Unidrive S III ENT 40701601-1 with KARL STORZ-SCB, DrillCut-X II Shaver Handpiece, High Speed EC Micromotor II for use with INTRA Drill Handpieces
K150969 · Karl Storz Endoscopy America, Inc. · ERL · Jan 8, 2016 · Ear, Nose, Throat
Device Facts
Record ID
K150969
Device Name
Unidrive S III ENT 40701601-1 with KARL STORZ-SCB, DrillCut-X II Shaver Handpiece, High Speed EC Micromotor II for use with INTRA Drill Handpieces
Applicant
Karl Storz Endoscopy America, Inc.
Product Code
ERL · Ear, Nose, Throat
Decision Date
Jan 8, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The UNIDRIVE® S III ENT system consists of an active control unit used in conjunction with the High-Speed Micro Motor and DrillCut-X® II Shaver handpiece. The system is intended for use by qualified surgeons to provide controlled cutting, drilling, debriding, sawing, and shaving for the ablation, excision, removal, or transection of tissue or bone during head, neck, ENT, or otoneurological surgical procedures.
Device Story
UNIDRIVE S III ENT is a motorized surgical system comprising an active control unit, high-speed micro motor, and DrillCut-X II shaver handpiece. Used by surgeons in head, neck, ENT, or otoneurological procedures to cut, drill, debride, saw, and shave tissue or bone. System provides controlled mechanical power to attached handpieces and blades/burs. Operates as a reusable surgical tool; requires cleaning and sterilization between uses. Output is mechanical action at the surgical site, controlled by the surgeon via the console. Benefits include precise tissue/bone removal during complex ENT surgeries.
Clinical Evidence
Bench testing only. No clinical or animal data required. Performance testing included system verification and validation, cleaning and sterilization validation (SAL 10^-6), and manual cleaning effectiveness for residual protein and hemoglobin.
Technological Characteristics
Motorized, reusable surgical system. Components: active control unit, high-speed micro motor, shaver handpiece, blades, and burs. Sterilization: pre-vacuum steam sterilization for reusable instruments. Connectivity: system-based control unit. Materials: reusable components cleaned/sterilized; disposable tubing/burrs are single-use.
Indications for Use
Indicated for qualified surgeons performing ablation, excision, removal, or transection of tissue or bone during head, neck, ENT, or otoneurological surgical procedures.
Regulatory Classification
Identification
An ear, nose, and throat electric or pneumatic surgical drill is a rotating drilling device, including the handpiece, that is intended to drive various accessories, such as an ear, nose, and throat bur (§ 874.4140), for the controlled incision or removal of bone in the ear, nose, and throat area.
Predicate Devices
KSEA UNIDRIVE ENT System and Accessories (K053262)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of a human figure in profile, with three overlapping heads suggesting community and connection. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 8, 2016
Karl Storz Endoscopy America, Inc. Jennifer Chambers Regulatory Affairs Specialist 2151 E. Grand Ave El Segundo, California 90245
Re: K150969
Trade/Device Name: Unidrive S IIi Ent 40701601-1 With Karl Storz-scb, Drillcut-x II Shaver Handpiece, High Speed Ec Micromotor II For Use With Intra Drill Handpieces Regulation Number: 21 CFR 874.4250 Regulation Name: Ear, Nose, And Throat Electric Or Pneumatic Surgical Drill Regulatory Class: Class II Product Code: ERL Dated: December 4, 2015 Received: December 8, 2015
Dear Ms. Chambers:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet
address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also. please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go
to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet
address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Eric A. Mann -S
Malvina Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
Device Name UNIDRIVE S III ENT
Indications for Use (Describe)
The UNIDRIVE® S III ENT system consists of an active control unit used in conjunction with the High-Speed Micro Motor and DrillCut-X® II Shaver handpiece. The system is intended for use by qualified surgeons to provide controlled cutting, drilling, debriding, sawing, and shaving for the ablation, excision, removal, or transection of tissue or bone during head, neck, ENT, or otoneurological surgical procedures.
Type of Use (Select one or both, as applicable)
|X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for STORZ. The logo is in blue and consists of the word "STORZ" in large, bold letters, with a circle in the middle of the "O". Below the word "STORZ" is the text "KARL STORZ - ENDOSKOPE" in smaller letters.
