Ambu aScope 3 Slim 3.8/1.2 and Ambu aScope 4 Broncho Slim 3.8/1.2; Ambu aScope 3 Regular 5.0/2.2 and Ambu aScope 4 Broncho Regular 5.0/2.2; Ambu aScope 3 Large 5.8/2.8 and Ambu aScope 4 Broncho Large 5.8/2.8; Ambu aView Monitor
Applicant
Ambu A/S
Product Code
EOQ · Ear, Nose, Throat
Decision Date
Mar 28, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4680
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The aScope 3 and aScope 4 Broncho endoscopes have been designed to be used with the aView monitor, endotherapy accessories and other ancillary equipment for endoscopy within the airways and tracheobronchial tree. The aScope 3 system and aScope 4 Broncho system are for use in a hospital environment. The aScope 3 and aScope 4 Broncho are single-use designed for use in adults.
Device Story
System comprises sterile, single-use flexible bronchoscopes (aScope 3/4) and reusable aView monitor. Bronchoscope features maneuverable distal tip, LED light source, camera, and working channel for accessory insertion. Used in hospital settings by clinicians for airway/tracheobronchial tree visualization and intervention. Input: optical signals from distal tip camera. Transformation: image processing for display on aView monitor. Output: real-time video feed, snapshots, or video recordings. Clinicians use visual output to guide endoscopic procedures, navigate airways, and perform interventions via working channel. Benefits include single-use sterility, reducing cross-contamination risk, and providing maneuverable visualization for diagnostic/therapeutic airway access.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including ISO 8600-1, ISO 8600-3, ISO 8600-4, and ISO 594-1 compliance, bending angle/endurance testing, depth of field, image quality validation, aging performance, sterile packaging integrity, and electrical safety/compatibility (IEC 60601-1, IEC 60601-2-18, IEC 60601-1-2).
Technological Characteristics
Flexible endoscope with maneuverable tip; distal LED light source and camera; working channel for accessories. Sterilized via Ethylene Oxide. Reusable aView monitor for image display/recording. Connectivity includes external monitor output. Standards: ISO 8600-1, ISO 8600-3, ISO 8600-4, ISO 594-1, IEC 60601-1, IEC 60601-2-18, IEC 60601-1-2.
Indications for Use
Indicated for endoscopy within the airways and tracheobronchial tree in adults in a hospital environment. Compatible with specific ETT, DLT, and endoscopic accessory sizes.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
Vision-Science BRS-5000 Video Bronchoscope (K091768)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 28, 2018
Ambu A/S % Sanjay Parikh Director, QA/RA Ambu, Inc. 6230 Old Dobbin Lane, Suite 250 Columbia, MD 21045
Re: K173727
Trade/Device Name: Ambu® aScope™ 3 Slim 3.8/1.2 and Ambu® aScope™ 4 Broncho Slim 3.8/1.2; Ambu® aScope™ 3 Regular 5.0/2.2 and Ambu® aScope™ 4 Broncho Regular 5.0/2.2; Ambu® aScope™ 3 Large 5.8/2.8 and Ambu® aScope™ 4 Broncho Large 5.8/2.8; Ambu® aView Monitor Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible or Rigid) and Accessories Regulatory Class: Class II Product Code: EOQ Dated: February 23, 2018 Received: February 26, 2018
Dear Sanjay Parikh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely yours,
Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known) K173727
#### Device Name
Ambu® aScope™ 3; Ambu® aScope™ 4 Broncho; Ambu® aView™ Monitor
#### Indications for Use (Describe)
The aScope 3 and aScope 4 Broncho endoscopes have been designed to be used with the aView monitor, endotherapy accessories and other ancillary equipment for endoscopy within the airways and tracheobronchial tree.
The aScope 3 system and aScope 4 Broncho system are for use in a hospital environment.
