PENTAX Medical Ultrasound Video Bronchoscope EB19-J10U
K183516 · Pentax of America, Inc. · EOQ · Sep 6, 2019 · Ear, Nose, Throat
Device Facts
Record ID
K183516
Device Name
PENTAX Medical Ultrasound Video Bronchoscope EB19-J10U
Applicant
Pentax of America, Inc.
Product Code
EOQ · Ear, Nose, Throat
Decision Date
Sep 6, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4680
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The PENTAX Medical Ultrasound Video Bronchoscope EB19-J10U is intended to provide optical visualization of, ultrasonic visualization of, and therapeutic access to, the Pulmonary Track including but not restricted to organs, tissues, and subsystem: Nasal Passage, Pharynx, Larynx, Trachea, Bronchial Tree (including access beyond the stem), and underlying areas. The instrument is introduced per orally when indications consistent with the requirement for procedure are observed in adult and pediatric patient populations.
Device Story
The PENTAX Medical Ultrasound Video Bronchoscope EB19-J10U is a flexible endoscope used for diagnostic and therapeutic procedures in the pulmonary tract. It features a CCD for optical imaging and a convex array ultrasound transducer for ultrasonic imaging. The device connects to a video processor and an ultrasound scanner (e.g., Arietta 70, Noblus, or Preirus). During procedures, the physician inserts the scope per-orally; the device provides real-time endoscopic and ultrasound images to the connected systems. The control body allows for angulation, suction, and the introduction of biopsy or other therapeutic instruments through a working channel. The device aids in visualizing internal structures and guiding interventions, potentially improving diagnostic accuracy and therapeutic outcomes. It is intended for use by trained clinicians in clinical settings.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including system compatibility, optical characteristics, electrical safety, EMC, biocompatibility (ISO 10993), and validation of reprocessing and sterilization processes.
Technological Characteristics
Flexible bronchoscope with CCD optical sensor and convex array ultrasound transducer. Features include angulation control, suction, and working channel. Connects to external video processors and ultrasound scanners. Biocompatible materials per ISO 10993. Sterilization via ethylene oxide for accessories. Software-controlled operation.
Indications for Use
Indicated for optical and ultrasonic visualization and therapeutic access to the pulmonary tract (nasal passage, pharynx, larynx, trachea, bronchial tree) in adult and pediatric patients.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
Predicate Devices
PENTAX Ultrasound Video Bronchoscope EB-1970UK + HI VISION PREIRUS (K131946)
Reference Devices
PENTAX Medical EPK-3000 Video Imaging System (K172156)
PENTAX Medical EPK-i5010 Video Processor with EB Family of Scopes (K143727)
PENTAX Medical EPK-i7010 Video Processor with EB Family of Scopes (K173679)
Hitachi Arietta70 Diagnostic Ultrasound System (K134016)
Hitachi HI VISION Preirus Diagnostic Ultrasound Scanner (K093466)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 6, 2019
PENTAX of America, Inc. % Beryl Jeanne Associate Regulatory Specialist NAMSA Inc. 400 Highway 169. Suite 500 Minneapolis, Minnesota 55426
Re: K183516
Trade/Device Name: PENTAX Medical Ultrasound Video Bronchoscope EB19-J10U Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible or Rigid) And Accessories Regulatory Class: Class II Product Code: EOO, ITX Dated: August 1, 2019 Received: August 5, 2019
Dear Beryl Jeanne:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Srinivas Nandkumar, Ph.D. Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K183516
Device Name
PENTAX Medical Ultrasound Video Bronchoscope EB19-J10U
Indications for Use (Describe)
The PENTAX Medical Ultrasound Video Bronchoscope EB19-J10U is intended to provide optical visualization of, ultrasonic visualization of, and therapeutic access to, the Pulmonary Track including but not restricted to organs, tissues, and subsystem: Nasal Passage, Pharynx, Larynx, Trachea, Bronchial Tree (including access beyond the stem), and underlying areas. The instrument is introduced per orally when indications consistent with the requirement for procedure are observed in adult and pediatric patient populations.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Indications for Use 510(k) Number: K183516 Device Name: PENTAX Medical Ultrasound Video Bronchoscope EB19-J10U
| System: | ARIETTA 70 | | | | | | |
|--------------------------|--------------------------------------------------------------------------------------------------|---|-------------------|-----|-----|------------------|----------------------|
| Probe: | EB19-J10U | | | | | | |
| | Intended use: Diagnostic ultrasound imaging or fluid flow analysis if the human body as follows. | | | | | | |
