FUJIFILM Ultrasonic Endoscope

K182825 · Fujifilm Corporation · EOQ · Feb 22, 2019 · Ear, Nose, Throat

Device Facts

Record IDK182825
Device NameFUJIFILM Ultrasonic Endoscope
ApplicantFujifilm Corporation
Product CodeEOQ · Ear, Nose, Throat
Decision DateFeb 22, 2019
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 874.4680
Device ClassClass 2
AttributesTherapeutic

Indications for Use

EB-530US is intended for the observation, diagnosis, and endoscopic treatment of the trachea, bronchial tree and surrounding organs using ultrasonic images. It is used with a FUJIFILM ultrasonic processor, video processor, light source, other peripheral equipment and endoscopic accessories. It is not intended for use on children and infants.

Device Story

Ultrasonic bronchoscope; emits ultrasound waves; scans reflected signals to generate ultrasonic images. Used in clinical settings by physicians for visualization and treatment of trachea, bronchial tree, and surrounding organs. Operates in conjunction with FUJIFILM ultrasonic processor, video processor, and light source. Output displayed to clinician for real-time guidance during endoscopic procedures; facilitates diagnosis and treatment of pulmonary conditions.

Clinical Evidence

Bench testing only. Performance testing confirmed equivalence to predicate across parameters including field of view, forceps channel diameter, axial/lateral resolution, bending capability, suction rate, and penetration depth. Biocompatibility testing leveraged data from reference device EC-600WL v2 per ISO 10993 standards. Electrical safety verified per ANSI/AAMI ES 60601-1 and IEC 60601 series.

Technological Characteristics

Ultrasonic bronchoscope; utilizes ultrasound transducer for imaging. Materials include new epoxy resin. Compliant with IEC 60601-1 (ed 3.1) and IEC 60601-2-37 (ed 2.1). Compatible with VP-7000 video processor, BL-7000 light source, and SU-1 ultrasonic processor. Storage/transport range: -20°C to 60°C, 10-85% RH. Biocompatibility per ISO 10993-1, -5, -10.

Indications for Use

Indicated for observation, diagnosis, and endoscopic treatment of trachea, bronchial tree, and surrounding organs in adults. Contraindicated for use on children and infants.

Regulatory Classification

Identification

A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. February 22, 2019 FUJIFILM Corporation % Jeffrev Wan Specialist, Regulatory Affairs FUJIFILM Medical Systems U.S.A., Inc. 419 West Ave. Stamford, Connecticut 06902 Re: K182825 Trade/Device Name: FUJIFILM Ultrasonic Endoscope Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible or Rigid) and Accessories Regulatory Class: Class II Product Code: EOQ, ITX Dated: January 22, 2019 Received: January 23, 2019 Dear Jeffrey Wan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device. please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely yours. # Srinivas Nandkumar -S for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ | 510(k) Number (if known) | K182825 | |-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name | FUJIFILM Ultrasonic Endoscope EB-530US | | Indications for Use (Describe) | EB-530US is intended for the observation, diagnosis, and endoscopic treatment of the trachea, bronchial tree and surrounding organs using ultrasonic images. It is used with a FUJIFILM ultrasonic processor, video processor, light source, other peripheral equipment and endoscopic accessories. It is not intended for use on children and infants. | | Type of Use (Select one or both, as applicable) | <span> <svg height="12" width="12"> <path d="M1 1 L11 11 M1 11 L11 1" stroke="black" stroke-width="2"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (7/17) Page 1 of 1 {4}------------------------------------------------ #### 510(k) SUMMARY ## FUJIFILM Ultrasonic Endoscope EB-530US Date: October 3, 2018 #### Submitter's Information: FUJIFILM Corporation 798 Miyanodai Kaisei-Machi Ashigarakami-Gun, Kanagawa, 258-8538, Japan FDA Establishment Registration Number: 3001722928 ## Contact Person: Jeffrey Wan Specialist, Regulatory Affairs Telephone: (201) 675-8947 E-Mail: jeffrey.wan@fujifilm.com ## Identification of the Proposed Device: | Proprietary/Trade Name: | FUJIFILM Ultrasonic Endoscope EB-530US | |-------------------------|--------------------------------------------------------------------------------------------------------------| | Common Name: | Ultrasonic Endoscope | | Device Class: | Class II | | Review Panel: | Ear, Nose & Throat | | Classification: | Bronchoscope and accessories, 21 C.F.R. § 874.4680<br>Diagnostic ultrasonic transducer, 21 C.F.R. § 892.1570 | | Product Code: | EOQ, ITX | #### Predicate Device: Fujifilm Ultrasonic Endoscope EB-530US (K121035) #### Reference Device: Fujifilm Video Colonoscope Model EC-600WL v2 (K160196) #### Intended Use / Indications for Use: EB-530US is intended for the observation, diagnosis, and endoscopic treatment of the trachea, bronchial tree and surrounding organs using ultrasonic images. It is used with a FUJIFILM ultrasonic processor, video processor, light source, other peripheral equipment and endoscopic accessories. It is not intended for use on children and infants. {5}------------------------------------------------ # Device Description: FUJIFILM Ultrasonic Endoscope EB-530US is an ultrasonic bronchoscope that emits ultrasound waves and scans the reflected signals to provide ultrasonic images when used in combination with an ultrasonic processor. # Technoloqical Characteristics: The proposed devices EB-530US differ from the predicate devices in the following minor modifications: - . Compatibility with Video Processor VP-7000 and Light Source BL-7000 (K163675). - Compatibility with Ultrasonic Processor SU-1 (K153206). ● - . Compliance to IEC60601-1 edition 3.1 and IEC60601-2-37 edition 2.1 to support compatibility with VP-7000, BL-7000, and SU-1. - . Expansion of transport and storage conditions to a temperature range of -20°C to 60°C and a humidity range of 10 to 85% RH. - Use of a new epoxy resin in the manufacturing process. Both the new and original epoxy ● resins can be used for repair. ## Performance Data: Electrical safety of the proposed device was evaluated using the following standards: ANSI/AAMI ES 60601-1:2012, IEC 60601-1-2:2007, IEC 60601-1-6:2013, IEC 60601-2-18:2009, and IEC 60601-2-37:2015. The proposed device EB-530US was adopted into biocompatibility testing conducted on reference device EC-600WL v2 using the following consensus standards: ISO 10993-1:2009, ISO 10993-5:2009, and ISO 10993-10:2010. Storage and transportation testing was conducted on the proposed device EB-530US to validate the expanded storage and transportation conditions. Fujifilm conducted the following performance testing on the proposed device EB-530US to ensure that the modified device performs equivalently to the predicate device: - . Field of view . - Forceps channel diameter . - . Axial resolution - . Bending capability - Viewing direction ● - Lateral resolution . - Suction rate - . Resolution - Penetration depth - Working length . - LG output ● In all cases, the device met the pre-defined acceptance criteria for the test. {6}------------------------------------------------ ## Substantial Equivalence: The company's EB-530US has the same intended use and indications for use as the previously cleared predicate EB-530US (K121035). In addition, the proposed device has similar technological characteristics and principles of operation as the predicate. The minor differences between the proposed device and their predicate device do not raise new or additional questions of safety or effectiveness of the proposed device. Thus, the proposed device EB-530US is substantially equivalent to the predicate device. ## Conclusions: The modified EB-530US is substantially equivalent to the predicate EB-530US and conforms to applicable medical device safety and performance standards.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...