The aScope 3 and aScope 4 Broncho are single-use designed for use in adults. They have been evaluated for the following endotracheal tubes (ETT), double lumen tubes (DLT) and endoscopic accessories (EA) sizes:
| | Minimum ETT<br>inner diameter | Minimum<br>DLT size | EA minimum working<br>channel width |
|----------------------------------------------|-------------------------------|---------------------|-------------------------------------|
| aScope 3 Slim and<br>aScope 4 Broncho Slim | 5.0 mm | 35 Fr | Up to 1.2 mm |
| aScope 3 Regular<br>aScope 4 Broncho Regular | 6.0 mm | 41 Fr | Up to 2.0 mm |
| aScope 3 Large<br>aScope 4 Broncho Large | 7.0 mm | - | Up to 2.6 mm |
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This 510(k) summary has been prepared in accordance with 21 CFR 807.87(h) and the 510(k) has been prepared in accordance with 21 CFR 807.92.
| Submitter | Ambu A/S<br>Baltorpbakken 13<br>DK-2750 Ballerup<br>Denmark<br>Tel.: +45 72 25 20 00<br>Fax.: +45 72 25 20 50 | | | | | | | | | | | | |
|-----------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|
| Contact Person | Name: Maja Brøns<br>Job Title: Regulatory Affairs Professional<br>Address: Ambu A/S, Baltorpbakken 13, DK-2750 Ballerup<br>Telephone number: +45 72 25 22 16<br>Fax number: +45 72 25 20 50 | | | | | | | | | | | | |
| Date Summary<br>Prepared | December 4, 2017 | | | | | | | | | | | | |
| Device Trade<br>Name | Ambu® aScope™ 3 Slim 3.8/1.2<br>Ambu® aScope™ 3 Regular 5.0/2.2<br>Ambu® aScope™ 3 Large 5.8/2.8<br>Ambu® aScope™ 4 Broncho Slim 3.8/1.2<br>Ambu® aScope™ 4 Broncho Regular 5.0/2.2<br>Ambu® aScope™ 4 Broncho Large 5.8/2.8<br>Ambu® aView™ Monitor | | | | | | | | | | | | |
| Device Common<br>Name | Flexible Bronchoscope | | | | | | | | | | | | |
| Device<br>Classification | Bronchoscope (flexible or rigid) and accessories<br>Product Codes: EOQ<br>21 CFR 874.4680<br>Class II | | | | | | | | | | | | |
| Legally<br>Marketed<br>devices to which<br>the device is<br>substantially<br>equivalent | Predicate Manufacturer Trade Name 510k number A+B+C Ambu A/S Ambu® aScope™ 3 K161656 Reference D Vision-Science BRS-5000 Video<br>Bronchoscope K091768 | | | | | | | | | | | | |
{4}------------------------------------------------
The Ambu® aScope™ 3 system consists of: Description of the Device Ambu® aScope™ 3 Slim 3.8/1.2 . Ambu® aScope™ 3 Regular 5.0/2.2 ● Ambu® aScope™ 3 Large 5.8/2.8 ● Ambu® aView™ Monitor . The Ambu® aScope™ 4 Broncho System Ambu® aScope™ 4 Broncho Slim 3.8/1.2 . Ambu® aScope™ 4 Broncho Regular 5.0/2.2 ● Ambu® aScope™ 4 Broncho Large 5.8/2.8 . Ambu® aView™ Monitor ● Ambu® aScope™ 3 and Ambu® aScope™ 4 Broncho is a sterile single use flexible bronchoscope and Ambu Ambu® aView™ is a reusable monitor. Ambu® aScope™ 3 and Ambu® aScope™ 4 Broncho endoscopes have the following physical and performance characteristics: . Maneuverable tip controlled by the user Flexible insertion cord ● Camera and LED light source at the distal tip ● Working channel ● . Sterilized by Ethylene Oxide . For single use The differences between sizes in both the Ambu® aScope™ 3 endoscope and Ambu® aScope™ 4 Broncho endoscope are as follows: Distal end outer diameter . ● Insertion tube outer diameter ● Working channel inner diameter ● Anqulation range Ambu® aView™ Monitor has the following physical and performance characteristics: ● Displays the image from Ambu® aScope™ 3 endoscope and Ambu® aScope™ 4 Broncho endoscope on the screen . Can record snapshots or video of image from Ambu® aScope™ 3 and Ambu® aScope™ 4 Broncho endoscope . Can connect to an external monitor . Reusable device The aScope 3 and aScope 4 Broncho endoscopes have been Indications for designed to be used with the aView monitor, endotherapy Use accessories and other ancillary equipment for endoscopy within the airways and tracheobronchial tree.