| | Clinical Application | | Mode of Operation | | | | |
| General<br>(TrackI only) | Specific<br>(TrackI & III) | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler |
| Ophthalmic | Ophthalmic | | | | | | |
| | Fetal | | | | | | |
| | Abdominal | | | | | | |
| | Intra-operative (Spec.) | | | | | | |
| | Intra-operative (Neuro.) | | | | | | |
| | Fetal Imaging Laparoscopic | | | | | | |
| & Other | Pediatric | | | | | | |
| | Small Organ (Spec.) | | | | | | |
| | Neonatal Cephalic | | | | | | |
| | Adult Cephalic | | | | | | |
| | Trans-rectal | | | | | | |
| | Trans-vaginal | | | | | | |
| | Trans-urethral | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | |
| | Intra-luminal | | | | | | |
| | Endoscopy | P | P | P | | P | P |
| | Cardiac Adult | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | |
| | Trans-esophageal (card.) | | | | | | |
| | Other (Spec.) | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | |
| | Other (Spec.) | | | | | | |
N=new indication: P=previously cleared by FDA, K131946 E=added under Appendix E
{4}------------------------------------------------
## Indications for Use 510(k) Number: K183516 Device Name: PENTAX Medical Ultrasound Video Bronchoscope EB19-J10U
| System: | Noblus | | | | | | |
|--------------------------|--------------------------------------------------------------------------------------------------|-------------------|---|-----|-----|------------------|----------------------|
| Probe: | EB19-J10U | | | | | | |
| | Intended use: Diagnostic ultrasound imaging or fluid flow analysis if the human body as follows. | | | | | | |
| | Clinical Application | Mode of Operation | | | | | |
| General<br>(TrackI only) | Specific<br>(TrackI & III) | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler |
| Ophthalmic | Ophthalmic | | | | | | |
| | Fetal | | | | | | |
| | Abdominal | | | | | | |
| | Intra-operative (Spec.) | | | | | | |
| | Intra-operative (Neuro.) | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | |
| | Pediatric | | | | | | |
| | Small Organ (Spec.) | | | | | | |
| | Neonatal Cephalic | | | | | | |
| | Adult Cephalic | | | | | | |
| | Trans-rectal | | | | | | |
| | Trans-vaginal | | | | | | |
| | Trans-urethral | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | |
| | Intra-luminal | | | | | | |
| | Endoscopy | P | P | P | | P | P |
| | Cardiac Adult | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | |
| | Trans-esophageal (card.) | | | | | | |
| | Other (Spec.) | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | |
| | Other (Spec.) | | | | | | |
N=new indication; P=previously cleared by FDA, K160559 E=added under Appendix E
{5}------------------------------------------------
## Indications for Use 510(k) Number: K183516 Device Name: PENTAX Medical Ultrasound Video Bronchoscope EB19-J10U
| System: | HI VISION Preirus | | | | | | |
|--------------------------|--------------------------------------------------------------------------------------------------|---|-------------------|-----|-----|------------------|----------------------|
| Probe: | EB19-J10U | | | | | | |
| | Intended use: Diagnostic ultrasound imaging or fluid flow analysis if the human body as follows. | | | | | | |
| | Clinical Application | | Mode of Operation | | | | |
| General<br>(TrackI only) | Specific<br>(TrackI & III) | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler |
| Ophthalmic | Ophthalmic | | | | | | |
| | Fetal | | | | | | |
| | Abdominal | | | | | | |
| | Intra-operative (Spec.) | | | | | | |
| | Intra-operative (Neuro.) | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | |
| | Pediatric | | | | | | |
| | Small Organ (Spec.) | | | | | | |
| | Neonatal Cephalic | | | | | | |
| | Adult Cephalic | | | | | | |
| | Trans-rectal | | | | | | |
| | Trans-vaginal | | | | | | |
| | Trans-urethral | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | |
| | Intra-luminal | | | | | | |
| | Endoscopy | P | P | P | | P | P |
| | Cardiac Adult | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | |
| | Trans-esophageal (card.) | | | | | | |
| | Other (Spec.) | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | |
| | Other (Spec.) | | | | | | |
N=new indication; P=previously cleared by FDA, K162447 E=added under Appendix E
{6}------------------------------------------------
# 510(k) Summary
| 510(k) Number | K183516 |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter | PENTAX of America, Inc. |
| Primary Contact | William Goeller<br>Vice President, Quality Assurance and Regulatory Affairs<br>PENTAX of America. Inc.<br>3 Paragon Drive<br>Montvale, New Jersey 07645-1782 |
| Preparation Date | June 21, 2019 |
| Trade Name | PENTAX Medical Ultrasound Video Bronchoscope EB19-J10U |
| Classification<br>Names | Bronchoscope (flexible or rigid) and accessories;<br>Diagnostic ultrasonic transducer |