#### 7. 510(k) Summary
Traditional Premarket Notification Submission (510(k)) Summary Prepared in accordance with 21 CFR 807.92
#### 7.1 Submitter Information
| Sponsor name: | KARL STORZ Endoscopy America, Inc. |
|-----------------------------|------------------------------------|
| Sponsor address: | 2151 E. Grand Ave. |
| | El Segundo, CA 90245 USA |
| Sponsor telephone: | 424-218-8100 |
| Sponsor fax: | 424-218-8519 |
| Establishment Registration: | 3010202439 |
| Contact person: | Jennifer Michelle Chambers<br>MPA, MBA, CCRP, CQA, PMP, RAC (US) |
|------------------------|------------------------------------------------------------------|
| Contact title: | Regulatory Affairs Specialist |
| Email direct: | jennifer.chambers@karlstorz.com |
| Telephone direct: | 424-218-8457 |
| Date summary prepared: | March 31, 2015 |
| Date summary updated: | January 5, 2016 |
#### 7.2 Device Name
| Trade (proprietary): | UNIDRIVE S III ENT |
|----------------------|------------------------------------------------------------------|
| Common (usual): | Drill, Surgical, ENT (Electric Or Pneumatic) Including Handpiece |
| Classification: | 21 CFR 874.4250 (Class II) |
| FDA Device Code: | ERL: Ear, nose, and throat electric or pneumatic surgical drill |
| Review Panel: | 77 (Ear Nose & Throat) |
#### 7.3 Substantially Equivalent Predicate Device
| Device Name: | KSEA UNIDRIVE ENT System and Accessories |
|----------------|------------------------------------------|
| Device 510(k): | K053262 |
#### 7.4 Device Description
The UNIDRIVE® S III ENT is a motorized surgical device system, consisting of an active control unit used in combination with: the High-Speed Micro Motor and attaching handpiece and the DrillCut-X® II Shaver handpiece (with optional attaching handle). The surgical device system provides controlled cutting, drilling, debriding, sawing, and shaving for the ablation, excision, removal, or transection of tissue or bone during head, neck, ENT, or otoneurological surgical procedures.
{4}------------------------------------------------
#### 7.5 Intended Use
The UNIDRIVE® S III ENT system consists of an active control unit used in conjunction with the High-Speed Micro Motor and DrillCut-X® II Shaver handpiece. The system is intended for use by qualified surgeons to provide controlled cutting, drilling, sawing, and shaving for the ablation, excision, removal, or transection of tissue or bone during head, neck, ENT, or otoneurological surgical procedures.
#### 7.6 Technological Characteristics
The UNIDRIVE® S III ENT system and the predicate UNIDRIVE ENT system are both motorized, reusable surgical device systems, used in conjunction with drills, handpieces, blades, and burs. Both systems provide controlled cutting, drilling, sawing, and removal of bone during head, neck, ENT, and or otoneurological surgical procedures. The performance of the new device has undergone system verification and validation testing to ensure it does not introduce new issues of safety or effectiveness.
#### 7.7 Performance Characteristics
Nonclinical performance characteristics evaluated in support of the substantial equivalence of the UNIDRIVE® S III ENT include system verification and validation testing. Study results demonstrate adherence to design specifications. Experiments vielded consistent results. Consistent test conditions were maintained throughout the study. Performance testing reports are provided in the Performance Testing (Bench) of this 510(k). Cleaning and sterilization validations were conducted for patient-contacting components.
#### 7.8 Cleaning and Sterilization
Reusable instruments (i.e. DrillCut-X® II Shaver handpiece, High-Speed Micro Motor and handpieces, plus compatible Straight Shaver Blades) are delivered non-sterile and must be cleaned and sterilized prior to the initial use and before each subsequent use. Compatible sinus burrs and disposable tubings are sterile, single-use products and must be disposed of after patient use. Sterility efficacy demonstrated a sterility assurance level of 106 in a pre-vacuum steam sterilizer. Manual cleaning effectiveness was demonstrated with sufficient recovery efficiency for residual protein and residual hemoglobin.
#### 7.9 Animal and Clinical Performance Data:
Animal and Clinical performance data are not required to demonstrate substantial equivalence for this type of device.
#### 7.10 Conclusion
Based on the information provided in this premarket notification. KARL STORZ concludes that the UNIDRIVE® S III ENT is safe, effective, and substantially equivalent to the predicate UNIDRIVE ENT in its indication for use, device design, materials, performance characteristics, and operational principles.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.