{5}------------------------------------------------
Special 510(K) Application Ambu aScope 3 System & aScope 4 Broncho System
The aScope 3 system and aScope 4 Broncho system is for use in a hospital environment.
The aScope 3 system and aScope 4 Broncho system are for use in a hospital environment. The aScope 3 and aScope 4 Broncho are single-use devices designed for use in adults.
hey have been evaluated for the following endotracheal tubes (ETT), double lumen tubes (DLT) and endoscopic accessories (EA) sizes:
| | Minimum ETT<br>inner<br>diameter | Minimum<br>DLT size | EA<br>minimum<br>working<br>channel<br>width |
|----------------------------------------------|----------------------------------|---------------------|----------------------------------------------|
| aScope 3 Slim and<br>aScope 4 Broncho Slim | 5.0 mm | 35 Fr | Up to 1.2 mm |
| aScope 3 Regular<br>aScope 4 Broncho Regular | 6.0 mm | 41 Fr | Up to 2.0 mm |
| aScope 3 Large<br>aScope 4 Broncho Large | 7.0 mm | - | Up to 2.6 mm |
Summary of the technological characteristics in comparison to the predicate and referenc devices
Ambu® aScope™ 4 is similar to the predicate and reference devices described in K161656 and K091768.
The product line extension Ambu® aScope™ 4 Broncho are similar to the predicate and reference devices in the following areas:
| and reference<br>devices | They all have the same intended use They are all single-use devices delivered sterile They are all flexible endoscopes with a maneuverable tip They are all video endoscopes with a camera located in the distal tip to provide an image on a separate monitor They all use an LED-light source located in the distal tip They all have suction functionality They all have a working channel for insertion of endoscopic accessories They all have the same insertion tube working length They all have equivalent inner and outer diameters in their corresponding sizes. |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance<br>Data -Bench | The following data are described for the product line extension<br>Ambu® aScope™ 4 Broncho System in the premarket notification: Declaration of Conformity with the product specific standards ISO 8600-1, ISO 8600-3, ISO 8600-4 and ISO 594-1 Performance tests to document the properties of bending angle, endurance and radius of the bending section, the depth of field and clearer image quality design validation test Aging Performance Test Sterile Packaging Integrity Test Electrical Compatibility according to IEC 60601-1-2 Electrical Safety according to IEC 60601-1 and IEC 60601-2-18 |
Results: All tests were passed.
{6}------------------------------------------------
| Performance<br>Data – Clinical | Not applicable. |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion | Based on the indication for use, technological characteristics,<br>performance data and comparison to predicate and reference<br>devices it is concluded that the functionality and intended use of<br>Ambu® aScope™ 3, Ambu® aScope™ 4 Broncho and Ambu® aView™<br>Monitor is equivalent to the predicate and reference devices.<br><br>It is concluded that Ambu® aScope™ 3, Ambu® aScope™ 4 Broncho<br>and Ambu® aView™ Monitor are as safe and effective and perform as<br>well as or better (regarding distal bending section/angulation range,<br>depth of field and exposure for clearer visibility of the image) than<br>the chosen legally marketed predicate and reference devices. |
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.