| Device<br>Classification | Class II |
| Regulations | 21 CFR 874.4680;<br>21 CFR 892.1570 |
| Product Codes | EOQ, ITX |
| Predicate Device | PENTAX Ultrasound Video Bronchoscope EB-1970UK + HI VISION PREIRUS<br>(K131946, decision date April 25, 2014) |
| Reference<br>Devices | PENTAX Medical EPK-3000 Video Imaging System (K172156)<br>PENTAX Medical EPK-i5010 Video Processor with EB Family of Scopes (K143727)<br>PENTAX Medical EPK-i7010 Video Processor with EB Family of Scopes (K173679)<br>Hitachi Arietta70 Diagnostic Ultrasound System (K134016)<br>Hitachi Noblus™ Diagnostic Ultrasound Scanner (K142368)<br>Hitachi HI VISION Preirus Diagnostic Ultrasound Scanner (K093466) |
{7}------------------------------------------------
| Device Description | |
|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The PENTAX Medical Ultrasound Video Bronchoscope EB19-J10U connects with a video processor and an ultrasound scanner, both of which are software controlled devices.<br><br>The endoscope has a flexible insertion tube, a control body, PVE connector, and scanning unit connector. The PVE connector attaches to the video processor and has connections for illumination and video signals. The ultrasound umbilical connector attaches to the ultrasound scanner unit.<br><br>The control body includes remote buttons for functions assigned from the video processor. It also includes controls for up/down angulation or neutral position, suction control, and ports for manual balloon insufflation/evacuation and accessory inlet.<br><br>The endoscope contains light carrying bundles to illuminate the body cavity, a change couple device to collect endoscopic image data, and a convex array ultrasound transducer to collect ultrasonic image data. The instrument contains a working channel through which biopsy devices, or other devices may be introduced. The video processor contains a lamp that provides white light focused at the endoscope PVE connector light guide prong. The endoscope light carrying bundles present the light to the body cavity and the CCD collects endoscopic image data. Image data and other screen display information are formatted and presented to the video outputs of the video processor for display.<br><br>The ultrasound transducer delivers ultrasonic pulses, reflections of the pulses are received and the signals are passed to the ultrasound scanner for processing and display. The instrument is immersible (with the use of supplied cleaning accessories). EB19-J10U is connected to the ultrasound scanners Arietta 70 and Noblus via the scanning unit connector of the endoscope directly to the probe connector of the scanning unit. In order to connect to the Preirus scanning unit, junction box PUN-JBP1 is required to connect the scanning unit connector to the probe connector.<br><br>The instrument is immersible (with the use of supplied cleaning accessories). |
{8}------------------------------------------------
| | System:<br>Probe: | The PENTAX Medical Ultrasound Video Bronchoscope EB19-J10U is intended to<br>provide optical visualization of, ultrasonic visualization of, and therapeutic access to, the<br>Pulmonary Track including but not restricted to organs, tissues, and subsystem: Nasal<br>Passage, Pharynx, Larynx, Trachea, Bronchial Tree (including access beyond the stem),<br>and underlying areas. The instrument is introduced per orally when indications consistent<br>with the requirement for procedure are observed in adult and pediatric patient populations.<br>ARIETTA 70<br>EB19-J10U<br>Intended use: Diagnostic ultrasound imaging or fluid flow analysis if the human body as follows. | | | | | | | |
|------------------------------------------|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|---|-----|-----|------------------|----------------------|--|
| | | Clinical Application | Mode of Operation | | | | | | |
| | General<br>(Track I only) | Specific<br>(Track I & III) | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | |
| | | Ophthalmic Ophthalmic | | | | | | | |
| | | Fetal | | | | | | | |
| | | Abdominal | | | | | | | |
| | Fetal Imaging<br>& Other | Intra-operative (Spec.) | | | | | | | |
| | | Intra-operative (Neuro.) | | | | | | | |
| Intended Use /<br>Indications for<br>Use | | Laparoscopic | | | | | | | |
| | | Pediatric | | | | | | | |
| | | Small Organ (Spec.) | | | | | | | |
| | | Neonatal Cephalic | | | | | | | |
| | | Adult Cephalic | | | | | | | |
| | | Trans-rectal | | | | | | | |
| | | Trans-vaginal | | | | | | | |
| | | Trans-urethral | | | | | | | |
| | | Trans-esoph. (non-Card.) | | | | | | | |
| | | Musculo-skel. (Convent.) | | | | | | | |
| | | Musculo-skel. (Superfic.) | | | | | | | |
| | | Intra-luminal | | | | | | | |
| | | Endoscopy | P | P | P | | P | P | |
| | | Cardiac Adult | | | | | | | |
| | Cardiac | Cardiac Pediatric | | | | | | | |
