The E.G. Scan™ II Esophagoscope System is intended for use in endoscopic access and examination of the larynx, esophagus and gastroesophageal junction.
Device Story
E.G. Scan™ II Esophagoscope System is a transnasal endoscopic system for diagnostic visualization. The system comprises a disposable esophagoscope probe, a controller, and E.G. View™ software. The probe captures images of the larynx, esophagus, and gastroesophageal junction, transmitting data to the controller for processing and upload. The controller outputs the processed image data to the E.G. View™ software, which displays the images for clinical diagnosis. The device is intended for use by healthcare professionals in a clinical setting. By providing direct visualization of the esophageal tract, the system assists clinicians in identifying esophageal diseases, facilitating diagnostic decision-making and patient management.
Clinical Evidence
Bench testing only. Biocompatibility testing (cytotoxicity, sensitization, and irritation) performed per ISO 10993-1. Electrical safety and EMC testing performed per IEC 60601-1, IEC 60601-2-18, and IEC 60601-1-2.
Technological Characteristics
Transnasal esophagoscope system. Disposable probe. Components include probe, controller, and E.G. View™ software. Conforms to IEC 60601-1 (general safety), IEC 60601-2-18 (endoscopic equipment safety), and IEC 60601-1-2 (EMC). Patient contact parts tested for biocompatibility per ISO 10993-1.
Indications for Use
Indicated for endoscopic access and examination of the larynx, esophagus, and gastroesophageal junction in patients requiring such diagnostic procedures.
Regulatory Classification
Identification
An esophagoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the esophagoscope and is intended to examine or treat esophageal malfunction symptoms, esophageal or mediastinal disease, or to remove foreign bodies from the esophagus. When inserted, the device extends from the area of the hypopharynx to the stomach. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.
Modified Flexible Trans-Nasal Esophagoscope with Digital Video Processor and Disposable EndoSheath® System (K072088)
Submission Summary (Full Text)
{0}------------------------------------------------
120702
JUN 2 0 2012
510/lx) Submission -- E.G. Scan™ II Esophagoscope System
# 5′10(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: Dec. 06. 2011
1. Company and Correspondent making the submission:
Name - IntroMedic Co., Ltd.
Address ~ Suite 1104, E&C Venture Dream Tower 6-Cha, 197-28 Guro-Dong, Guro-Gu,
Seoul, 152-719 Korea
Telephone - +82-2-801-9300
Fax – +82-2-801-9330
Contact - JinYoung, Lee
Internet – http://www.intromedic.com
- 2. Device :
| <b>Proprietary name</b> | : E.G. Scan™ II Esophagoscope System |
|----------------------------|--------------------------------------|
| <b>Common Name</b> | : Esophagoscope System |
| <b>Classification Name</b> | : Esophagoscope System |
3. Predicate Device :
| Manufacturer : | IntroMedic Co., Ltd, |
|-----------------|---------------------------------|
| Device : | E.G. Scan™ Esophagoscope System |
| 510(k) Number : | K111030 |
| Manufacturer | : Vision-Science, Inc. |
|---------------|--------------------------------------------------------------------------------------------------------------------|
| Device | : Modified Flexible Trans-Nasal Esophagoscope with<br>Digital Video Processor and Disposable EndoSheath®<br>System |
| 510(k) Number | : K072088 |
4. Classifications Names & Citations :
. 21CFR874.4710, EOX, Esophagoscope System, Class2
IntroMedic Co., Ltd.
{1}------------------------------------------------
## 510(k) Submission - E.G. Scan™ II Esophagoscope System
5. Description :
#### 5.1 Introduction
E.G. Scan™ II Esophagoscope System and its accessories are used for diagnosis of patients. E.G. Scan™ II Probe takes pictures of the esophagus of human and sends image data to E.G. Scan™ II Controller. E.G. Scan™ II Controller processes and converts image data and upload. E.G. View™ image displaying software displays the image for diagnosis.
The E.G Scan™ II Esophagoscope Probe is disposable.
#### 5.2 General Technology
E.G. Scan™ II Esophagoscope system is a transnasal esophagoscope designed to capture images of the esophagus. Captured images are viewed via the E.G. View™ Software for diagnosis of diseases related to the esophagus.
- 6. Indication for use :
The E.G. Scan™ II Esophagoscope System is intended for use in endoscopic access and examination of the larynx, esophagus and gastroesophageal junction.
- 7. Comparison with predicate device :
The E.G. Scan™ II Esophagoscope System and predicate device are substantially equivalent in the areas of design, indication for use, technological characteristics, function, application and safety and effectiveness. This was determined by reviewing the information provided in the 510(k) in comparison to the content specified in the FDA guidance documents.
## 8. Safety, EMC and Performance Data :
The E.G. Scan™ II Esophagoscope System has the same device characteristics as the predicate device, E.G. Scan™ Esophagoscope System of IntroMedic Co., Ltd.; intended use, material, design and use concept are similar. The biocompatibility of the patient contact parts has been demonstrated through the cytotoxicity, sensitization and irritation testing by ISO 10993-1 Biological evaluation of medical devices. The E.G. Scan™ II Esophagoscope System conforms to IEC 60601-1 Medical electric equipment, Part 1: General requirements for safety, IEC 60601-2-18
IntroMedic Co., Ltd.
{2}------------------------------------------------
K120702
# 510(k) Submission ~ E.G. Scan™ II Esophagoscope System
Medical electrical equipment-Part 2: Particular requirements for the safety of endoscopic equipment and IEC 60601-1-2 Medical electric equipment, General requirements for safety collateral standard electromagnetic compatibility.
#### 9. Conclusions :
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification IntroMedic Co., Ltd. concludes that The E.G. Scan™ II Esophagoscope System is safe and effective and substantially equivalent to predicate devices as described herein.
IntroMedic Co., Ltd.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle with three stripes representing the three branches of government. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the perimeter of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
IntroMedic Company, Limited % Mr. Marc M. Mouser Underwriters Laboratories, Inc. 2600 NW Lake Road Camas, WA 98607
JUN 2 0 2012
Re: K120702
Trade/Device Name: E.G. Scan™ II Esophagoscope System Regulation Number: 21 CFR 874.4710 Regulation Name: Esophagoscope (flexible or ridgid) and accessories Regulatory Class: II Product Code: EOX Dated: May 25, 2012 Received: June 1. 2012
Dear Mr. Mouser:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{4}------------------------------------------------
Page 2 – Mr. Marc M. Mouser
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
E. E. Ivanov
Malvina B. Eydelman M.D. Director Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices Office of Device Evaluation Health Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K120702
# Indications for Use
510(k) Number(if known):
Device Name: E.G. Scan™ II Esophagoscope System
Indications for Use:
E.G. Scan™ II Esophagoscope System is intended for use in endoscopic access and examination of the larynx, esophagus and gastroesophageal junction.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use_ (Part 21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)
Page 1 of 1
S.M. Mofizur
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number_
× Prescription Use (Per 21 CFR 801.109)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.