been designed to be used with ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree. Do not use this device for any purpose other than its intended use.
Device Story
ViziShot 2 FLEX is a manual aspiration needle device used with ultrasound endoscopes for transbronchial needle aspiration (TBNA) of tracheobronchial lesions. Device consists of handle, sheath, needle, and removable stylet. Physician inserts flexible catheter into bronchoscope working channel; locks handle to adapter biopsy valve. Needle advanced to target site under real-time ultrasound visualization; echogenic tip facilitates guidance. Physician performs puncture and applies suction via syringe at proximal handle to obtain tissue sample. After sampling, vacuum released, handle unlocked, and device withdrawn. Used in endoscopy/bronchoscopy suites. Provides tissue samples for cytopathological or microbiological examination; aids in diagnosis of lesions. Benefits include precise, ultrasound-guided tissue acquisition.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing and design verification. Performance tests included sheath/needle insertion/withdrawal force, bronchoscope angulation, activation force, plastic deformation, penetration force, transmission force, device/handle durability, and echogenicity. Biocompatibility testing performed per ISO 10993-4, -5, -10, and -11.
Technological Characteristics
Manual aspiration needle; 19G needle size. Materials: Stainless steel, PTFE, PEBAX, Nitinol. Features: Adjustable/lockable needle and sheath, echogenic needle tip, aspiration capability. Sterilization: Ethylene oxide (EO) per ISO 11135-1. Packaging per ISO 11607-1. No software or electronic components.
Indications for Use
Indicated for ultrasound-guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree in patients requiring tissue sampling via ultrasound endoscope.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 6, 2017
Olympus Surgical Technologies America Mary Patella Senior Specialist, Regulatory Affairs 136 Turnpike Road Southborough, Massachusetts 01772-2104
Re: K163469
Trade/Device Name: Vizishot 2 Flex Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible or Rigid) and Accessories Regulatory Class: Class II Product Code: KTI Dated: March 9, 2017 Received: March 10, 2017
Dear Mary Patella:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/8 description: The image contains a signature and a name. The name reads "Tina Kiang-S". The signature is a series of looping lines to the left of the name.
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name ViziShot 2 FLEX
#### Indications for Use (Describe)
The ViziShot 2 FLEX has been designed to be used with ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree. Do not use this device for any purpose other than its intended use.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary Gyrus ACMI ViziShot 2 FLEX
# General Information
| Manufacturer: | Olympus Surgical Technologies<br>America<br>Gyrus ACMI, Inc.<br>136 Turnpike Rd.<br>Southborough, MA 01772-2104<br>Phone: 508-804-2600<br>Fax: 508-804-2624 |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment Registration Number: | 3003790304 |
| Contact Person: | Mary Anne Patella<br>Senior Specialist, Regulatory Affairs<br>508-804-2771<br>Maryanne.patella@olympus-osta.com |
| Date Prepared: | December 16, 2016 |
Device Description
| Classification Name: | Bronchoscope (flexible or rigid) and<br>accessories |
|----------------------|-----------------------------------------------------|
| CFR Citation Number: | 21 CFR 874.4680 |
| Product Code: | KTI |
| Classification: | Class II |
| Review Panel | Ear Nose & Throat |
| Trade Name: | ViziShot 2 FLEX |
| Generic/Common Name: | Aspiration Needle |
Predicate Devices
| Spiration, Inc. ViziShot FLEX | K152922 |
|-------------------------------|---------|
|-------------------------------|---------|
{4}------------------------------------------------
# Product Description
The ViziShot 2 FLEX is intended for use with compatible ultrasound endoscopes for Transbronchial Needle Aspiration (TBNA) of submucosal and extramural lesions of the tracheobronchial tree. The device is supplied sterile and intended for single patient use.
The device consists of a handle, sheath, needle, and stylet. The sheath and needle are attached to the handle, and the removable stylet is located within the needle. Note that although, the device has a component called a needle, the device is often referred to as a needle as well.
Prior to a procedure, the flexible catheter portion is inserted into a bronchoscope's working channel (2.2mm) and advanced forward until fully inserted. The handle is then affixed to the channel port of the endoscope via a lever mechanism that locks onto the Adapter Biopsy Valve.
The needle is advanced through the bronchoscope to the sampling site while visualizing both the target and the needle in real time with ultrasound. The handle facilitates advancement of the needle during puncture of the targeted biopsy site. The sample is obtained by penetrating the lesion with the needle while applying suction at the proximal end of the handle. After completing the sampling, the vacuum from the syringe is released to atmosphere, the handle unlocked from the bronchoscope, and the catheter and needle pulled out from the working channel. The removed tissue can then be prepared for cytopathological or microbiological examination and testing.
The ViziShot 2 FLEX is available in one model only (NA-U403SX-4019), with a needle size of 19 gauge (19G). The two required accessories, the Adapter Biopsy Valve and the Merit Syringe with Stopcock, are packaged with the ViziShot 2 FLEX.
### Intended Use
The ViziShot 2 FLEX has been designed to be used with ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree. Do not use this device for any purpose other than its intended use.
The Intended Use is identical to that of the predicate device.
### Technological Characteristics
The technological characteristics of the ViziShot 2 FLEX are the same as those of the predicate ViziShot FLEX. That is they have substantially the same design, material, and energy source. They are both designed to obtain a sample from lung tissue or
{5}------------------------------------------------
lymph node tissue under ultrasound visualization. Both devices have the following technological characteristics:
- Needle length adjustable and lockable ●
- Sheath length adjustable and lockable
- Secure attachment to scope with controlled device orientation ●
- Manual needle movement through bronchoscope to target site
- Flexible needle
- Stylet resistant to kinking
- Echogenic needle tip visible with ultrasound ●
- Sharp needle tip ●
- Aspiration capability ●
### Comparison of Technological Characteristics
The ViziShot 2 FLEX has the same basic technological characteristics as the predicate ViziShot FLEX cleared under K152922. The subject and predicate device operate in the same manner to obtain a tissue biopsy using an ultrasound endoscope. The indications for use of the ViziShot 2 FLEX are the same as that of the predicate.