| | | Trans-esophageal (card.) | | | | | | | |
| | | Other (Spec.) | | | | | | | |
| | Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | | Other (Spec.) | | | | | | | |
| | | N=new indication: P=previously cleared by FDA, K131946<br>E=added under Appendix E | | | | | | | |
{9}------------------------------------------------
| | Clinical Application | | | | | Mode of Operation | | | |
|--------------------------|---------------------------|----------------------------|---|---|-----|-------------------|------------------|----------------------|--|
| | General<br>(TrackI only) | Specific<br>(TrackI & III) | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | |
| Ophthalmic | Ophthalmic | | | | | | | | |
| | Fetal | | | | | | | | |
| | Abdominal | | | | | | | | |
| | Intra operative (Spec.) | | | | | | | | |
| | Intra operative (Neuro.) | | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | | |
| | Pediatric | | | | | | | | |
| | Small Organ (Spec.) | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | | | | | |
| | Trans rectal | | | | | | | | |
| | Trans vaginal | | | | | | | | |
| | Trans urethral | | | | | | | | |
| | Trans esoph. (non Card.) | | | | | | | | |
| | Musculo skel. (Convent.) | | | | | | | | |
| | Musculo skel. (Superfic.) | | | | | | | | |
| | Intra luminal | | | | | | | | |
| | Endoscopy | | P | P | P | | P | P | |
| | Cardiac Adult | | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | | |
| | Trans esophageal (card.) | | | | | | | | |
| | Other (Spec.) | | | | | | | | |
| | Peripheral vessel | | | | | | | | |
| Peripheral<br>Vessel | Other (Spec.) | | | | | | | | |
{10}------------------------------------------------
| | Clinical Application | | | | Mode of Operation | | |
|--------------------------|----------------------------|---|---|-----|-------------------|------------------|----------------------|
| General<br>(TrackI only) | Specific<br>(TrackI & III) | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler |
| | Ophthalmic Ophthalmic | | | | | | |
| | Fetal | | | | | | |
| | Abdominal | | | | | | |
| | Intra-operative (Spec.) | | | | | | |
| | Intra-operative (Neuro.) | | | | | | |
| | Fetal Imaging Laparoscopic | | | | | | |
| & Other | Pediatric | | | | | | |
| | Small Organ (Spec.) | | | | | | |
| | Neonatal Cephalic | | | | | | |
| | Adult Cephalic | | | | | | |
| | Trans-rectal | | | | | | |
| | Trans-vaginal | | | | | | |
| | Trans-urethral | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | |
| | Intra-luminal | | | | | | |
| | Endoscopy | P | P | P | | P | P |
| | Cardiac Adult | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | |
| | Trans-esophageal (card.) | | | | | | |
| | Other (Spec.) | | | | | | |
| Peripheral | Peripheral vessel | | | | | | |
| Vessel | Other (Spec.) | | | | | | |
| | | | | | | | |
{11}------------------------------------------------
| Technological<br>Characteristics | The subject and predicate devices have the following identical technological<br>characteristics: image supply, angulation control function, angulation lock function,<br>elevator control function, instrument channel inlet, air/water feeding function,<br>illumination, direction of view, field of view, depth of field, tip angulation, insertion tube<br>width, insertion tube working length, minimum visible distance of biopsy forceps, type of<br>CCD, both use software, scan direction, scan system, and scan angle. |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The subject and predicate devices have the following different but equivalent<br>technological characteristics: suction control functions, distal end width, minimum<br>instrument channel width, and acoustic frequency. A summary of the differences between<br>the equivalent technological characteristics of the subject and predicate devices are as<br>follows: |
| | Suction control function: The subject device accessory suction control valve is single-use.<br>The predicate device accessory suction control valve is reusable. |
| | Distal end width: The predicate device distal end width is 0.1mm smaller than the<br>predicate device. |
| | Minimum instrument channel width: The predicate device minimum instrument channel<br>width is 0.2mm larger than the predicate device. |
| | Acoustic frequency: The predicate device acoustic frequency is 3MHz greater than the<br>predicate device. |
| | These differences do not raise new questions of safety or effectiveness. |
{12}------------------------------------------------
| Testing<br>Summary | Sterilization (Accessories Provided Sterile) |
|-------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | • Sterilization |
| | • Packaging |
| | • Shelf Life |
| | Reprocessing |
| | • Soil accumulation |
| | • Cleaning validation |
| | • Rinsing validation study after cleaning process…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.