The ViziShot 2 FLEX handle profile design has been changed from the predicate device to improve ergonomics; as well as to utilize Design for Manufacturing principles to make the manufacturing assembly more efficient. The resulting device functionality, look, and interaction with the physician are equivalent to those of the predicate device ViziShot FLEX. A detailed comparison of the ViziShot 2 FLEX and the ViziShot FLEX is provided in the following table.
| | Predicate Device (K152922) | Modified Device |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name/<br>Characteristics | ViziShot FLEX | ViziShot 2 FLEX |
| Indications for Use | The ViziShot FLEX has been<br>designed to be used with<br>ultrasound endoscopes for<br>ultrasound guided fine needle<br>aspiration (FNA) of<br>submucosal and extramural<br>lesions of the<br>tracheobronchial tree. Do not<br>use this device for any<br>purpose other than its<br>intended use. | The ViziShot 2 FLEX has been<br>designed to be used with<br>ultrasound endoscopes for<br>ultrasound guided fine needle<br>aspiration (FNA) of<br>submucosal and extramural<br>lesions of the tracheobronchial<br>tree. Do not use this device for<br>any purpose other than its<br>intended use. |
| Use Conditions | Surgical suite, endoscopy or<br>bronchoscopy suite, used with<br>a bronchoscope | Identical |
| | Predicate Device (K152922) | Modified Device |
| Device Name/<br>Characteristics | ViziShot FLEX | ViziShot 2 FLEX |
| Mechanics of Action | Manual | Identical |
| Mode of Action | Single/multiple puncture and<br>aspirate | Identical |
| General design | Handle, Sheath, Needle,<br>Stylet | Similar<br>The handle profile design has<br>been changed to improve<br>ergonomics and utilize Design<br>for Manufacturing principles.<br>Changes include:<br>• Handle shape and added<br>rubber finger grip<br>• Sheath adjuster rotates<br>• Snap bonds instead of<br>screw and glue |
| Patient Contacting<br>Materials | Stainless Steel, PTFE,<br>PEBAX, Nitinol | Stainless Steel, PTFE, PEBAX,<br>Nitinol |
| Biocompatible | Yes | Yes |
| Product Specification<br>and Properties | Engineering bench testing confirms that the modified ViziShot 2<br>FLEX meets its critical product specifications, which are<br>substantially equivalent to those of the predicate ViziShot<br>FLEX. | |
| Sterilization | EO | EO |
| Single Use Only | Yes | Identical |
| Working OD (mm) | 2.08 | Identical |
| Catheter Length (cm) | 70 | Identical |
| Needle Gauge | 19G | Identical |
| Typical Needle Length<br>(mm) | 20 | Identical |
| Max Needle Length<br>(mm) | 40 | Identical |
| Stylet OD (in) | 0.0205 | Equivalent<br>0.0204 |
| Stylet Surface Finish | Polished | Identical |
| Accessories | Syringe with stopcock<br>Adapter biopsy valve<br>Devices provided with device | Identical |
| Packaging | Needle assembly in tray with | Needle assembly, syringe, |
| | Predicate Device (K152922) | Modified Device |
| Device Name/<br>Characteristics | ViziShot FLEX | ViziShot 2 FLEX |
| | snap downs. Tray placed in<br>pouch. Pouch placed in shelf<br>box (dust cover) for sterilization.<br>After sterilization, the sterile<br>pouched syringe and sterile<br>pouched adapter will be added<br>to the dust cover. | adapter placed in tray with<br>snap downs and tyvek lid.<br>Tray placed in shelf box<br>prior to sterilization. |
{6}------------------------------------------------
.
{7}------------------------------------------------
# Summary of Non-Clinical Testing
# Biocompatibility:
Biocompatibility testing on all patient contacting surfaces has been performed in compliance to relevant requirements of ISO-10993. Biocompatibility testing included the following tests:
- ISO 10993-4: 2002 Biological evaluation of medical devices Part 4: ● Selection of tests for interactions with blood
- ISO 10993-5: 2009 Biological evaluation of medical devices - Part5: Tests for in vitro cytotoxicity
- ISO 10993-10: 2010 Biological evaluation of medical devices. Tests for ● irritation and sensitization
- . ISO 10993-11:2006. Biological evaluation of medical devices. Tests for systemic toxicity
### Sterilization:
The ViziShot 2 FLEX will be delivered in a sterile state and is intended for single patient use only. Sterilization (ethylene oxide) and packaging of the device was validated using the following standards:
- . ANSI/AAMI/ISO 11607-1:2006 Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems
- . ANSI/AAMI/ISO 11135-1:2014 Sterilization of health-care products ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices
Packaging integrity and performance testing on devices that had undergone accelerated aging support a labeled one year shelf life.
{8}------------------------------------------------
# Bench testing:
During design verification, the output of the design process was evaluated against the physical and performance specifications. The following performance tests were conducted:
- Sheath and Needle Insertion and Withdrawal Force
- Stylet Insertion and Withdrawal Force ●
- Bronchoscope Angulation ●
- Activation Force
- Plastic Deformation Angle
- Penetration Force
- Transmission Force
- Device Durability/Handle Durability
- Bronchoscope Adapter Sliding Force
- Handle Durability
- Sheath to Handle Joint Strength ●
- . Echogenicity
### Conclusion:
In summary, the ViziShot 2 FLEX is substantially equivalent to the predicate device and presents no new questions of safety or efficacy.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.